Browse hierarchy Pathology (PA) Subpart B — Biological Stains 21 CFR 864.1870 Product Code PFG K131508 — VYSIS D7S486/CEP 7 FISH PROBE KIT
VYSIS D7S486/CEP 7 FISH PROBE KIT
K131508 · Abbott Molecular, Inc. · PFG · Sep 13, 2013 · Hematology
Device Facts
Record ID K131508
Device Name VYSIS D7S486/CEP 7 FISH PROBE KIT
Applicant Abbott Molecular, Inc.
Product Code PFG · Hematology
Decision Date Sep 13, 2013
Decision SESE
Submission Type Traditional
Regulation 21 CFR 864.1870
Device Class Class 2
Intended Use
The Vysis D7S486/CEP 7 FISH Probe Kit is a device intended for specimen characterization, and detects the LSI D7S486 probe target on chromosome 7q31 and the CEP 7 probe target on chromosome 7p11.1-q11.1 in bone marrow and peripheral blood specimens from patients with acute myeloid leukemia or myelodysplastic syndrome. The assay results are intended to be interpreted by a qualified pathologist or cytogeneticist. This device is not intended for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening. The use of this device for diagnosis, prognosis, monitoring, or risk assessment has not been established.
Device Story
The Vysis D7S486/CEP 7 FISH Probe Kit is an in vitro diagnostic tool for detecting chromosomal abnormalities in bone marrow and peripheral blood. It utilizes fluorescence in situ hybridization (FISH) to target the LSI D7S486 locus (7q31) and the CEP 7 centromeric region (7p11.1-q11.1). The kit includes DNA probes labeled with SpectrumOrange and SpectrumGreen fluorophores, hybridization buffer, and counterstain. In the laboratory, specimens are prepared on slides, denatured, and hybridized with the probes. After washing, a pathologist or cytogeneticist uses a fluorescence microscope to visualize signal patterns. A typical cell shows two orange and two green signals (2R2G). A 1R1G pattern indicates monosomy 7; a 1R2G pattern indicates 7q deletion. The device provides objective data on chromosomal status to assist in specimen characterization; it is not intended for clinical diagnosis, prognosis, or therapeutic decision-making.
Clinical Evidence
No clinical trials were conducted for this submission. Evidence consists of analytical performance data (specificity, sensitivity, reproducibility) and literature review. Analytical specificity was 100% (200/200 loci). Analytical sensitivity was 98.1% for bone marrow and 98.5% for peripheral blood. Reproducibility studies across three sites and three lots showed high agreement with known status. Literature references (Vance et al., Cherry et al., Tefferi et al.) support the detection of 7q- and monosomy 7 in AML, MDS, and myelofibrosis patients.
Technological Characteristics
The device uses FISH technology with fluorophore-labeled DNA probes (SpectrumOrange/SpectrumGreen). Components include LSI D7S486/CEP 7 probes, hybridization buffer, DAPI II counterstain, NP-40, and 20X SSC. It requires a fluorescence microscope for visualization. The LSI D7S486 probe is ~308 kb. The CEP 7 probe targets the D7Z1 alpha satellite sequence. The assay is manual, performed on microscope slides using standard cytogenetic procedures.
Indications for Use
Indicated for specimen characterization in patients with acute myeloid leukemia (AML) or myelodysplastic syndrome (MDS) using bone marrow or peripheral blood specimens. Not for high-risk uses including therapy selection, therapeutic response prediction, or disease screening.
Regulatory Classification
Identification An early growth response 1 (EGR1) gene fluorescence in-situ hybridization (FISH) test system for specimen characterization is a device intended to detect the EGR1 probe target on chromosome 5q in bone marrow specimens from patients with acute myeloid leukemia (AML) or myelodysplastic syndrome (MDS). The assay results are intended to be interpreted only by a qualified pathologist or cytogeneticist. These devices do not include automated systems that directly report results without review and interpretation by a qualified pathologist or cytogeneticist. These devices also do not include any device intended for use to select patient therapy, predict patient response to therapy, or to screen for disease as well as any device with a claim for a particular diagnosis, prognosis, monitoring, or risk assessment.
Special Controls
EGR1 Gene FISH test system for specimen characterization must comply with the following special controls: 1) Premarket notification submissions must include the following information: i) A detailed description of all probes included in the kit ii) Purpose of each probe iii) Probe molecular specificity iv) Probe specificity v) Probe limits vi) Probe sensitivity vii) Specification of required ancillary reagents, instrumentation and equipment viii) Specification of the specimen collection, processing, storage and slide preparation methods ix) Specification of the assay procedure x) Specification of control elements that are incorporated into the recommended testing procedures xi) Specification of risk mitigation elements: description of all additional procedures. methods, and practices incorporated into the directions for use that mitigate risks associated with testing xii) Specification of the criteria for test result interpretation and reporting xiii) Device analytical sensitivity data xiv) Device analytical specificity data xv) Device reference limit data xvi) Device precision/reproducibility data xvii) Device stability data to include: A) Real-time Stability B) Freeze-Thaw Stability C) Transport and Temperature Stability D) Post-Hybridization Signal Stability E) Photostability of Probe xviii) Documentation that demonstrates the clinical validity of the device. The documentation must include data from clinical studies, a minimum of two peer-reviewed published literature references using the specific device seeking marketing clearance, or both. Documentation for peer-reviewed published literature references cited must include the following elements: A) Documentation that the sponsor's probe was used in the literature reference B) Number & type of specimens C) Target population studied. Target population must include the intended use population D) Upper reference limit F) Range of positive probe results 2) 21 CFR 809.10(b)(12) compliant labeling must include a statement summarizing the data identified in subparagraphs (1)(xiii)-(xviii) and a description of the studies supporting the information, including the pre-specified acceptance criteria for these performance studies, justification for the prespecified acceptance criteria, and whether the prespecified acceptance criteria were met. 3) 21 CFR 809.10 compliant labeling must include: i) A warning that reads "The assay results are intended to be interpreted only by a qualified pathologist or cytogeneticist." ii) A warning that reads "This device is not for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening." iii) A warning that reads "The use of this device for diagnosis, monitoring or risk assessment has not been established."
