MULTICHEM U

K131993 · Techno-Path Manufacturing , Ltd. · JJY · Oct 15, 2013 · Clinical Chemistry

Device Facts

Record IDK131993
Device NameMULTICHEM U
ApplicantTechno-Path Manufacturing , Ltd.
Product CodeJJY · Clinical Chemistry
Decision DateOct 15, 2013
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

Multichem U is intended for use as an assayed urine quality control to monitor the precision of laboratory testing procedures for the analytes listed in the package insert. Two levels of Controls (Level 1 and Level 2) are supplied in liquid form. The following analytes are listed in the package insert: Amylase, Calcium, Chloride, Creatinine Enzymatic, Creatinine Picrate, Glucose, Magnesium, Microalbumin, Phosphorous, Potassium, Sodium, Urea Nitrogen, Uric Acid, and Urinary Protein.

Device Story

Multichem U is a liquid, human urine-based quality control material; contains purified biochemical extracts, chemicals, drugs, preservatives, and stabilizers. Used in clinical laboratories to monitor precision of urine chemistry testing procedures. Supplied in two levels (Level 1 and Level 2) to cover analytical ranges. Healthcare providers use the control by running it through clinical chemistry systems (e.g., Abbott ARCHITECT c8000) alongside patient samples; results are compared against established value assignment ranges to verify system performance and ensure accuracy of patient test results. Benefits include improved reliability of diagnostic testing and early detection of analytical drift.

Clinical Evidence

Bench testing only. Value assignment performed on Abbott ARCHITECT c8000 systems using 32 data points (16 runs, 2 replicates). Stability validated per CLSI EP25A; open vial stability established at 30 days at 2-8°C; accelerated shelf-life testing supports 24-month stability at 2-8°C. Drift limits set at 10% for most analytes, with 20% for others.

Technological Characteristics

Human urine-based matrix; liquid form. Contains purified biochemical material, chemicals, drugs, preservatives, and stabilizers. Storage: 2° to 8°C. Open vial stability: 30 days at 2° to 8°C. Analyte concentration adjusted using commercially available sources. No traceability to higher-order reference materials claimed.

