NON-STERILE POWDER-FREE BLUE NITRILE EXAMINATION GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUGS

K101822 · Sgmp Co., Ltd. · LZC · Nov 19, 2010 · General Hospital

Device Facts

Record IDK101822
Device NameNON-STERILE POWDER-FREE BLUE NITRILE EXAMINATION GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUGS
ApplicantSgmp Co., Ltd.
Product CodeLZC · General Hospital
Decision DateNov 19, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Intended Use

A disposable medical glove to be worn on the hand of the healthcare and similar personnel to prevent contamination between healthcare personnel and patient.

Device Story

Disposable, non-sterile, powder-free blue nitrile examination glove; worn by healthcare personnel to prevent cross-contamination between patient and examiner. Device tested for permeation resistance against various chemotherapy drugs per ASTM D6978-05. Input: physical barrier requirement during clinical procedures. Output: protective barrier for hands. Used in clinical settings; operated by healthcare professionals. Output informs clinical decision-making by providing a barrier against patient-to-provider contamination and specific chemotherapy drug exposure. Benefits include reduced risk of infection transmission and protection against hazardous drug contact, excluding specific agents with low breakthrough times.

Clinical Evidence

Bench testing only. No clinical data. Testing included ASTM D6319-00a-05 (physical properties/dimensions), ASTM D6124-06 (residual powder), and ASTM D6978-05 (chemotherapy drug permeation). Biocompatibility testing included primary skin irritation in rabbits and guinea pig maximization/sensitization tests. Breakthrough detection times provided for 11 chemotherapy drugs; Carmustine and Thiotepa showed low breakthrough times (0.49 and 2.61 minutes, respectively), while others exceeded 240 minutes.

Technological Characteristics

Nitrile rubber material. Powder-free. Blue color. Non-sterile. Dimensions and physical properties meet ASTM D6319-00a-05. Residual powder meets ASTM D6124-06. Chemotherapy resistance tested per ASTM D6978-05. No software or electronic components.

