Z CARTRIDGE IOL DELIVERY SYSTEM

K123888 · Aaren Scientific, Inc. · MSS · Apr 15, 2013 · Ophthalmic

Device Facts

Record IDK123888
Device NameZ CARTRIDGE IOL DELIVERY SYSTEM
ApplicantAaren Scientific, Inc.
Product CodeMSS · Ophthalmic
Decision DateApr 15, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4300
Device ClassClass 1

Intended Use

The Aero Z Cartridge IOL Delivery System is indicated to fold and insert Aaren Scientific IOLs that have the use of this inserter in their approved labeling.

Device Story

Aero Z Cartridge IOL Delivery System is a two-part surgical instrument for ophthalmic use. It consists of a reusable titanium alloy handpiece (push-style or push-and-turn screw-style) and a single-use, sterile polypropylene cartridge containing a glyceryl monostearate (GMS) lubricity agent. The device is operated by a surgeon in a clinical setting to fold and inject intraocular lenses (IOLs) into the posterior chamber of the eye. The surgeon loads the IOL into the cartridge, which is then inserted into the handpiece; the plunger mechanism advances the lens through the nozzle into the eye. This system facilitates minimally invasive lens delivery, reducing surgical trauma and improving patient outcomes during cataract or refractive lens exchange procedures.

Clinical Evidence

Bench testing only. Testing conducted per ISO 10993-1:2009 confirmed the device is non-cytotoxic, non-sensitizing, and non-irritating. Testing per ISO 11979-3 demonstrated that the delivery system does not significantly impact the optical or mechanical performance, dimensions, or cosmetic appearance of Aaren Scientific IOL Models EC-3 and EC-3 PAL.

Technological Characteristics

Reusable titanium alloy handpiece; single-use polypropylene cartridge with glyceryl monostearate (GMS) lubricity agent. Mechanism: syringe-like push or push-and-turn screw-style plunger. Sterilization: handpiece via flash autoclave/steam; cartridge via Ethylene Oxide (EtO).

Indications for Use

Indicated for folding and insertion of Aaren Scientific intraocular lenses (IOLs) into the posterior chamber of the eye, specifically models EC-3 and EC-3 PAL, for patients requiring IOL implantation.

