K120994 · Becton, Dickinson and Company · MDB · Aug 7, 2012 · Microbiology
Device Facts
Record ID
K120994
Device Name
BD BACTEC Plus PRIME Aerobic/F Culture Vials
Applicant
Becton, Dickinson and Company
Product Code
MDB · Microbiology
Decision Date
Aug 7, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2560
Device Class
Class 1
Indications for Use
The BD BACTEC Plus PRIME Aerobic/F medium is used in a qualitative procedure for the aerobic culture and recovery of microorganisms (bacteria and yeast) from blood. The principal use of this medium is with the BD BACTEC fluorescent series instruments.
Device Story
The BD BACTEC Plus PRIME Aerobic/F is a blood culture medium vial containing enriched soybean casein digest broth, a chemical CO2 sensor, and a resin blend (two hydrophobic adsorbing resins, one cationic exchange resin) for antimicrobial neutralization. The device is used in clinical laboratories. A blood sample (3-10 mL) is inoculated into the vial and placed in a BD BACTEC fluorescent-series instrument. The instrument incubates the vial at 35°C and monitors the CO2 sensor every ten minutes via fluorescence. Increased fluorescence indicates CO2 production from microbial growth. The instrument provides a positive reading to alert the clinician to the presumptive presence of viable microorganisms. The resin blend enhances recovery of organisms in the presence of antibiotics. The output allows clinicians to identify bloodstream infections and guide antimicrobial therapy.
Clinical Evidence
No clinical data; bench testing only. Analytical studies evaluated 246 paired sets for percent recovery (99.2% agreement) and 180 paired sets for microbial detection limits. Antimicrobial neutralization was tested against 11 drugs at MIC levels (99 sets) and 5 drugs at peak serum levels (150 sets), showing improved recovery rates for the subject device in peak serum challenges. False positive rate was 0% (120 bottles).
Technological Characteristics
Glass vial containing culture medium with two hydrophobic absorbing resins and 0.085% SPS. Detection via chemical sensor monitoring CO2-dependent fluorescence. Compatible with BD BACTEC fluorescent series instruments. Incubation at 35°C (± 1.5°C) for up to 120 hours. Standalone media component used within a networked instrument system.
Indications for Use
Indicated for the aerobic culture and recovery of microorganisms (bacteria and yeast) from human blood samples using BD BACTEC fluorescent series instruments.
Regulatory Classification
Identification
A microbial growth monitor is a device intended for medical purposes that measures the concentration of bacteria suspended in a liquid medium by measuring changes in light scattering properties, optical density, electrical impedance, or by making direct bacterial counts. The device aids in the diagnosis of disease caused by pathogenic microorganisms.
Predicate Devices
BD BACTEC Plus Aerobic /F medium (k083572)
Related Devices
K113558 — BD BACTEC Plus Aerobic/F Culture Vials (plastic) · Becton, Dickinson and Company · Mar 16, 2012
K161810 — BD BACTEC Standard/10 Aerobic/F Culture Vials Soybean-Casein Digest Broth in a Plastic Vial · Becton Dickinson · Feb 9, 2017
K141810 — BD BACTEC PLUS ANAEROBIC/F (PLASTIC) · Becton, Dickinson and Company · Dec 10, 2014
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k120994
B. Purpose for Submission:
To obtain a substantial equivalent determination for a premarket notification for the BD BACTEC Plus PRIME Aerobic/F blood culture medium.
C. Measurand:
Aerobic bacteria and yeast
D. Type of Test:
Liquid culture medium for recovery of microorganisms (bacteria and yeast) from blood using fluorescent instruments to detect increased $\mathrm{CO}_{2}$
E. Applicant:
Becton Dickinson and Company
F. Proprietary and Established Names:
BD BACTEC Plus PRIME Aerobic /F
G. Regulatory Information:
1. Regulation section:
21 CFR 866.2560
2. Classification:
Class I
3. Product code:
MDB
{1}
4. Panel:
83 Microbiology
H. Intended Use:
1. Intended use:
The BD BACTEC Plus PRIME Aerobic/F medium is used in a qualitative procedure for the aerobic culture and recovery of microorganisms (bacteria and yeast) from blood. The principal use of this medium is with the BD BACTEC fluorescent series instruments.
2. Indications for use:
The BD BACTEC Plus Aerobic/F medium is used in a qualitative procedure for the aerobic culture and recovery of microorganisms (bacteria and yeast) from blood. The principal use of this medium is with the BD BACTEC fluorescent series instruments.
