MANUKA FILL

K121227 · Links Medical Products, Inc. · FRO · May 23, 2012 · SU

Device Facts

Record IDK121227
Device NameMANUKA FILL
ApplicantLinks Medical Products, Inc.
Product CodeFRO · SU
Decision DateMay 23, 2012
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Intended Use

MANUKA FILL wound dressings are sterile, single-use, wound care dressings for use in moist wound management. MANUKA FILL Wound Dressings may be used Over-The-Counter for: - Minor Abrasions - Lacerations - Minor Cuts - Minor Scalds and Burns Under the supervision of a healthcare professional, MANUKA FILL wound dressings may be used for: - Leg Ulcers - Pressure Ulcers - 1st and 2nd Degree Burns (Superficial and Partial Thickness) - Diabetic Foot Ulcers - Surgical Wounds - Traumatic Wounds

Device Story

MANUKA FILL wound dressing consists of sterile Leptospermum scoparium honey packaged in low density polyethylene (LDPE) tubes. Device intended for moist wound management. Applied topically to wound site by patient (OTC for minor wounds) or healthcare professional (for chronic/surgical wounds). Honey provides moist environment conducive to healing. Single-use, gamma-irradiated device.

Clinical Evidence

Bench testing only. Biocompatibility testing (cytotoxicity, skin irritation, sensitization, wound healing) performed per ISO 10993-1 standards. Additional testing included sterilization validation, shelf-life (accelerated and real-time), and packaging validation. All acceptance criteria met.

Technological Characteristics

Primary material: Leptospermum scoparium honey. Packaging: Low density polyethylene (LDPE) tubes. Sterilization: Gamma irradiation. Single-use form factor. No electronic or software components.

