EUROIMMUN ANTI-ENA POOL ELISA (IGG)

K112996 · Euroimmun US · LLL · Apr 9, 2013 · Immunology

Device Facts

Record IDK112996
Device NameEUROIMMUN ANTI-ENA POOL ELISA (IGG)
ApplicantEuroimmun US
Product CodeLLL · Immunology
Decision DateApr 9, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5100
Device ClassClass 2

Indications for Use

The EUROIMMUN Anti-ENA Pool ELISA (IgG) is intended for the qualitative determination of IgG class antibodies against nuclear antigens (mixture of nRNP/Sm, Sm, SS-A (SS-A 60/Ro-52), SS-B, Scl-70 and ribosomal P proteins) in human serum. It is used as an aid in the diagnosis of mixed connective tissue diseases (MCTD), systemic lupus erythematosus, Sjögren's syndrome and progressive systemic sclerosis, in conjunction with other laboratory and clinical findings.

Device Story

EUROIMMUN Anti-ENA Pool ELISA (IgG) is a qualitative enzyme immunoassay for human serum. Device uses 96-well microtiter plates coated with a mixture of nuclear antigens (nRNP/Sm, Sm, SS-A, SS-B, Scl-70, ribosomal P). Procedure involves two-step incubation: patient sample incubation followed by wash, then anti-human IgG peroxidase-labeled conjugate incubation, wash, and TMB substrate addition. Stop solution added; optical density measured at 450nm using a microwell plate reader. Results expressed as OD ratio; ratio ≥ 1.0 indicates positive result. Used in clinical laboratories by trained personnel to aid diagnosis of autoimmune connective tissue diseases. Two-step design minimizes matrix effects and interference compared to one-step assays. Provides clinicians with serological evidence to support diagnosis of MCTD, SLE, Sjögren's syndrome, and systemic sclerosis.

Clinical Evidence

Clinical study evaluated 432 characterized samples. Overall sensitivity 60.9% (95% CI: 54.7-66.9%) and specificity 96.5% (95% CI: 92.5-98.7%). Sensitivity by condition: MCTD 100%, SLE 55.3%, Scleroderma 45.5%, Sjögren's 72.7%. Specificity confirmed against various autoimmune and infectious disease controls. Analytical performance included precision/reproducibility studies (intra-assay and inter-assay) and interference testing (hemoglobin, triglycerides, bilirubin, rheumatoid factor) showing no significant interference.

Technological Characteristics

Qualitative ELISA; 96-well microtiter plate format. Antigen mixture: nRNP/Sm, Sm, SS-A (60/Ro-52), SS-B, Scl-70, ribosomal P. Detection: Anti-human IgG-peroxidase conjugate with TMB substrate. Requires microwell plate reader (450nm/620-650nm). Manual/automated processing. No software algorithm; rule-based thresholding (Ratio 1.0).

Indications for Use

Indicated for qualitative detection of IgG antibodies against nuclear antigens in human serum to aid in the diagnosis of mixed connective tissue disease (MCTD), systemic lupus erythematosus, Sjögren’s syndrome, and progressive systemic sclerosis in patients suspected of these conditions.

