RANDOX IMMUNOASSAY PREMIUM PLUS CONTROL LEVELS 1,2 AND LEVEL 3 AND IMMUNOASSAY PREMIUM PLUS TRI-LEVEL CONTROL

K140522 · Randox Laboratories Limited · JJY · May 22, 2014 · Clinical Chemistry

Device Facts

Record IDK140522
Device NameRANDOX IMMUNOASSAY PREMIUM PLUS CONTROL LEVELS 1,2 AND LEVEL 3 AND IMMUNOASSAY PREMIUM PLUS TRI-LEVEL CONTROL
ApplicantRandox Laboratories Limited
Product CodeJJY · Clinical Chemistry
Decision DateMay 22, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

This product is intended for in vitro diagnostic use as assayed quality control material to monitor the accuracy and reproducibility of analytes listed in the package insert. This in vitro diagnostic device is intended for prescription use only.

Device Story

Randox Immunoassay Premium Plus Controls are lyophilized human serum-based quality control materials used to monitor the performance of clinical chemistry and immunoassay systems. The device is supplied in three levels (Level 1, 2, and 3) and a Tri-Level configuration. It is intended for use by laboratory professionals in clinical settings to verify the accuracy and reproducibility of 49 specific analytes (e.g., hormones, therapeutic drugs, tumor markers, vitamins). The user reconstitutes the lyophilized serum with distilled water and processes it alongside patient samples on diagnostic analyzers. The resulting output—measured analyte concentrations—is compared against assigned values provided by the manufacturer to assess system performance. By identifying potential deviations in instrument, reagent, or technique, the device helps ensure the reliability of patient test results.

Clinical Evidence

Bench testing only. Stability studies (open-vial and frozen) confirmed that analytes remain within ≤10% deviation from fresh samples when stored according to instructions. Accelerated and real-time stability testing supported a 3-year shelf life. Value assignment was performed via consensus of external laboratories and in-house testing using statistical analysis (mean, SD, %CV).

Technological Characteristics

Lyophilized human serum matrix; requires reconstitution with distilled water. Analyte-specific quality control material. Storage: 2-8°C (unopened). Stability: 7 days at 2-8°C (reconstituted) for most analytes; specific analytes (C-Peptide, PTH, Thyroglobulin, ACTH) have shorter stability windows. No electronic components, software, or energy sources.