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must also include the following information:
(i) A detailed description of all probes included in the kit;
(ii) Purpose of each probe;
(iii) Probe molecular specificity;
(iv) Probe specificity;
(v) Probe limits;
(vi) Probe sensitivity;
(vii) Specification of required ancillary reagents, instrumentation, and equipment;
(viii) Specification of the specimen collection, processing, storage and slide preparation methods;
(ix) Specification of the assay procedure;
(x) Specification of control elements that are incorporated into the recommended testing procedures;
(xi) Specification of risk mitigation elements: Description of all additional procedures, methods, and practices incorporated into the directions for use that mitigate risks associated with testing;
(xii) Specification of the criteria for test result interpretation and reporting;
(xiii) Device analytical sensitivity data;
(xiv) Device analytical specificity data;
(xv) Device reference limit data;
(xvi) Device precision/reproducibility data;
(xvii) Device stability data to include:
(A) Real-time stability,
(B) Freeze-thaw stability,
(C) Transport and temperature stability,
(D) Post-hybridization signal stability,
(E) Photostability of probe, and
(xviii) Documentation that demonstrates the clinical validity of the device. The documentation must include data from clinical studies, a minimum of two peer-reviewed published literature references using the specific device seeking marketing clearance, or both. Documentation for the clinical studies and peer-reviewed published literature references cited must include the following elements:
(A) Documentation that the sponsor's probe was used in the literature reference,
(B) Number and type of specimens,
(C) Target population studied,
(D) Upper reference limit, and
(E) Range of positive probe results.
(2) Your § 809.10(b)(12) of this chapter compliant labeling must include a statement summarizing the data identified in paragraphs (b)(1)(xiii) through (xviii) of this section and a description of the studies supporting the information, including the pre-specified acceptance criteria for these performance studies, justification for the pre-specified acceptance criteria, and whether the pre-specified acceptance criteria were met.
(3) Your § 809.10 of this chapter compliant labeling must include:
(i) A warning that reads “The assay results are intended to be interpreted only by a qualified pathologist or cytogeneticist.”
(ii) A warning that reads “This device is not for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening.”
(iii) A warning that reads “The use of this device for diagnosis, monitoring or risk assessment has not been established.”
Predicate Devices
Vysis EGR1 FISH probe Kit- SC (K123951 )
Related Devices
DEN170070 — MLL (KMT2A) Breakapart FISH Probe Kit; AML1 (RUNX1) Breakapart FISH Probe Kit, P53 (TP53) Deletion FISH Probe Kit; EVI1 (MECOM) Breakapart FISH Probe Kit, Del(20q) Deletion FISH Probe Kit; AML1/ETO (RUNX1/RUNXIT1)) Translocation, Dual Fusi, CBFB (CBFB) /MYH11 Translocation, Dual Fusion FISH Probe Kit,Del(5q)Deletion FISH Probe Kit; Del(7q)Deletion FISH Probe Kit · Cytocell, Ltd. · Dec 21, 2018
K091960 — VYSIS EGR1 FISH PROBE KIT (VYSIS LSI EGR1 SPECTRUM/ORANGE/D5S23, D5S721, SPECTRUM GREEN PROBE), MODEL 4N37-020 · Abbott Molecular, Inc. · Aug 29, 2011
DEN130010 — VYSIS EGR1 FISH PROBE KIT - SC (SPECIMEN CHARACTERIZATION) · Abbott Molecular, Inc. · Jul 29, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(k) Summary
510(k) Number - K131508
Device Name: Vysis D7S486/CEP 7 FISH Probe Kit (List No. 04N78-020)
#### Purpose of the Submission
The purpose of this 510(k) is to gain clearance to market the Vysis D7S486/CEP 7 FISH Probe Kit (List No. 04N78-020).
#### Official Correspondent to the File
| Name | Dr. Nancy Bengtson |
|------------|-----------------------------------------------|
| Title: | Associate Director, Global Regulatory Affairs |
| Telephone: | (224) 361-7087 |
| Fax: | (847) 775-6777 |
| Email: | Nancy.Bengtson@abbott.com |
| Name: | Ms. Pamela L. Swatkowski |
| Title: | Director, Global Regulatory Affairs |
| Telephone: | (224) 361-7013 |
| Fax: | (847) 775-6777 |
| Email: | pamela.swatkowski@abbott.com |
Address: Abbott Molecular Inc. 1300 E. Touhy Avenue Des Plaines, IL 60018
# Date of Preparation
August 5, 2013
#### Manufacturer
Abbott Molecular Inc. is the legal manufacturer of the Vysis D7S486/CEP 7 FISH Probe Kit (List No. 04N78-020).
| Name: | Timothy Zurow, PhD |
|------------|--------------------------------------|
| Title: | Director of Manufacturing Operations |
| Telephone: | (224) 361-7379 |
| Fax: | (224) 361-7038 |
| Email: | timothy.zurow@abbott.com |
{1}------------------------------------------------
Address: Abbott Molecular Inc. 1300 E. Touhy Avenue Des Plaines. IL 60018
#### Establishment Registration No .: 3005248192
#### Intended Use/Indications For Use
The Vysis D7S486/CEP 7 FISH Probe Kit is a device intended for specimen characterization, and detects the LSI D7S486 probe target on chromosome 7q31 and the CEP 7 probe target on chromosome 7p11.1-g11.1 in bone marrow and peripheral blood specimens from patients with acute myeloid leukemia or myelodysplastic syndrome. The assay results are intended to be interpreted by a qualified pathologist or cytogeneticist. This device is not intended for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening. The use of this device for diagnosis, prognosis, monitoring or risk assessment has not been established.