Regulatory Classification

Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k112996 B. Purpose for Submission: New device C. Measurand: IgG antibodies to nRNP/Sm, Sm, SS-A (SS-A 60/Ro-52), SS-B, Scl-70, ribosomal P proteins D. Type of Test: Qualitative, enzyme-linked immunosorbent assay (ELISA) E. Applicant: EUROIMMUN US Inc. F. Proprietary and Established Names: EUROIMMUN Anti-ENA Pool ELISA (IgG) G. Regulatory Information: 1. Regulation section: 21 CFR §866.5100 Antinuclear Antibody Immunological Test System 2. Classification: Class II 3. Product code: LLL — Extractable antinuclear antibody, antigen and control 4. Panel: Immunology (82) H. Intended Use: 1. Intended use(s): The EUROIMMUN Anti-ENA Pool ELISA (IgG) is intended for the qualitative determination of IgG class antibodies against nuclear antigens (mixture of nRNA/Sm, Sm, SS-A (SS-A 60/Ro-52), SS-B, Scl-70, and ribosomal P proteins) in human serum. It is used as an aid in the diagnosis of mixed connective tissue disease (MCTD), systemic lupus erythematosus, Sjögren’s syndrome and progressive systemic sclerosis, in conjunction with other laboratory and clinical findings. 2. Indication(s) for use: Same as Intended Use {1} 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Dual (450 nm and 620-650 nm reference) wavelength spectrophotometer (microwell plate reader) I. Device Description: The EUROIMMUN Anti-ENA Pool ELISA (IgG) contains the following components: microplate strips coated with a mixture of nuclear antigens (nRNA/Sm, Sm, SS-A (SS-A 60/Ro-52), SS-B, Scl-70, and ribosomal P proteins), calibrator, positive and negative controls, horseradish peroxidase (HRP)-conjugated rabbit anti-human IgG, sample buffer, 10x wash buffer, TMB/H₂O₂ (3,3',5,5'-tetramethylbenzidine/hydrogen peroxide) chromogenic substrate solution, and 0.5 M sulphuric acid stop solution. J. Substantial Equivalence Information: 1. Predicate device name(s) and Predicate 510(k) number(s): AESKULISA® ANA HEp-2, k081104 2. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Detection of IgG class antibodies against nuclear antigens | Same | | Indications for Use | Aids in the diagnosis of mixed connective tissue disease (MCTD), systemic lupus erythematosus, Sjögren’s syndrome and progressive systemic sclerosis, in conjunction with other laboratory and clinical findings | Aids in the diagnosis of certain systemic rheumatic diseases and should be used in conjunction with other serological tests and clinical findings. | | Assay Measurement | Qualitative | Same | | Methodology | ELISA | Same | | Assay platform | 96-well microtiter plate | Same | | IgG Enzyme conjugate | HRP-labeled anti-human IgG | Same | | Substrate | TMB | Same | | Matrix | Serum | Same | | Reported results | Optical density (OD) ratio | Same | | Cut-off | OD ratio 1.0 | Same | {2} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Antigen Mixture | nRNA/Sm, Sm, SS-A (SS-A 60/Ro-52), SS-B, Scl-70, and ribosomal P proteins | dsDNA, histones, SS-A (Ro), SS-B (La), Sm, snRNP/Sm, Scl-70, Jo-1 and centromeric antigens and lysed HEp-2 cells | | Calibrators and Controls | 1 calibrator 2 controls: 1 positive, 1 negative | 3 controls: 1 positive, 1 cut-off, 1 negative | | Sample buffer | Ready to use | 5x concentrate | | Wash buffer | 10x concentrate | 50x concentrate | | Stop solution | 0.5 M sulphuric acid | 1 M hydrochloric acid | | Sample dilution | 1:201 | 1:101 | # K. Standard/Guidance Document Referenced (if applicable): Guidance for Industry and FDA Staff: Recommendations for Anti-Nuclear Antibody (ANA) Test System Premarket (510(k)) Submissions (January 22, 2009) DIN EN 13640: 2002 Stability testing of in vitro diagnostic reagent; German version EN 13640: 2002 German and English texts # L. Test Principle: Antibodies present in patient samples bind to the antigen mixture-coated microtiter wells during