ELECSYS CK-MB CALCHECK 5

K133284 · Roche Diagnostics · JJX · Nov 22, 2013 · Clinical Chemistry

Device Facts

Record IDK133284
Device NameELECSYS CK-MB CALCHECK 5
ApplicantRoche Diagnostics
Product CodeJJX · Clinical Chemistry
Decision DateNov 22, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

For use in calibration verification and for use in the verification of the assay range established by the Elecsys CK-MB reagent on the indicated Elecsys and cobas e immunoassay analyzers.

Device Story

Elecsys CK-MB CalCheck 5 is a lyophilized, human serum-based control containing human creatine kinase MB isoenzyme (CK-MB). Used in clinical laboratory settings by trained personnel to verify calibration and assay range of Elecsys CK-MB reagents on Roche immunoassay analyzers (Elecsys 2010, MODULAR ANALYTICS E170, cobas e 411, 601, 602). Product is reconstituted with distilled/deionized water before use. Healthcare providers use the resulting verification data to ensure analyzer performance remains within established specifications, supporting accurate clinical decision-making for cardiac diagnostic testing.

Clinical Evidence

No clinical data. Bench testing only. Value assignment conducted via duplicate runs on at least three Elecsys 2010 analyzers; median value used for assignment. Stability testing (real-time and open-vial) supports 36-month shelf-life at 2-8°C and 4-hour open-vial stability at 20-25°C.

Technological Characteristics

Lyophilized human serum matrix containing human CK-MB. Five-level concentration format. Reconstituted with 1.0 mL distilled/deionized water. Standardized against Elecsys CK-MB STAT assay (traceable to Abbott IMx assay). Compatible with Elecsys and cobas e immunoassay analyzers.

