K Number
K113747

Validate with FDA (Live)

Date Cleared
2012-08-24

(248 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Randox Opiates Assay: The Randox Laboratories Ltd. Opiates Assay is an in vitro diagnostic test for the qualitative and semi-quantitative analysis of Opiates in human urine. The cut off for both the qualitative and semi-quantitative application is 2000ng/ml for morphine to which the assay is calibrated. Qualitative and semi-quantitative results can be utilized in the diagnosis and treatment of Opiate use or overdose. The Randox Opiates Assay has been developed for use on the RX series analysers, which includes the RX Daytona and the RX Imola. This in vitro diagnostic device is intended for prescription use only.

The semi-quantitative mode is for purposes of (1) Enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GCMS, or (2) Permitting laboratories to establish quality control procedures.

This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatograph/Mass Spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

Randox Multidrug Calibrator Set: The Randox Multidrug Calibrator Set is for use on human urine samples. They are liquid calibrators containing Methamphetamine, Secobarbital, Methadone, and Morphine. There are 5 levels of calibrator. They have been developed for use in the calibration of Amphetamine, Barbiturates, Methadone and Opiate assays on the RX series analysers, which includes the RX Daytona and RX Imola. This in vitro diagnostic device is intended for prescription use only.

Randox Multidrug Controls, Level 1 & 2: The Randox Multidrug Controls, level 1 and 2 are for use on human urine samples. They are liquid controls containing Methamphetamine, Secobarbital, Methadone, and Morphine. There are 2 levels of controls. They have been developed for use in the quality control of Amphetamine, Barbiturates, Methadone and Opiate assays on the RX series analysers, which includes the RX Daytona and RX Imola. This in vitro diagnostic device is intended for prescription use only.

Device Description

Not Found

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text, filled in as much as possible given the limitations of the document:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document describes an FDA 510(k) clearance for the Randox Opiates Assay, Randox Multidrug Calibrator Set, and Randox Multidrug Controls. The core of this clearance hinges on "substantial equivalence" to legally marketed predicate devices. The document explicitly states the "cut off for both the qualitative and semi-quantitative application is 2000ng/ml for morphine to which the assay is calibrated." While it doesn't provide a detailed table of "acceptance criteria" in the format one might see for a diagnostic device's sensitivity/specificity, this cutoff is a key performance parameter and an implied acceptance criterion for the assay's function.

Acceptance Criterion (Implied)Reported Device Performance
Qualitative/Semi-quantitative Opiates assay cut-off (for morphine)2000 ng/mL
Calibration for Amphetamine, Barbiturates, Methadone, and Opiate assays using Multidrug Calibrator Set on RX series analyzersDeveloped for use on RX series analyzers (RX Daytona and RX Imola)
Quality control for Amphetamine, Barbiturates, Methadone, and Opiate assays using Multidrug Controls on RX series analyzersDeveloped for use on RX series analyzers (RX Daytona and RX Imola)
Intended for prescription use onlyDevice is intended for prescription use only
Provides preliminary analytical test result, requiring confirmatory method for confirmed analytical resultAssay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result (e.g., GC/MS).
Semi-quantitative mode for specific purposes (dilution, QC)Purposes explicitly stated: (1) Enabling laboratories to determine an appropriate dilution for confirmation (e.g., GC/MS), or (2) Permitting laboratories to establish quality control procedures.

Study Information:

The provided text is an FDA 510(k) clearance letter. These letters affirm "substantial equivalence" based on information submitted by the manufacturer. They typically do not contain the detailed study protocols, results, and specific performance metrics of the underlying validation studies. The letter refers to the "referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent (for the referenced above and have determined the device is substantially equivalent to legally marketed predicate devices." This implies that the manufacturer submitted data demonstrating equivalence in performance characteristics to existing, cleared assays.

Since this is a clearance letter, specific details about the studies themselves are not present. However, we can infer some information from the nature of the device:

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not specified in the provided document.
  • Data Provenance: Not specified in the provided document. As the manufacturer is based in the United Kingdom, it's possible some or all of the data originated there, but this is not confirmed. The clinical context (human urine samples, diagnosis and treatment of Opiate use/overdose) suggests that the data would be from human subjects. The fact that GC/MS is mentioned as a confirmatory method indicates a rigorous approach to ground truth establishment.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not available in the provided document. For drug testing, the "ground truth" is typically established by a gold standard confirmatory method like Gas Chromatography-Mass Spectrometry (GC/MS), rather than human expert consensus on initial assay results.

4. Adjudication Method for the Test Set

  • This information is not available in the provided document. Given that GC/MS is the preferred confirmatory method, the "adjudication" would be based on the objective results of that method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. An MRMC study is typically for evaluating observer performance with imaging or subjective diagnostic tasks. This device is an in-vitro diagnostic test for chemical analysis (opiates in urine), which is an objective measurement. Therefore, an MRMC study would not be applicable or performed for this type of device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Yes, implicitly. The Randox Opiates Assay is an in-vitro diagnostic device run on automated analyzers (RX Daytona, RX Imola). Its performance characteristics (like sensitivity, specificity, cutoff) would be evaluated in a standalone manner, i.e., the instrument and reagent system's ability to correctly identify and quantify opiates in urine samples without direct human interpretation of the assay result itself. Human involvement comes in collecting the sample, running the test, and interpreting the clinical significance of the result, especially requiring confirmation for positive preliminary results.

7. The Type of Ground Truth Used

  • Confirmatory Analytical Method, specifically Gas Chromatograph/Mass Spectrometry (GC/MS). The document explicitly states: "A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatograph/Mass Spectrometry (GC/MS) is the preferred confirmatory method."

