LIQUICHEK URINE CHEMISTRY CONTROL
K121794 · Bio-Rad, Diagnostics Grp. · JJY · Jul 11, 2012 · Clinical Chemistry
Device Facts
| Record ID | K121794 |
| Device Name | LIQUICHEK URINE CHEMISTRY CONTROL |
| Applicant | Bio-Rad, Diagnostics Grp. |
| Product Code | JJY · Clinical Chemistry |
| Decision Date | Jul 11, 2012 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1660 |
| Device Class | Class 1 |
Intended Use
Liquichek Urine Chemistry Control is intended for use as an assayed quality control urine to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.
Device Story
Liquichek Urine Chemistry Control is an assayed quality control material prepared from human urine with added human/animal constituents, chemicals, preservatives, and stabilizers. It is provided in liquid form for laboratory use. The device is used by laboratory personnel to monitor the precision of testing procedures for specific analytes. By comparing observed results against the provided mean values and ±3SD ranges, clinicians assess the performance and reliability of their testing instrumentation and reagents. This monitoring helps ensure the accuracy of patient diagnostic testing. The device is stored at 2°C to 8°C.
Clinical Evidence
No clinical data. Bench testing only. Stability studies confirmed open vial stability of 30 days at 2-8°C and shelf life of 24 months at 2-8°C. Mean values and ±3SD ranges were derived from replicate analyses.
Technological Characteristics
Human urine matrix with added human/animal constituents, chemicals, preservatives, and stabilizers. Liquid form. Storage at 2°C to 8°C. Analyte list includes Amylase, Calcium, Chloride, Creatinine, Glucose, Magnesium, Microalbumin, Phosphorus, Potassium, Total Protein, Sodium, Urea Nitrogen, and Uric Acid.
Indications for Use
Indicated for use as an assayed quality control urine to monitor the precision of laboratory testing procedures for specific analytes (Amylase, Calcium, Chloride, Creatinine, Glucose, Magnesium, Microalbumin, Phosphorus, Potassium, Total Protein, Sodium, Urea Nitrogen, Uric Acid) in clinical laboratory settings.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
- Liquichek Urine Chemistry Control (K020817)
Related Devices
- K971954 — LIQUICHEK URINE CHEMISTRY CONTROL (CAT.NO. 397,398) · Bio-Rad · Jun 11, 1997
- K070848 — LIQUICHEK URINALYSIS CONTROL (BI-LEVEL), CONTROL LEVEL1, LEVEL 2,MINIPAK, 435,436,437,435X · Bio-Rad, Diagnostics Grp. · Apr 25, 2007
- K072835 — LIQUICHEK MICROALBUMIN CONTROL LEVEL 1, LEVEL 2, CONTROL MINIPAK · Bio-Rad, Diagnostics Grp. · Dec 11, 2007
- K082107 — CLINIQA LIQUID QC URINE CHEMISTRY CONTROLS, LEVELS 1 AND 2 · Cliniqa Corporation · Sep 17, 2008
- K990317 — ABBOT LIQUID URINE CHEMISTRY CONTROL, MODEL 1E96-01/1E97-01 · Bio-Rad · Mar 18, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
K121794
Page 1 of 3
Bio-Rad Laboratories Premarket Notification Section 510 (k) Liquichek Urine Chemistry Control
## Summary of Safety and Effectiveness Liquichek Urine Chemistry Control
#### 1.0 Submitter
Bio-Rad Laboratories 9500 Jeronimo Road, Irvine, Callfornia 92618-2017 Telephone: (949) 598-1200 (949) 598-1557 Fax:
### Contact Person
Suzanne Parsons Regulatory Affairs Manager Telephone: (949) 598-1467
### Date of Summary Preparation
June 15th, 2012
#### 2.0 Device Identification
| Product Trade Name: | Liquichek Urine Chemistry Control |
|---------------------|---------------------------------------------|
| Common Name: | Multi-Analyte Controls, All Kinds (Assayed) |
| Classifications: | Class I, reserved |
| Product Code: | JJY |
| Regulation Number: | 21 CFR 862.1660 |
#### Device to Which Substantlal Equivalence is Claimed 3.0
Liquichek Urine Chemistry Control Bio-Rad Laboratories Irvine, California
510 (k) Number: K020817
#### Description of Device 4.0
Liquichek Urine Chemistry Control is prepared from human urine with added constituents of human and animal origin, chemicals, preservatives and stabilizers. The control is provided in liquid form for conventence
#### Value Assignment 5.0
The mean values and the corresponding ±3SD ranges printed in the insert were derived from replicate analyses and are specific for this lot of product. The tests listed were performed by the manufacturer and/or independent laboratories using manufacturer supported reagents and a representative sampling of this lot of product. It is recommended that each laboratory establish its own acceptable ranges and use those provided only as guides. Laboratory established ranges may vary from those listed during the life of this control. Variations over time and between laboratories may be caused by differences in laboratory technique, instrumentation and reagents, or by manufacturer test method modifications.