#### Trade Name
Vysis D7S486/CEP 7 FISH Probe Kit (List No. 04N78-020)
#### Common Name
Vysis D7S486/CEP 7 Fluorescence In Situ Hybridization (FISH) Probe Kit
#### Classification
Class II
#### Regulation Number
21 CFR 864.1870
510(k) Summary Page 2 of 19
{2}------------------------------------------------
# Product Code
# PFG
# Predicate Device(s)
#### 81, Immunology; PDO, Vysis EGR1 FISH Probe Kit - SC (Specimen Characterization); 510(k) K123951 :
.
# Comparison with Predicate Device(s)
| Predicate | Device Item Being<br>Compared | Similarities | Differences: |
|-----------------------------------------------------------------------------------------------------|--------------------------------------------------|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| For specimen<br>characterization and<br>detection of a specific<br>probe target on a<br>chromosome. | Intended Use | Same | |
| Specimen characterization<br>and detection of LSI<br>EGR1 probe target on<br>chromosome 5q | Indications for Use | | Specimen<br>characterization and<br>detection of LSI<br>D7S486 (7q31)<br>probe target on<br>chromosome 7q and<br>the LSI D7S486 and<br>CEP 7 (7p11.1-<br>q11.1) probe targets<br>on chromosome 7. |
| Acute myeloid leukemia<br>(AML) patients | Patient Population | Same | In addition<br>myelodysplastic<br>syndrome |
| Vysis EGR1 FISH Probe<br>Kit – SC (Specimen<br>Characterization) | Device Name | | Vysis D7S486/CEP<br>7 FISH Probe Kit |
| 04N37-001 | List Number | | 04N78-020 |
| PDO | PFG | | |
| LSI EGRI | Probe Target | | LSI D7S486<br>CEP 7 |
| Deletion | Type of atypical FISH<br>signal pattern detected | Same | In addition loss of<br>chromosome |
{3}------------------------------------------------
| DNA Fluorophore labeled<br>Probes<br>Vysis LSI/WCP<br>Hybridization buffer,<br>DAPI II<br>NP-40<br>20XSSC | Kit components | Same | |
|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|--------------|----------------------------------------|
| Predicate | Device Item Being<br>Compared | Similarities | Differences |
| FISH assay method | FISH procedure | Same | |
| 6% | Upper reference limit | | 4.5% (monosomy 7)<br>6.5% (loss of 7q) |
| Fluorescence In Situ<br>Hybridization | Technology | Same | |
| Bone Marrow | Specimen Type | Same | Addition of<br>Peripheral Blood |
| Analytical Sensitivity<br>Analytical Specificity<br>Reproducibility | Analytical<br>Performance | Same | |
| Shortest dated component<br>24 months<br>24 months<br>12 months<br>36 months<br>60 months | Shelf Life<br>Kit<br>• Probe<br>• Hybridization<br>buffer<br>• DAPI II<br>• NP-40<br>• 20X SSC | Same | Probe dating<br>12 months |
# Device Description
The Vysis D7S486/CEP 7 FISH Probe Kit is for specimen characterization and detects the LSI D7S486 (7q31) probe target on chromosome 7q31 and CEP 7 probe target chromosome 7p11.1-q11.1 in bone marrow and peripheral blood specimens.
DNA Probe Description
Vysis LSI D7S486 SpectrumOrange/ CEP 7 SpectrumGreen Probes:
{4}------------------------------------------------
The SpectrumOrange labeled LSI D7S486 probe is approximately 308 kb in length (chr7:115983468-115675366; February 2009 Assembly UCSC Human Genome Browser).
The SpectrumGreen labeled CEP 7 probe targets the D7Z1 alpha satellite sequence at the centromere of chromosome 7.
The Vysis D7S486/CEP 7 FISH Probe Kit (List No. 04N78-020) consists of a mixture of two DNA FISH probes and four general reagents sufficient to process 20 assays.
- . Vysis LSI D7S486 SpectrumOrange/ CEP 7 SpectrumGreen Probes
- . Vysis LSI/WCP Hybridization Buffer
- . DAPI II Counterstain
- NP-40 .
