K Number
K102089

Validate with FDA (Live)

Date Cleared
2010-10-06

(62 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Classified by FDA's General Hospital and Personal Use Device panel as Class I, 21 CFR 880.6250, Power Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.

AI/ML Overview

This is not a medical device that utilizes AI/ML. The provided text is a 510(k) summary for "Powder Free Vinyl Patient Examination Gloves," which is a Class I medical device. Therefore, information related to AI/ML specific criteria such as multi-reader multi-case studies, effect size of human readers with AI assistance, standalone algorithm performance, or training set details is not applicable and cannot be extracted from this document.

However, I can provide the acceptance criteria and the studies that prove the device meets these criteria based on the provided text.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device PerformanceStudy Reference Section
Physical and DimensionsConformance to ASTM D-5250-06e1"All testing meets requirements for physical and dimensions testing conducted on gloves."Section 7
Freedom from Pinhole DefectsFDA 1000 ml. Water Fill Test requirements (AQL 2.5, inspection level I)"The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, meeting these requirements."Section 7
Biocompatibility - Primary Skin IrritationNo primary skin irritant reactions"Primary Skin Irritation...testing was conducted with results showing no primary skin irritant...reactions."Section 7
Biocompatibility - Skin SensitizationNo sensitization reactions (allergic contact dermatitis)"...Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no...sensitization reactions."Section 7
Residual PowderNo more than 2 mg powder per glove (based on ASTM D-6124-06)"...our gloves meet our 'powder-free' claims (contain no more than 2 mg powder per glove)."Section 7
General EquivalenceAs safe, as effective, and performs as well as or better than the predicate device (Better Care Plastic Products Co., Ltd Powder Free Vinyl Patient Examination Glove)"Grand Work Plastic Products Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves conform fully to ASTM D-5250-06e1 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the 'substantial equivalence' products cited. The device herein mentioned is as safe, as effective, and performs as well as or better than the predicate device."Section 10

2. Sample size used for the test set and the data provenance

  • Physical and Dimensions Testing: "Inspection level S-2, AQL 4.0." (Specific numerical sample size not provided, but AQL and inspection level are specified which determine the sample size. The data provenance is internal testing by the manufacturer.)
  • FDA 1000 ml. Water Fill Test: "samplings of AQL 2.5, inspection level I." (Specific numerical sample size not provided, but AQL and inspection level are specified. The data provenance is internal testing by the manufacturer.)
  • Primary Skin Irritation and Skin Sensitization: The document does not specify the sample size for human or animal subjects used in these tests. The testing was conducted, implying it's a prospective study for the purpose of this submission.
  • Residual Powder Test: The document does not specify the sample size. The data provenance is internal testing by the manufacturer.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This device is a physical product (examination gloves), and the "ground truth" is established through standardized physical, chemical, and biological testing rather than expert-based interpretations.

4. Adjudication method for the test set

Not applicable. The testing methods for physical properties, pinholes, and biocompatibility are objective measurements against established standards, not subjective interpretations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML powered device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm-based device.

7. The type of ground truth used

The "ground truth" for the device's performance is based on objective measurements against established national and international standards for medical gloves:

  • ASTM D-5250-06e1 (Standard Specification for Vinyl Patient Examination Gloves) for physical and dimensional properties.
  • FDA 1000 ml. Water Fill Test for barrier integrity (pinholes).
  • Primary Skin Irritation and Skin Sensitization testing for biocompatibility.
  • ASTM D-6124-06 for Residual Powder.

8. The sample size for the training set

Not applicable. There is no AI/ML algorithm or training set for this type of device. The "training" for manufacturing would involve adherence to quality control processes and standards.

9. How the ground truth for the training set was established

Not applicable. As there is no AI/ML algorithm or training set, this question is not relevant.

{0}------------------------------------------------

OCT 6 2010

EXHIBIT #1 Page 1 of 2

510(K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(K) number is: K102089

    1. Owner's Identification:
      Ms. Yuli Wu Grand Work Plastic Products Co., Ltd. Donggao Industrial Zone, Zanhuang, Hebei, CHINA 050000 Date Summary Prepared: June 4, 2010
  1. Name of the Device: Grand Work Plastic Products Co., Ltd.

Powder Free Vinyl Patient Examination Gloves

3. Predicate Device Information:

Better Care Plastic Products Co., Ltd Powder Free Vinyl Patient Examination Glove

4. Device Description:

Classified by FDA's General Hospital and Personal Use Device panel as Class I, 21 CFR 880.6250, Power Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.

5. Intended Use:

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

6. Comparison to Predicate Devices:

Grand Work Plastic Products Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves is substantially equivalent in safety and effectiveness to the Better Care Plastic Products Co., Ltd's Powder Free Vinyl Patient Examination Glove.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a sequence of handwritten characters, which appear to be alphanumeric. The sequence reads 'K102089'. The characters are written in a simple, clear style, with each digit and letter easily distinguishable. The image seems to be a close-up of the characters, possibly extracted from a larger document or label.

EXHIBIT #I Page 2 of 2

7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows:

The standards used for Grand Work Plastic Products Co., Ltd.'s glove production are based on ASTM D-5250-06e1. All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, meeting these requirements.

Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

A Residual Powder Test that based on ASTM D-6124-06 for powder at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).

8. Labeling:

There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.

9. Discussion of Clinical Tests Performed:

Not Applicable - There is no hypoallergenic Claim.

10. Conclusions:

Grand Work Plastic Products Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves conform fully to ASTM D-5250-06e1 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. The device herein mentioned is as safe, as effective, and performs as well as or better than the predicate device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Grand Work Plastic Products Company, Limited C/O Ms. Kathy Liu Gloveco Incorporated 3973 Schafer Avenue Chino, California 91710

CT 6 2010

Re: K102089 ·

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: June 4, 2010 Received: July 27, 2010

Dear Ms. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Ms. Liu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Attachment A

INDICATION FOR USE

K102089

510 (k) NUMBER (IF KNOW): APPLICANT: DEVICE NAME:

Grand Work Plastic Products Co., Ltd. Powder Free Vinyl Patient Examination Gloves

INDICATIONS FOR USE:

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/ OR

Over-The-Counter-Use _ 4 (21CFR 801Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

O.H. F. Cluverius-Will

Concurrent of CDRH, Office Dasser & eral Hospital Division of Anesthesiology

Division of Anoothoo
Infection Control, Dental Device

510(k) Number: .

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.