5.5 / 6.5MM SPARTAN PEEK SURTURE IMPLANT WITH #2 MAGNUMWIRE / WITH NEEDLES, EXTRACTION TOOL, 5.5 / 6.5MM PUNCH TAP

K110781 · Arthrocare Corp. · MBI · May 27, 2011 · Orthopedic

Device Facts

Record IDK110781
Device Name5.5 / 6.5MM SPARTAN PEEK SURTURE IMPLANT WITH #2 MAGNUMWIRE / WITH NEEDLES, EXTRACTION TOOL, 5.5 / 6.5MM PUNCH TAP
ApplicantArthrocare Corp.
Product CodeMBI · Orthopedic
Decision DateMay 27, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The ArthroCare Spartan PEEK Suture Implant is indicated for use in fixation of soft tissue to bone. Examples of such procedures include: Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis, and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction, and midfoot reconstruction Foot: Hallux valgus reconstruction Elbow: Tennis elbow repair, biceps tendon reattachment Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions

Device Story

ArthroCare Spartan PEEK Suture Implant System consists of fully-threaded, corkscrew-shaped PEEK suture anchor (5.5mm diameter) preconfigured with MagnumWire sutures; mounted on disposable delivery driver. Associated instruments (Punch Tap, Extraction Tool) used for implantation/removal. Device used by surgeons in clinical/OR settings for soft tissue-to-bone fixation. Implant provides mechanical anchor point for sutures to secure soft tissue during orthopedic repairs (e.g., rotator cuff, ligament reconstruction).

Clinical Evidence

Bench testing only. In vitro testing performed involving insertion of anchors in simulated human bone substrate, including ultimate strength and cyclic loading tests.

Technological Characteristics

Material: PEEK (polyether-etherketone). Form factor: Fully-threaded, corkscrew-shaped anchor, 5.5mm diameter. Delivery: Disposable driver, preconfigured with MagnumWire sutures. Sterile supply.

Indications for Use

Indicated for soft tissue to bone fixation in shoulder, ankle, foot, elbow, and knee orthopedic procedures requiring suture anchoring.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K110781 (112) # MAY 2 7 2011 #### 510(k) Summary # ArthroCare Corporation Spartan™ PEEK Suture Implant System ### General Information | Submitter Name/Address: | ArthroCare Corporation<br>680 Vaqueros Avenue<br>Sunnyvale, CA 94085-2936 | |------------------------------------|---------------------------------------------------------------------------| | Establishment Registration Number: | 2951580 | | Contact Person: | Valerie Defiesta-Ng<br>Director, Regulatory Affairs | | Date Prepared: | March 18, 2011 | **Device Description** | Trade Name: | ArthroCare Spartan PEEK Suture Implant System | |------------------------|-------------------------------------------------------------------------| | Generic/Common Name: | Screw, Fixation, Bone | | Classification Name: | Smooth or Threaded Metallic Bone Fixation<br>Fastener (21 CFR 888.3040) | | Device Classification: | Class II, 21 CFR 888.3040<br>Product Code MBI | **Predicate Device** | ArthroCare 5.5mm Spartan PEEK Suture<br>Implant System | K110164 (February 7, 2011) | |--------------------------------------------------------|----------------------------| | ArthroCare 5.5mm Spartan PEEK Suture<br>Implant System | K102262 (October 21, 2010) | ## Product Description The ArthroCare Spartan PEEK (polyether-etherketone) Suture Implant is a fullythreaded, corkscrew shape anchor available in 5.5mm diameter size. The suture anchor comes preconfigured with MagnumWire® sutures for traditional knot tying and is mounted on a disposable delivery driver. The device is supplied sterile and is available with our without needles. Associated instruments (Punch Tap and Extraction Tool) for implantation and removal of the implant are available separately and altogether are referred to as the ArthroCare Spartan PEEK Suture Implant System. {1}------------------------------------------------ ## Intended Uses/Indications for Use The ArthroCare Spartan PEEK Suture Implant is indicated for use in fixation of soft tissue to bone. Examples of such procedures include: Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis, and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction, and midfoot reconstruction Foot: Hallux valgus reconstruction Elbow: Tennis elbow repair, biceps tendon reattachment Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions ### Substantial Equivalence In establishing substantial equivalence to the predicate. ArthroCare compared the intended use, device design, technology, and device components of the subject device with the predicate device. Bench performance testing has been completed to demonstrate the substantial equivalence of the ArthroCare Spartan PEEK Suture Implant in accordance with FDA Guidance Document Testing for Bone Anchors. The in vitro testing performed involved insertion of the anchors in a simulated human bone substrate followed by both ultimate strength and cyclic loading tests. The performance testing and device comparison demonstrate that the subject device is substantially equivalent to the predicate device, and is safe and effective for its intended use. #### Summary of Safety and Effectiveness The proposed modifications to the Spartan PEEK Suture Implant System are not substantial changes, and do not significantly affect the safety or efficacy of the proposed device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a bird-like figure with three curved lines representing wings or snakes. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 ArthoCare Corporation % Ms. Valerie Defiesta-Ng Director, Regulatory Affairs 680 Vaqueros Avenue Sunnyvale, California 94085-2936 MAY 2 7 2011 Re: K110781 Trade/Device Name: ArthoCare® Spartan™ PEEK Suture Implant System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: May 13, 2011 Received: May 16, 2011 Dear Ms. Ng: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ Page 2 - Ms. Valerie Defiesta-Ng or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. E.Id Keith for Mark Director Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # 110781 (111) # Indications for Use 510(k) Number (if known): Device Name: ArthroCare® Spartan™ PEEK Suture Implant System Indications For Use: The ArthroCare Spartan PEEK Suture Implant is indicated for use in fixation of soft tissue to bone. Examples of such procedures include: Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis, and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction, and midfoot reconstruction Foot: Hallux valgus reconstruction Elbow: Tennis elbow repair, biceps tendon reattachment Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Eiel Keitt (Division Sign-Off) (Division Sign-On) Division of Surgical, Orthopedic, Division of Son of Societive Devices K110781 510(k) Number Page 1 of
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%