- . 20X SSC
# Summary and Explanation of the Test
Deletion of chromosome 7q and loss of a complete chromosome 7 (monosomy 7) are recurring abnormalities in several hematologic malignancies. Vance et al2 demonstrated the Vysis LSI D7S486 SpectrumOrange/CEP 7 SpectrumGreen Probes detected deletion of the LSI D7S486 probe target on chromosome 7q or loss of chromosome 7 (monosomy 7) in a series of bone marrow and blood specimens from untreated acute myeloid leukemia (AML) patients. In the study, 179 bone marrow specimens and 47 peripheral blood specimens were tested. Among the 179 bone marrow specimens tested, 4 bone marrow specimens were atypical for the 1R2G FISH pattern associated with a deletion of the locus specific identifier (LSI) D7S486 probe target on chromosome 7q with a range of 58 to 93% atypical nuclei. Among the 179 bone marrow specimens tested. 5 bone marrow specimens were atypical for the 1R1G FISH pattern associated with monosomy 7 with a range of 70 to 96% atypical nuclei. Among the 47 peripheral blood specimens tested, 1 peripheral blood specimen contained the 1R1G atypical FISH pattern in 11% of nuclei . An additional 3 peripheral blood specimens were atypical for the 1R2G FISH pattern with a range of 37 - 96% atypical nuclei. Cherry et als reported the Vysis LSI D7S486
{5}------------------------------------------------
SpectrumOrange/CEP 7 SpectrumGreen Probes detected 7g- in 3 myelodysplastic syndrome (MDS) patients with a range of 22.5 to 44% atypical bone marrow nuclej and also detected monosomy 7 in 2 patients with 23 and 87.5% atypical nuclei. Tefferi et al1 described 2 myelofibrosis with myeloid metaplasia (MMM) patients who had 9% and 16% bone marrow nuclei with 7q- (1R2G) atypical signal patterns. A peripheral blood specimen from the MMM patient with 16% atypical bone marrow nuclei had 27% of nuclei with the IR2G atypical signal pattern. The peripheral blood specimen from the patient with 9% 1R2G bone marrow was determined to be typical. However, the author stated that the percent of atypical nuclei may have been below detection limits in this peripheral blood specimen.
The Vysis D7S486/CEP 7 FISH probe kit uses FISH DNA probe technology to detect the probe target for LSI D7S496 (7g31) on chromosome 7q and the probe target for CEP 7 (7p11.1 – q11.1) on chromosome 7.
### Technological Description of the Device
FISH is a technique that allows visualization of specific nucleic acid sequences within a cellular preparation. Specifically, FISH involves precise annealing of a single-stranded, fluorophore-labeled DNA probe to a complementary target sequence. Hybridization of the probe with the cellular DNA site is visible by direct detection using fluorescence microscopy. Interpretation of FISH results should be made utilizing appropriate controls and analytical techniques as well as taking into consideration other clinical and diagnostic test data.5
Bone marrow and peripheral blood cell specimens attached to microscope slides using standard cytogenetic procedures are used for this assay. The resulting specimen DNA is denatured to single-stranded form and subsequently allowed to hybridize with the probes of the Vysis D7S486/CEP 7 FISH Probe Kit. Following hybridization, the unbound probe is removed by a series of washes, and the nuclei are counterstained with DAPI II, a DNAspecific stain that fluoresces blue. Hybridization of the Vysis LSI D7S486 SpectrumOrange/CEP 7 SpectrumGreen Probes is viewed using a fluorescence
{6}------------------------------------------------
microscope equipped with appropriate excitation and emission filters, allowing visualization of the orange and green fluorescent signals.
In a cell with typical copy numbers of the Vysis LSI D7S486 SpectrumOrange/CEP 7 SpectrumGreen probe targets, two orange (2R) signals (D7S486) and two green (2G) signals (CEP 7) will be expected.
A 1 orange, 1 green (1R1G) pattern is expected in a cell having only one copy of chromosome 7. A 1 orange, 2 green (1R2G) pattern is expected in a cell with loss of 7q.
# Summary of Nonclinical Studies
#### Analytical Specificity
Analytical specificity is defined as the percentage of signals that hybridize to the correct locus and no other location.56 The analytical specificity of the Vysis LSI D7S486 SpectrumOrange//CEP 7 SpectrumGreen Probes for their respective chromosome target loci was established using metaphase chromosomes prepared from peripheral blood cultures of 4 male and 1 female karyotypically normal specimen slides. The hybridization location of each FISH signal on chromosomes of 100 consecutive metaphase nuclei was evaluated by 1 technologist for a total of 200 target loci per probe.
For each probe, the number of metaphase chromosome FISH signals hybridized to the correct locus and the number of metaphase chromosome FISH signals hybridized to the incorrect locus were enumerated.
For each probe, the specificity was calculated as the number of metaphase chromosome FISH signals hybridized to the correct locus divided by the total number of metaphase chromosome FISH signals hybridized and multiplied by 100 to give a percentage.
The analytical specificity of the Vysis LSI D7S486/CEP 7 FISH Probe Kit was 100%, as shown in Table 1.
{7}------------------------------------------------
| Table 1. Analytical Specificity | | | | | | |
|---------------------------------|----------------------------|----------------------------------------------|--------------------------------|--------------------|-----------------------------------|--|
| | | No. of Metaphase Chromosome<br>Signals | | | | |
| Probe | Correct<br>Target<br>Locus | Hybridized to the<br>Correct Target<br>Locus | Total<br>Hybridized<br>Signals | Specificity<br>(%) | 95%<br>Confidence<br>Interval (%) | |
| LSI D7S486 | 7q31 | 200 | 200 | 100 | (98,100) | |
| CEP 7 | 7p11.1-q11.1 | 200 | 200 | 100 | (98,100) | |
# Analytical Sensitivity
Analytical sensitivity is defined as the percentage of scoreable interphase nuclei with the expected typical signal pattern. The expected typical interphase signal pattern for the probes in the Vysis D7S486/CEP 7 FISH Probe Kit is 2 orange (2R) and 2 green (2G) signals per nucleus.