Indications for Use

Indicated for use as an assayed urine quality control to monitor the precision of laboratory testing procedures for specific analytes (Amylase, Calcium, Chloride, Creatinine, Glucose, Magnesium, Microalbumin, Phosphorous, Potassium, Sodium, Urea Nitrogen, Uric Acid, Urinary Protein). Intended for professional laboratory use.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ k131993 ### 510(k) Summary Multichem U OCT 1 5 2013 #### 1.0 Submitter and Contact: Paula McCarthy Head of Quality and Regulatory Affairs Technopath Fort Henry Business Park Ballina, Co. Tipperary Ireland Telephone: + 353 61 335844 Fax: + 353 61 203034 Email: paula.mccarthy@techno-path.com Stephanie Garth Quality and Regulatory Affairs Consultant Global Compliance Plus 325 Big Elm St. Highland Village, TX 75077 Telephone: 469 230 0959 Email: sgarth@globalcomplianceplus.com #### 2.0 Date Submitted: June 28, 2013 #### 3.0 Device Identification Proprietary Names: Common Name: Classification: Product Code: JJY Regulation Number: Multichem U Multi-Analyte Controls, (Assayed and unassayed) Class 1, Reserved 21 CFR 862.1660 #### 4.0 Legally Marketed Predicate Device | - Candidate(s) | Predicate | Manufacturer | Document<br>. Number | |----------------|-----------------------------------------|-------------------------|----------------------| | Multichem U | Liquichek TM Urine Chemistry<br>Control | Bio-Rad<br>Laboratories | K020817 | The Multichem U Control is substantially equivalent to the Bio-Rad product listed above, currently in commercial distribution. #### 5.0 Device Description Technopath has developed two levels of a urine control prepared from human urine to which purified biochemical material (extracts of human and animal origin), chemicals, drugs/preservatives and stabilizers have been added. The control is used in liquid form for convenience. The two levels of control are available to allow performance monitoring within the analytical range. The following kit configurations are available: Model UC 201A.10 with Level 1 control; 15 vials with 10 mL contents Model UC 202A.10 with Level 2 control; 15 vials with 10 mL contents Model 05P80-10, with Bi-Level controls; 6 Level 1 vials and 6 Level 2 vials; vials have 5mL contents {1}------------------------------------------------ The serum from each donor contributing urine for this material has been tested by United States Food and Drug Administration (FDA) approved methods and found to be negative for antibodies to HIV and HCV, and nonreactive for HBsAg. #### 6.0 Intended Use Multichem U is intended for use as an assayed urine quality control to monitor the precision of laboratory testing procedures for the analytes listed in the package insert. Two levels of Controls (Level 1 and Level 2) are supplied in liquid form. The following analytes are listed in the package insert: Amylase, Calcium, Chloride, Creatinine Enzymatic, Creatinine Picrate, Glucose, Magnesium, Microalbumin, Phosphorous, Potassium, Sodium, Urea Nitrogen, Uric Acid, and Urinary Protein. #### 7.0 Comparison to the Predicate Multichem U control claims to be substantially equivalent to Liquichek™ Urine Chemistry Control. The control has same/similar design and modes of operation. The key features are summarized in the following table. | Characteristics: | Predicate Device: LiquichekTM<br>Urine Chemistry | Proposed Device: | | | | |------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--| | | Control | : Multichem U | | | | | Similarities | | | | | | | Intended Use: | Liquichek™ Urine Chemistry<br>Control is intended for use as an<br>assayed quality control urine to<br>monitor the precision of laboratory<br>procedures listed in the package<br>insert. | Multichem U is intended for use as an<br>assayed urine quality control to monitor<br>the precision of laboratory testing<br>procedures for the analytes listed in the<br>package insert. Two levels of Controls<br>(Level 1 and Level 2) are supplied in<br>liquid form. | | | | | Form: | Liquid | Liquid | | | | | Matrix: | Human urine based | Human urine based | | | | | Storage<br>(Closed/Shelf-<br>Life) | 2° to 8°C Until expiration date | 2° to 8°C Until expiration date | | | | | Open Vial | 30 days at 2° to 8°C | 30 days at 2° to 8°C | | | | | Differences | | | | | | | Analytes<br>(Assayed) | Contains:<br>Amylase<br>Calcium<br>Chloride<br>Cortisol<br>Creatinine | Contains:<br>Amylase<br>Calcium<br>Chloride<br>Creatinine Enzymatic<br>Creatinine Picrate | | | | {2}------------------------------------------------ | Characteristics | Predicate Device: Liquichek™ Urine Chemistry Control | Proposed Device: Multichem U | |-----------------|------------------------------------------------------|------------------------------| | | Glucose | Glucose | | | hCG (Qualitative) | | | | Magnesium | Magnesium | | | Microalbumin | Microalbumin | | | Osmolality | | | | pH | | | | Phosphorus | Phosphorous | | | Potassium | Potassium | | | Pregnancy | | | | Protein-Total | | | | Sodium | Sodium | | | Specific Gravity | | | | Urea | | | | Urea Nitrogen | Urea Nitrogen | | | Uric Acid | Uric Acid | | | | Urinary Protein | #### 8.0 Performance Characteristics #### 8.1 Value Assignment Summary Value assignment testing was performed utilizing internal procedures and protocols to determine typical values that would be seen for the product across Abbott ARCHITECT c8000 clinical chemistry systems with the associated reagent test systems: For the Multichem U control, 2 reagent lots and 2 calibrator lots were used to incorporate reagent and calibrator variation where available. 