K Number
K113100

Validate with FDA (Live)

Date Cleared
2012-02-06

(110 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTMD6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

Test Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time (Minutes)
Cisplatin, 1.0mg/ml (1,000ppm)>240
Cyclophosphamide (Cytoxan), 20mg/ml (20,000ppm)>240
Carmustine (BCNU), 3.3mg/ml (3,300ppm)0.53
Dacarbazine (DTIC), 10.0mg/ml (10,000ppm)>240
Doxorubicin Hydrochloride, 2.0mg/ml (2,000ppm)>240
Etoposide (Toposar), 20.0mg/ml (20,000ppm)>240
Fluorouracil, 50.0mg/ml (50,000ppm)>240
Paclitaxel (Taxol), 6.0mg/ml (6,000ppm)>240
Thiotepa, 10.0mg/ml (10,000ppm)75.54

Please note that the following drugs have low permeation time of less than 240 minutes: Carmustine (BCNU), 3.3mg/ml (3,300ppm) Thiotepa, 10.0mg/ml (10,000ppm)

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)

Device Description

Not Found

AI/ML Overview

This document describes a 510(k) premarket notification for "Blue Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs (Non-Sterile)". The primary purpose of this submission is to demonstrate substantial equivalence to legally marketed predicate devices for general patient examination glove use, and additionally, to provide data on the gloves' resistance to permeation by chemotherapy drugs.

As such, the information provided focuses on the testing for chemotherapy drug permeation.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the test standard ASTM D6978-05, which is a standard practice for assessing medical gloves for permeation by chemotherapy drugs. While the document doesn't explicitly state a "pass/fail" threshold for breakthrough time (e.g., all drugs must be > X minutes), it does highlight drugs with permeation times less than 240 minutes, implying that longer breakthrough times are generally more desirable. The reported performance is the minimum breakthrough detection time for each drug.

Test Chemotherapy Drug and ConcentrationAcceptance Criteria (Implicit)Reported Device Performance (Minimum Breakthrough Detection Time)
Cisplatin, 1.0mg/ml (1,000ppm)Longer breakthrough time is desirable.>240 minutes
Cyclophosphamide (Cytoxan), 20mg/ml (20,000ppm)Longer breakthrough time is desirable.>240 minutes
Carmustine (BCNU), 3.3mg/ml (3,300ppm)Longer breakthrough time is desirable.0.53 minutes
Dacarbazine (DTIC), 10.0mg/ml (10,000ppm)Longer breakthrough time is desirable.>240 minutes
Doxorubicin Hydrochloride, 2.0mg/ml (2,000ppm)Longer breakthrough time is desirable.>240 minutes
Etoposide (Toposar), 20.0mg/ml (20,000ppm)Longer breakthrough time is desirable.>240 minutes
Fluorouracil, 50.0mg/ml (50,000ppm)Longer breakthrough time is desirable.>240 minutes
Paclitaxel (Taxol), 6.0mg/ml (6,000ppm)Longer breakthrough time is desirable.>240 minutes
Thiotepa, 10.0mg/ml (10,000ppm)Longer breakthrough time is desirable.75.54 minutes

Note on Acceptance Criteria: The document explicitly notes: "Please note that the following drugs have low permeation time of less than 240 minutes: Carmustine (BCNU), 3.3mg/ml (3,300ppm) Thiotepa, 10.0mg/ml (10,000ppm)". This indicates that while the device was tested, a breakthrough time of less than 240 minutes for these specific drugs is acknowledged and disclosed, rather than being an outright "failure". The acceptance criterion is therefore the performance according to the specified ASTM standard, with the understanding that users would be informed about the permeation characteristics.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size (number of gloves or individual tests) used for the chemotherapy drug permeation testing. It only lists the results for each drug.