Regulatory Classification

Identification

An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ S10(K) Summary # APR 1 5 2013 ### Submitter's Name, Address, Telephone Number, Contact Person, and Date 5.1 Summary Prepared | A. | Company Name: | Aaren Scientific Inc. | |----|------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | B. | Company Address: | 4290 E. Brickell St., Bldg. A.<br>Ontario, CA 91761 | | C. | Company Phone: | 1 (909) 937-1033 | | D. | Contact Person: | Robert K. Sheehan<br>Vice President, Regulatory Affairs and<br>Quality Systems<br>Aaren Scientific Inc.<br>4290 East Brickell St., Bldg. A | Ontario, CA 91761 USA Phone +1 (909) 937-1033 ext. 108 Email: rksheehan@aareninc.com #### 5.2 Date Summary Prepared January 5, 2013 Name of Device, Including Trade Name and Classification Name 5.3 A. Device Trade Name: Aero Z Cartridge IOL Delivery System B. Common Name: Small Incision Injector ். Classification Name(s): Intraocular lens guide D. Classification Regulation: 21 CFR 886.4300 Class I E. Product Code: MSS F. Advisory Panel: Ophthalmic {1}------------------------------------------------ ### 5.4 Predicate Devices The Aero Z Cartridge IOL Delivery System is substantially equivalent to the Monarch III IOL Delivery System (C Cartridge) device cleared under 510(k) K112977 with respect to indications for use, materials and mechanism of action. #### Device Description ર્સ્ડ The Aero Z Cartridge IOL Delivery System is a two-part device comprised of a reusable handpiece and single use disposable cartridge. The single use disposable cartridge is , fabricated from polypropylene with a glyceryl monostearate (GMS) agent incorporated for lubricity. It is intended for single use only and is provided sterile. The reusable handpiece is titanium, is supplied non-sterile and is designed to accommodate the cartridge and lens for delivery of the lens into the eye. The reusable handpiece is supplied in both a push-and-turn and screw type configuration. #### 5.6 Indications for Use The Aero Z Cartridge IOL Delivery System is indicated to fold and insert Aaren Scientific IOLs that have the use of this inserter in their approved labeling. #### 5.7 Comparison of Technological Characteristics The Aero Z Cartridge IOL Delivery System has the same indications for use, materials and mechanism of action as the predicate device (Alcon Research Monarch III IOL Delivery System, C Cartridge) device. {2}------------------------------------------------ | Table 3.1: Comparison of Z Cartridge IOL Delivery System with Predicate Device | | | |--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Aaren Scientific<br>Z Cartridge IOL Delivery<br>System | Alcon Research<br>Monarch® III IOL Delivery<br>System (C Cartridge) | | 510(k) Number | K123888 | K112977 | | Substantial Equivalence Characteristics | | | | Intended Use | The Z Cartridge IOL Delivery<br>System is indicated to fold and<br>insert Aaren Scientific Models<br>EC-3 and EC-3 PAL intraocular<br>lenses into the posterior<br>chamber of the eye | Folding and injection of<br>AcrySof® intraocular lenses<br>into the posterior chamber of<br>the eye | | Anatomical Site of Use | Posterior chamber of the eye | Posterior chamber of the eye | | Components | Reusable handpiece and single-<br>use, sterile Cartridge | Reusable handpiece and single-<br>use, sterile Cartridge | | Handpiece | | | | Material | Titanium alloy | Titanium alloy | | Lens Injecting Mechanism | Syringe-like (Push style) and<br>Push-and-turn rotational<br>(Screw style) | Push and turn | | Configuration | Barrel and plunger assembly,<br>the barrel has a chamber to<br>accept the Cartridge and the<br>plunger advances the lens for<br>injection | Barrel and plunger assembly,<br>the barrel has a chamber to<br>accept the Cartridge and the<br>plunger advances the lens for<br>injection | | Sterilization | Flash autoclave or steam<br>sterilization by user | Flash autoclave or steam<br>sterilization by user | | Cartridge | | | | Material | Polypropylene with a glycerol<br>monostearate (GMS) lubricity<br>agent | Polypropylene with a<br>polyvinylpyrrolidone (PVP)<br>coating on the inner lumen | | Loading Chamber | Lumen with wings for folding | Internal Cartridge<br>geometry | | Configuration | Lens loading and folding area<br>connected to a lens injecting<br>nozzle | Lens loading and folding area<br>connected to a lens injecting<br>nozzle | | Sterilization | Ethylene oxide (EtO) | Ethylene oxide (EtO) | . Table 5.1: Comparison of Z Cartridge IOL Delivery System with Predicate Device . {3}------------------------------------------------ #### Brief Summary of Nonclinical Tests and Results 5.8 In accordance with ISO 10993-1:2009, cytotoxicity, sensitization and irritation testing were performed on the IOL Delivery System. The device was found to be non-cytotoxic. non-sensitizing, and non-irritating. Testing conducted in accordance with ISO 11979-3 demonstrated that the Aero Z Cartridge IOL Delivery System can deliver IOL Models EC-3 and EC-3 PAL without significantly impacting the optical or mechanical performance, dimensions or the cosmetic appearance of the lens. Based on the indications for use and the technological characteristics, it can be concluded that the Aero Z Cartridge IOL Delivery System is substantially equivalent to the predicate device. #### Substantial Equivalence 5.8 On the basis of the same indications for use, materials and mechanism of action, the Aero Z Cartridge IOL Delivery System is substantially equivalent to the predicate device (Monarch III IOL Delivery System, C Cartridge). {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Public Health Service April 15, 2013 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Aaren Scientific, Inc. % Mr. Robert Sheehan . Vice President of Regulatory Affairs and Quality Systems 4290 East Brickell Street, Bldg A Ontario, CA 91761-1569 Re: K123888 Trade/Device Name: AERO™ Z Cartridge IOL Delivery System Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular lens guide Regulatory Class: Class I (reserved) Product Code: MSS Dated: March 7, 2013 Received: March 11, 2013 Dear Mr. Sheehan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {5}------------------------------------------------ Page 2 - Mr. Robert Sheehan CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Kesia YAlexander -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure - {6}------------------------------------------------ Aero Z Cartridge IOL Delivery System Device Name: Indications for Use: The Aero Z Cartridge IOL Delivery System is indicated to fold and insert Aaren Scientific IOLs that have the use of this inserter in their approved labeling. Prescription Use X AND/OR (21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) Tieuvi Hī 2013.04.05 1:5 5:07 -04'00' (Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices 510(k) Number: K123888
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%