The analytical sensitivity of the Vysis LSI D7S486 SpectrumOrange//CEP 7 SpectrumGreen Probes was established using interphase nuclei prepared from 25 bone marrow and 25 peripheral blood specimens that were either karyotypically normal individuals or patients lacking monosomy 7 and loss of 7q. The orange and green signal patterns of nuclei for each of the 25 specimens were evaluated by 2 technologists. Each technologist evaluated 100 nuclei per specimen for a total of 200 nuclei per specimen and 5000 scoreable nuclei for each of the specimen types.
The analytical sensitivity was calculated as the percentage of scoreable interphase nuclei with the expected 2R2G signal pattern.
The Vysis D7S486/CEP 7 FISH Probe Kit has an analytical sensitivity of 98.1% for bone marrow and 98.5% for peripheral blood as shown in Table 2.
{8}------------------------------------------------
| Table 2. Analytical Sensitivity | | | | | | |
|---------------------------------|-----------|-----------------------------------------|---------------------|------------------------|-------------|--|
| | | No. of Interphase<br>Chromosome Signals | | Analytical Sensitivity | | |
| Probe | Specimena | With<br>Expected<br>Signal<br>Pattern | Scoreable<br>Nuclei | Point<br>Estimate | 95% CIb | |
| D7S486 | BMb | 4903 | 5000 | 98.1 | (97.6,98.4) | |
| D7S486 | PBb | 4923 | 5000 | 98.5 | (98.1,98.8) | |
4 25 Karyotypically normal specimens for both specimen types.
b BM: Bone Marrow; PB: Peripheral Blood; CI: Confidence interval
# Verification of Upper Reference Limit
The upper reference limit is defined as the maximum quantity of scoreable interphase nuclei with a specific atypical signal pattern at which a specimen is considered karyotypically normal for that signal pattern. The upper reference limit is expressed in terms of a percentage or the actual number of a specific atypical nuclear FISH signal pattern per the standard number of nuclei tested.
The upper reference limit for monosomy 7 is 4.5% or 9 1R1G patterns per 200 scoreable interphase nuclei, and the upper reference limit for loss of 7q is 6.5% or 13 1R2G patterns per 200 scoreable interphase nuclei. Specimens exceeding 9 1R1G patterns and/or 13 l R2G patterns per 200 scoreable nuclei are considered atypical for monosomy 7 and/or loss of q arm on chromosome 7 with the Vysis D7S486/CEP 7 FISH probe target.
The Vysis D7S486/CEP 7 assay was performed on interphase nuclei from 25 bone marrow and 25 peripheral blood specimens from either karyotypically normal individuals or patients lacking monosomy 7 and loss of 7q. The signal patterns of 200 nuclei for each specimen type were evaluated by each of 2 technologists scoring 100 nuclei per specimen.
Among the 25 karyotypically normal specimens for both peripheral blood and bone marrow, none produced 1R1G and 1R2G signals above the 4.5% and 6.5% upper reference limits.
{9}------------------------------------------------
#### Reproducibility
Two replicates of the assay were run on 2 high-positive, 2 low-positive, and 2 negative panel members at 3 sites on 5 non-consecutive days for each specimen type. The positive panel members for the site-to-site study were obtained by mixing positive cells with normal cells, for each of the specimen types, to obtain the desired levels of positivity. Results shown in Tables 3 through 6 show the overall agreement with the negative/positive status of the test panel members.
All 3 sites demonstrated 100% agreement with the known status of the negative and high positive panel members for both specimen types and both signal patterns. For the bone marrow specimens, the low positives demonstrated 88% (del(7q)) and 97% (monosomy 7) agreement. For the peripheral blood, the low positives demonstrated 95% (del(7q)) and 93% (monosomy 7) agreement.
| Table 3. Overall Agreement, Site-to-Site - For Specimen type BM and signal Del | | | |
|--------------------------------------------------------------------------------|--|--|--|
| (7q) | | | |
and the control of the control of the country of the country of the country of
| | Number | | | |
|---------------|--------|-----------|-------|-------------------|
| Category | Agreea | Disagreeb | Total | Percent Agreement |
| Negative | 60 | 0 | 60 | 100% |
| Low Positive | 53 | 7 | 60 | 88% |
| High Positive | 60 | 0 | 60 | 100% |
ª Agree is number of concordant slides.
b Disagree is number of discordant slides.
# Table 4. Overall Agreement, Site-to-Site - For Specimen type BM and signal Monosomy 7 ·
| | Number | | | |
|---------------|--------|-----------|-------|-------------------|
| Category | Agreea | Disagreeb | Total | Percent Agreement |
| Negative | 60 | 0 | 60 | 100% |
| Low Positive | 58 | 2 | 60 | 97% |
| High Positive | 60 | 0 | 60 | 100% |
ª Agree is number of concordant slides.
b Disagree is number of discordant slides.
{10}------------------------------------------------
# Table 5. Overall Agreement, Site-to-Site – For Specimen type PB and signal Del (7q)
| | Number | | | |
|---------------|--------|-----------|-------|-------------------|
| Category | Agreea | Disagreeb | Total | Percent Agreement |
| Negative | 60 | 0 | 60 | 100% |
| Low Positive | 57 | 3 | 60 | 95% |
| High Positive | 60 | 0 | 60 | 100% |
4 Agree is number of concordant slides.
b Disagree is number of discordant slides.
# Table 6. Overall Agreement, Site-to-Site - For Specimen type PB and signal Monosomy 7
| | Number | | | |
|---------------|--------|-----------|-------|-------------------|
| Category | Agreea | Disagreeb | Total | Percent Agreement |
| Negative | 60 | 0 | 60 | 100% |
| Low Positive | 56 | 4 | 60 | 93% |
| High Positive | 60 | 0 | 60 | 100% |
ª Agree is number of concordant slides.
b Disagree is number of discordant slides.
