(118 days)
GAL-1E Blood Glucose Monitoring System: The GAL-IE Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is indicated for lay use by people with diabetes, as an aid to monitoring levels in Diabetes Mellitus and should only be used by a single patient and it should not be shared. It is not indicated for the diagnosis or screening of diabetes or for neonatal use.
GAL-1E Blood Glucose Test Strips: The GAL-IE Blood Glucose Test Strips are to be used with the GAL-1E Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from fingertips, palm, or forearm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). They are not indicated for the diagnosis or screening of diabetes or for neonatal use.
GAL-1E Multi Blood Glucose Monitoring System: The GAL-1E Multi Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is indicated to be used for multiple patients in a clinical setting by healthcare professionals, as an aid to monitoring levels in Diabetes Mellitus. This system is only used with single-use, auto-disabling lancing device. It is not indicated for the diagnosis or screening of diabetes or for neonatal use.
GAL-1E Multi Blood Glucose Test Strips: The GAL-IE Multi Blood Glucose Test Strips are to be used with the GAL-IE Multi Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from fingertips, palm, or forearm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). They are not indicated for the diagnosis or screening of diabetes or for neonatal use. They are indicated in a clinical setting to be used for multiple patients by healthcare professionals. This system is only used with single-use, auto-disabling lancing devices.
The GAL-1E blood glucose meter and GAL-1E test strips are used for testing of blood glucose. There is one version for self-testers at home and a second version ("GAL-1E Multi") for professional use. Contrex Plus III Glucose Control Solutions are used for quality control testing of the systems.
Here's a breakdown of the acceptance criteria and study information for the GAL-1E Blood Glucose Monitoring System, based on the provided 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document summarizes an accuracy study but does not explicitly state numerical acceptance criteria in a table. It only states that the "Results demonstrate substantial equivalence to the predicate system." This implies that the performance met the expected standards for a substantially equivalent device. Without specific numerical targets, we can infer that the device's accuracy was comparable to the predicate (GAL-1C Blood Glucose Monitoring System).
A typical accuracy study for a blood glucose monitor would compare the device's readings to a laboratory reference method. Acceptance criteria often follow ISO standards (e.g., ISO 15197), which may require:
- A certain percentage of results to be within ±X mg/dL or ±Y% of the reference value.
- Error Grid Analysis (EGA) to show that a high percentage of results fall within clinically acceptable zones (A and B).
Since these specific numerical criteria and detailed performance data are not provided in the summary, we can only report the general conclusion:
| Acceptance Criteria (Inferred from "Substantial Equivalence") | Reported Device Performance |
|---|---|
| Clinically acceptable accuracy compared to reference method. | Demonstrated substantial equivalence to the predicate system (GAL-1C). |
2. Sample Size Used for the Test Set and Data Provenance
- Sample size: Not explicitly stated. The summary mentions an "accuracy study" and "user study data" but does not provide the number of participants or samples.
- Data Provenance: Not explicitly stated. The company is based in China (Taiwan), but tests could be conducted globally. The summary does not specify if the data was retrospective or prospective.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- This information is not provided in the summary. For blood glucose monitors, the ground truth is typically established by comparing the device readings to a laboratory reference analyzer, not directly by "experts" in the context of interpretation.
4. Adjudication Method for the Test Set
- This information is not applicable and not provided in the summary. Adjudication methods (like 2+1, 3+1) are typically used for subjective clinical assessments (e.g., image interpretation) where multiple experts might disagree. For objective measurements like blood glucose, the reference method provides the "ground truth."
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not mentioned. This type of study is typically for evaluating devices that assist human readers in making clinical decisions (e.g., AI for radiology). Blood glucose meters are standalone measurement devices and do not involve human "readers" in that sense.
6. Standalone (Algorithm Only) Performance
- Yes, a standalone performance study was done. The "accuracy study" and "user study" mentioned relate to the standalone performance of the device in measuring blood glucose levels. The 510(k) summary explicitly states that the device was tested to demonstrate its accuracy in quantitative measurement of glucose.
7. Type of Ground Truth Used
- The summary states "An accuracy study was conducted with professional testing." For blood glucose monitors, the ground truth is established by comparison to a laboratory reference method (e.g., a YSI analyzer), rather than expert consensus, pathology, or outcomes data. While not explicitly named, "professional testing" would imply a validated laboratory method.
