K Number
K101630

Validate with FDA (Live)

Date Cleared
2010-09-07

(89 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. This device also provides protection against exposure to certain chemotherapy drugs.

Device Description

Powder Free Nitrile Examination Gloves (6Mil), Tested For Use With Chemotherapy Drugs (Blue); Sterile, Powder Free Nitrile Examination Gloves, Tested For Use With Chemotherapy Drugs (Blue)

AI/ML Overview

This document describes the 510(k) premarket notification for Powder Free Nitrile Examination Gloves (6Mil) and Sterile, Powder Free Nitrile Examination Gloves, both tested for use with chemotherapy drugs (Blue). The key information revolves around the gloves' resistance to permeation by specific chemotherapy drugs.

Here's an analysis based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" as a single numerical threshold for each drug, but rather presents the average breakthrough detection time for each chemotherapy drug. The implication is that a longer breakthrough time is better, with "> 240 minutes" being the best reported performance for many drugs, suggesting that a 240-minute duration without breakthrough is considered acceptable for practical use. The specific performance for some drugs which are noted as having "extremely low permeation time" for the Sterile glove indicates they do not meet a high standard of protection for those particular drugs.

Chemo Drug (Concentration)Acceptance Criteria (Implied)Powder Free Nitrile (6mil) Performance (Average Breakthrough Time in Minutes)Sterile, Powder Free Nitrile Performance (Average Breakthrough Time in Minutes)
Carmustine (BCNU) (3.3 mg / ml)Longer breakthrough time desired; >240 likely ideal111.90.51 (Noted as "extremely low")
Thiotepa (10.0 mg / ml)Longer breakthrough time desired; >240 likely ideal162.336.5 (Noted as "extremely low")
Fluorouracil (50.0 mg / ml)Longer breakthrough time desired; >240 likely ideal> 240> 240
Etoposide (Toposar) (20.0 mg / ml)Longer breakthrough time desired; >240 likely ideal> 240> 240
Cyclophosphamide (Cytoxan) (20.0 mg / ml)Longer breakthrough time desired; >240 likely ideal> 240> 240
Mitomycin C (0.5 mg / ml)Longer breakthrough time desired; >240 likely ideal> 240> 240
Methotrexate (25.0 mg / ml)Longer breakthrough time desired; >240 likely ideal> 240> 240
Paclitaxel (Taxol) (6.0 mg / ml)Longer breakthrough time desired; >240 likely ideal> 240> 240
Doxorubicin Hydrochloride (2.0 mg / ml)Longer breakthrough time desired; >240 likely ideal> 240> 240
Dacarbazine (DTIC) (10.0 mg / ml)Longer breakthrough time desired; >240 likely ideal> 240> 240
Cisplatin (1.0 mg / ml)Longer breakthrough time desired; >240 likely ideal> 240> 240
Mitoxantrone (2.0 mg / ml)Longer breakthrough time desired; >240 likely ideal> 240> 240
Ifosfamide (50.0 mg / ml)Longer breakthrough time desired; >240 likely ideal> 240> 240
Vincristine Sulfate (1.0mg / ml)Longer breakthrough time desired; >240 likely ideal> 240> 240

The acceptance criteria for glove performance against chemotherapy drugs generally follow standards like ASTM D6978, which specifies a minimum breakthrough time. While not explicitly stated as "acceptance criteria," the consistent reporting of "> 240 minutes" for many drugs suggests that 240 minutes is the target threshold for extended protection. The note concerning Carmustine and Thiotepa for the Sterile gloves indicates that their performance of < 30 minutes / 60 minutes is considered unacceptably low for prolonged exposure.

2. Sample Size Used for the Test Set and Data Provenance

The document does not provide details on:

  • The specific sample size (number of gloves or repetitions) used for testing each chemotherapy drug.
  • The data provenance (e.g., country of origin of the data, retrospective or prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not applicable to this type of device (nitrile examination gloves) and study. The "ground truth" for chemical permeation is established through laboratory testing using standardized methods (likely an ASTM standard, e.g., ASTM D6978 - Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs), not through expert consensus or interpretation of images.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication is typically relevant for studies involving human interpretation or subjective assessments. Chemical permeation testing is an objective laboratory measurement.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

An MRMC study is not applicable to this device. This type of study assesses the performance of human readers, often with and without AI assistance, in interpreting medical images or data. The current device is a physical product (gloves), and its performance is measured through objective chemical tests.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This concept is not applicable to this device. "Standalone performance" refers to the performance of an AI algorithm without human involvement. The current device is a physical product, not an AI or software algorithm. The study described is a laboratory test of its physical properties.

7. Type of Ground Truth Used

The ground truth used is based on objective laboratory measurements of chemical permeation, likely following a recognized standard such as ASTM D6978. The "ground truth" is the measured average breakthrough detection time of the chemotherapy drug through the glove material.

8. Sample Size for the Training Set

This information is not applicable. The study described is not a machine learning or AI-based study that would require a training set. It is a laboratory performance test of a physical product.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no training set for this type of device and study.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Cho Sow Fong Quality Assurance/ Regulatory Affairs Manager PT. Smart Glove Indonesia JL Raya Medan-Lubuk Pakam KM19, Tanjong Morawa, Deli Serdang Sumatera U Indonesia 20362

SEP 0 7 2010

Re: K101630

Trade/Device Name: Powder Free Nitrile Examination Gloves (6Mil), Tested For Use With Chemotherapy Drugs (Blue) Sterile, Powder Free Nitrile Examination Gloves, Tested For Use

Chemotherapy Drugs (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: August 23, 2010 Received: August 26, 2010

Dear Mr. Fong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2- Mr. Fong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing.(21.CFR Part 807); labeling (21 CER Part.801); medical device.reporting. (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:

  1. Powder Free Nitrile Examination Gloves (6Mil), Tested For Use With Chemotherapy Drugs (Blue)

  2. Sterile, Powder Free Nitrile Examination Gloves, Tested For Use With Chemotherapy Drugs (Blue)

Indication For Use:

This glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. This device also provides protection against exposure to certain chemotherapy drugs.

Chemotherapy Drugs Permeation (Average Breakthrough detection time in minutes)

Powder Free NitrileExamination Gloves (6mil)Sterile, Powder Free NitrileExamination Gloves
Carmustine (BCNU) (3.3 mg / ml)111.90.51
Thiotepa (10.0 mg / ml)162.336.5
Fluorouracil (50.0 mg / ml)> 240> 240
Etoposide (Toposar) (20.0 mg / ml)> 240> 240
Cyclophosphamide (Cytoxan) (20.0 mg / ml)> 240> 240
Mitomycin C (0.5 mg / ml)> 240> 240
Methotrexate (25.0 mg / ml)> 240> 240
Paclitaxel (Taxol) (6.0 mg / ml)> 240> 240
Doxorubicin Hydrochloride (2.0 mg / ml)> 240> 240
Dacarbazine (DTIC) (10.0 mg / ml)> 240> 240
Cisplatin (1.0 mg / ml)> 240> 240
Mitoxantrone (2.0 mg / ml)> 240> 240
Ifosfamide (50.0 mg / ml)> 240> 240
Vincristine Sulfate (1.0mg / ml)> 240> 240

Please note that the following drugs have extremely low permeation time of less than 30 minutes / 60 minutes (Carmustine and Thiotepa) for Sterile, Powder Free Nitrile Examination Gloves.

AND/OR Prescription Use (Part 21 CFR 801 Subpart D).

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office Of Device Evaluation (ODE

Elizabeth F. Claverie-Williams

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K101630

SEP 0 7 2010

「 ここで、 「 」 」

Page 1 of

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.