F10
K110630 · Mediana Co., Ltd. · KNG · Jun 24, 2011 · Obstetrics/Gynecology
Device Facts
| Record ID | K110630 |
| Device Name | F10 |
| Applicant | Mediana Co., Ltd. |
| Product Code | KNG · Obstetrics/Gynecology |
| Decision Date | Jun 24, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.2660 |
| Device Class | Class 2 |
| Attributes | 3rd-Party Reviewed |
Intended Use
The F10 Fetal Doppler measures heart rate, which is displayed on a LCD display and provides fetal heart sounds. The fetal heart rate is measured using Doppler ultrasound.
Device Story
Pocket-size fetal Doppler; uses 2MHz PW Doppler ultrasound transducer to irradiate pregnant abdomen; detects Doppler frequency signals from fetal heart; analyzes signals to calculate fetal heart rate (FHR); outputs FHR on LCD screen and provides audible fetal heart sounds via built-in speaker. Used in clinical settings by healthcare providers to assess fetal well-being. Provides real-time monitoring; aids in clinical decision-making regarding fetal health.
Clinical Evidence
Bench testing only. Performance data provided includes electrical, mechanical, and environmental safety testing per EN/IEC 60601-1 and EMC testing per EN/IEC 60601-1-2. No clinical data presented.
Technological Characteristics
Pocket-size fetal Doppler; 2MHz PW Doppler ultrasound transducer; LCD display; built-in speaker; battery-powered. Complies with EN/IEC 60601-1 (safety) and EN/IEC 60601-1-2 (EMC).
Indications for Use
Indicated for fetal heart rate monitoring and fetal heart sound auscultation in pregnant women using Doppler ultrasound.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
Predicate Devices
- Edan Instruments, Inc. Ultrasonic Pocket Doppler, Sonotrax (K040480)
- Bistos Co., Ltd. Ultrasonic Pocket Doppler, BT-200 (K05219)
Related Devices
- K142799 — Model BT-220L and BT-220C · Bistos Co., Ltd. · Jul 9, 2015
- K100885 — FETAL MONITOR MODEL: BT-200T · Bistos Co., Ltd. · Jun 7, 2010
- K182526 — Fetal Doppler (Models: FD-231B, FD-231D, FD-640B, FD-640D, FD-200B, FD-200D, FD-591B, and FD-591D) · Vcomin Technology Limited · May 29, 2019
- K172780 — SD1 Ultrasonic Pocket Doppler · Edan Instruments, Inc. · Jun 6, 2018
- K252669 — Fetal Doppler U8-25, U9-25 · Zhongshan Xiaolan Town Senlan Electronic Factory · Jan 8, 2026
Submission Summary (Full Text)
{0}------------------------------------------------
# JUN 2 4 2011
510(k) Submission - F10
## 510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
### Submitter
Mediana Co., Ltd. Dongwha Medical Instrument Complex, 1650-1, Donghwa-ri, Munmak-eup, Wonju-si, Gangwon-do, Korea Tel) +82-70-7092-9964 Fax) +82-2-542-7447 Company Contact: Amy M.H. Kim Date Summary Prepared: October 26, 2010
#### Device Name
| Trade Name: | F10 Fetal Doppler |
|----------------------|--------------------------------------------------|
| Common Name: | Monitor, Ultrasonic, Fetal |
| Classification Name: | Monitor, Ultrasonic, Fetal (21CFR 884.2660, KNG) |
| Classification: | Class 2 |
#### Predicate Device (Legally Marketed Devices)
The predicate devices for Fetal Doppler, Model F10 are:
- 0 Edan Instruments, Inc. Ultrasonic Pocket Doppler, Sonotrax cleared by FDA through 510(k) No. K040480 (Decision Date - May. 25. 2004), and
- 0 Bistos Co., Ltd. Ultrasonic Pocket Doppler, BT-200 cleared by FDA through 510(k) No. K05219 (Decision Date - Oct. 4. 2005)
#### Device Description
F10 is a pocket-size fetal Doppler that measures the fetal heat rate and outputs the fetal heart sound through built-in speaker. By measuring fetal heart rate(FHR), you are able to predict fetal well-being. F10 irradiates fetal wave to the abdomen of a pregnant woman to detect the Doppler frequency signal and analyze, and displays the heart rate on LCD screen. The device also provides the heart sound from the heart of fetus
{1}------------------------------------------------
#### Indications for Use
The F10 Fetal Doppler measures heart rate, which is displayed on a LCD display and provides fetal heart sounds. The fetal heart rate is measured using Doppler ultrasound.