Indications for Use

Indicated for in vitro diagnostic use as assayed quality control material to monitor accuracy and reproducibility of specific analytes in clinical chemistry and immunoassay systems. Prescription use only.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # MAY 2 2 2014 Image /page/0/Picture/1 description: The image shows the word "RANDOX" in all capital letters. The font is bold and sans-serif. The letters are black against a white background. 510 (k) Summarv Randox Immunoassay Premium Plus Controls Levels 1, 2 and 3, KJYJSZ 2 Immunoassay Premium Plus Tri-Level Control 1. SAFETY AND EFFECTIVENESS AS REQUIRED BY 21 CFR 807.92 STATEMENT This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirement 21 CFR 807.92. #### 2. SUBMITTER NAME AND ADDRESS Name: Dr. Pauline Armstrong Address: Randox Laboratories Limited 55 Diamond Road, Crumlin, County Antrim, BT29 4QY, United Kingdom. Telephone: +44 (0) 28 9442 2413 Fax: +44 (0) 28 9445 2912 E-mail: Pauline.Armstrong@randox.com Date of Summary Preparation: May 15, 2014 3. 510k NUMBER, DEVICE PROPRIETARY NAME, COMMON NAME, PURPOSE FOR SUBMISSION, REGULATORY CLASSIFCATION, PANEL, PRODUCT CODE AND 21 CFR NUMBER 510k No: K140522 Device Proprietary Name: Randox Immunoassay Premium Plus Controls Levels 1, 2 & 3, Immunoassay Premium Plus Tri-Level Control. Common Name: Randox Immunoassay Premium Plus Controls Levels 1, 2 & 3, Immunoassay Premium Plus Tri-Level Control. Purpose for Submission: New Device Regulatory Classification: Multi-analyte Controls, All kinds (Assayed and Unassayed) Classification: Class 1. Reserved Panel: Clinical Chemistry Product Code: JJY 21 CFR Number: 21 CFR 862.1660 1 | Page {1}------------------------------------------------ # RANDOX ### 510 (k) Summary Randox Immunoassay Premium Plus Controls Levels 1, 2 and 3, Immunoassay Premium Plus Tri-Level Control #### 4. PREDICATE DEVICE PROPRIETARY NAMES AND 510 (k) NUMBERS Predicate Device Proprietary Name: Randox Immunoassay Controls Levels 1, 2 and 3 510 (k) Numbers: K040379 #### 5. INTENDED USE This product is intended for in vitro diagnostic use as assayed quality control material to monitor the accuracy and reproducibility of analytes listed in the package insert. #### 6. DEVICE DESCRIPTION Randox Immunoassay Premium Plus Controls are manufactured at three levels, Level 1, Level 2 and Level 3. The materials are supplied as Level 1, Level 2 , Level 3 and a Tri-Level configuration. Each 5 ml vial of lyophilized serum is reconstituted with exactly 5 ml of distilled water at +20 to 25° C. The base matrix used for the manufacture of Randox Immunoassay Premium Plus Controls is Human Serum with added chemicals. Human source material from which this product has been derived and has been tested at the donor level for the Human Immunodeficiency Virus (HIV1 & HIV2) antibody, Hepatitis B surface antigen (HbsAg) and the Hepatitis C virus (HCV) antibody and were found to be non-reactive based on FDA approved methods. However, since no method can offer complete assurance as to the absence of infectious agents, this material and all patient samples should be handled as though capable of transmitting infectious diseases and disposed of accordinaly. {2}------------------------------------------------ # RAND◎X ## 510 (k) Summary Randox Immunoassay Premium Plus Controls Levels 1, 2 and 3, Immunoassay Premium Plus Tri-Level Control ## 7. PREDICATE DEVICE COMPARISON TABLE ## TABLE 1: COMPARISON OF RANDOX IMMUNOASSAY PREMIUM PLUS CONTROLS LEVELS 1, 2 AND 3 WITH THE PREDICATE DEVICE | CHARACTERISTICS | RANDOX IMMUNOASSAY PREMIUM<br>PLUS CONTROLS LEVELS 1, 2 AND 3<br>( <i>New Device</i> ) | RANDOX IMMUNOASSAY CONTROLS<br>LEVELS 1, 2 AND 3<br>K040379<br>( <i>Predicate Device</i> ) | |-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | INTENDED USE | This product is intended for <i>in vitro</i><br>diagnostic use as assayed quality control<br>material to monitor the accuracy and<br>reproducibility of analytes listed in the<br>package insert. | The Randox Immunoassay Controls<br>Levels 1, 2 and 3 are intended for <i>in vitro</i><br>diagnostic use in the quality control of<br>Immunoassays on clinical chemistry and<br>Immunoassay systems. | | FORMAT | Lyophilised | Lyophilised | | MATRIX | Human Serum | Human Serum | | ANALYTES | 49 Analytes<br>25-OH Vitamin D, ACTH, AFP,<br>Aldosterone(Levels 2 & 3 only), Amikacin,<br>Androstenedione, Beta-2 Microglobulin, C-Peptide,<br>CA15-3, CA19-9, CA125, Carbamazepine, CEA,<br>Cortisol, DHEA-S, Digoxin, Ferritin, Folate, Free T3,<br>Free T4, FSH, Gentamicin, GH, IgE, Insulin, LH,<br>Oestradiol, Paracetamol, PTH, Phenytoin, Primodone,<br>Progesterone, Prolactin, PSA Free, PSA Total, Salicylic<br>Acid, SHBG, T-Uptake, Testosterone, Theophylline,<br>Thyroglobulin, TSH, Tobramycin, Total Beta hCG, Total<br>T3, Total T4, Valproic Acid, Vancomycin, Vitamin B12. | 33 Analytes<br>AFP, Beta-2 Microglobulin, CA15-3, CA19-9, CA125,<br>Carbamazepine, CEA, Cortisol, DHEA-S, Digoxin,<br>Ferritin, Folate, Free T3, Free T4 FSH, GH, IgE, LH,<br>Oestradiol, 17-OH-Progesterone, Phenytoin,<br>Progesterone, Prolactin, PSA Free, PSA Total,<br>Testosterone, Theophylline, TSH, Tobramycin, Total<br>Beta hCG, Total T3, Total T4, Valproic Acid, Vitamin B12. | | STORAGE<br>(Unopened) | 2 to 8 °C<br>Until expiration date | 2 to 8 °C<br>Until expiration date | | OPEN VIAL CLAIM | Reconstituted