1 1 2012 JUL
{1}------------------------------------------------
#### 6.0 Intended Use
Liquichek Urine Chemistry Control is intended for use as an assaved quality control urine to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.
#### 7.0 Comparison of the new device with the Predicate Device
Liquichek Urine Chemistry Control claims substantial equivalence to the Liquichek Urine Chemistry Control currently in commercial distribution (K020817). Table 1 (below) contains comparison information of similarities and differences between the new and predicate device to which substantial equivalence is claimed.
| Characteristics | Liquichek Urine Chemistry Control<br>( <i>New Device</i> ) | Liquichek Urine Chemistry Control<br>( <i>Predicate Device, K020817</i> ) |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Similarities | | |
| Intended Use | Liquichek Urine Chemistry Control is<br>intended for use as an assayed quality<br>control urine to monitor the precision of<br>laboratory testing procedures for the<br>analytes listed in this package insert. | Liquichek Urine Chemistry Control is<br>intended for use as an assayed quality<br>control urine to monitor the precision of<br>laboratory testing procedures for the<br>analytes listed in this package insert. |
| Matrix | Human Urine | Human Urine |
| Form | Liquid | Liquid |
| Open Vial Stability | 30 days at 2°C to 8 °C | 30 days at 2°C to 8 °C |
| Storage unopened (Shelf life) | 2 to 8°C until expiration date | 2 to 8°C until expiration date |
| Differences | | |
| Fill Volume | For Level 1 and 2 - 12 x 2.5 mL<br>For Bilevel MiniPak - 2 x 2.5 mL | For Level 1 and 2 - 12 x 10 mL<br>For Bilevel MiniPak - 2 x 10 mL |
| Analytes | Contains:<br>Amylase<br>Calcium<br>Chloride<br>Creatinine<br>Glucose<br>Magnesium<br>Micoalbumin (Albumin)<br>Phosphorus<br>Potassium<br>Protein, Total<br>Sodium<br>Urea Nitrogen (BUN)<br>Uric Acid<br><br>Does not contain<br>Urea<br>Cortisol<br>Osmolality<br>pH<br>Pregnancy<br>Specific Gravity | Contains:<br>Amylase<br>Calcium<br>Chloride<br>Cortisol<br>Creatinine<br>Glucose<br>Magnesium<br>Micoalbumin (Albumin)<br>Osmolality<br>Phosphorus<br>pH<br>Pregnancy<br>Potassium<br>Protein, Total<br>Specific Gravity<br>Sodium<br>Urea<br>Urea Nitrogen (BUN)<br>Uric Acid |
Table 1. Similarities and Differences between new and predicate device.
{2}------------------------------------------------
Bio-Rad Laboratories Premarket Notification Section 510 (k) Liquichek Urine Chemistry Control
#### 8.0 Statement of Supporting Data
Stability studies have been performed for Liquichek Urine Chemistry control to determine the open vial and shelf life claims. Product claims are as follows:
Open Vial:
Shelf Life Stability
30 days at 2 to 8°C. 24 Months at 2°C to 8°C
#### Conclusion 9.0
Liquichek Urine Chemistry control is intended to be used for the same purpose as the predicate device. It has the human urine matrix and performs similarly as the predicate device.
All supporting data is retained on file at Blo-Rad Laboratories.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration
10903 New Hampshire Avenue Silver Spring, MD 20993
Bio-Rad Laboratories c/o Suzanne Parsons 9500 Jeronimo Road Irvine, CA 92618-2017
Re: k121794 JUL 11 2012
Trade Name: Liquichek Urine Chemistry Control Regulation Number: 21 CFR §862.1660 Regulation Name: Quality Control Material (assayed) Regulatory Class: Class I, reserved Product Codes: JJY Dated: June 18, 2012 Received: June 19, 2012
Dear Suzanne Parsons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
## Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 17 you desire specific ad rive to: Device Evaluation and Safety at (301) 796-5450. Also, commor the regulation entitled, "Misbranding by reference to premarket notification" (21 please note the rogulation regarding postmarket surveillance, please contact CDRB13 Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) Office of But romance and garding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical
Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance...
You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely yours,
signature
Countney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety
Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Form
K121794 510(k) Number (if known):
Device Name:
Liquichek Urine Chemistry control
Indications for Use:
Liquichek Urine Chemistry Control is intended for use as an assayed quality control urine to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.
The following analytes are listed in the package insert:
- Amylase
- Calcium
- Chloride
- Creatinine
- Glucose
- Magnesium
- Microalbumin (Albumin)
-
- Phosphorus
- Potassium
- Protein, total
- Sodium
- Urea Nitrogen (BUN)
- Uric Acid
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 14121794 510(K)
Page 1 of 1