Indications for Use

Indicated for moist wound management in patients with minor abrasions, lacerations, cuts, scalds, and burns (OTC use); or leg ulcers, pressure ulcers, 1st/2nd degree burns, diabetic foot ulcers, surgical wounds, and traumatic wounds (professional supervision).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K121227 page 1 of 3 MAY 2 3 2012 Image /page/0/Picture/2 description: The image shows a logo with the letters "lmp" in a stylized, interconnected design. The letters are bold and outlined, with rounded corners and a retro aesthetic. The background is black, and the letters are white, creating a high-contrast visual. The logo has a slightly distressed or aged appearance, adding to its vintage feel. Links Medical Products Inc # 510(k) SUMMARY ### Submitted by: ું મું Owner's Name: Address: Contact: Telephone: Fax: E-mail: Links Medical Products, Inc. 9247 Research Drive Irvine, CA 92618 Tom Buckley, Chief Executive Officer 949-753-0001 949-753-7412 tbuckley@linksmed.com #### Contact Person: Company Address Contact: Telephone Fax: E-mail: Date Prepared: Trade Name: Common Name: Classification Name: Device Class: Product Code: Predicate Device: Predicate 510(k) #: Device Description: NanoSmart, Inc. 29442 Pointe Royale Laguna Niguel, CA 92677 James Smith, Ph.D. 949-340-7261 949-340-7141 irsmith00@cox.net April 17, 2012 MANUKA FILL wound dressings Wound Dressing Dressing, Wound, Drug Unclassified FRO Manukapli wound dressings (Manuka Medical, Ltd.) K092689 MANUKA FILL wound dressings are sterile, single-use wound care dressings for use in moist wound management. The primary device is Leptospermum scoparium honey from New Zealand that is harvested and processed under controlled conditions. This honey is sealed into low density polyethylene tubes (LDPE) before sterilization using gamma irradiation. {1}------------------------------------------------ Image /page/1/Figure/0 description: The image contains handwritten text. The top line reads 'K121227'. The second line reads 'posi 2 of 3'. The text appears to be written in black ink on a white background. Image /page/1/Figure/1 description: The image shows a logo with the letters "LMP" in a stylized, blocky font. The letters are outlined in white and filled in with black. The logo has a retro or vintage feel due to the font style and the slightly distressed appearance of the image. #### Links Medical Products Inc. #### Intended Use: MANUKA FILL wound dressings are sterile, single-use wound care dressings for use in moist wound management. MANUKA FILL Wound Dressings may be used Over-The-Counter for: - s Minor Abrasions - > Lacerations - > Minor Cuts - Minor Scalds and Burns A Under the supervision of a healthcare professional, MANUKA FILL wound dressings may be used for: - > Leg Ulcers - Pressure Ulcers s - 1st and 2nd Degree Burns (Superficial and Partial A Thickness) - A Diabetic Foot Ulcers - a Surgical Wounds - Traumatic Wounds A The technical characteristics of MANUKA FILL wound dressing are substantially equivalent to the predicate device. The devices are similar in function, composition, and intended use. Leptospermun scoparium honey is the primary ingredient for MANUKA FILL and the predicate device. Both MANUKA FILL and the predicate device are provided as a single-use device in individually-sterilized packaging. Standard biocompatibility tests were performed on the MANUKA FILL wound dressings; including cytotoxicity, skin irritation, sensitization, and wound healing studies. All tests were performed in accordance with US FDA General Program Memorandum #G95-1 Part-10993-1 of the International Standard Organization (ISO) Standard (Biological Evaluation of Medical Devices) by North American Science Associates, Inc. (NAMSA). The MANUKA FILL wound dressing met the acceptance criteria for all tests conducted and is considered biocompatible under the conditions tested. # LINKS MEDICAL PRODUCTS, INC. 9247 Research Drive Irvine, CA 92618 Technology Comparison: Nonclinical Testing: ર-2 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a logo with the letters 'LMP' in a stylized, interconnected design. The letters are bold and outlined, with the 'L' and 'M' sharing a common base. The 'P' is a rounded shape connected to the 'M'. The logo has a retro or vintage feel due to its design and the image quality. K 12 12 27 Page 3 of 3 Links Medical Products Inc. Additional testing included sterilization validation, shelf-life under accelerated and real-time conditions, and packaging validation. All acceptance criteria were met for all tests conducted. Conclusion of Comparison: MANUKA FILL and the predicate device were both demonstrated to be biocompatible and meet performance requirements for sterility, shelf-life, and packaging. Based upon technological characteristics and nonclinical performance data, MANUKA FILL wound dressings are substantially equivalent and as safe and effective as the currently-marketed predicate device. ## Links Medical Products, Inc. 9247 Research Drive Irvine, CA 92618 5-3 {3}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Links Medical Products, Incorporated % Nano Smart Incorporated James Smith, Ph.D. Consultant 29442 Pointe Royale Laguna Niguel, California 92677 WAY 2 3 2012 Re: K121227 Trade/Device Name: MANUKA FILL wound dressings Regulatory Class: Unclassified Product Code: FRO Dated: April 16 , 2012 Received: April 23, 2012 Dear Dr. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {4}------------------------------------------------ #### Page 2 - Dr. Smith or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page 1 of 1 # INDICATIONS FOR USE 510(k) Number (if known): Device Name: MANUKA FILL wound dressings Indications for Use: MANUKA FILL wound dressings are sterile, single-use, wound care dressings for use in moist wound management. MANUKA FILL wound dressings may be used Over-The-Counter for: - > Minor Abrasions - A Lacerations - A Minor Cuts - > Minor Scalds and Burns Under the supervision of a healthcare professional, MANUKA FILL wound dressings may be used for: - > Leg Ulcers - > Pressure Ulcers - A 1st and 2nd Degree Burns (Superficial and Partial Thickness) - > Diabetic Foot Ulcers - Surgical Wounds A - Traumatic Wounds Prescription Use X (Part 21 CFR 801 Subpart D) ### AND/OR Over-The-Counter Use X (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Kroneberger MM (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K121227
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%