3. Special conditions for use statement(s):
Prescription use
4. Special instrument requirements:
BD BACTEC fluorescent series instrument
I. Device Description:
The sample to be tested is inoculated into one or more vials which are inserted into the BACTEC fluorescent series instrument for incubation and periodic reading. Each vial contains a chemical sensor which can detect increases in $\mathrm{CO}_{2}$ produced by the growth of microorganisms. The sensor is monitored by the instrument every ten minutes for an increase in its fluorescence, which is proportional to the amount of $\mathrm{CO}_{2}$ present. A positive reading indicates the presumptive presence of viable microorganisms in the vial. Detection is limited to microorganisms that will grow in a particular type of medium.
Resins have been described for the treatment of blood specimens both prior to and after their inoculation into culture media. Resins have been incorporated into BACTEC culture media to enhance recovery of organisms without a need for special processing.
{2}
J. Substantial Equivalence Information:
1. Predicate device name(s):
BD BACTEC Plus Aerobic /F medium
2. Predicate k number(s):
k083572
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Qualitative procedure for the aerobic culture and recovery of microorganisms (bacteria and yeast) from blood, with the BD BACTEC fluorescent series instrument | Same |
| Specimen type | Human blood | Same |
| Maximum blood to broth ratio | 1:4 | Same |
| Instrumentation | BD BACTEC fluorescent series | Same |
| Detection Technology | Continuous monitoring; incorporate chemical sensor for detection of CO2 increases produced by the growth of aerobic bacteria and yeast | Same |
| Incubation | 35°C (± 1.5°C) up to 120 hrs | Same |
| Vial | Glass | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Resin blend formulation | Two hydrophobic absorbing resins | One hydrophobic absorbing resins |
| Resin blend weight per bottle | Decreased | -- |
| Percent oxygen in head space | changed | -- |
| Sodium polyanetholsulfonate (SPS) concentration | 0.085% | 0.05% |
{3}
4
K. Standard/Guidance Document Referenced (if applicable):
Not applicable
L. Test Principle:
If microorganisms are present in the test sample inoculated into the BACTEC vial, CO₂ will be produced when the organisms metabolize the substrates present in the vial. Increases in the fluorescence of the vial sensor caused by the higher amount of CO₂ are monitored by the BACTEC fluorescent series instrument. Analysis of the rate and amount of CO₂ increase enables the BACTEC fluorescent series instrument to determine if the vial is positive, i.e., that the test sample contains viable organisms.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The reproducibility study was evaluated by time to detection and recovery, in the Instrument Time to Detection and Recovery studies across three lots. Results exhibited no statistically significant difference.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The range of time-to-detection in hours was ≤ 72 hours for each of the organisms listed below. The QC study was performed on three lots and the QC specifications were met.
Aerobic Medium Organisms
| Neisseria meningitidis | Candida glabrata |
| --- | --- |
| ATCC 13090 | ATCC 66032 |
| Haemophilus influenzae | Staphylococcus aureus |
| ATCC 19418 | ATCC 25923 |
| Streptococcus pneumoniae* | Escherichia coli |
| ATCC 6305 | ATCC 25922 |
| Streptococcus pyogenes | Alcaligenes faecalis |
| ATCC 19615 | ATCC 8750 |
| Pseudomonas aeruginosa | |
| ATCC 27853 | |
*CLSI recommended strain
{4}
d. Detection limit:
# Microbial Detection Limit study
The test includes 15 strains tested at two blood volumes, each with two inoculum levels over three lots for a total of 180 paired sets:
15 strains x 2 blood volumes x 2 inoculum levels x 3 lots = 180
Microbial Detection Limit Comparison summary
| Condition | Number of Bottles |
| --- | --- |
| Growth and detection in both the new and predicate device | 89 |
| Growth and detection in the predicate device only | 30 |
| Growth and detection in the new device only | 23 |
| No growth and detection in either the new or predicate devices | 38 |
| McNemar p-value | 0.3363 |
Of the 30 cultures that were positive only in the predicate device, 23 of those that were negative cultures in the new device had low inoculum (0 to 1 CFU per bottle) levels. The remaining seven failures were at the 1 to 10 CFU per bottle inoculum level. The organisms were Cryptococcus neoformans, Micrococcus luteus, Staphylococcus epidermidis (3 and $10\mathrm{mL}$ of blood), Streptococcus pneumoniae (two replicates at $3\mathrm{mL}$ of blood) and Streptococcus sanguinis.
Of the 23 cultures that were positive only in the new device, 19 had low inoculum (0 to 1 CFU per bottle) levels. The remaining four failures were at the 1 to 10 CFU per bottle inoculum level. The four that failed to detect in the 1 to 10 CFU range included Candida glabrata, Micrococcus luteus, Staphylococcus epidermidis and Streptococcus pneumoniae.