8. Sample Size for the Training Set
- The provided document does not contain information about a "training set" in the context of machine learning. The GAL-1E system uses "the same test algorithm as the predicate meter," meaning the core algorithm wasn't necessarily "trained" anew for this specific submission, but rather inherited and validated. Therefore, the concept of a training set size for a machine-learning algorithm is not directly applicable or discussed here. The development of the original algorithm would have involved its own validation, but that's not part of this 510(k) submission.
9. How the Ground Truth for the Training Set Was Established
- As the concept of a "training set" for a machine learning algorithm is not directly applicable to this submission (refer to point 8), information on how its ground truth was established is not available in the provided text. The device's algorithm is described as identical to its predicate, implying its core function and associated "ground truth" establishment would have been part of the predicate's development.
{0}------------------------------------------------
MAR 2 8 2012
.
| 510(k) Summary | |
|---|---|
| Submitter: | Apex Biotechnology Corp. |
| No. 7, Li-Hsin Road V, Hsinchu Science Park | |
| Hsinchu, 30078 | |
| CHINA (TAIWAN) | |
| Contact Person: | Lisa Liu |
| Assistant Manager of Quality Assurance Department | |
| Apex Biotechnology Corp. | |
| No. 7, Li-Hsin Road V, Hsinchu Science Park | |
| Hsinchu, 30078 | |
| CHINA (TAIWAN) | |
| email: lisliu@apexbio.com | |
| Phone: 011-886-3-5641952, ext 567 | |
| Fax: 011-886-3-5678302 | |
| Date Prepared: | March 22, 2012 |
| Trade Names: | GAL-1E Blood Glucose Monitoring System |
| GAL-1E Blood Glucose Test Strips | |
| GAL-1E Multi Blood Glucose Monitoring System | |
| GAL-1E Multi Blood Glucose Test Strips | |
| Classification: | Glucose test system, 21 CFR 862.1345, Class II |
| Product Codes: | CGA, NBW |
| Predicate Device: | GAL-1C Blood Glucose Monitoring System (K102816) |
| GAL-1C Blood Glucose Test Strip (K102816) | |
| Device Description: | The GAL-1E blood glucose meter and GAL-1E test strips are used fortesting of blood glucose. There is one version for self-testers at homeand a second version ("GAL-1E Multi") for professional use. Contrex |
| Plus III Glucose Control Solutions are used for quality control testing ofthe systems. | |
| Intended Use: | |
| GAL-1E Blood Glucose Monitoring System | |
| The GAL-1E Blood Glucose Monitoring System is intended for the | |
| quantitative measurement of glucose in fresh capillary whole blood | |
| samples drawn from the fingertips, forearm, or palm. Alternative site | |
| testing should be performed only during steady-state (when glucose is | |
| not changing rapidly). Testing is done outside the body (In Vitro | |
| diagnostic use). It is indicated for lay use by people with diabetes, as an | |
| aid to monitoring levels in Diabetes Mellitus and should only be used by | |
| a single patient and it should not be shared. It is not indicated for the | |
| diagnosis or screening of diabetes or for neonatal use. | |
| GAL-1E Blood Glucose Test Strips | |
| The GAL-1E Blood Glucose Test Strips are to be used with the GAL- | |
| 1E Blood Glucose Meter to quantitatively measure glucose in capillary | |
| whole blood taken from fingertips, palm, or forearm. Alternative site | |
| testing should be performed only during steady-state (when glucose is | |
| not changing rapidly). They are not indicated for the diagnosis or | |
| screening of diabetes or for neonatal use. | |
| GAL-1E Multi Blood Glucose Monitoring System | |
| The GAL-1E Multi Blood Glucose Monitoring System is intended for | |
| the quantitative measurement of glucose in fresh capillary whole blood | |
| samples drawn from the fingertips, forearm, or palm. Alternative site | |
| testing should be performed only during steady-state (when glucose is | |
| not changing rapidly). Testing is done outside the body (In Vitro | |
| diagnostic use). It is indicated to be used for multiple patients in a | |
| clinical setting by healthcare professionals, as an aid to monitoring | |
| levels in Diabetes Mellitus. This system is only used with single-use, | |
| auto-disabling lancing device. It is not indicated for the diagnosis or | |
| screening of diabetes or for neonatal use. | |
| GAL-1E Multi Blood Glucose Test Strips | |
| The GAL-1E Multi Blood Glucose Test Strips are to be used with the | |
| GAL-1E Multi Blood Glucose Meter to quantitatively measure glucose | |
| in capillary whole blood taken from fingertips, palm, or forearm. | |
| Alternative site testing should be performed only during steady-state | |
| (when glucose is not changing rapidly). They are not indicated for the | |
| diagnosis or screening of diabetes or for neonatal use. They are | |
