SOFTECH PLUS ETC02 CANNULA
K132946 · Teleflexmedical, Inc. · CCK · Jun 4, 2014 · Anesthesiology
Device Facts
| Record ID | K132946 |
| Device Name | SOFTECH PLUS ETC02 CANNULA |
| Applicant | Teleflexmedical, Inc. |
| Product Code | CCK · Anesthesiology |
| Decision Date | Jun 4, 2014 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 868.1400 |
| Device Class | Class 2 |
| Attributes | Therapeutic, Pediatric |
Intended Use
The Hudson RCI SOFTECH® Plus ETCO2 Nasal Cannula is an adjunct to oxygen therapy with its primary function being that of delivering low flow oxygen to a patient while providing a means to sample expired gas. It is intended for use in patients requiring oxygen therapy to improve blood oxygen levels while monitoring expired gas to determine ventilatory rate.
Device Story
SOFTECH Plus ETCO2 Cannula is a non-sterile, single-patient-use nasal cannula; functions as an adjunct to oxygen therapy. Device delivers low-flow oxygen through one nasal prong while simultaneously sampling exhaled gas from the other prong. Used in clinical settings for patients requiring oxygen therapy and ventilatory rate monitoring. Healthcare providers connect the device to oxygen supply and gas monitoring equipment; output allows clinicians to assess patient oxygenation and respiratory status. Benefits include simultaneous oxygen delivery and CO2 monitoring.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ETCO2 performance testing under simulated conditions at various oxygen flow rates, biocompatibility testing (cytotoxicity, sensitization, irritation, genotoxicity, implantation), and DEHP testing.
Technological Characteristics
Materials: Polyvinylchloride (PVC), Polypropylene, Polycarbonate. Non-DEHP PVC used. Split nasal prong design. Connectivity: Luer lock connectors for gas sampling. Single-patient use, non-sterile. No software or electronic components.
Indications for Use
Indicated for infant, pediatric, and adult patients requiring oxygen therapy to improve blood oxygen levels and monitoring of expired gas to determine ventilatory rate.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
Predicate Devices
- Double Lumen Oxygen Delivery Gas Sampling Nasal Cannula, Hudson RCI Bi-Flo (K961150)
Related Devices
- K143469 — SOFTECH Plus ETCO2 Cannula · Teleflexmedical, Inc. · Feb 12, 2015
- K120518 — END TIDAL CO2 NASAL CANNULA · Care Fusion · May 4, 2012
- K063813 — VENTLAB NASAL CO2 SAMPLING/O2 DELIVERY CANNULA · Ventlab Corp. · Mar 26, 2008
- K131410 — CO2/O2 NASAL CANNULA (WITH 4 (10CM) O2 AND CO2 TUBE), CO2/O2 NASAL CANNULA (WITH 8 (240CM) O2 AND 4 (10CM) CO2 TUBE), · Southmedic, Inc. · Jan 13, 2014
- K213767 — Endure ETCO2/02 Nasal Cannula · Endure Industries, Inc. · Sep 12, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Additional Information Response
## 510(k) SUMMARY
### A. Name, Address, Phone and Fax Number of Applicant
JUN 0 4 2014
Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-433-8050 919-433-4996 Fax:
### B. Contact Person
Angela Bouse Senior Regulatory Affairs Specialist
### C. Date Prepared
June 3, 2014
### D. Device Name
| Trade Name: | SOFTECH® Plus ETCO2 Cannula |
|--------------------------|----------------------------------------------|
| Classification Name: | Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase |
| Product Code: | CCK |
| Additional Product Code: | CAT |
| Regulation Number: | 868.1400 |
| Classification: | II |
| Classification Panel: | Anesthesiology |
### E. Predicate Device
This submission demonstrates substantial equivalence to the predicate device Double Lumen Oxygen Delivery Gas Sampling Nasal Cannula, Hudson RCI Bi-Flo -K961150.
## F. Device Description
The SOFTECH Plus ETCO2 Cannula is a non-sterile disposable, single patient use device that acts as an adjunct to oxygen therapy with its primary function providing a means to deliver low flow oxygen, while sampling part of the patients exhaled gas. The SOFTECH Plus ETCO2 Cannula has a split nare blank with oxygen delivery through one nasal prongs while allowing sampling of the patient's exhaled gas from the corresponding nasal prongs.
{1}------------------------------------------------
## G. Indications for Use
The Hudson RCI SOFTECH® Plus ETCO2 Nasal Cannula is an adjunct to oxygen therapy with its primary function being that of delivering low flow oxygen to a patient while providing a means to sample expired gas. It is intended for use in patients requiring oxygen therapy to improve blood oxygen levels while monitoring expired gas to determine ventilatory rate.