8. The Sample Size for the Training Set

  • This information is not available in the provided document. For an IVD assay, the "training set" would refer to the samples used during product development and optimization to establish assay parameters like the 2000 ng/mL cutoff, linearity, precision, etc. The specific number of samples used for this development is not disclosed in the FDA clearance letter.

9. How the Ground Truth for the Training Set Was Established

  • This information is not available in the provided document. However, given the nature of the device, the ground truth for optimizing the assay would have been established using precisely prepared samples with known concentrations of opiates (spiked controls) and/or clinically relevant samples (patient urine) that have been thoroughly characterized by a gold-standard method like GC/MS.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is composed of three thick, curved lines that form the shape of the bird's body and wings. The text is in a simple, sans-serif font and is evenly spaced around the circle.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Randox Laboratories Limited c/o Pauline Armstrong 55 Diamond Road Crumlin, County Antrim, BT29 4QY United Kingdom

K113747 Re:

Trade Name: Randox Opiates Assay Randox Multidrug Calibrator Set Randox Multidrug Controls, Level 1 and Level 2 Regulation Number: 21 CFR § 862.3650 Regulation Name: Opiate test system

Regulatory Class: Class II Product Codes: DJG, DLJ, LAS Dated: July 5, 2012 Received: July 13, 2012

Dear Dr Armstrong:

We have reviewed your Section 510(k) premarket notification of intent to market the device is We have reviewed your Sector 310(x) pecimatice is substantially equivalent (for the referenced above and have detemmed the device is succedicate devices marketed in
indications for use stated in the enclosure) to legally marketed in
the enclose of use stated indications for use stated in the enclosure) to regality mannel date of the Medical Device in interstate commerce prior to May 25, 1976, the enactive with the provisions of
Amendments, or to devices that have been recurse approval of a premarks Amendments, or to devices that have been recurse worker approval of a premaikst
the Federal Food, Drug, and Cosmetic Act (Act ) that do not require approval of a premailse the Federal Food, Drug, and Cosmetic Act (Act) that do horio sports of the general
approval application (PMA). You may, therefore, market the device, subject to the general approval application (PMA). The general controls provisions of the Act include controls provisions of the Act. Lie general controls provides, good manufacturing practice,
requirements for annual registration, listing of devices, good manufacturing requirements for annual registration, noming and adulteration.
Iabeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III)
Controls (see and collicion) control litianal controls. Existing major regulat If your device is classified (see above) into elife) in (Openar or regulations affecting
(PMA), it may be subject to such additional controls. Existings (CFR), Parts 800 to (PMA), it may be subject to such additions: Existing institutions (CFR), Parts 800 to 895.
your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 your device can be found in Title 21, Code of Federal Negallatons (Cris)
In addition, FDA may publish further announcements concerning your device in the Escient Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not
he months and and to mination that your device complies with other requirement Please be advised that FDA's issuance of a substance complies with other requirement's
mean that FDA has made a determination that your device with other requires. mean that FDA his made a determination into continues and processor in the Federal agencies.
of the Act or any Federal des and regulations administered by not limited to: reg of the Act or any Federal stantes and regulations administered by out not limited to: registration
You must comply with all the Act's requirements, including, but not limited You must comply with all the Act S requirements, including of the more
and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and go and listing (21 CFR Part 807); labeling (21 CFR rats 803 mln.) (21 CFR 803); and good
reporting (reporting of medical device-related adverse events) (21 CFR 803); regulation reporting (reporting of medical device-related adverse evenis) (21 UN 000), and go
manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

AUG 2 4 2012

{1}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire specific ad view 10. Jour es . Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 proase note the regulation villance, "Milion news arrestlance, please contact CDRB is Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) Office of Surveinance and Droinears of (SD) of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...

You may obtain other general information on your responsibilities under the Act from the Tou may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.html

Sincerely yours,

V

Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indication for Use

510(k) Number (if known): K113747

Device Name: OPIATES ASSAY AND THE RANDOX MULTIDRUG CALIBRATOR SET AND MULTIDRUG CONTROLS LEVEL 1 & LEVEL 2

Indication For Use:

Randox Opiates Assay: The Randox Laboratories Ltd. Opiates Assay is an in vitro diagnostic test for the qualitative and semi-quantitative analysis of Opiates in human urine. The cut off for both the qualitative and semi-quantitative application is 2000ng/ml for morphine to which the assay is calibrated. Qualitative and semi-quantitative results can be utilized in the diagnosis and treatment of Opiate use or overdose. The Randox Opiates Assay has been developed for use on the RX series analysers, which includes the RX Daytona and the RX Imola. This in vitro diagnostic device is intended for prescription use only.

The semi-quantitative mode is for purposes of (1) Enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GCMS, or (2) Permitting laboratories to establish quality control procedures.

This assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatograph/Mass Spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

Randox Multidrug Calibrator Set: The Randox Multidrug Calibrator Set is for use on human urine samples. They are liquid calibrators containing Methamphetamine, Secobarbital, Methadone, and Morphine. There are 5 levels of calibrator. They have been developed for use in the calibration of Amphetamine, Barbiturates, Methadone and Opiate assays on the RX series analysers, which includes the RX Daytona and RX Imola. This in vitro diagnostic device is intended for prescription use only.

Randox Multidrug Controls, Level 1 & 2: The Randox Multidrug Controls, level 1 and 2 are for use on human urine samples. They are liquid controls containing Methamphetamine, Secobarbital, Methadone, and Morphine. There are 2 levels of controls. They have been developed for use in the quality control of Amphetamine, Barbiturates, Methadone and Opiate assays on the RX series analysers, which includes the RX Daytona and RX Imola. This in vitro diagnostic device is intended for prescription use only.

Prescription Use マ (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use _ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Signature

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) < 113742

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).