| indicated in a clinical setting to be used for multiple patients by | |
| healthcare professionals. This system is only used with single-use, auto- | |
| disabling lancing devices. |
{1}------------------------------------------------
510(k) Summary (Continued)
{2}------------------------------------------------
510(k) Summary (Continued)
| Comparison ofTechnologicalCharacteristics: | The GAL-1E and GAL-1E Multi meters have been modified relative tothe predicate by orienting the Liquid Crystal Display (LCD) verticallyand rearranging its icons, adding a strip ejection button, plus altering themeter case to accommodate the LCD change. The GAL-1E and GAL-IE Multi meters use the same test algorithm as the predicate meter. TheGAL-IE and GAL-IE Multi test strips are identical to their predicatedevices. |
|---|---|
| Non-ClinicalTesting: | Testing was conducted as follows: EMC and Electrical Safety, droptesting, disinfection performance (robustness of meter to multiplecleanings and disinfections), software verification and validation, andlinearity testing with validation of Lo/Hi detection. Disinfection testingwith Dispatch wipes was done using a Hepatitis B test model. Resultsdemonstrate substantial equivalence to the predicate system. |
| Clinical Testing | An accuracy study was conducted with professional testing. User studydata were discussed. Results demonstrate substantial equivalence to thepredicate system. |
| Conclusion: | Clinical and non-clinical testing demonstrated that the GAL-IE, andGAL-1E Multi, systems perform in a substantially equivalent manner tothat of the predicate. We conclude that the GAL-1E and GAL-1E Multisystems are substantially equivalent to the predicate system. |
.
.
.
.
:
{3}------------------------------------------------
Image /page/3/Picture/11 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract image of an eagle with three lines representing its head and wings. The seal is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
10903 New Hampshire Avenue Silver Spring, MD 20993
MAR 2 8 2012
Apex Biotechnology Corp. c/o Dr. Bruce A. MacFarlane Senior Principal Scientist 4050 Olson Memorial Hwy, Suite 450 Minneapolis, MN, 55422 USA
Re: K113547 GAL-1E Blood Glucose Monitoring System and GAL-1E Multi Blood Trade/Device Name: Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: II Product Code: CGA, NBW Dated: March 22, 2012 Received: March 23, 2012
Dear Dr. MacFarlane,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical
Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm
Sincerely yours,
signature
Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use Statement
510(k) Number (if known): K113547
Device Name: GAL-1E Blood Glucose Monitoring System
Indications for Use:
The GAL-IE Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is indicated for lay use by people with diabetes, as an aid to monitoring levels in Diabetes Mellitus and should only be used by a single patient and it should not be shared. It is not indicated for the diagnosis or screening of diabetes or for neonatal use.
The GAL-IE Blood Glucose Test Strips are to be used with the GAL-1E Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from fingertips, palm, or forearm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). They are not indicated for the diagnosis or screening of diabetes or for neonatal use.
Over-The-Counter Use Prescription Use AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Signature
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) k113(4)
Page 1 of 2
{6}------------------------------------------------
Indications for Use Statement
510(k) Number (if known): K113547
Device Name: GAL-1E Multi Blood Glucose Monitoring System
Indications for Use:
The GAL-1E Multi Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is indicated to be used for multiple patients in a clinical setting by healthcare professionals, as an aid to monitoring levels in Diabetes Mellius. This system is only used with single-use, auto-disabling lancing device. It is not indicated for the diagnosis or screening of diabetes or for neonatal use.
The GAL-IE Multi Blood Glucose Test Strips are to be used with the GAL-IE Multi Blood Glucose Meter to quantitatively measure glucose in capillary whole blood taken from fingertips, palm, or forearm. Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly). They are not indicated for the diagnosis or screening of diabetes or for neonatal use. They are indicated in a clinical setting to be used for multiple patients by healthcare professionals. This system is only used with single-use, auto-disabling lancing devices.
Prescription Use Over-The-Counter Use X × AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) k 113547
Page 2 of 2
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.