Patient Population: Infant, Pediatric, Adult
# H. Technological Characteristics and Material Comparison to the predicate
The proposed SOFTECH Plus ETCO2 Cannula is substantially equivalent to the predicate device listed above in that the indications for use, the intended use, and fundamental scientific technology remain unchanged. Table 1 summarizes the differences between the proposed and predicate devices.
| Features | Predicate<br>K961150<br>Double Lumen Oxygen<br>Delivery Gas Sampling<br>Nasal Cannula, Hudson<br>RCI Bi-Flo | Proposed<br>SOFTECH® Plus ETCO2 Cannula | Performance<br>Testing |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Nasal Prongs<br>(Nares) | Dual-channel nasal prong<br>that allows oxygen delivery<br>and gas sampling from both<br>nares | Split nasal prong that directs oxygen<br>flow into one nare and samples<br>expired gas from the other nare | ETCO2<br>Performance<br>Testing with<br>Simultaneous<br>Oxygen Delivery |
| Oxygen supply<br>tubing and CO2<br>sampling line<br>length | Oxygen supply tubing: 7 ft<br>CO2 sampling line: 7 in | Oxygen supply tubing: 7 ft and 14 ft<br>CO2 sampling line: 2 in, 7 ft and 14 ft | ETCO2<br>Performance<br>Testing with<br>Simultaneous<br>Oxygen Delivery |
| Materials | | | |
| Nares | Polyvinylchloride | Polyvinylchloride | Biocompatibility<br>Testing: |
| Bolo Slide | Polycarbonate or<br>Polyvinylchloride | Polyvinylchloride | Testing:<br>Cytotoxicity, |
| Tubing Clip | None | Polypropylene | Sensitization, |
| Lariat Tubing | Polyvinylchloride | Polyvinylchloride | Irritation, |
| Oxygen Supply<br>Tubing | Polyvinylchloride | Polyvinylchloride | Genotoxicity,<br>Implantation |
| CO2 Sampling<br>Line | Polyvinylchloride | Polyvinylchloride | |
| Oxygen<br>Connector | Polyvinylchloride | Polyvinylchloride | |
| Tubing Connector | Polyvinylchloride | Polyvinylchloride | |
| Male and Female<br>Luer Lock<br>Connectors | Acrylic | Polycarbonate | |
| 22mm Oxygen | None | Polypropylene | |
| Table 1 - Differences Between the Proposed and Predicate Devices<br> |
|----------------------------------------------------------------------|
|----------------------------------------------------------------------|
Teleflex Medical, Inc.
{2}------------------------------------------------
Additional Information Response
## K132946 – SOFTECH Plus ETCO2 Cannula Tab 006 - Section 011 - 510(k) Summary
| Tubing Adaptor | | | |
|----------------|---------------|---------------|--------------|
| Adhesive | Cyclohexanone | Cyclohexanone | |
| PVC | PVC with DEHP | Non-DEHP PVC | DEHP Testing |
## I. Performance Data
The proposed device was tested for ETCO2 sampling at various oxygen flow rates, biocompatibility, and DEHP testing. The test results demonstrate that the device is substantially equivalent to the predicate device.
The tests performed are summarized in Table 2 below.
| | 114 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 | Table 2 - Performance Testing Summary | |
|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|--|
| | | | |
| General Description |
|--------------------------------------------------------------------------------------------------------------------------------|
| ETCO2 performance in simulated conditions at various oxygen flow rates |
| Biocompatibility testing as per ISO 10993-1: cytotoxicity, sensitization, irritation,<br>genotoxicity and implantation testing |
| DEHP Testing |
## J. Conclusion
The device data and test results demonstrate that the device is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo features a stylized human figure with three arms reaching upwards, symbolizing health, hope, and well-being. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA", which is arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
#### June 4, 2014
Teleflex Medical, Inc. Angela Bouse Senior Regulatory Affairs Specialist 2917 Weck Drive Research Triangle Park, NC 27709
Re: K132946
Trade/Device Name: SOFTECH® Plus ETCO2 Cannula Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK, CAT Dated: May 02, 2014 Received: May 05, 2014
Dear Ms. Bouse:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of general voluless provincitive, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Ms. Angela Bouse
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
MarySuBunner-S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K132946
Device Name
Hudson RCI SOFTECH® Plus ETCO2 Nasal Cannula
Indications for Use (Describe)
The Hudson RCI SOFTECH® Plus ETCO2 Nasal Cannula is an adjunct to oxygen therapy with its primary function being that of delivering low flow oxygen to a patient while providing a means to sample expired gas. It is intended for use in patients requiring oxygen therapy to improve blood oxygen levels while monitoring expired gas to determine ventilatory rate.
Patient Population: Infant, Pediatric, Adult
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
Paul Career For Comment FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR SECTION CONSE
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Todd D. Courtney -S
2014.06.03 16:33:20 -0400
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# ·DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.