an incubation step. The unbound sample is washed away and anti-human IgG enzyme conjugate is added. After washing away unbound enzyme conjugate, the remaining conjugate is incubated with chromogenic substrate solution. The reaction is stopped by adding "stop solution" and color intensity is measured as OD units by spectrophotometer. The ODs are proportional to the amount of bound conjugate, which in turn is proportional to the amount of antibodies bound to the antigen mixture in microtiter wells. Results are expressed as ratios which is a ratio of the OD of the control or patient sample to the OD value of the calibrator. # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: Intra- and inter-assay reproducibility of the device was evaluated using 16 and 14 serum samples with different concentrations, respectively. Intra-assay reproducibility was assessed from 16 or 20 determinations on one day. Inter-assay reproducibility was based on 26-30 results from 10 runs performed on 5 days with 2 runs per day. Each sample was assayed in triplicate for each run. For all samples, $100\%$ of the results corresponded to the expected results. {3} Intra-assay reproducibility | Sample | Mean Ratio | Ratio Range | Expected Result | % Correct Result | | --- | --- | --- | --- | --- | | 1 | 0.4 | 0.3 - 0.4 | Negative | 100% | | 2 | 0.8 | 0.7 - 0.8 | Negative | 100% | | 3 | 1.4 | 1.3 - 1.4 | Positive | 100% | | 4 | 2.4 | 2.2 - 2.6 | Positive | 100% | | 5 | 5.7 | 5.4 - 6.0 | Positive | 100% | | 6 | 8.3 | 7.8 - 9.0 | Positive | 100% | | 7 | 0.1 | 0.1 | Negative | 100% | | 8 | 7.4 | 6.8 - 7.6 | Positive | 100% | | 9 | 2.8 | 2.5 - 3.0 | Positive | 100% | | 10 | 4.9 | 4.7 - 5.2 | Positive | 100% | | 11 | 2.5 | 2.1 - 2.7 | Positive | 100% | | 12 | 8.3 | 8.0 - 8.6 | Positive | 100% | | 13 | 10.3 | 9.5 - 11.1 | Positive | 100% | | 14 | 2.1 | 2.0 - 2.3 | Positive | 100% | | 15 | 5.6 | 4.9 - 6.2 | Positive | 100% | | 16 | 2.8 | 2.5 - 3.0 | Positive | 100% | Inter-assay reproducibility | Sample | Mean Ratio | Ratio Range | Expected Result | % Correct Result | | --- | --- | --- | --- | --- | | 1 | 0.4 | 0.3 - 0.5 | Negative | 100% | | 2 | 0.8 | 0.7 - 0.9 | Negative | 100% | | 3 | 1.3 | 1.2 - 1.4 | Positive | 100% | | 4 | 2.7 | 2.5 - 3.0 | Positive | 100% | | 5 | 5.6 | 4.9 - 6.1 | Positive | 100% | | 6 | 8.4 | 7.6 - 9.1 | Positive | 100% | | 7 | 0.2 | 0.2 - 0.3 | Negative | 100% | | 8 | 6.5 | 5.3 - 7.3 | Positive | 100% | | 9 | 2.0 | 1.5 - 2.8 | Positive | 100% | | 10 | 3.5 | 2.6 - 4.9 | Positive | 100% | | 11 | 1.8 | 1.1 - 2.1 | Positive | 100% | | 12 | 6.8 | 5.4 - 8.9 | Positive | 100% | | 13 | 7.1 | 4.5 - 9.8 | Positive | 100% | | 14 | 2.0 | 1.6 - 2.4 | Positive | 100% | Lot-to-lot reproducibility was investigated using 15 sera (3 negative and 12 positive) with a range of ratio values. Each sample/lot combination was tested between 6 and 11 times. The expected result was determined from the sample mean. All of the results correctly matched the expected results. | Sample ID | Mean Ratio | Ratio Range | Lots | Runs/Lot | Total Replicates | Expected Result | Correct Result | | --- | --- | --- | --- | --- | --- | --- | --- | | 1 | 1.1 | 1.1 - 1.2 | 3 | 2 | 6 | Positive | 100% | | 2 | 0.9 | 0.8 - 0.9 | 3 | 2 | 6 | Negative | 100% | | 3 | 0.1 | 0.1 - 0.2 | 11 | 1 | 11 | Negative | 100% | | 5 | 2.7 | 2.5 - 3.3 | 9 | 1 | 9 | Positive | 100% | {4} | Sample ID | Mean Ratio | Ratio Range | Lots | Runs/Lot | Total Replicates | Expected Result | Correct Result | | --- | --- | --- | --- | --- | --- | --- | --- | | 6 | 3.6 | 3.0 - 4.2 | 11 | 1 | 11 | Positive | 100% | | 7 | 5.0 | 4.1 - 5.7 | 11 | 1 | 11 | Positive | 100% | | 8 | 8.5 | 7.0 - 9.8 | 9 | 1 | 9 | Positive | 100% | | 9 | 0.2 | 0.1 - 0.3 | 3 | 2 | 6 | Negative | 100% | | 10 | 6.3 | 5.5 - 7.6 | 3 | 2 | 6 | Positive | 100% | | 11 | 2.4 | 1.9 - 3.0 | 3 | 2 | 6 | Positive | 100% | | 