K Number
K101270

Validate with FDA (Live)

Date Cleared
2011-02-10

(280 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The flush syringe is intended for flushing of IV catheters and IV tubing. SwabCap® is intended for use on a swab-able luer access valve as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between the line accesses. SwabCap® will disinfect the valve five (5) minutes after application and act as a physical barrier to contamination for up to ninety-six (96) hours under normal conditions if not removed. The SwabFlush™ product is provided as: 1. 10 ml fill in 10 ml saline flush syringe with integrated SwabCap®. 2. 5 ml fill in 10 ml saline flush syringe with integrated SwabCap®. 3. 3 ml fill in 10 ml saline flush syringe with integrated SwabCap®.

Device Description

SwabFlush" is a saline intravenous flush syringe that is combined with an integrated SwabCap®. Each SwabFlush" syringe contains a sterile, non-pyrogenic isotonic solution of 0.9% Sodium Chloride Injection USP, 9 mg NaCl per ml, with an osmolarity of 0.3 l mOsm/mL, pH 4.5 - 7.0. A SwabCap® is a luer access valve cap that contains 70% isopropyl alcohol as a disinfectant. A SwabCap® is designed to securely fit on swab-able luer access valves. The product is intended for single-use and is provided sterile. The SwabFlush" provides the healthcare professional with immediate access to the SwabCap luer access valve disinfectant cap after an IV flush.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Excelsior Medical Corporation's SwabFlush™ device, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance CriterionReported Device Performance
Flush Syringe Component:
ISO 7886-1: Sterile Hypodermic Syringes for single use standardsMet all standards. (K101270 p.3)
ISO 594-1&-2: Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipmentMet all standards. (K101270 p.3)
Biocompatibility requirements per ISO 10993: Biological evaluation of medical devicesMet all requirements. (K101270 p.3)
Sterility Assurance LevelSterilization validations performed to achieve the desired sterility assurance level. (K101270 p.3)
Ability to hold and apply a SwabCap® without affecting disinfectant contentFunctions to hold and apply a SwabCap® while not affecting the disinfectant content. (K101270 p.3)
Saline flush compliance with 0.9% Sodium Chloride Injection USP monographVerified to comply. (K101270 p.3)
Shelf life stability (including sterility)Controlled aging performed to verify all product requirements, including sterility, throughout the shelf life. (K101270 p.3)
Packaging integrity (drop, vibration, compression)Ship testing performed on representative final packaged product for drop, vibration, and compression as specified by applicable ISTA or ASTM standards. All results met the established criteria. (K101270 p.3)
Syringe functional characteristics (e.g., injection force, liquid/air leakage)Meets the same requirements as predicate syringes for syringe functional characteristics. (K101270 p.2)
Integrated SwabCap® Component:
Disinfection of luer access valveWill disinfect the valve five (5) minutes after application. (K101270 p.2) (This performance is based on the predicate SwabCap® K083508, as the integrated SwabCap® is identical).
Physical barrier to contaminationActs as a physical barrier to contamination between line accesses for up to ninety-six (96) hours under normal conditions if not removed. (K101270 p.2) (This performance is based on the predicate SwabCap® K083508, as the integrated SwabCap® is identical).
Clinical Need (Overall Device)Design Validation confirmed that SwabFlush™ meets the clinical need expected for a typical flush procedure in a simulated clinical setting. (K101270 p.3)

Study Information

  1. Sample Size Used for the Test Set and Data Provenance:

    • The document mentions "All results met the criteria established in the testing" for non-clinical tests (K101270 p.3). However, specific numerical sample sizes for each test mentioned (e.g., ISO standards, biocompatibility, sterilization validations, ship testing, controlled aging) are not provided.
    • For the "Design Validation" to verify clinical need in a "simulated clinical setting," the sample size of participants or test runs is not specified.
    • Data Provenance: This information is not explicitly stated. Given it's a 510(k) submission, the studies would likely be conducted by the manufacturer, Excelsior Medical Corporation. It doesn't specify if the studies were retrospective or prospective, though "Design Verification" and "Design Validation" generally imply prospective testing. There's no mention of the country of origin of the data beyond the manufacturer's location in Neptune, NJ.
  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    • This information is not provided in the document. The studies mentioned are primarily non-clinical and simulated clinical setting evaluations. If expert assessment was used for "Design Validation," the number and qualifications of experts are not described.
  3. Adjudication Method for the Test Set:

    • This information is not provided. Given the nature of the tests (compliance with ISO standards, physical tests, chemical verification), adjudication methods like 2+1 or 3+1 typically used for subjective diagnostic interpretations are not applicable here. The "criteria established in the testing" (K101270 p.3) implies objective pass/fail metrics.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