Indications for Use

Indicated for use as an assayed control for calibration verification and assay range verification of the Elecsys CK-MB reagent on Elecsys and cobas e immunoassay analyzers. Prescription use only.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K133284 # NOV 2 2 2013 | Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. | | | | | | | | | |----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|--------------------------|--------------|------------------|-----------------|------------------------------------------------------------------------------|---------------|-----| | Submitter name, address, contact | Roche Diagnostics<br>9115 Hague Road, P.O. Box 50416<br>Indianapolis, IN 46250-0416 | | | | | | | | | | | Contact Person: Kelli Turner<br>Phone: 317-521-4515<br>Fax: 317-521-2324<br>Email: kelli.turner@roche.com | | | | | | | | | | | Secondary Contact: Colleen Adams<br>Phone: 317-521-3577<br>Fax: 317-521-2324<br>Email: colleen.adams@roche.com | | | | | | | | | | | Date Prepared: November 22, 2013 | | | | | | | | | | Device Name | | | | | | | | | | | | <table><tr><td>Proprietary name:</td><td>Elecsys CK-MB CalCheck 5</td></tr><tr><td>Common name:</td><td>CK-MB CalCheck 5</td></tr><tr><td>Classification:</td><td>21 CFR 862.1660, Single (specified) analyte controls (assayed and unassayed)</td></tr><tr><td>Product Code:</td><td>JJX</td></tr></table> | Proprietary name: | Elecsys CK-MB CalCheck 5 | Common name: | CK-MB CalCheck 5 | Classification: | 21 CFR 862.1660, Single (specified) analyte controls (assayed and unassayed) | Product Code: | JJX | | Proprietary name: | Elecsys CK-MB CalCheck 5 | | | | | | | | | | Common name: | CK-MB CalCheck 5 | | | | | | | | | | Classification: | 21 CFR 862.1660, Single (specified) analyte controls (assayed and unassayed) | | | | | | | | | | Product Code: | JJX | | | | | | | | | | Device Description | The Elecsys CK-MB CalCheck 5 is a lyophilized product consisting of human creatine kinase MB isoenzyme (CK-MB) in a human serum matrix. During manufacture, the analyte is spiked into the matrix at the desired concentration levels.<br><br>All human material should be considered potentially infectious. All products derived from human blood are prepared exclusively from the blood of donors tested individually and shown to be free from HBsAg and antibodies to HCV and HIV. The testing methods applied were FDA-approved or cleared in compliance with the European Directive 98/79/EC, Annex II, List A. However, as no testing method can rule out the potential risk of infection with absolute certainty, the material should be handled with the same level of care as a patient specimen. In the event of exposure, the directives of the responsible health authorities should be followed. | | | | | | | | | | | Continued on next page | | | | | | | | | # 510(k) Summary , i Continued on next page {1}------------------------------------------------ | | Intended use | For use in calibration verification and for use in the verification of the assay<br>range established by the Elecsys CK-MB reagent on the indicated Elecsys and<br>cobas e immunoassay analyzers. | | | |--|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | | Predicate<br>device | The Elecsys CK-MB CalCheck 5 is substantially equivalent to other products<br>in commercial distribution intended for similar use. We claim equivalency to<br>the currently marketed Elecsys CK-MB CalCheck 5 (K093582). | | | | | Performance<br>Characteristics | The Elecsys CK-MB CalCheck 5 was evaluated for value assignment and<br>stability. See the following sections for details. | | | | | Traceability | The Elecsys CK-MB CalCheck 5 was standardized against the Elecsys CK-<br>MB STAT assay, which is traceable to the Abbott IMx assay and linearized<br>using human recombinant CK-MB from Seradyn. | | | {2}------------------------------------------------ # 510(k) Summary, Continued The following table compares the CK-MB CalCheck 5 with the predicate Substantial device, Elecsys CK-MB CalCheck 5 (K093582). Equivalence Comparison | Characteristic | Elecsys CK-MB CalCheck 5<br>(Candidate) | Elecsys CK-MB CalCheck 5<br>(K093582) | |----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------| | Intended Use | The Elecsys CK-MB CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys CK-MB reagent on the indicated Elecsys and cobas e immunoassay analyzers. | Same | | Analyte | Human CK-MB | Same | | Matrix | Human serum matrix | Same | | Levels | Five | Same | | Format | Lyophilized | Same | | Handling | Reconstitute Check 1, Check 2, Check 3, Check 4, and Check 5 with exactly 1.0 mL distilled or deionized water. Allow to stand closed for 15 minutes, then mix gently by inversion to ensure homogeneity. | Same | | Stability | Unopened:<br>Store at 2-8°C until expiration date<br>Opened:<br>20-25°C: 4 hours | Same | Continued on next page {3}------------------------------------------------ # 510(k) Summary, Continued · | Value Assignment | Value assignment testing was conducted and passed pre-defined acceptance criteria. For each Elecsys CK-MB CalCheck 5 lot manufactured, the CalChecks are run in duplicate on at least three Elecsys 2010 analyzers. The assigned value of each CalCheck is defined as the median value obtained over at least 6 determinations (duplicate runs on at least 3 analyzers) of the respective CalCheck. The assigned range is calculated as ±27% of assigned value for level 2 and as ±30% of the assigned value for levels 3 through 5. The label states that each laboratory should establish appropriate acceptance criteria when using this product for its intended use. | |------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The same value assignment procedure is performed on the MODULAR ANALYTICES E170. The assigned values obtained are compared to those obtained on the Elecsys 2010. The median value obtained on the additional analyzer must be within 10% of the master platform assigned value. After this acceptance criterion is met, the assigned values from the master platform are deemed valid for the MODULAR ANALYTICS E170, Elecsys 2010, cobas e 411, cobas e 601, and cobas e 602 immunoassay analyzers. | | Stability | Real-time and open-vial stability tests were conducted to establish the shelf-life and open-vial claims. | | | Open-Vial Stability After Reconstitution: | | | Testing was performed and the data support the package insert claim that reconstituted Elecsys CK-MB CalCheck 5 is stable up to 4 hours at 20-25°C. | | | Shelf-Life Stability: | | | Real-time testing done at 2-8°C supports a claim of 36 months. | . {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (USA). The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the bird-like figure. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002 November 22, 2013 ROCHE DIAGNOSTICS KELLI TÜRNER REGULATORY AFFAIRS PRINCIPAL 9115 HAGUE ROAD, P.O. BOX 50416 INDIANAPOLIS IN 46250-0416 Re: K133284 Trade/Device Name: Elecsys CK-MB CalCheck 5 Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material Regulatory Class: I, reserved Product Code: JJX Dated: October 24, 2013 Received: October 25, 2013 ## Dear Ms. Turner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2-Ms. Turner If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Carol C. Benson -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) k133284 ## Device Name Elecsys CK-MB CalCheck 5 ### Indications for Use (Describe) The Elecsys CK-MB CalCheck 5 is an assayed control for use in calibration and for use in the verification of the assay range established by the Elecsys CK-MB reagent on the indicated Elecsys and cobas e innunoassay analyzers. Type of Use (Select one or both, as applicable) [X] Prescription Use (Part 21 CFR 801 Subpart D) [ ] Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%