12 | 3.9 | 2.9 - 5.2 | 3 | 2 | 6 | Positive | 100% | | 13 | 2.2 | 1.9 - 2.7 | 3 | 2 | 6 | Positive | 100% | | 14 | 6.8 | 5.9 - 8.2 | 3 | 2 | 6 | Positive | 100% | | 15 | 8.2 | 7.0 - 9.7 | 3 | 2 | 6 | Positive | 100% | | 16 | 2.0 | 1.8 - 2.3 | 3 | 2 | 6 | Positive | 100% | b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): i. **Traceability**: A recognized international standard or reference material for anti-nuclear antibodies is not available. Results of this assay are based on OD reading expressed as in ratios. ii. **Calibrator and Controls**: The OD values of the calibrator and the ratios of the positive and negative controls must lie within the limits stated in the quality control certificate provided with the relevant test kit lot. iii. **Stability**: Three lots of all kit reagents were tested in real time stability studies. Original sealed products were demonstrated to be stable for 12 months when stored at 2-8°C. Opened products were demonstrated to be stable for 12 months when stored at 2-8°C. The reconstituted wash buffer is stable for up to 28 days (4 weeks). d. Detection limit: Not applicable e. Analytical specificity: i. **Interference**: Four samples with different ENA concentrations (ratios from 0.7 to 8.4) were spiked with varying concentrations of hemoglobin up to 1,000 mg/dL, triglycerides up to 2,000 mg/dL, or bilirubin up to 40 mg/dL. Recoveries of all spiked sample/interferent combinations compared to the measurements for the unspiked samples were calculated. The individual recovery of all samples was within the range of 91% to 105%. Six samples with ENA concentration ratios from 1.1 to 11.7 were spiked with rheumatoid factor at 500 IU/mL. Recovery in relation to the original unspiked sample was calculated and found to be between 96% and 106%. {5} ii. Cross Reactivity: The reactivity of the EUROIMMUN Anti-ENA Pool ELISA (IgG) was verified using twelve human reference sera from the CDC ANA reference panel. Samples characterized as homogenous/rim/nDNA (No. 1), nucleolar/U3 RNP (Fibrillarin) (No. 6), centromere (No. 8), and PM-Scl (No. 11) tested negative. All samples with target antigens (Nos. 2, 3, 4, 5, 7, 9, and 12) included in the kit tested positive. CDC sample characterized as anti-Jo-1 positive (No. 10) was also positive. Further testing was conducted to determine whether anti-Jo-1 reactivity was due to the presence of Ro-52 antibodies in the sample. Five samples positive for both anti-Jo-1 and anti-Ro-52 (but negative for nRNA/Sm, Sm, SS-A 60, SS-B, Scl70, and ribosomal P proteins) were positive with the EUROIMMUN Anti-ENA Pool ELISA (IgG). Two anti-Jo-1 positive samples that were negative for anti-Ro-52 antibodies tested negative with the EUROIMMUN Anti-ENA Pool ELISA (IgG). No cross reactivity to Jo-1 is expected. Cross reactivity was investigated in other disease conditions using 82 clinically and serologically characterized samples. The samples were as follows: 10 celiac disease with antibodies against gliadin and tissue transglutaminase, 17 Wegener's granulomatosis with anti-neutrophil cytoplasmic antibodies (ANCA), 39 rheumatoid arthritis with anti-CCP antibodies, and 16 infectious diseases antibody positive. All samples except 2 Wegener's granulomatosis samples were negative with the EUROIMMUN Anti-ENA Pool ELISA (IgG). f. Assay cut-off: The assay cut-off is OD ratio of 1.0. Results ≥ 1.0 are positive, and results < 1.0 are negative. 