SANTEVIEW
K111312 · Sehyun Co., Ltd. · MQB · Jul 13, 2011 · Radiology
Device Facts
| Record ID | K111312 |
| Device Name | SANTEVIEW |
| Applicant | Sehyun Co., Ltd. |
| Product Code | MQB · Radiology |
| Decision Date | Jul 13, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1680 |
| Device Class | Class 2 |
Intended Use
The SanteView-3001C provides digital image capture for conventional film/screen radiographic examinations. The device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Device Story
SanteView-3001C is a solid-state digital flat panel X-ray detector; 440x440mm imaging area. Device intercepts X-ray photons; scintillator emits visible light; photo-detector array converts light to electrical signals; signals converted to digital values for monitor display. Used in clinical settings for general radiographic procedures; operated by radiology staff. Output provides digital images for diagnostic review by physicians; replaces traditional film/screen systems to facilitate digital workflow and image capture.
Clinical Evidence
No clinical data. Safety and effectiveness established via bench testing and EMC testing per EN/IEC 60601-1-2. Concurrence study performed to compare performance against predicate devices.
Technological Characteristics
Solid-state X-ray imager; 440x440mm imaging area; scintillator-based photo-detector array. Connectivity: digital output for monitor display. Standards: EN/IEC 60601-1-2 (EMC).
Indications for Use
Indicated for digital image capture in general purpose diagnostic radiographic examinations, replacing conventional film/screen systems. Not indicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- PaxScan 4030 Medical Digital Imaging System (K024147)
- CXDI-60C (K091545)
- Pixium 4600 (Philips Bucky Vision) (K982795)
Related Devices
- K091435 — DIGITAL RADIOGRAPHY MODEL CXDI-55G · Canon, Inc. · May 29, 2009
- K150929 — CareView 1500Cw X-ray Flat Panel Detectors · Careray Digital Medical System Co., Ltd. · Aug 4, 2015
- K120020 — VIVIX-S · Vieworks Co., Ltd. · May 4, 2012
- K232753 — EXPD 4343D; EXPD 3643D · DRTECH Corporation · Nov 16, 2023
- K153492 — CareView 1800L · Careray Digital Medical System Co., Ltd. · Dec 24, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
1115
This summary of 510(K) -- safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 05/06/2011
- 1. Company making the submission
| | Submitter |
|---------|------------------------------------------------------------------------------------------------|
| Name | SEHYUN Co., Ltd. |
| Address | #709, 1 dong, Innoplex #343, Wonchon-dong, Youngtong-gu,<br>Suwon, Gyeonggi-do, 443-823, Korea |
| Phone | +82-31-8002-2000 |
| Fax | +82-33-8002-2017 |
- 2. U.S Agent/Contact Person
َ ُ ُ َ َ Consulting Group 951 Starbuck St. Unit J, Fullerton, CA 92833 Priscilla Chung Phone: 714-869-3080 Fax: 714-409-3357 Email: info@LKconsultinggroup.com
- 3. Device
- · Trade Name: SanteView-3001C
- Common Name: Digital Flat Panel X-ray Detector
- · Classification: Class II
- · Classification regulation: 21 CFR 892.1650
- · Product Code: MQB
## 4. Predicate Device:
•PaxScan 4030 Medical Digital Imaging System by Varian Medical System (K024147) .CXDI-60C by Canon USA, Inc.(K091545)
•Pixium 4600 (cleared under the name, Philips Bucky Vision) by Trixell (K982795)
- 5. Description:
The SanteView-3001C is a solid state x-ray imager which has 440x440mm
510(k) Submission.
510(k) summary , 1 / 2 page
{1}------------------------------------------------
imaging area. The SanteView-3001C intercepts x-ray photons and the scintillator emits visible spectrum photons that illuminate an array of photo-detectors that create an electrical signals. After the electrical signals are generated, it is converted to digital value, and the images will be displayed on monitors.
6. Indication for use:
The SanteView-3001C provides digital image capture for conventional film/screen radiographic examinations. The device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for ·· mammography applications.
7. Comparison to substantially equivalent devices:
The proposed and the predicate devices utilize similar technology and materials, and are similar in design and construction.
- 8. Safety, EMC and Performance Data:
EMC testing was conducted in accordance with standard EN/IEC 60601-1-2: All test results were satisfactory. The result of concurrence study shows that the new device is safe and effective as the predicate device
#### 9. Conclusion
SEHYUN Co., Ltd. concludes that the SanteView-3001C is substantially equivalent to the currently marketed devices. It does not introduce new indications for use, has the same technological characteristics and does not introduce new potential hazards or ·safety risks.
ﺮ ۰۰
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
JUL 1 3 2011
SEHYUN Co., Ltd. % Ms. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group 951 Starbuck St., Unit J FULLERTON CA 92833
Re: K111312
Trade/Device Name: SanteView-3001C Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: MOB Dated: May 6, 2011 Received: May 10, 2011
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 5.10(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K111312
Device Name: SanteView-3001C
Indications for Use:
The SanteView-3001C provides digital image capture for conventional film/screen radiographic examinations. The device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S. Posth
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Number K141312
Page 1 of 1