    • No, an MRMC comparative effectiveness study was not done or described. The document focuses on demonstrating substantial equivalence to predicate devices through technological characteristics and non-clinical as well as simulated clinical testing of the device itself, rather than comparing human reader performance with and without AI assistance.
  5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Yes, the tests described (Design Verification and Design Validation) represent standalone performance of the device (SwabFlush™) itself. These tests evaluate the syringe's ability to meet various technical specifications and the combined device's ability to function as intended, independent of human operators beyond standard usage. However, it's not an "algorithm only" evaluation as the device is a physical product.
  6. The Type of Ground Truth Used:

    • The "ground truth" for the non-clinical tests was based on established objective standards and specifications:
      • ISO 7886-1 and ISO 594-1&-2 standards.
      • ISO 10993 for biocompatibility.
      • Defined Sterility Assurance Levels.
      • 0.9% Sodium Chloride Injection USP monograph requirements.
      • ISTA or ASTM standards for ship testing.
      • Internal product design requirements and established criteria.
    • For the "Design Validation," the ground truth was "the clinical need expected for a typical flush procedure in a simulated clinical setting" (K101270 p.3), which implies the device's functional performance under intended use conditions.
  7. The Sample Size for the Training Set:

    • As this is a physical medical device (syringe with an integrated cap) and not an AI or algorithm-based device, there is no "training set" in the conventional sense of machine learning. The device design and manufacturing processes are validated through the non-clinical and simulated clinical testing mentioned.
  8. How the Ground Truth for the Training Set Was Established:

    • Since there is no "training set" for an AI model, this question is not applicable. The device's design is based on established engineering principles, materials science, and medical device regulations, not on a data-driven training process. The "ground truth" for its development and testing aligns with the predefined standards and functional requirements outlined above.

{0}------------------------------------------------

K101270 p1of3

Excelsior Medical Corporation Traditional 510(k)

SwabFlush

510(k) Summary

FEB 1 n 2011

Manufacture Name:Excelsior Medical Corporation
Contact Name:John Linfante
Postal Address:1933 Heck AvenueNeptune, NJ 07753
Phone Number:732-643-6088
Fax:732-776-7600
Title:VP, Regulatory Affairs and QualityAssurance
Date:February 4, 2011
Device Proprietary Name:SwabFlush™
Device Common or Usual Name:Flush Syringe with Device DisinfectantCap
Classification Name:Saline Vascular Access Flush
Classification Code:NGT/LKB
Regulation Number:General Hospital

Predicate Devices:

Substantial equivalence is claimed to the following devices as related to intended use, technology and design characteristics:

  • . Excelsior Disposable Syringe w/ Normal Saline (0.9% Sodium Chloride), K962938 (Excelsior Medical Corporation)
  • Excelsior Saline pre-filled Syringe(s) in Sterile Field Packaging, K082837 . (Excelsior Medical Corporation)
  • SwabCap®, K083508 (Excelsior Medical Corporation) .

Description of the Device

SwabFlush" is a saline intravenous flush syringe that is combined with an integrated SwabCap®.

Each SwabFlush" syringe contains a sterile, non-pyrogenic isotonic solution of 0.9% Sodium Chloride Injection USP, 9 mg NaCl per ml, with an osmolarity of 0.3 l mOsm/mL, pH 4.5 - 7.0.

A SwabCap® is a luer access valve cap that contains 70% isopropyl alcohol as a disinfectant. A SwabCap® is designed to securely fit on swab-able luer access valves. The product is intended for single-use and is provided sterile.

The SwabFlush" provides the healthcare professional with immediate access to the SwabCap luer access valve disinfectant cap after an IV flush.

{1}------------------------------------------------

K101270 p.2 of 3

SwabFlush"

Intended Use/Indications for Use

The flush syringe is intended for flushing IV catheters and IV tubing. The integrated SwabCap® is intended for use on swab-able luer access valves as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between the line accesses. SwabCap® will disinfect the valve five (5) minutes after application and act as a physical barrier to contamination for up to ninety-six (96) hours under normal conditions if not removed.

Technological Characteristics

The SwabFlush" syringe has the same technological characteristics as the predicates. SwabFlush" syringe and predicate syringes are all single use, disposable prefilled saline syringes.