Thirty-eight species failed to detect in both the new and predicate devices. Thirty-three of the failures were at the lowest dilution (0 to 1 CFU per bottle) and 5 were at the 1 to 10 CFU per bottle dilution: Escherichia coli, Haemophilus parainfluenzae, and Streptococcus pneumoniae.
The claim for the new BD BACTEC Plus PRIME Aerobic /F bottle was 10-100 CFU/bottle. This data is therefore acceptable for the claimed microbial detection level.
{5}
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
## 2. Comparison studies:
Performance of the BD BACTEC Plus PRIME Aerobic/F medium was evaluated on three lots during internal analytical studies to demonstrate comparable performance to the predicate device – the BD BACTEC Plus Aerobic/F medium. All studies utilized the BD BACTEC FX fluorescent-series instrument operating on the released software (4.10a).
## Percent Recovery (Sensitivity) study
A total of 246 paired sets at 10 to 100 CFU per bottle were evaluated in the Percent Recovery comparison. The study included six replicate paired sets per organism and there were a total of 41 organisms. The six replicate paired sets are comprised of three lots with two blood volumes per lot for both the new and predicate devices. There were 244 (99.2%) positive in both the new and predicate devices.
| Condition | Number of Bottles |
| --- | --- |
| Detected in both the new and predicate devices | 244 per device |
| Detected in the predicate device only | 2 |
| Detected in the new device only | 0 |
| Not detected in either device | 0 |
The two recovery failures observed were with Kingella kingae. Subsequent testing of five isolates of $K$ kingae, including the false negative isolate noted above, under the same test conditions and over three lots all demonstrated growth and recovery as expected.
## Instrument Time to Detection
The Instrument Time to Detection (TTD) study included blood volumes, and organisms. The TTD study included data from both the Percent Recovery and Microbial Detection Limit studies. Each organism was inoculated at appropriate inoculum levels, across two blood volumes (3 mL and 10 mL) over three lots of media. The Microbial Detection Limit study was conducted at the lower inoculum levels (replicates at 0 - 1 CFU, 1 - 10 CFU per bottle); the Percent Recovery study at 10 - 100 CFU per bottle.
{6}
The following organisms had a mean TTD difference that was more rapid in the predicate device: Abiotrophia defectiva (2.39 h), Acinetobacter lwoffii (0.97), Cardiobacterium hominis (3.11 h), Haemophilus influenzae (3.70 h), Streptococcus sanguinis (8.12 h), Candida albicans (0.75 h), Enterococcus faecalis (0.62 h), Granulicatella spp. (1 h), Kingella kingae (4.38 h), Micrococcus luteus (1.06 h), Pediococcus acidilactici (1.08 h) and Streptococcus pneumoniae (0.81 h). Leuconostoc spp. exhibited a mean TTD difference increase of 10.00 h compared to the predicate device (mean TTD of 69.21 h), an increase of 14.45% in the new device.
## False Positive Rate Study
False positivity was assessed with bottles inoculated with freshly drawn human blood (i.e. 2, 4, 6, 8 and 10 mL). No organisms were added to the bottle. There were 120 bottles per device across three lots. No positive results were observed.
## False Negative Rate Study
The data for this study was generated from the Instrument Time to Detection, Percent Recovery (Sensitivity), and Microbial Detection Limit studies. Bottles that are instrument negative at 120 hours and found to be positive when subcultured are classified as false negative bottles.
A total of 93 paired sets were evaluated to determine the false negatives, with the following results:
| Detected in predicate device only | 32 |
| --- | --- |
| Detected in new device only | 23 |
| Detected in neither devices (Negatives) | 38 paired sets |
There were a total of 12 false negative bottles with instrument negative, subculture positive. There were eight false negative results with the BACTEC Plus PRIME Aerobic/F: Micrococcus luteus, Cryptococcus neoformans and six replicates of Haemophilus parainfluenzae biotype I. There were four false negative results with the predicate device: two replicates of Haemophilus parainfluenzae biotype I, Cryptococcus neoformans and Candida glabrata.
There were two false negative bottles with both instrument and subculture negative: two replicates of Kingella kingae with 3 mL of blood (plate count 31 CFU) in the new BACTEC Plus PRIME Aerobic/F media. Subsequent testing of five additional isolates of K. kingae, including the false negative isolate noted above, at inoculum 50-100 CFU and across three lots all demonstrated recovery of Kingella kingae.
There was no statistically significant difference in the false negative rate between the new BACTEC Plus PRIME Aerobic/F media and the predicate device.
