STERITEC SMART-READ BIOLOGICAL TEST PACK WITH INSTANT READOUT INTEGRATOR

K111150 · Steritec Products Mfg Co, Inc. · FRC · Jul 8, 2011 · General Hospital

Device Facts

Record IDK111150
Device NameSTERITEC SMART-READ BIOLOGICAL TEST PACK WITH INSTANT READOUT INTEGRATOR
ApplicantSteritec Products Mfg Co, Inc.
Product CodeFRC · General Hospital
Decision DateJul 8, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

SteriTec Smart Read Biological Test Pack with Instant Readout Integrator is designed specifically for testing of 3 or 4 minutes at 132°C (270°F) and 3 minutes at 135°C (275°F) in pre-vacuum steam sterilizers. The test pack consists of a self-contained biological indicator containing Geobacillus stearothermophilus inside a small package of porous and nonporous materials, simulating the biological indicator 16 towel test packs defined by AAMI ST79. When used as directed, the Chemical Indicator Record Card gives visible indication that sterilizing conditions were met.

Device Story

Device is a self-contained biological indicator test pack for pre-vacuum steam sterilizers. Input: steam sterilization cycle conditions (132°C/135°C). Operation: Geobacillus stearothermophilus spores within a package of porous/nonporous materials simulate AAMI ST79 16-towel test packs. Output: visual color change on an integrated Chemical Indicator Record Card confirming sterilization parameters met. Used by sterile processing staff in clinical settings to validate sterilizer performance. Benefits: provides assurance of sterilization efficacy, ensuring patient safety by confirming instruments are processed correctly.

Clinical Evidence

Bench testing only. Performance validated against AAMI ST79 requirements for 16-towel test packs using Geobacillus stearothermophilus biological indicators and chemical integration.

Technological Characteristics

Self-contained biological indicator containing Geobacillus stearothermophilus; integrated chemical indicator record card; porous and nonporous packaging materials; designed for pre-vacuum steam sterilization cycles at 132°C and 135°C.

Indications for Use

Indicated for monitoring pre-vacuum steam sterilization cycles at 132°C (3 or 4 minutes) and 135°C (3 minutes). Used in healthcare facilities to verify sterilization conditions.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Jonathan Rutigliano Director, Regulatory Affairs SteriTec Products Manufacturing Company, Incorporated 74 Inverness Drive East Englewood, Colorado 80112 - 8 2011 . IIII Re: K111150 Trade/Device Name: Smart-Read Biological Test Pack with Instant Readout Integrator 135, Model No.: Bl 108 Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: FRC Dated: June 8, 2011 Received: June 9, 2011 Dear Mr. Rutigliano: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ ## 2 Page- Mr. Rutigliano Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffice s/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Anthony D. Martin Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: SteriTec Smart-Read Biological Test Pack with Instant Readout Integrator 135, Model No.: Bl 108 ## Indications For Use: SteriTec Smart Read Biological Test Pack with Instant Readout Integrator is designed specifically for testing of 3 or 4 minutes at 132°C (270°F) and 3 minutes at 135°C (275°F) in pre-vacuum steam sterilizers. The test pack consists of a self-contained biological indicator containing Geobacillus stearothermophilus inside a small package of porous and nonporous materials, simulating the biological indicator 16 towel test packs defined by AAMI ST79. When used as directed, the Chemical Indicator Record Card gives visible indication that sterilizing conditions were met. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use × (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE) Mapieolangulni ion Sian-Off) Division of Anesthesiology, General Hospital nfection Control, Dental Devices 510(k) Number: KII1/SQ
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...