The mean and the standard deviation (SD) of the percentage of cells with the 1R2G and
1R1G signal patterns were calculated.
{11}------------------------------------------------
The analysis of variance components for the site-to-site study is shown in Tables 7 through 10.
| and signal Del (7q) | | | Within-<br>Day<br>(Comp.) | Between-<br>Day<br>(Comp.) | Between-<br>Site<br>(Comp.) | Total |
|---------------------|----|------|---------------------------|----------------------------|-----------------------------|-------|
| Sample | N | Mean | SD | SD | SD | SD |
| Negative 1 | 30 | 0.5 | 0.27 | 0.32 | 0.06 | 0.43 |
| Negative 2 | 30 | 0.4 | 0.38 | 0.14 | 0.24 | 0.47 |
| Low Positive 1 | 30 | 10.5 | 2.11 | 1.41 | 1.14 | 2.78 |
| Low Positive 2 | 30 | 10.0 | 3.19 | 0.00 | 2.33 | 3.95 |
| High Positive 1 | 30 | 43.9 | 6.76 | 0.00 | 7.62 | 10.19 |
| High Positive 2 | 30 | 42.0 | 4.61 | 0.00 | 5.44 | 7.13 |
Table 7. Site-to-Site Analysis of Variance Components – For Specimen type BM and signal Del (7g)
# Table 8. Site-to-Site Analysis of Variance Components – For Specimen type BM and signal Monosomy 7
| | | | Within-<br>Day<br>(Comp.) | Between-<br>Day<br>(Comp.) | Between-<br>Site<br>(Comp.) | Total |
|-----------------|----|------|---------------------------|----------------------------|-----------------------------|-------|
| Sample | N | Mean | SD | SD | SD | SD |
| Negative 1 | 30 | 0.5 | 0.67 | 0.14 | 0.49 | 0.84 |
| Negative 2 | 30 | 0.3 | 0.37 | 0.00 | 0.25 | 0.44 |
| Low Positive 1 | 30 | 8.9 | 2.61 | 0.00 | 1.19 | 2.87 |
| Low Positive 2 | 30 | 9.3 | 2.43 | 0.00 | 0.00 | 2.43 |
| High Positive 1 | 30 | 48.3 | 6.20 | 1.30 | 8.06 | 10.25 |
| High Positive 2 | 30 | 43.9 | 3.78 | 3.97 | 4.19 | 6.90 |
{12}------------------------------------------------
| Sample | N | Mean | Within-Day (Comp.) SD | Between-Day (Comp.) SD | Between-Site (Comp.) SD | Total SD |
|-----------------|----|------|-----------------------|------------------------|-------------------------|----------|
| Negative 1 | 30 | 0.4 | 0.53 | 0.27 | 0.38 | 0.71 |
| Negative 2 | 30 | 0.4 | 0.42 | 0.25 | 0.00 | 0.49 |
| Low Positive 1 | 30 | 10.4 | 2.46 | 0.00 | 2.03 | 3.19 |
| Low Positive 2 | 30 | 12.5 | 2.57 | 0.00 | 0.38 | 2.60 |
| High Positive 1 | 30 | 42.1 | 3.26 | 2.05 | 5.82 | 6.98 |
| High Positive 2 | 30 | 52.8 | 4.07 | 1.82 | 2.15 | 4.95 |
Table 9. Site-to-Site Analysis of Variance Components - For Specimen type PB and signal Del (7q)
# Table 10. Site-to-Site Analysis of Variance Components – For Specimen type PB and signal Monosomy 7
| | | | Within-<br>Day<br>(Comp.) | Between-<br>Day<br>(Comp.) | Between-<br>Site<br>(Comp.) | Total |
|-----------------|----|------|---------------------------|----------------------------|-----------------------------|-------|
| Sample | N | Mean | SD | SD | . SD | SD |
| Negative 1 | 30 | 0.3 | 0.65 | 0.00 | 0.32 | 0.72 |
| Negative 2 | 30 | 0.4 | 0.47 | 0.20 | 0.25 | 0.57 |
| Low Positive 1 | 30 | 9.1 | 2.61 | 0.00 | 0.55 | 2.67 |
| Low Positive 2 | 30 | 6.9 | 2.35 | 0.52 | 0.27 | 2.42 |
| High Positive 1 | 30 | 44.8 | 3.77 | 0.00 | 5.35 | 6.55 |
| High Positive 2 | 30 | 38.8 | 3.90 | 3.17 | 1.63 | 5.29 |
Using the same panel members from the site-to-site study, 4 replicates of the assay were run on 2 high-positive, 2 low-positive, and 2 negative panel members using 3 different lots of probe at a single site for each specimen type. The overall agreement with the known negative/positive status of the test panel members is shown in Tables 11 through 14.