7
{7}
# Antimicrobial Neutralization Capability
The ability of the medium to recover organisms in the presence of eleven antimicrobials at minimum inhibitory concentration (MIC) levels was evaluated. Each antimicrobial was tested with three organisms that would be appropriately treated with the drug and each organism was susceptible to the drug at the concentration inoculated into the bottle. The bottles were then assessed for recovery. There were a total of 99 paired sets for this study.
The antimicrobial classes evaluated in this study were: Aminoglycosides, Carbapenems, Glycopeptides, 3rd gen. Cephalosporins, Fluoroquinolones, Tetracyclines, 4th gen. Cephalosporins, Monobactams, Triazoles, Glycylcycline, +Penicillins / β-lactamase inhibitors
MIC Level Drug-Bug Recovery Pair Wise Comparison
| | Predicate Device | | | |
| --- | --- | --- | --- | --- |
| | | Detected | Not Detected | Total |
| New Device | Detected | 94 | 1 | 95 |
| | Not Detected | 2 | 2 | 4 |
| | Total | 96 | 3 | 99 |
MIC Level Drug-Bug Recovery Comparison
| | | Predicate | | New Device | |
| --- | --- | --- | --- | --- | --- |
| Drug | Class | Detected | Not Detected | Detected | Not Detected |
| Aztreonam | Monobactam | 6 | 3 | 5 | 4 |
| Ciprofloxacin | Floroquinolones | 9 | 0 | 9 | 0 |
| Cefotaxime | 3rd gen. Cephalosporin | 9 | 0 | 9 | 0 |
| Fluconazole | Triazole | 9 | 0 | 9 | 0 |
| Cefepime | 4th gen. Cephalosporin | 9 | 0 | 9 | 0 |
| Gentamicin | Aminoglycoside | 9 | 0 | 9 | 0 |
| Meropenem | Carbapenem | 9 | 0 | 9 | 0 |
| Tetracyclin | Tetracyclin | 9 | 0 | 9 | 0 |
| Tigecycline | Glycylcycline | 9 | 0 | 9 | 0 |
| Piperacillin-Tazobactam | Penicillin – β-lactamase inhibitor | 9 | 0 | 9 | 0 |
| Vancomycin | Glycopetide | 9 | 0 | 9 | 0 |
| | Total | 96 | 3 | 95 | 4 |
{8}
The data demonstrated that in the presence of Aztreonam at MIC level, the ability of the new medium to recover organisms was 55.6% (5/9), and 66.7% (6/9) for the predicate medium. Monobactams (Aztreonam) and Carbapenems were not included in the predicate device when cleared. The performance of the carbapenem was acceptable.
A subset of five commonly-used antimicrobials representative of their classes was further evaluated at peak serum levels, concentrations at an average of 20x the MIC of the test organisms. Each antimicrobial was tested with organisms that would be appropriately treated with the drug and each organism was susceptible to the drug at the concentration inoculated into the bottle. There were 150 paired sets of which 66 pair sets were not detected in both the predicate and the new devices:
Peak Serum Level Drug-Bug Challenge Summary
| | Predicate | | | |
| --- | --- | --- | --- | --- |
| | | Detected | Not Detected | Total |
| New Device | Detected | 32 | 44 | 76 |
| | Not Detected | 8 | 66 | 74 |
| | Total | 40 | 110 | 150 |
Peak Serum Level Drug-Bug Challenge Results by Drug
| | Predicate | | New Device | |
| --- | --- | --- | --- | --- |
| Drug | Detected | Not Detected | Detected | Not Detected |
| Amoxicillin – Clavulanate | 3 | 27 | 12 | 18 |
| Cefotaxime | 5 | 25 | 10 | 20 |
| Cefepime | 12 | 18 | 20 | 10 |
| Gentamicin | 17 | 13 | 21 | 9 |
| Meropenem | 3 | 27 | 13 | 17 |
| Total | 40 | 110 | 76 | 74 |
The overall recovery rate for the BD BACTEC Plus PRIME Aerobic/F medium was 51% vs. 27% for the BD BACTEC Plus Aerobic/F medium. The target organism concentrations for these studies were 10-100 CFU/bottle, and 7 mL of blood were used.
a. Method comparison with predicate device:
Performance of the new BD BACTEC Plus PRIME Aerobic /F Plastic was compare to that of the BACTEC Plus Aerobic/F medium.
{9}
b. Matrix comparison:
BD BACTEC Plus culture media, human blood volume, common bloodstream pathogens
3. Clinical studies:
Not Applicable, analytical seeded studies for comparison between the new BD BACTEC Plus PRIME Aerobic /F and the BACTEC Plus Aerobic/F medium.
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Seeded studies performed by BD have shown equivalent performance of the BD BACTEC Plus PRIME Aerobic /F medium compared to the BD BACTEC Plus Aerobic/F medium.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.