The SwabFlush" syringe is similar in design and uses the same types of materials as the Excelsior Medical Disposable Syringe with Normal Saline (K962938) and Saline prefilled Syringe(s) in Sterile Field Packaging (K082837). The SwabFlush" syringe, like the predicate syringes, has a standard male luer lock fitting which is intended to connect to standard female luer lock fittings, including luer access valves. The saline contained in the SwabFlush" syringe and predicate syringes meet the same requirements for 0.9% Sodium Chloride Injection, USP. Both the SwabFlush™ syringe and predicate syringes meet the same requirements for syringe functional characteristics such as, but not limited to, injection force and liquid/air leakage. The SwabFlush" syringe is filled, sterilized, and packaged using the same equipment and processes of the Disposable Syringe with Normal Saline (K962938). The SwabFlush " syringe and the Disposable Syringe with Normal Saline (K962938) are intended for the same user and indicated for the same use.

The integrated SwabCap® within the SwabFlush™ syringe is identical to the predicate device. There have been no changes to the intended use, design, materials or processing of SwabCap® as cleared under K083508.

Technological Modifications

The SwabFlush" has two unique technological modifications from the predicate devices. The first is a modification of the thumb press end of the plunger. Instead of a flat surface, the end has an open cup feature that allows a single fully assembled and packaged SwabCap® to be integrated with the syringe. The second modification is in the form of rounded recesses molded into the opening of the syringe barrel. This feature prevents a fully injected syringe plunger from rotating. This anti-rotational feature allows the fully injected syringe to become a tool for SwabCap® application and facilitates its installation to the luer access device.

Substantial Equivalence Discussion

The subject device is similar to the predicate devices based on intended use, technology and design. No changes have been made to the design of the SwabCap® as cleared by the FDA for integration into the syringe plunger.

{2}------------------------------------------------

K101270

SwabFlush™ p³-g

Non Clinical Testing Data to Demonstrate Safety

All of the non clinical testing was conducted in accordance to the requirements of design control. The following is a summary of the non clinical data that demonstrate the SwabFlush" syringe is safe and effective. It consists of the following:

DESIGN VERIFICATION

Excelsior has demonstrated that the SwabFlush"" syringe meets all ISO 7886-1: Sterile Hypodermic Syringes for single use standards and ISO 594-1&-2: Conical fittings with a 6 % (Luer) taper for syringes, needles and certain other medical equipment.

The device was confirmed to meet all biocompatibility requirements of ISO 10993: Biological evaluation of medical devices.

Sterilization validations were performed to validate the effectiveness of the sterilization methods to achieve the desired sterility assurance level.

Additional tests performed ensure the SwabFlush" syringe functions to hold and apply a SwabCap®, while not affecting the disinfectant content of the SwabCap®. The saline flush was verified to comply with 0.9% Sodium Chloride Injection USP monograph requirements. The verification also included the controlled aging of SwabFlush" to verify all product requirements including sterility throughout the shelf life. Ship testing was performed on a representative final packaged product for drop, vibration, and compression as specified by applicable ISTA or ASTM standards.

All results met the criteria established in the testing.

DESIGN VALIDATION

A Design Validation was done to verify that the SwabFlush™ meets the clinical need expected for a typical flush procedure in a simulated clinical setting.

Conclusion

Based on the information provided in this 510(k) premarket notification, SwabFlush" is substantially equivalent in terms of safety and effectiveness to the predicate devices identified above.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the HHS logo, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the logo.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Excelsior Medical Corporation C/O Mr. John Linfante Vice President Regulatory Affairs/Quality Assurance Excelsior Medical Corporation 1933 Heck Avenue Neptune. New Jersey 07753

FEB 1 0 2011

· Re: K101270

Trade/Device Name: SwabFlush™ Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: NGT Dated: January 25, 2011 Received: January 27, 2011

Dear Mr. Linfante:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. :

{4}------------------------------------------------

Page 2- Mr. Linfante

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices

Office of Device Evaluation Canter for Devices and Radiological Health

{5}------------------------------------------------

: : : ·

Indications for Use

510(k) Number:K101270
Device Name:SwabFlush™
Indication for Use:The flush syringe is intended for flushing of IV catheters and IV tubing.
SwabCap® is intended for use on a swab-able luer access valve as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between the line accesses. SwabCap® will disinfect the valve five (5) minutes after application and act as a physical barrier to contamination for up to ninety-six (96) hours under normal conditions if not removed.
The SwabFlush™ product is provided as:1. 10 ml fill in 10 ml saline flush syringe with integrated SwabCap®.2. 5 ml fill in 10 ml saline flush syringe with integrated SwabCap®.3. 3 ml fill in 10 ml saline flush syringe with integrated SwabCap®.
Prescription Use (21 CFR Part 801 Subpart D)X And/Or Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off

Anesthesiology, General Hospital, Infection Control and Dental Devices

510(k)

Signature

(Division Sign-Off)

Division of Anesthesiology, General Hospital
Infection Control, Dental Devices

Page 1 of 1

510(k) Number: K101270

、「

. '

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).