POWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED)
K110969 · Ppp Medical and Safety Products Company Limited · LYZ · Jun 10, 2011 · General Hospital
Device Facts
| Record ID | K110969 |
| Device Name | POWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED) |
| Applicant | Ppp Medical and Safety Products Company Limited |
| Product Code | LYZ · General Hospital |
| Decision Date | Jun 10, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.6250 |
| Device Class | Class 1 |
Intended Use
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable vinyl patient examination gloves; worn on examiner's hand or finger; intended to prevent cross-contamination between patient and examiner. Used in clinical settings by healthcare providers. Device is non-sterile and clear (non-colored).
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing, including physical property testing, pinhole leak testing, and biocompatibility assessments (skin irritation and dermal sensitization) per ASTM and ISO standards.
Technological Characteristics
Material: Vinyl. Standards: ASTM D5250-06e1 (dimensions, physical properties), ASTM D6124-06 (powder residual), 21 CFR 800.20 (pinholes), AAMI/ANSI/ISO 10993-10 (biocompatibility). Powder residual <2mg/glove. Non-sterile.
Indications for Use
Indicated for use as a non-sterile, disposable patient examination glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
- Powder-Free Vinyl Patient Examination Gloves, Clear(Non-colored), ZHANG JIA GANG FENGYUAN PLASTIC PRODUCTS CO LTD. (K091663)
Related Devices
- K100963 — POWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED) · Hebei Shenzichen Plastic & Rubber Products Co., LT · Aug 6, 2010
- K102431 — POWDER-FREE VINYL PATIENT EXAMINATION GLOVES · Tangshan Fuqiang Glove Co., Ltd. · Oct 4, 2010
- K070076 — POWDER FREE VINYL PATINET EXAMINATION GLOVES, CLEAR (NON-COLORED) · Jiangsu Cureguard Glove Co., Ltd. · Feb 22, 2007
- K100419 — POWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED) · Tangshan Yongkang Glove Co., Ltd. · Apr 6, 2010
- K191292 — Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) · Anhui Powerguard Technology Co., Ltd. · Nov 5, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
# Section C 510(k) Summary (21 CFR 807.92)
.
.. ... ... .
。・
, 1
# 510(K) Summary
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: < ( / 0 9 6 9 " (applicant leave blank)
Premarket Notification [510(k)] Summary
Panel:
Product Code:
# [(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :
| Submitter's name : | PPP Medical and Safety Products Company<br>Limited |
|--------------------------------|----------------------------------------------------|
| Submitter's address : | Xinlian,Houqiao Town, Wuxi,Jiangsu,214106,China |
| Phone number : | 0086-510-88725378 |
| Fax number : | 0086-510-88725378 |
| Name of contact person: | Mr. Zou Xiaojun |
| Date the summary was prepared: | Feb 9, 2011 |
### [(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known
| Device Name: | Powder Free Vinyl Patient Examination Gloves,<br>Clear (non-colored) |
|-------------------------|--------------------------------------------------------------------------------------------------------|
| Proprietary/Trade name: | Powder Free Vinyl Patient Examination Gloves,<br>Clear (non-colored)<br>Other clients private labeling |
| Common Name: | Exam gloves |
| Classification Name: | Patient examination glove |
| Device Classification: | I |
| Regulation Number: | 21 CFR 880.6250 |
[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .
LYZ
General Hospital
(80)
Class I* Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06e1.
{1}------------------------------------------------
Predicate device: Powder-Free Vinyl Patient Examination Gloves, Clear(Non-colored), ZHANG JIA GANG FENGYUAN PLASTIC PRODUCTS CO LTD.. K091663 .
### [(a)(4)] A description of the device
Device Description: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06e1.
#### [(a)(5)] The summary describes the intended use of the device
Device Intended Use: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
# [(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.
The Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.
| Characteristics | Standard | Device performance |
|-----------------------|----------------------------------------------------------------------|-----------------------------------------|
| Dimension | ASTM standard D 5250-06 el. | Meets |
| Physical Properties | ASTM standard D 5250-06 el. | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Powder Residual | ASTM standard D 5250-06 el<br>and D6124-06 | <2mg/glove |
| Biocompatability | Primary Skin Irritation in rabbits<br>AAMI / ANSI / ISO 10993-10 | Passes<br>Not a Primary Skin Irritation |
| | Dermal sensitization in the guinea pig<br>AAMI / ANSI / ISO 10993-10 | Passes<br>Not a Dermal sensitization |
### [(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), meet requirements per ASTM D5250-06 e1, per ASTM D6124-06, per 21 CFR 800.20 and FDA recognition number 2-152: AAMI / ANSI / ISO 10993-10:2002/Amd. 1:2006.
#### [{b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence _
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
# [(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe.as effective. and performed as well or better than the legally marketed device identified in (a)(3).
It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims.
It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is as safe, as effective, and performs as well as the predicate device, Powder-Free Vinyl Patient Examination Gloves, Clear(Non-colored), ZHANG JIA GANG
FENGYUAN PLASTIC PRODUCTS CO LTD.. K091663 .
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
PPP Medical and Safety Products Company, Limited C/O Mr. Chu Xiaoan Beijing Easy-Link Company Room 1606, Building 1, Jianxiang Yuan No. 209 Bei Si Huan Zhong Road Haidian District Beijing CHINA 100083
JUN 1 0 2011
Re: K110969
Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-Colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: February 9, 2011 Received: May 27, 2011
Dear Ms. Xiaoan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Mr. Xiaoan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE
PPP Medical and Safety Products Company Limited Applicant:
510(k) Number (if known): *
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) Device Name:
Indications For Use:
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizith F. Clamie Well
Division Sian-Offi Division of Anesthesiology, General Hospital nfection Control, Dental Devices
510(k) Number:
.
・
·
.