HYDRAVIEW GUIDEWIRE
K141831 · Brivant Limited · DQX · Sep 4, 2014 · Cardiovascular
Device Facts
| Record ID | K141831 |
| Device Name | HYDRAVIEW GUIDEWIRE |
| Applicant | Brivant Limited |
| Product Code | DQX · Cardiovascular |
| Decision Date | Sep 4, 2014 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1330 |
| Device Class | Class 2 |
Intended Use
The HydraView Polymer Guidewires are intended for use in the coronary and peripheral vasculature to facilitate the selective placement of interventional devices with compatible guidewire lumen and compatible coronary venous leads.
Device Story
HydraView Polymer Guidewire is a single-use, disposable device; 190cm length, 0.014" diameter stainless steel core wire with PTFE coating. Core wire undergoes centreless grinding to create a reduced-diameter distal segment, providing three levels of support (ES, DS, EDS). Operated manually by physicians in clinical settings to navigate coronary and peripheral vasculature; facilitates placement of interventional devices and coronary venous leads. Device provides mechanical support and guidance; output is physical navigation of the vasculature. Benefits include precise delivery of interventional tools. No AI/ML or software components.
Clinical Evidence
Bench testing only. No clinical data. Testing included dimensional verification, tensile strength, torque strength/response, coating adherence, catheter/lead compatibility, tip stiffness, biocompatibility, radiopacity, and particulate testing. Results demonstrate equivalence to predicate devices.
Technological Characteristics
Stainless steel core wire, PTFE coating, polymer distal tip. 190cm length, 0.014" diameter. Manual operation. Single-use. Biocompatibility established via identical materials/processes to existing approved products.
Indications for Use
Indicated for patients requiring selective placement of interventional devices in coronary and peripheral vasculature. Contraindicated for use in cerebral vasculature and for patients not acceptable for percutaneous intervention.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
- Streamer Guidewire (K083094)
- Charter Guidewire (K103377)
Related Devices
- K060551 — BRIVANT GUIDEWIRE · Brivant Limited · Jun 7, 2006
- K992237 — MEDTRONIC GT2 FUSION GUIDE WIRES · Medtronic Vascular · Sep 9, 1999
- K131710 — MERIT HYDROPHILIC GUIDE WIRE · Merit Medical Systems, Inc. · Jul 29, 2013
- K173532 — Raider Guidewire · Vascular Solutions, Inc. · Dec 15, 2017
- K152497 — PW Guidewires · Tatara Vascular, LLC · Jan 22, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an emblem featuring a stylized human figure with three faces in profile, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 4, 2014
Brivant Limited % Kenneth Walsh Regulatory Affairs Group Leader Parkmore West Business Park Galway, Ireland
Re: K141831
> Trade/Device Name: HydraView Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DOX Dated: July 4, 2014 Received: July 7, 2014
Dear Kenneth Walsh,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M. A. Hilleman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K141831
Device Name HydraView Polymer Guidewire
Indications for Use (Describe)
The HydraView Polymer Guidewires are intended for use in the coronary and peripheral vasculature to facilitate the selective placement of interventional devices with compatible guidewire lumen and compatible coronary venous leads.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Lake Region Medical. The logo consists of a stylized graphic to the left of the words "Lake Region" in a bold, serif font. Below "Lake Region" is the word "Medical" with a registered trademark symbol.