serum is stable for 7 days<br>at +2°C to +8°C if kept capped in original<br>container and free from contamination or<br>4 weeks frozen once at -20°C. C-<br>Peptide is stable for 1 day at<br>+2°C to +8°C. Parathyroid hormone<br>(PTH) and Thyroglobulin should be tested<br>within 4 hours of reconstitution, when<br>stored at +2°C to +8°C or within 2 weeks<br>when stored at -20°C. ACTH should be<br>assayed within 4 hours of reconstitution,<br>when stored at +2°C to +8°C. | Reconstituted serum is stable for 7 days<br>at +2°C to +8°C if kept capped in original<br>container and free from contamination or<br>4 weeks frozen once at -20°C. The PSA<br>in this sera will be stable for 2 days at<br>+2°C to +8°C. | | SIZE | 12 x 5ml | 10 x 5ml | | SHELF LIFE | 36 months | 36 months | {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "RANDOX" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The word appears to be a logo or brand name. ### 510 (k) Summary Randox Immunoassay Premium Plus Controls Levels 1, 2 and 3, Immunoassay Premium Plus Tri-Level Control ### 8. SUMMARY OF STABILITY STUDIES #### Open vial stabiltiy Open vial stability of the Randox Immunoassay Premium Plys Controls levels 1, 2 & 3 was assessed by opening a set of Immunoassay Premium Plus Controls levels 1, 2 & 3 and reconstituting them according to the package insert. Samples were reconstituted and stored at +2 to +8°C for 7 days and tested for the quoted analytes except C-Peptide stable for 1 day, Thyroglobulin, ACTH and Parathyroid hormone (PTH) should be tested within 4 hours of reconstitution. The acceptance criteria state the percentage deviation of reconstituted to fresh should be ≤10%. The tables below show the summary of the open vial stability. #### Results Reconstituted Stability Day 7 | Control | Lot<br>number | Reconstituted Stability at Day 7 | |------------------------------------------------------|---------------------------|----------------------------------| | Randox Immunoassay Premium Plus<br>Control levels 1 | 972EC | All analytes pass | | Randox Immunoassay Premium Plus<br>Control levels 2 | 974EC | All analytes pass | | Randox Immunoassay Premium Plus<br>Control levels 3 | 977EC | All analytes pass | | Randox Immunoassay Premium Plus<br>Tri-Level Control | 972EC,<br>974EC,<br>977EC | All analytes pass | #### Reconstituted Stability Day 1 for C-Peptide | Control | Lot<br>number | Reconstituted Stability at Day 1 | |------------------------------------------------------|---------------------------|----------------------------------| | Randox Immunoassay Premium Plus<br>Control levels 1 | 972EC | Pass | | Randox Immunoassay Premium Plus<br>Control levels 2 | 974EC | Pass | | Randox Immunoassay Premium Plus<br>Control levels 3 | 977EC | Pass | | Randox Immunoassay Premium Plus<br>Tri-Level Control | 972EC,<br>974EC,<br>977EC | Pass | {4}------------------------------------------------ # RANDOX ### 510 (k) Summary Randox Immunoassay Premium Plus Controls Levels 1, 2 and 3, Immunoassay Premium Plus Tri-Level Control Reconstituted Stability 4 Hours for ACTH, Thyroglobulin & Parathyroid Hormone. | Control | Lot number | Reconstituted Stability at 4 Hours | |----------------------------------------------------------|------------------------|------------------------------------| | Randox Immunoassay<br>Premium Plus Control<br>levels 1 | 972EC | All analytes pass | | Randox Immunoassay<br>Premium Plus Control<br>levels 2 | 974EC | All analytes pass | | Randox Immunoassay<br>Premium Plus Control<br>levels 3 | 977EC | All analytes pass | | Randox Immunoassay<br>Premium Plus Tri-Level<br>Control | 972EC, 974EC,<br>977EC | All analytes pass | The data demonstrates that the Randox Immunoassay Premium Plus Controls levels 1, 2 & 3 are stable for 7 days reconstituted and stored at + 2 - 8ºC. C-Peptide is stable for 1 day at +2℃ to +8℃. Parathyroid hormone (PTH) Thyroglobulin and ACTH should be tested within 4 hours of reconstitution, when stored at +2°C to +8°C ### Frozen stabiltiy Frozen stability of the Randox Immunoassay Premium Plus Controls levels 1, 2 & 3 was assessed by opening a set of Immunoassay Premium Plus Controls levels 1, 2 & 3 and reconstituting them according to the package insert. Samples were reconstituted and stored at -18 to -24°C for 28 days. After this period, the controls were thawed and a fresh vial was tested alongside the stability vials. The acceptance criteria state the percentage deviation of reconstituted to fresh should be ≤10%. No frozen stability claim is made for ACTH, Aldosterone and C-Peptide. {5}------------------------------------------------ # RANDOX #### 510 (k) Summary Randox Immunoassay Premium Plus Controls Levels 1, 2 and 3, Immunoassay Premium Plus Tri-Level Control The table below shows the summary of the Frozen stability Frozen stability after 4 weeks at -20°C | Control | Lot number | Reconstituted Stability at Day 7 | |---------------------------------------------------------|------------------------|----------------------------------| | Randox Immunoassay<br>Premium Plus Control<br>levels 1 | 972EC | All analytes pass | | Randox Immunoassay<br>Premium Plus Control<br>levels 2 | 974EC | All analytes pass | | Randox Immunoassay<br>Premium Plus Control<br>levels 3 | 977EC | All analytes pass | | Randox Immunoassay<br>Premium Plus Tri-Level<br>Control | 972EC, 974EC,<br>977EC | All analytes