SAFE'N'SOUND STAKED PASSIVE DELIVERY SYSTEM - CONE VERSION
K141664 · Rexam Healthcare · MEG · Oct 2, 2014 · General Hospital
Device Facts
| Record ID | K141664 |
| Device Name | SAFE'N'SOUND STAKED PASSIVE DELIVERY SYSTEM - CONE VERSION |
| Applicant | Rexam Healthcare |
| Product Code | MEG · General Hospital |
| Decision Date | Oct 2, 2014 |
| Decision | SESE |
| Submission Type | Abbreviated |
| Regulation | 21 CFR 880.5860 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
Single use device that is indicated for use as an accessory with sterile 1 mL-long staked needle prefilled ISO Standard glass syringes to aid in the protection of healthcare professionals, patients who self-inject doctor prescribed medications, and individuals that assist self-injecting patients, from accidental needle sticks. The devices can be used on a wide range of patients including children and adults.
Device Story
Passive needle-shielding accessory for 1 mL-long staked needle prefilled glass syringes; consists of subassembly (body, cone, sleeve, spring) and plunger rod. Operates via spring-activated mechanism triggered upon injection completion to fully contain needle. Used by healthcare professionals, patients, or caregivers in clinical or home settings. Provides physical barrier to prevent accidental needle sticks; enhances visibility of medication volume. Design modifications include split body (body and cone), redesigned proximal body, reduced syringe clips for lower insertion force, and stiffer polypropylene plunger rod.
Clinical Evidence
Bench testing and simulated clinical use studies performed. Biocompatibility testing met ISO 10993-5 and ISO 10993-10 requirements. Simulated clinical testing confirmed safe and effective needle shielding post-use.
Technological Characteristics
Molded plastic subassembly (body, cone, sleeve) and plunger rod; spring-actuated passive needle shielding. Polypropylene resin used for plunger rod. Compatible with 1 mL-long staked needle prefilled ISO Standard glass syringes (max needle length 1/2"). Biocompatibility per ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for use as an accessory with 1 mL-long staked needle prefilled ISO Standard glass syringes to prevent accidental needle sticks. Intended for healthcare professionals, patients self-injecting prescribed medications, and caregivers. Suitable for all patient populations, including children and adults.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
- Safe'n'Sound® Passive Delivery System (K101233)
Related Devices
- K150562 — Safe'n'Sound Staked Passive Delivery System - Plajex version · Nemera · Jun 2, 2015
- K122558 — ULTRASAFE PASSIVE NEEDLE GUARD · Safety Syringes, Inc. · Sep 20, 2012
- K112936 — SAFE'SOUND LUER LOCK PASSIVE DEVLIVERY SYSTEM · Rexam Consummer Plastics Inc.D/B/A Rexam Pharma · Jan 20, 2012
- K123743 — ULTRASAFE PLUS PASSIVE NEEDLE GUARD · Safety Syringes, Inc. · Mar 26, 2013
- K101233 — SAFE'N'SOUND PASSIVE DELIVERY SYSTEM · Rexam Pharma · Feb 18, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 2, 2014
Nemera Ms. Beatrice Grand Demars Regulatory Scientist 600 Deerfield Parkway Buffalo Grove, IL 60089
Re: K141664
Trade/Device Name: Safe'n'Sound® Staked Passive Delivery System - Cone Version Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Accessory Regulatory Class: II Product Code: MEG Dated: September 1, 2014 Received: September 5, 2014
Dear Ms. Grand-Demars:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runner DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141664
Device Name
Safe'n'Sound® Staked Passive Delivery System - Cone version
Indications for Use (Describe)
Single use device that is indicated for use as an accessory with sterile 1 mL-long staked needle prefilled ISO Standard glass syringes to aid in the protection of healthcare professionals, patients who self-inject doctor prescribed medications, and individuals that assist self-injecting patiental needle sticks. The devices can be used on a wide range of patients including children and adults.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
5 - Page 1 of 2
| Assigned 510(k) number: | K141664 |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company: | Nemera<br>600 Deerfield Parkway<br>Buffalo Grove, IL 60089<br>Phone: (800) 537-0178<br>Fax: (847) 325-3795 |
| Contact: | Beatrice GRAND DEMARS |
| Date Prepared: | September 11, 2014 |
| Trade/Proprietary Name: | Safe'n'Sound® Staked Passive Delivery System -<br>Cone version |
| Classification Name: | Antistick piston syringe accessory |
| Classification/Product Code: | 21 CFR 880.5860, Class II, Product Code MEG |
| Predicate Device: | K101233 Safe'n'Sound® Passive Delivery System<br>by Rexam Healthcare |
| Device Description: | This submission is provided for modifications to the<br>predicate device, which includes design changes<br>and a plunger rod raw material change. For details,<br>please refer to Attachment 1 for Design<br>Modifications Rationale. The Safe'n'Sound®<br>Staked Passive Delivery System – Cone Version is a<br>single use anti-needle stick accessory for use with<br>sterile 1 mL-long staked needle prefilled ISO<br>Standard glass syringes. It fits with 1 mL-long<br>staked syringes with a maximum needle length of<br>1/2" and consists of a subassembly (body, cone,<br>sleeve, and spring) and a loose plunger rod. |
| Intended Use: | Single use device that is indicated for use as an<br>accessory with sterile 1 mL-long staked needle<br>prefilled ISO Standard glass syringes to aid in the<br>protection of healthcare professionals, patients who<br>self-inject doctor prescribed medications, and<br>individuals that assist self-injecting patients, from<br>accidental needle sticks. The devices can be used on<br>a wide range of patients including children and |
| | |
| Technological Characteristic<br>Comparison Summary to<br>Predicate Device: | The Safe'n'Sound® Staked Passive Delivery System<br>– Cone version is similar to the predicate device in<br>general technological features and principle of<br>operation. Both are molded plastic subassemblies<br>and plunger rod. Both have a spring that activates<br>upon injection completion to fully contain the<br>needle. However, the Safe'n'Sound® Staked Passive<br>Delivery System - Cone version body part is split<br>into two (2) components (Body and Cone). The<br>proximal end of the body is slightly redesigned.<br>These changes enable an enhanced view of the unit<br>of measure. Sleeve design changes, including<br>syringe clips (2 of 6) are reduced in size so that<br>insertion force of the prefilled syringe is reduced.<br>Change to a new polypropylene resin for the<br>plunger rod, which is slightly stiffer to improve<br>rigidity. These design changes have proven to be<br>insignificant based upon design and process<br>validations, bench testing, biocompatibility testing<br>and simulated clinical use studies performed. |
| Performance Testing: | Bench testing has been performed on the<br>Safe'n'Sound® Staked Passive Delivery System -<br>Cone version. It confirmed the product functions as<br>intended and is substantially equivalent to the<br>predicate device. Biocompatibility testing<br>performed demonstrates that the product meets ISO<br>10993-5 and ISO 10993-10 requirements. |
| Simulated Clinical Testing: | Simulated clinical use testing has been performed. It<br>confirmed that the Safe'n'Sound® Staked Passive<br>Delivery System - Cone version could be used<br>safely and effectively to shield needles inside the<br>protection device after use. |
| Conclusion: | Based upon the design, technology, performance,<br>functional testing, and intended use, the<br>Safe'n'Sound® Staked Passive Delivery System -<br>Cone version is substantially equivalent to predicate<br>devices currently marketed under the Federal Food,<br>Drug and Cosmetic Act. The Safe'n'Sound® Staked<br>Passive Delivery System - Cone version raises no<br>new issues of safety or effectiveness. |
adults.
# Section 5: 510(k) Summary
{4}------------------------------------------------