Product: HydraView Guidewire Submission Type: Special 510(k) Submission Date: July 04, 2014
## 510(k) Summary
| Device Name | HydraView Polymer Guidewire | | | |
|---------------------------------------------------|------------------------------------------------------------------------------------------|-----------------------------|---------|---------------|
| Submitters name | Brivant Limited, | | | |
| | Parkmore West Business Park | | | |
| | Galway, | | | |
| | Ireland | | | |
| Application | Kenneth Walsh | | | |
| Correspondent | Regulatory Affairs Group Leader | | | |
| | Lake Region Medical Limited | | | |
| | Tel: +353 91 385037 | | | |
| | Fax: +353 91 766598 | | | |
| Summary Preparation<br>Date | 4th July 2014 | | | |
| Device Name &<br>Classification | Trade Name: | HydraView Polymer Guidewire | | |
| | Common Name: | Guidewire | | |
| | Classification Name: | Catheter, Guidewire | | |
| | Device Classification: | Class II, 21 CFR §870.1330 | | |
| | Product Code: | DQX | | |
| Intended Use | Intended Use: | | | |
| | The HydraView Polymer Guidewires are intended for use in the coronary and peripheral | | | |
| | vasculature to facilitate the selective placement of interventional devices with | | | |
| | compatible guidewire lumen and compatible coronary venous leads. | | | |
| | Contraindications: | | | |
| | The HydraView Polymer Guidewires are not intended for use in the cerebral vasculature. | | | |
| | Patients judged not acceptable for percutaneous intervention. | | | |
| Device Description | The HydraView Guidewire is a disposable medical device designed for single use only. It | | | |
| | consists of a PTFE coated 190cm, 0.014" diameter stainless steel core wire, one end of | | | |
| | which is reduced in diameter over a 30cm approx. segment in a progressive fashion | | | |
| | through a centreless grinding operation. The profile of this reduced section affords the | | | |
| | product a reduced area of stiffness and is varied to produce 3 unique levels of support; | | | |
| | ES, DS and EDS. | | | |
| Predicate Devices | | Device/Manufacturer | 510k | Date |
| | Streamer Guidewire/ Brivant Limited | | K083094 | 7th July 2009 |
| | Charter Guidewire/ Brivant Limited | | K103377 | 18th May 2011 |
| Principle of Operation | The HydraView guidewire is operated manually by a manual process | | | |
| Comparison of<br>Technological<br>Characteristics | This HydraView Guidewire has the following differences from the primary predicate | | | |
| | device: | | | |
| | - Change in the polymer material at the distal tip of the device. | | | |
| | - Minor dimensional changes to the tip profile. | | | |
| | - Minor change to indications for use statement | | | |
| | No other changes are proposed to the device. | | | |
| | In vitro bench testing data is available to support a determination of substantial | | | |
| | equivalence between the Hydraview Guidewire models and the predicate (refer to | | | |
| | performance testing below). | | | |
| | The results of these tests provide reasonable assurance that the proposed device has | | | |
| | been designed and tested to assure conformance to the requirements for its intended | | | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Lake Region Medical. The logo consists of a stylized symbol resembling interconnected circles on the left, followed by the words "Lake Region" in a bold, serif font. Below "Lake Region" is the word "Medical" in a smaller font, with the registered trademark symbol next to it. A thin black line runs horizontally beneath the logo.
| | use and performs comparably to the existing devices in the range. The Hydraview<br>Guidewire design does not introduce any new issues of safety and effectiveness such<br>that the proposed Hydraview Guidewire is considered substantially equivalent to the<br>predicate devices. |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Testing<br>(non-clinical) | In vitro bench tests were carried out to demonstrate equivalence of the modified design<br>to the predicate marketed devices i.e Streamer and Charter with reference to the FDAs<br>guidance document "Coronary and Cerebrovascular Guidewire Guidance, Jan 1995". |
| | The following bench tests were performed:<br>- Dimensional Verification<br>- Tensile Strength<br>- Torque strength<br>- Torque response/Torqueability<br>- Coating adherence/Integrity<br>- Catheter/Lead compatibility<br>- Tip Stiffness<br>- Biocompatibility testing<br>- Radiopacity testing<br>- Particulate Testing |
| | The results from these performance evaluations demonstrate the proposed HydraView<br>design is substantially equivalent to the predicate Streamer Guidewire and that the<br>materials of construction are identical to the predicate Charter Guidewire<br><br>The changes made to the device do not raise any new questions regarding safety and<br>effectiveness. |
| | Biological Safety of the device has been established through successful use of the same<br>materials and manufacturing process in current 510(k) approved Lake Region Medical<br>products. |
| Conclusions | Based on safety and performance testing, technological characteristics and the<br>indications for use for the device, the HydraView Guidewire has been demonstrated to<br>be appropriate for its intended use and is considered to be substantially equivalent to<br>the predicate devices. |