K Number
K100790

Validate with FDA (Live)

Date Cleared
2010-12-21

(274 days)

Product Code
Regulation Number
864.8625
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

COULTER® LIN-X Linearity Controls are intended to assess calibration and verify the reportable range of Coulter® Cellular Analysis Systems listed in the TABLE OF EXPECTED RESULTS in conjunction with specific COULTER reagents. Refer to Instructions for Use.

Device Description

COULTER® LIN-X Linearity Controls are stabilized human blood components whose white blood cells (WBC), red blood cells (RBC), hemoglobin (HGB) and platelets (PLT) concentrations span the instrument's reportable range. Results from repeated measurements for each concentration are compared to the established expected range to assess the instrument's calibration and verify the reportable range.

AI/ML Overview

Here's an analysis of the provided text regarding the COULTER® LIN-X Linearity Control, focusing on acceptance criteria and supporting studies.

Based on the provided K100790 submission for the COULTER® LIN-X Linearity Control, the device is a quality control product, not an AI or diagnostic device that would typically have performance metrics like sensitivity, specificity, or AUC for detecting a condition. Therefore, the acceptance criteria and study designs are geared towards demonstrating the stability and proper assignment of values for a control material, not diagnostic accuracy.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

For a linearity control, acceptance criteria are generally related to the control material maintaining its specified performance characteristics (e.g., stability, assigned values) over its shelf life and under various conditions. The "performance" is that it "demonstrated acceptable results" within those criteria.

Acceptance Criteria CategoryReported Device Performance SummarySpecific Study Information from Document
Open and Closed Vial StabilityAcceptable stability over shelf life."Evaluated open and closed vial stability of 3 lots of Level 6 of LIN-X Linearity Control over the shelf life of the product on a DxH™ 300 and DxH™ 300C... LIN-X Linearity Control demonstrated acceptable results."
Value Assignment ProcessEstablished process for generating assigned values."Value assignment of Level 6 WBC for each lot of LIN-X Linearity Control is determined on validated systems using specific COULTER reagents. Assigned Values are confirmed by multiple analysis of the control product."
Range Determination ProcessEstablished WBC expected range for Level 6."The expected range for WBC on Level 6 was calculated by a bio-statistician based upon the assigned values and expected ranges of the WBC parameters from Levels 1-5 and 7."

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Test Set Sample Size: For the stability study, it states "3 lots of Level 6". The specific number of individual measurements or vials tested within those lots is not provided. For the value assignment and range determination process, the text refers to "multiple analysis" but does not give a specific number of samples.
  • Data Provenance: The document does not explicitly state the country of origin of the data. Given the manufacturer (Beckman Coulter, Inc.) is based in Miami, Florida, USA, and the submission is to the FDA, it is highly probable the studies were conducted in the USA. The studies are prospective in nature, as they involve testing newly manufactured lots and establishing ranges/stability for a new or modified product.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • For the Range Determination Process, "The expected range for WBC on Level 6 was calculated by a bio-statistician". The specific number of biostatisticians or their qualifications (beyond being a "bio-statistician") are not provided.
  • For Value Assignment, the text indicates determination "on validated systems" and confirmation by "multiple analysis," implying internal expert processes rather than external expert consensus studies. No specific number or qualification of experts is listed in the typical sense of a clinical diagnostic study.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This type of adjudication method is not applicable to this device. The device is a linearity control, not a diagnostic tool requiring expert review of findings. The "adjudication" (if one could even call it that) comes from internal laboratory processes and statistical analysis by a biostatistician to establish expected ranges and confirm assigned values.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices (especially those involving image interpretation or complex decision-making, where human performance can be aided by AI). The COULTER® LIN-X Linearity Control is a laboratory quality control product, not a diagnostic or AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone algorithm performance study was not done. This device is a biochemical control, not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for a linearity control is its assigned value and established expected range. This is determined through:

  • Validated system measurements: Values are determined on "validated systems" using "specific COULTER reagents."
  • Multiple analysis: Assigned values are confirmed by "multiple analysis of the control product."
  • Statistical calculation: The expected range is calculated by a "bio-statistician" based on assigned values and expected ranges of other levels.