The data provenance is not explicitly stated as "country of origin" or "retrospective/prospective". However, the manufacturer is Maxter Glove Manufacturing Sdn. Bhd. in Malaysia, suggesting the testing likely occurred in or was commissioned by the manufacturer, possibly in Malaysia or a testing lab accredited to perform ASTM D6978-05. The testing itself is a prospective performance test of the device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of device (patient examination gloves) and the specific test (chemotherapy drug permeation) do not involve "experts" establishing a ground truth in the traditional sense of clinical diagnosis or interpretation. The "ground truth" is established by the standardized laboratory test procedure itself (ASTM D6978-05), which objectively measures the breakthrough time of the chemotherapy drugs through the glove material. The "truth" is the measured time according to the standard.

4. Adjudication Method for the Test Set

Not applicable. This is a laboratory performance test, not a subjective assessment requiring adjudication. The results are quantitative measurements.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a performance test for a patient examination glove, not an AI-assisted diagnostic or interpretive device involving human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a physical device (glove), not an algorithm or AI. The testing performed is an objective measurement of its physical properties.

7. The type of ground truth used

The ground truth used is objective laboratory measurement based on a standardized test method (ASTM D6978-05). This standard details how to measure the permeation of chemotherapy drugs through medical glove materials, providing a quantifiable breakthrough time.

8. The sample size for the training set

Not applicable. This is a physical device undergoing performance testing, not a machine learning model that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As this is not a machine learning model, there is no training set or associated ground truth for a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Yap P. Geeh Assistant Technical Sales Manager MAXTER GLOVE MANUFACTURING SDN BHD Lot 6070. Jalan Haji Abdul Manan 610 Miles Off Jalan Meru Klang, Selangor MALAYSIA 41050

FEB - 6 2012

Re: K113100

Trade/Device Name: Blue Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs (Non-Sterile) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA, LZC Dated: January 16, 2012 Received: January 19, 2012

Dear Mr. Geeh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

{1}------------------------------------------------

Page 2 - Mr. Geeh

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

$\$$ fa Anthony D. lvatson

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo for MAXTER GLOVE MANUFACTURING SDN. BHD. The logo features a black silhouette of a cat-like animal on the left side. To the right of the animal is the text "MAXTER" in bold, capitalized letters. Below the company name is the text "GLOVE MANUFACTURING SDN. BHD. (229652-H)".

LOT 6070

Jalan Haji Abdu! Manan, 6th Miles Off Jalan Meru, 41050 Klang, Selangor, Malaysia. Tel: 603-33929888 (8 Lines) Fax: 603-33923328 E-MAIL: maxter@tm_net.my

INDICATIONS FOR USE

Applicant:

Maxter Glove Manufacturing Sdn. Bhd.

510(K) Number (if known):

K113100

Device Name:

Blue Powder Free Nitrile Examination Gloves Tested For Use With Chemotherapy Drugs (Non-Sterile)

Indications For Use:

A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTMD6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

Test Chemotherapy Drug and ConcentrationMinimum Breakthrough DetectionTime (Minutes)
Cisplatin, 1.0mg/ml (1,000ppm)>240
Cyclophosphamide (Cytoxan), 20mg/ml (20,000ppm)>240
Carmustine (BCNU), 3.3mg/ml (3,300ppm)0.53
Dacarbazine (DTIC), 10.0mg/ml (10,000ppm)>240
Doxorubicin Hydrochloride, 2.0mg/ml (2,000ppm)>240
Etoposide (Toposar), 20.0mg/ml (20,000ppm)>240
Fluorouracil, 50.0mg/ml (50,000ppm)>240
Paclitaxel (Taxol), 6.0mg/ml (6,000ppm)>240
Thiotepa, 10.0mg/ml (10,000ppm)75.54

Please note that the following drugs have low permeation time of less than 240 minutes: Carmustine (BCNU), 3.3mg/ml (3,300ppm) Thiotepa, 10.0mg/ml (10,000ppm)

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Elijah D. Williams

neurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K113100

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.