#### Comparison with predicate device
Mediana Co., Ltd. believes that the F10 Fetal Doppler is substantially equivalent to the Sonotrax of Edan Instruments, Inc. and BT-200 of Bistos Co., Ltd.
The F10 Fetal Monitors described in this 510(k) submissions has the same intended use and similar technical characteristics as the Sonotrax of Edan Instruments, Inc., and the BT-200 of Bistos Co., Ltd..
- The similarities are as below;
'Mode of operation', 'System characteristics', 'FHR Monitoring', 'Transducer Type', 'Ultrasound Frequency', 'FHR Detection Method', 'FHR Range', 'Display method', and 'Intended use', etc.
- The differences are as below;
'Ultrasound Frequency' - Ultrasound frequency has a little difference. But there is not a difference to basic function.
'Battery type, voltage and life' – Battery type, voltage and life have a little difference. But it has not effect on efficiency, effectiveness and safety.
#### Safety, EMC and Performance Data
Electrical, mechanical, environmental safety and performance testing according to standard EN/IEC 60601-1 was performed, and EMC testing was conducted in accordance with standard EN/IEC 60601-1-2(2001). All test results were satisfactory.
#### Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Mediana Co., Ltd. concludes that F10 are safe and effective and substantially equivalent to predicate devices as described herein.
Mediana Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA.
12-A1-2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mediana Co., LTD. % Mr. Marc M. Mouser Engineering Leader & FDA Office Coordinator, Program Reviewer Underwriters Laboratories, Inc. 2600 NW Lake Road CAMAS WA 98607
JUN 2 4 2011
Re: KI10630
> Trade Name: F-10 Fetal Doppler Regulation Number: 21 CFR §884.2660 Regulation Name: Fetal Ultrasonic monitor and accessories Regulatory Class: II Product Code: KNG Dated: June 16, 2011 Received: June 16, 2011
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic For (110) that be device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply of any I outstar of and regulations, but not limited to: registration and listing (21 CFR Part with an the s requirements) in medical device reporting (reporting of medical device-related
{3}------------------------------------------------
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default:htm.
Sincerely yours,
Hubert Lenoir MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health.
Enclosure
{4}------------------------------------------------
## Indications for Use
KI10630 510(k) Number (if known):
Device Name: __ F10 Fetal Doppler
Indications for Use:
The F10 Fetal Doppler measures heart rate, which is displayed on a LCD display and provides fetal heart sounds. The fetal heart rate is measured using Doppler ultrasound.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Helene Kenny
rision Sign-Off) ductive, Gastro-Renal, and
{5}------------------------------------------------
## F10 Fetal Doppler
## 510(k) Submission
# K110630
Appendix F
# Diagnostic Ultrasound Indications For Use Form
Fill out one form for each ultrasound system and each transducer.
## 2MHz PW DOPPLER FETAL PROBE – MODEL : F10
Intended use : Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | N | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(specify) | | | | | | | | | | |
| Intraoperative | | | | | | | | | | |
| Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(specify) | | | | | | | | | | |
| Neonatal | | | | | | | | | | |
| Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Pheperial | | | | | | | | | | |
| Vascular | | | | | | | | | | |
| Lapaloscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; E=added under Appendix E
Additional Comments :
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
Prescription Use(Per 21 CFR 801.109)
Division Sign-Off ion of Reproductive. Gastro-Renal, and 510(k) Number