2 replicates from 16 runs were performed to give a total of 32 data points. Distinct runs, with minimum gaps of 2 hours were performed and a minimum of 8 calibration events were performed to incorporate variation from calibration and environmental sources. Value assignment ranges were established at the pre-determined criteria of 20% around the grand mean to most analytes; however, a 10% range is applied to the Potassium, Sodium and Chloride. #### 8.2 Stability Testing Summary Stability studies have been performed to determine the open vial stability and shelf-life for this For open vial stability, Technopath utilized internal procedures and two protocols control. methods (Classical [Forward] method and Isochronous - Staggered Start [Backwards / Backended] method) from CLSI EP25A entitled "Evaluation of Stability of In Virro Diagnostic Reagents." To minimize variation, where possible, one lot of reagent, calibrator and reference/control was used for the entire study, per analyte. Testing was performed over multiple days on 1 Abbott ARCHITECT c8000 clinical chemistry system with the associated reagent test systems. All Multichem U analytes from open vial and freshly thawed vial samples were tested in replicates of 3 at each time points were tested and the point of failure was determined by the maximum allowable drift (degradation), which was determined to be analyte specific. {3}------------------------------------------------ Product claims are as follows: Open Vial Stability: 30 days at 2° - 8°C Accelerated testing was carried out utilizing CLSI EP25A in order to support a shelf-life storage claim of 24 months at 2° to 8°C. The accelerated testing utilized three lots of controls. All data was generated using the Abbott ARCHITECT c8000 clinical chemistry system with the associated reagent test systems. For the Multichem U control, the Drift Limit was determined to be 10%. These results concluded that the Multichem U Controls product is predicted to be stable for in excess of 24 months when stored at 2° - 8° C. The real-time testing is not available at this time, but is on-going and will be performed on a minimum of 3 lots. #### 8.3 Traceability Summary The analytes contained within the Multichem U control (Level 1 and 2) are found endogenously in the base urine matrix and are adjusted to the required concentration through use of commercially available sources. Technopath does not claim traceability to higher order reference materials or methods. #### 9.0 Conclusion: The conclusions drawn from the nonclinical tests (discussed above) demonstrate that the Multichem U control is as safe, as effective, and performs as well as the predicate device. The submitted information in this premarket notification is complete and supports a substantial equivalence decision. {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002 October 15, 2013 Techno-Path Manufacturing Ltd. C/O Stephanie Garth Global Compliance Plus 325 Big Elm St. HIGHLAND VILLAGE TX 75077 Re: K131993 Trade/Device Name: Multichem U Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: I, reserved Product Code: JJY Dated: August 23, 2013 Received: August 26, 2013 Dear Ms. Garth: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Orug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisiopp of the I 11 The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbrandi, wand adulteration. Please note: CDRH does not evaluate information related to contractly and warranties. We remind you, however, that device labeling must be truthful and not mislerading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your cevile can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA over publish further announcements conceming your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2-Ms. Garth If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.lda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Carol C. Benson -S for Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use Form 510(k) Number (if known): k131993 Device Name: _________________________________________________________________________________________________________________________________________________________________ Indications for Use: : . .. . > Multichem U is intended for use as an assayed urine quality control to monitor the precision of laboratory testing procedures for the analytes listed in the package insert. Two levels of Controls (Level 1 and Level 2) are supplied in liquid form. The following analytes are listed in the package insert: - · Amylase - · Calcium - Chloride . - · Creatinine Enzymatic - Creatinine Picrate . - · Glucose - Magnesium - · Microalbumin . - Phosphorous - · Potassium - · Sodium - · Urea Nitrogen - * Uric Acid - . Urinary Protein | Prescription Use _X | |-----------------------------| | (Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use _ | |------------------------| | (21 CFR 801 Subpart C) | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) # Yung W. Chan -S Division Sign-Off Office of In Vitro Diagnostics and Radiological Health 510(k):k131993 Page 1 of _ 1 _
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%