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs; however, not approved for use with Carmustine (BCNU) or Thiotepa due to low breakthrough detection times.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K101822 APPENDIX M ( k) SUMMARY 210 ( As Required by 21 section 807.92 (c) ) 1. Submitted For : SGMP Company Limited 181 Moo 6, Tambol Kampaengphet , Rattaphum , Songkhla 90180 Thailand, 2. Submitted By : Tucker & Associates Official Correspondent for SGMP Company Limited Janna P. Tucker , President - CEO 198 , Avenue de la D'emerald , Sparks, NV 89434-9550 Phone No : 775-342-2612 Fax No : 775-342-2613 Email : Tuckerjan@aol.com r # 3. Device Trade or Proprietary Name : Non-sterile , Powder-free Blue Nitrile Examination Gloves, Tested for use with Chemotherapy Drugs. - 4. Device Common Name : Examination Gloves ### 5. Device Classification Name : Patient Examination Gloves ( per 21CFR 880.6250) #### 6. Device Description : Non-sterile , Powder-free Blue Nitrile Examination Gloves , Tested for use with Chemotherapy Drugs. #### 7. Intended Use of Device : A disposable medical glove to be worn on the hand of the healthcare and similar personnel to prevent contamination between healthcare personnel and patient. ATCH 2 APPENDIX M revised 10-12-10, James Packer Now shows equivalence to K082957. NOV 1 9 2010 {1}------------------------------------------------ # K101822 The Non-sterile Powder Free Blue Nitrile Examination Gloves Tested for uses with The Non-sterile Powder Free Blue device intended for medical purposes that is world on the Chemotherapy Drugs , is a disposable device intended to models paper Chemotherapy Drug Permeation (Breakthrough Detection Time) in Minutes In Minutes of the March and with these gloves Chemotherapy Drug i cinicallon (Broadiness). The following chemical | Test Chemotherapy Drug | Average Breakthrough Detection Time (Minutes) | |----------------------------|-----------------------------------------------| | *Carmustine (BCNU) | 0.49 minutes | | Cisplatin | >240 minutes | | Cyclophosphamide (Cytoxan) | >240 minutes | | Doxorubicin Hydrochloride | >240 minutes | | Etoposide (Toposar) | >240 minutes | | Flurouracil | >240 minutes | | Paclitaxel (Taxol) | >240 minutes | | *Thiotepa | 2.61 minutes | | Vincristine Sulfate | >240 minutes | | Dacarbazine (DTIC) | >240 minutes | | Methotrexate | >240 minutes | *Please note that Carmustine (BCNU) and Thiotepa have extremely low Prease Bott that of 0.49 and 2.61 minute only. ﺮ {2}------------------------------------------------ KID/822 (1) = 1. ・ # · 8. Substantial Equivalence Discussion : . ... .. ﻨﻲ | Characteristic and<br>Parameter | SGMP Company Limited<br>Non-sterile , Powder-free Blue Nitrile<br>Examination Gloves.<br>Tested for use with Chemotherapy<br>Drugs | Siam Sempermed Corp.Ltd.<br>Non-sterile , Powder-free Nitrile<br>Examination Glove , Blue with<br>Polymer Coating, Tested for use<br>with Chemotherapy Drugs<br>510 K # K082957 | Substantial<br>Equivalence<br>(SE) | |---------------------------------|------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------| | Devise Class | 1 | 1 | SE | | Product Code | LZA | LZA | SE | | Glove Color | Blue | Blue | SE | | Dimensions | Meets ASTM D6319-00a-05 | Meets ASTM D6319-00a-05 | SE | | Physical<br>Properties | Meets ASTM D6319-00a-05 | Meets ASTM D6319-00a-05 | SE | | Freedom From<br>Pinholes | Meets ASTM D6319-00a-05 | Meets ASTM D6319-00a-05 | SE | | Powder-free | Meets ASTM D6124-06 | Meets ASTM D6124-06 | SE | | Biocompatibility<br>Test | Passes Primary Skin Irritation<br>in Rabbits | Passes Primary Skin Irritation<br>in Rabbits | SE | | | Passes Guinea Pig<br>Maximization | Passes Guinea Pig<br>Sensitization | SE | | Chemotherapy<br>Drugs Tests | Meets ASTM D6978-05<br>Cisplatin > 240 mins | Meets ASTM D6978-05<br>Cisplatin > 240 mins | SE | | | Cyclophosphamide > 240 mins<br>Doxorubicin Hydrochloride | Cyclophosphamide > 240 mins<br>Doxorubicin Hydrochloride | | | | > 240 mins<br>Etoposide > 240 mins | > 240 mins<br>Etoposide > 240 mins | | | | Flurouracil > 240 mins<br>Paclitaxel > 240 mins | Flurouracil > 240 mins<br>Paclitaxel > 240 mins | | | | Vencristine Sulfate > 240 mins<br>Dacarbazine > 240 mins | Vencristine Sulfate > 240 mins<br>Dacarbazine > 240 mins | | | | Methotrexate > 240 mins | Methotrexate > 240 mins | | . {3}------------------------------------------------ K101822 : . {4}------------------------------------------------ K101822 # CONCLUSION: The data presented indicates the Powder-Free, Blue Nitrile Examination Gloves, Tested I the una presence managers wabeling Claim, (Non-Sterile), K.101822 is equivalent 101 OSC With Channellapy Doles-free Nitrile Examination Gloves, Blue with Polymer Coating, Tested for use with Chemotherapy Drugs. It should be noted that testing for use with two chemotherapy drugs had extremely It should be loced use total use as follows: CARMUSTINE (BCNU) @ 0.49 minutes, and iow permeation unles. Therefore, these gloves are not approved for use when using those chemo drugs. These gloves do meet the following recognized standards unless otherwise noted: ASTM D6319-00a00A(2005), Standard Specification for Nitrile Gloves ISO 2859-1, Standard for Water Leak Test and/or ASTM D5151-06, Test Method for Detection of Holes in Medical Gloves ASTM D6124-06, Standard Test Method for Residual Powder on Medical Gloves. Biocompatibility Testing on White Rabbits and Guinea Pigs, Labeling meets FDA requirement Substantially equivalent to Siam Sempermed Corp. Ltd. K082957 ASTM D6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. {5}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular seal on the left and a stylized symbol on the right. The seal contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of a circle. The symbol on the right features a stylized human figure with three arms, representing the department's mission to protect the health of all Americans. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SGMP Company, Limited C/O Ms. Janna P. Tucker Tucker & Associates 198 Avenue De La D'Emerald Sparks. Nevada 89434 NOV 1 9 2010 Re: K101822 Trade/Device Name: Non-Sterile, Powder Free Blue Nitrile Examination Gloves Tested for use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA/LZC Dated: October 22, 2010 Received: October 21, 2010 Dear Ms. Tucker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # INDICATION FOR USE STATEMENT -00. | Applicant : SGMP Company Limited | | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------| | 510K NUMBER: K10 1822 | | | Device Name : Non-Sterile, Powder Free Blue Nitrile Examination Gloves , Tested for use with<br>Chemotherapy Drugs. | | | Indication For Use : | | | The Non-sterile Powder Free Blue Nitrile Examination Gloves Tested for use with<br>Chemotherapy Drugs , is a disposable device intended for medical purposes that is worn on the<br>examiner's hand or finger to prevent contamination between patient and examiner. | | | Chemotherapy Drug Permeation (Breakthrough Detection Time) in Minutes<br>The following chemical have been tested with these gloves. | | | Test Chemotherapy Drug | Average Breakthrough<br>Detection Time (Minutes) | | *Carmustine (BCNU) | 0.49 minutes | | Cisplatin | >240 minutes | | Cyclophosphamide (Cytoxan) | >240 minutes | | Doxorubicin Hydrochloride | >240 minutes | | Etoposide (Toposar) | >240 minutes | | Flurouracil | >240 minutes | | Paclitaxel (Taxol) | >240 minutes | | *Thiotepa | 2.61 minutes | | Vincristine Sulfate | >240 minutes | | Dacarbazine (DTIC) | >240 minutes | *Please note that Carmustine (BCNU) and Thiotepa have extremely low permeation times of 0.49 and 2.61 minute only. Prescription Use .................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................tabularLabels = []