FIL 1520

K101629 · Glenn Medical Systems, Inc. · CAH · Jul 9, 2010 · Anesthesiology

Device Facts

Record IDK101629
Device NameFIL 1520
ApplicantGlenn Medical Systems, Inc.
Product CodeCAH · Anesthesiology
Decision DateJul 9, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5260
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

Glenn Medical oxygen concentrator filter model FIL 1520 is intended to help remove air-borne contaminants, including air borne bacteria and other particulate debris from the air stream of an oxygen concentrator. The device is only used for Invacare Oxygen Concentrator.

Device Story

Glenn Oxygen Concentrator Filter (model FIL 1520) functions as a mechanical filtration component for Invacare oxygen concentrators. Device installed in air stream to capture airborne contaminants, bacteria, and particulate debris. Operates via passive filtration; no active processing or electronic components. Used in clinical or home settings where oxygen concentrators are deployed. Benefits patient by reducing delivery of particulate matter and bacteria during oxygen therapy.

Clinical Evidence

Bench testing only.

Technological Characteristics

Mechanical bacterial filter for oxygen concentrator air stream. Class II device (21 CFR 868.5260). Passive filtration technology. No electronic components, software, or energy source.

Indications for Use

Indicated for use with Invacare Oxygen Concentrators to remove airborne contaminants, bacteria, and particulate debris from the oxygen concentrator air stream.

Regulatory Classification

Identification

A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its body and wings. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Glenn Medical Systems, Incorporated C/O Mr. Daniel W. Lehtonen Responsible Third Party Official 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087 JUL -- 9 2010 Re: K101629 Trade/Device Name: Glenn Oxygen Concentrator Filter, model FIL 1520 Regulation Number: 21 CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Filter Regulatory Class: II Product Code: CAH Dated: June 30, 2010 Received: July 1, 2010 Dear Mr. Lehtonen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Page 2- Mr. Lehtonen Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Wh for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ FDA-Filter-04 ## Chapter 04 PREMARKET NOTIFICATION Indications for Use 510(k) Number (if known): Device Name: Glenn Oxygen Concentrator Filter, model FIL 1520 Indications For Use: Glenn Medical oxygen concentrator filter model FIL 1520 is intended to help remove air-borne contaminants, including air borne bacteria and other particulate debris from the air stream of an oxygen concentrator. The device is only used for Invacare Oxygen Concentrator. | Prescription Use | √ | |-----------------------------|---| | (Part 21 CFR 801 Subpart D) | | | AND/OR Over-The-Counter Use | | | (21 CFR 801 Subpart C) | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER NEEDED) Concurrence of CDRH, Office of Device, Evaluation (ODE) (Division Sign-Off) (Division Sign-On) Division of Anesthesiology, General Hospital Division of Anesthesiology, General Devices Division of ontrol, Dental Devices 510(k) Number: K101624 Issuing Date: Code: FDA-001 Version: A/0 Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...