2. Comparison studies: a. Method comparison with predicate device: Two hundred seventy-eight (278) clinically characterized samples from patients and control groups were tested with the predicate device and the EUROIMMUN Anti-ENA Pool ELISA (IgG). The samples were from 49 mixed connective tissue disease (MCTD), 26 systemic lupus erythrematosus (SLE), 29 Sjögren's syndrome, 22 systemic sclerosis, 20 polymyositis/dermatomyositis, 10 celiac disease, 17 Wegener's granulomatosis, 39 rheumatoid arthritis, and 16 infectious disease patients. Fifty healthy individuals were also included. | | Predicate | | | | | --- | --- | --- | --- | --- | | | | Positive | Negative | Total | | EUROIMMUN Anti-ENA Pool ELISA (IgG) | Positive | 135 | 1 | 136 | | | Negative | 3 | 139 | 142 | | | Total | 138 | 140 | 278 | Positive agreement: 135/138 = 97.8% (95% CI: 93.8% – 99.5%) Negative agreement: 139/140 = 99.3% (95% CI: 96.1% – 100%) 6 {6} Overall agreement: $274 / 278 = 98.6\%$ (95% CI: 96.4% - 99.6%) # Method comparison with individual ENA ELISAs A comparison study was performed with six monospecific autoantibody ELISAs to ensure that all targets are recognized by the EUROIMMUN Anti-ENA Pool ELISA (IgG). Samples from MCTD, SLE, Sjögren's syndrome, systemic sclerosis, rheumatoid arthritis, fibromyalgia, infectious diseases and healthy individuals. The combined positive percent agreement was $100\%$ (95%CI: 94.7%-100%), negative percent agreement was $91.7\%$ (95% CI: 80.0%-97.7%) and overall agreement of $96.6\%$ (95%CI: 91.4%-99.1%). There were two false positive samples each for the SSA/SSB and Scl-70 comparison. # b. Matrix comparison: Serum is the only sample specimen. # 3. Clinical studies: # a. Clinical Sensitivity: Clinical sensitivity was determined from 261 clinically characterized samples from patients with MCTD $(n = 44)$ , SLE $(n = 85)$ , systemic sclerosis $(n = 66)$ , and Sjögren's syndrome $(n = 66)$ . The EUROIMMUN Anti-ENA Pool ELISA (IgG) showed an overall sensitivity of $60.9\%$ (95% CI: $54.7 - 66.9\%$ ). The results are in the table below. | Condition | Total Samples | Anti-ENA Pool ELISA (IgG) | | | | --- | --- | --- | --- | --- | | | | Positive | % Positive | 95% CI | | MCTD | 44 | 44 | 100.0 | 92.0 – 100.0 | | SLE | 85 | 47 | 55.3 | 44.1 – 66.1 | | Systemic sclerosis | 66 | 30 | 45.5 | 33.1 – 58.2 | | Sjögren's syndrome | 66 | 48 | 72.7 | 60.4 – 83.0 | | Total | 261 | 159 | 60.9 | 54.7 – 66.9 | # b. Clinical specificity: Clinical specificity was determined from 171 clinically characterized samples from the following control groups: polymyositis/dermatomyositis $(n = 26)$ , celiac disease $(n = 21)$ , Wegener's granulomatosis $(n = 17)$ , rheumatoid arthritis $(n = 39)$ , other autoimmune diseases $(n = 52)$ , and bacterial/viral infections $(n = 16)$ . The EUROIMMUN Anti-ENA Pool ELISA (IgG) showed an overall specificity of $96.5\%$ (95% CI: $92.5 - 98.7\%$ ). The results are listed below. | Control Group | Total Samples | Anti-ENA Pool ELISA (IgG) | | | | --- | --- | --- | --- | --- | | | | Negative | % | 95% CI | | Polymyositis/dermatomyositis | 26 | 22 | 84.6 | 65.1 – 95.6 | | Celiac disease | 21 | 21 | 100.0 | 83.9 – 100.0 | | Wegener's granulomatosis | 17 | 15 | 88.2 | 63.6 – 98.5 | {7} | Control Group | Total Samples | Anti-ENA Pool ELISA (IgG) | | | | --- | --- | --- | --- | --- | | | | Negative | % | 95% CI | | Rheumatoid arthritis | 39 | 39 | 100.0 | 91.0 – 100.0 | | Other autoimmune diseases* | 52 | 52 | 100.0 | 93.2 – 100.0 | | Bacterial/viral infections | 16 | 16 | 100.0 | 79.4 – 100.0 | | Total | 171 | 165 | 96.5 | 92.5 – 98.7 | *Other autoimmune diseases includes autoimmune hepatitis (n=8), primary biliary cirrhosis (n=9), Grave’s disease (n=12), Hashimoto’s disease (n=11), and type I diabetes (n=12). c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: See Assay cut-off 5. Expected values/Reference range: The levels of anti-ENA antibodies (IgG) were analyzed in a panel of 200 samples from apparently healthy blood donors. With a cut-off ratio of 1.0, a prevalence of 1.5% was obtained. The mean ratio was 0.2 (SD = 0.37) and the values ranged from 0.0 to 5.0. Users of the kit should generate their own ranges, as stated in the product insert. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...