{13}------------------------------------------------
All 3 lots demonstrated 100% agreement with the know status of the negative and high positive panel members for both specimen types and both signal patterns. For the bone marrow specimens, the low positives demonstrated 88% (del(7q)) and 92% (monosomy 7) agreement. For the peripheral blood, the low positives demonstrated 100% (del(7q)) and 96% (monosomy 7) agreement.
| | Table 11. Overall Agreement, Lot-to-Lot - For Specimen type BM and signal Del |
|------|-------------------------------------------------------------------------------|
| (7q) | |
| Category | Number | | | Percent |
|---------------|--------|-----------|-------|---------------|
| | Agreea | Disagreeb | Total | Agreement (%) |
| Negative | 24 | 0 | 24 | 100% |
| Low Positive | 21 | 3 | 24 | 88% |
| High Positive | 24 | 0 | 24 | 100% |
ª Agree is number of concordant slides.
b Disagree is number of discordant slides
| Table 12. Overall Agreement, Lot-to-Lot – For Specimen type BM and signal |
|---------------------------------------------------------------------------|
| Monosomy 7 |
| Category | Number | | | Percent |
|---------------|--------|-----------|-------|---------------|
| | Agreea | Disagreeb | Total | Agreement (%) |
| Negative | 24 | 0 | 24 | 100% |
| Low Positive | 22 | 2 | 24 | 92% |
| High Positive | 24 | 0 | 24 | 100% |
4 Agree is number of concordant slides.
b Disagree is number of discordant slides
# Table 13. Overall Agreement, Lot-to-Lot - For Specimen type PB and signal Del (7q)
| Category | Number | | | Percent |
|---------------|--------|-----------|-------|---------------|
| | Agreea | Disagreeb | Total | Agreement (%) |
| Negative | 24 | 0 | 24 | 100% |
| Low Positive | 24 | 0 | 24 | 100% |
| High Positive | 24 | 0 | 24 | 100% |
* Agree is number of concordant slides.
b Disagree is number of discordant slides
{14}------------------------------------------------
# Table 14. Overall Agreement, Lot-to-Lot - For Specimen Type PB and signal Monosomy 7
| | | Number | | | |
|---------------|--------|-----------|-------|--------------------------|--|
| Category | Agreea | Disagreeb | Total | Percent<br>Agreement (%) | |
| Negative | 24 | 0 | 24 | 100% | |
| Low Positive | 23 | 1 | 24 | 96% | |
| High Positive | 24 | 0 | 24 | 100% | |
ª Agree is number of concordant slides.
b Disagree is number of discordant slides
#### , --
The analysis of variance components for the lot-to-lot study is shown in Tables 15 through 18.
| Table 15. Lot-to-Lot Analysis of Variance Components - For Specimen type BM<br>and signal Del (7q) | | | | | |
|----------------------------------------------------------------------------------------------------|----|------|----------------------|-----------------------|-------|
| | | | Within-Lot<br>(Comp) | Between-Lot<br>(Comp) | Total |
| Sample | N | Mean | SD | SD | SD |
| Negative 1 | 12 | 0.1 | 0.22 | 0.06 | 0.23 |
| Negative 2 | 12 | 0.2 | 0.28 | 0.00 | 0.28 |
| Low Positive 1 | 12 | 8.7 | 2.31 | 1.63 | 2.82 |
| Low Positive 2 | 12 | 11.3 | 1.58 | 0.00 | 1.58 |
| High Positive 1 | 12 | 47.8 | 5.65 | 2.49 | 6.17 |
| High Positive 2 | 12 | 48.9 | 3.53 | 2.66 | 4.42 |
{15}------------------------------------------------
| | | | Within-Lot<br>(Comp) | Between-Lot<br>(Comp) | Total |
|-----------------|----|------|----------------------|-----------------------|-------|
| Sample | N | Mean | SD | SD | SD |
| Negative 1 | 12 | 0.1 | 0.22 | 0.06 | 0.23 |
| Negative 2 | 12 | 0.0 | 0.14 | 0.00 | 0.14 |
| Low Positive 1 | 12 | 6.8 | 2.58 | 0.88 | 2.73 |
| Low Positive 2 | 12 | 8.1 | 1.16 | 0.00 | 1.16 |
| High Positive 1 | 12 | 43.9 | 6.55 | 4.86 | 8.16 |
| High Positive 2 | 12 | 38.5 | 4.03 | 3.71 | 5.48 |
Table 16. Lot-to-Lot Analysis of Variance Components – For Specimen type BM and signal Monosomy 7
Table 17. Lot-to-Lot Analysis of Variance Components - For Specimen type PB and signal Del (7q)
| | | | Within-Lot<br>(Comp) | Between-Lot<br>(Comp) | Total |
|-----------------|----|------|----------------------|-----------------------|-------|
| Sample | N | Mean | SD | SD | SD |
| Negative 1 | 12 | 0.1 | 0.22 | 0.06 | 0.23 |
| Negative 2 | 12 | 0.1 | 0.20 | 0.00 | 0.20 |
| Low Positive 1 | 12 | 11.6 | 2.14 | 0.98 | 2.35 |
| Low Positive 2 | 12 | 11.0 | 1.99 | 0.00 | 1.99 |
| High Positive 1 | 12 | 46.5 | 2.47 | 1.52 | 2.89 |
| High Positive 2 | 12 | 58.1 | 3.38 | 0.00 | 3.38 |
: : : .
{16}------------------------------------------------
| | | | Within-Lot<br>(Comp) | Between-Lot<br>(Comp) | Total |
|-----------------|----|------|----------------------|-----------------------|-------|
| Sample | N | Mean | SD | SD | SD |
| Negative 1 | 12 | 0.0 | 0.00 | 0.00 | 0.00 |
| Negative 2 | 12 | 0.0 | 0.14 | 0.00 | 0.14 |
| Low Positive 1 | 12 | 9.1 | 2.05 | 1.76 | 2.70 |
| Low Positive 2 | 12 | 5.7 | 1.22 | 0.00 | 1.22 |
| High Positive 1 | 12 | 37.1 | 2.87 | 0.90 | 3.01 |
| High Positive 2 | 12 | 32.0 | 3.42 | 1.69 | 3.81 |
Table 18. Lot-to-Lot Analysis of Variance Components – For Specimen type PB and signal Monosomy 7
·
.