pass | The data demonstrates that the Randox Immunoassay Premium Plus Controls levels 1, 2 & 3 are stable for 28 days reconstituted and stored frozen at -18 to -24°C. No frozen stability claim is made for ACTH. Aldosterone and C-Peptide. ## Real Time Testing & Accelerated Stability Testing Accelerated stability was used to predict the shelf life of control materials. The Randox Immunoassay Premium Plus Controls levels 1, 2 & 3 were stored at elevated temperature for a period of four weeks. Following storage at the elevated temperature (28-32 °C, 35-39 °C and 43-47 °C), the accelerated stability control is then tested alongside a fresh control and the percentage deviation is calculated The acceptance criteria state the percentage deviation to fresh should be ≤10%. The shelf life of these products was set at 3 years from the date of manufacture, based on the performance of the accelerated stability testing. The real time stability of the controls was monitored to verify and validate the predicted or desirable shelf life. The Randox Immunoassay Premium Plus Controls levels 1, 2 & 3 were stored at ultra frozen conditions -75 to -90°C. Following storage at the ultra frozen temperature, the controls were then tested alongside control material stored unopened at the routine storage temperature of +2 to +8°C at various timepoints and the percentage deviation is calculated. The acceptance criteria state the percentage deviation to controls stored at the routine temperature should be ≤10%. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the word "RANDOX" in a bold, sans-serif font. The letters are black against a white background. The "O" in "RANDOX" is stylized with a circle inside of a circle. #### 510 (k) Summary Randox Immunoassay Premium Plus Controls Levels 1, 2 and 3, Immunoassay Premium Plus Tri-Level Control Current Real Time studies support a 3 year shelf life. The table below shows the summary of the accelerated stability and real time testing. Accelerated stability and Real time testing | Control | Lot<br>number | Accelerated<br>Stability | Real time<br>testing | |---------------------------------------------------------|---------------------------------------|--------------------------------|----------------------| | Randox Immunoassay<br>Premium Plus Control<br>levels 1 | 852EC/<br>941EC | All analytes pass<br>at Week 4 | 36 Months | | Randox Immunoassay<br>Premium Plus Control<br>levels 2 | 854EC/<br>943EC | All analytes pass at<br>Week 4 | 36 Months | | Randox Immunoassay<br>Premium Plus Control<br>levels 3 | 857EC/<br>946EC | All analytes pass at<br>Week 4 | 36 Months | | Randox Immunoassay<br>Premium Plus Tri-Level<br>Control | 852/941EC,<br>854/943EC,<br>857/946EC | All analytes pass at<br>Week 4 | 36 Months | ### 9. SUMMARY OF VALUE ASSIGNMENT Each batch of Immunoassay Premium Plus is submitted to a number of external laboratories and values are assigned from a consensus of results obtained by these laboratories and/or by in-house testing. Statistical analysis including the mean, SD, and %CV were calculated. With each batch, an assigned value is calculated from the mean specific value and an analyte specific percentage range is applied. Average values should normally fall within the listed range. However, variations may be caused by instrument, reagent, and laboratory technique. Therefore the range provided should only be considered as a reference and it is recommended that each laboratory establish its own ranges. ### 10. CONCLUSION Testing results indicate that the proposed device is substantially equivalent to the predicate device. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Gf Silver Spring, MD 20993-0002 May 22, 2014 #### RANDOX LABORATORIES LIMITED PAULINE ARMSTRONG QA/RA MANAGER 55 DIAMOND RD. CRUMLIN, CO. ANTRIM BT29 40Y, UK Re: K140522 Trade/Device Name: Randox Immunoassay Premium Plus Control Levels 1.2 & 3 Immunoassay Premium Plus Tri-level Control Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: I. Reserved Product Code: JJY Dated: April 01. 2014 Received: April 4, 2014 Dear Dr. Pauline Armstrong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class 11 (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements. including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. {8}------------------------------------------------ Page 2-Dr. Pauline Armstrong If you desire specific advice for your device on our labeling regulations (2) CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resources/orYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Sincercly yours. # Courtney H. Lias -S Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {9}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use 510(k) Number (if known) K140522 Device Name Immunoassay Premium Plus Controls Levels 1, 2 & 3, Immunoassay Premium Plus Tri-Level Control. Indications for Use (Describe) This product is intended for in vitro diagnostic use as assayed quality control materials to monitor the accuracy and reproducibility of the analytes listed in the package insert. This in vitro diagnostic device is intended for prescription use only. Type of Use (Select one or both, as applicable) V Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) # Yung W. Chan -S This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 Sec PRA Statement below.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%