.
{17}------------------------------------------------
# Summary of Results from Cited Published Literature
Cited published literature may discuss device uses that have not been approved or cleared by FDA.
| Literature<br>Reference | Population<br>Studied | Number<br>and Type of<br>Specimens | Device Used | Observed D7S486/CEP 7 Results |
|-------------------------|-----------------------|-----------------------------------------------------------|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Vance et al1 | AMLa | 179 bone<br>marrow and<br>47 blood<br>specimensb | Vysis LSI<br>D7S486/CEP 7<br>probes | Overall 7q- (1R2G signal pattern)<br>was detected in 4/179 bone marrow<br>specimens and 3/47 peripheral blood<br>specimens.<br>Overall monosomy 7 (IR1G signal<br>pattern) was detected in 5/179 bone<br>marrow specimens and 1/47<br>peripheral blood specimens. |
| Cherry et al2 | MDSa | 48 bone<br>marrow<br>specimens | Vysis LSI<br>D7S486/CEP 7<br>probes | Overall 7q- (1R2G signal pattern)<br>was detected in 3/48 bone marrow<br>specimens.<br>Overall monosomy 7 (IRIG signal<br>pattern) was detected in 2/48 bone<br>marrow specimens. |
| Tefferi et al3 | MMMa | 42 bone<br>marrow and<br>peripheral<br>blood<br>specimens | Vysis LSI<br>D7S486/CEP 7<br>probes | Overall 7q- (1R2G signal pattern)<br>was detected in 2/42 bone marrow<br>specimens and 1/42 peripheral blood<br>specimens. |
Data from supporting literature
ª AML: acute myeloid leukemia; MDS: myelodysplastic syndrome; MMM: myelofibrosis with myeloid metaplasia
b Based on unpublished data.
{18}------------------------------------------------
# BIBLIOGRAPHY
- 1. Genome Bioinformatics Group of UC Santa Cruz. The UCSC Genome Browser. © The Regents of the University of California. Available at: http://genome.ucsc.edu/cgibin/hgGateway?hgsid=185806115&clade=mammal&org=Human&db=hg18. Accessed [April 30, 2013].
- 2. Vance GH. Kim H. Hicks GA, et al. Utility of interphase FISH to Stratify Patients into Cytogenetic Risk Categories at Diagnosis of AML in an Eastern Cooperative Oncology Group (ECOG) Clinical Ttrial (E1900). Leuk Res. 2007;31:605-09.
- 3. Cherry A, Brockman S, Paternoster, S et al. Comparison of interphase FISH and metaphse cytogenetics to study myelodysplastic syndrome: an Eastern Cooperative Oncology Group (ECOG) study. Leuk Res. 2003;27:1085-1090.
- 4. Tefferi A, Meyer R. Wyatt W, et al. Comparison of peripheral blood interphase cytogenetics with bone marrow karyotype analysis in myelofibrosis with myeloid metaplasia. Br J Haematol. 2001;115:316-319.
- 5. Wiktor AE, Van Dyke DL, Stupca PJ et al. Preclinical validation of fluorescence in situ hybridization assays for clinical practice. Genet Med. 2006;8:16-23.
- 6. AMERICAN COLLEGE OF MEDICAL GENETICS (ACMG). Standards and Guidelines for Clinical Genetics Laboratories. 2008 Edition.
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "U.S. Department of Health & Human Services" around the perimeter of the circle. Inside the circle is an abstract symbol that resembles a stylized human figure.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W() bookings Silver Spring, MD 20993-0002
September 13. 2013
ABBOTT MOLECULAR, INC. NANCY W. BENGTSON MANAGER. GLOBAL REGULATORY AFFAIRS 1300 E. TOUHY AVENUE DES PLAINES. IL 60018
Rc: K131508
Trade/Device Name: Vysis D7S486/CEP 7 FISH Probe Kit Regulation Number: 21 CFR 864.1870 Regulation Name: Early growth response 1 (EGFRI) gene fluorescence in-situ hybridization (FISH) test system for specimen characterization Regulatory Class: II Product Code: PFG Dated: August 13, 2013 Received: August 14, 2013
Dear Dr. Bengtson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRI does not evaluate information related to contract fiability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{20}------------------------------------------------
Page 2-Dr. Bengtson
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.hum.
Sincerely yours,
# Reena Philip -S
for
Maria M. Chan, Ph.D.
Director, Division of Immunology and Hematology Devices
Office of In Vitro Diagnostics and Radiological Health
Center for Devices and Radiological Health
Enclosure
{21}------------------------------------------------
# Indications for Use
510(k) Number (if known): K131508
Device Name: Vysis D7S486/CEP 7 FISH Probe Kit
Indications For Use:
The Vysis D7S486/CEP 7 FISH Probe Kit is a device intended for specimen characterization, and detects the LSI D7S486 probe target on chromosome 7q31 and the CEP 7 probe target on chromosome 7p11.1-q11.1 in bone marrow and peripheral blood specimens from patients with acute myeloid leukemia or myelodysplastic syndrome. The assay results are intended to be interpreted by a qualified pathologist or cytogeneticist. This device is not intended for high-risk uses such as selecting therapy, predicting therapeutic response or disease screening. The use of this device for diagnosis, prognosis, monitoring, or risk assessment has not been established.
Prescription Use × (Part 21 CFR 801 Subpart D)
:
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR) Donna M. Roscoe:-S 2013.09.13 13:23:06 -04'00'
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________