Essentially, the ground truth here is an empirically and statistically established reference value and range for the control material itself, rather than a clinical ground truth like pathology for a patient sample.

8. The sample size for the training set

This concept is not directly applicable. Linearity controls do not use "training sets" in the machine learning sense. The "training" for such a product involves:

  • Manufacturing process control: Ensuring consistent quality of the control material.
  • Extensive empirical testing: To establish stability and determine precise assigned values and expected ranges. The document mentions "3 lots" for stability, and "multiple analysis" for value assignment, but doesn't quantify this in terms of a "training set" size.

9. How the ground truth for the training set was established

As noted above, there isn't a "training set" in the conventional sense. The "ground truth" (assigned values and expected ranges) for the control is established through:

  • Standardized laboratory methods: Using validated instruments and COULTER reagents.
  • Repeat measurements: "multiple analysis" confirming assigned values.
  • Biostatistical analysis: A biostatistician calculates the expected range based on the empirically determined values and parameters from other known levels of the control. This is a scientific and statistical process for characterizing the properties of the control material itself.

{0}------------------------------------------------

K100790

510(k) Summary COULTER® LIN-X Linearity Control

1. Submitted By:

Lourdes Coba Staff Requlatory Affairs Specialist Beckman Coulter, Inc. 11800 SW 147 Avenue: M/C: 31-B06 Miami, Florida 33196-2500 Telephone: (305) 380-4079 FAX: (305) 380-4344

DEC 2 1 2010

2. Date Submitted:

March 19, 2010

3. Device Name(s):

3.1 Proprietary Names

COULTER® LIN-X Linearity Control

3.2 Classification Name

Hematology quality control mixture (21 CFR § 864.8625)

4. Predicate Device:

Candidate(s)PredicateManufacturerDocketNumber
COULTER® LIN-XLinearity Control(7 Levels)COULTER® LIN-XLinearity Control(12 Levels)Beckman Coulter, Inc.K081641

5. Description:

COULTER® LIN-X Linearity Controls are stabilized human blood components whose white blood cells (WBC), red blood cells (RBC), hemoglobin (HGB) and platelets (PLT) concentrations span the instrument's reportable range. Results from repeated measurements for each concentration are compared to the established expected range to assess the instrument's calibration and verify the reportable range.

Traditional 510(k) Submission Lin-X Linearity Control Beckman Coulter, Inc.

{1}------------------------------------------------

ે. Intended Use:

COULTER® LIN-X Linearity Controls are intended to assess calibration and verify the reportable range of Coulter® Cellular Analysis Systems listed in the TABLE OF EXPECTED RESULTS in conjunction with specific COULTER reagents. Refer to Instructions for Use.

7. Comparison to Predicate(s):

COULTER® LIN-X Linearity Control is essentially identical to the current LIN-X Linearity Control except that it only contains Levels 1-7 of the current product and Level 6 is additionally assayed for WBC. Levels 1-5 and Level 7 have not been modified.

8. Summary of Performance Data:

StudyStudy DesignStudy Results
Open andClosed VialStabilityEvaluated open and closed vial stabilityof 3 lots of Level 6 of LIN-X LinearityControl over the shelf life of the producton a DxH™ 300 and DxH™ 300C(pending 510(k) clearance K100489).The stability of Level 1-5 and Level 7 waspreviously cleared in K081641.LIN-X Linearity Controldemonstratedacceptable results.
ValueAssignmentProcessValue assignment of Level 6 WBC foreach lot of LIN-X Linearity Control isdetermined on validated systems usingspecific COULTER reagents. AssignedValues are confirmed by multipleanalysis of the control product.Established process forgenerating assignedvalues for LIN-XLinearity Control.
RangeDeterminationProcessThe expected range for WBC on Level 6was calculated by a bio-statistician basedupon the assigned values and expectedranges of the WBC parameters fromLevels 1-5 and 7.Established WBCexpected range forLevel 6 of LIN-XLinearity Control.

The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to products already in commercial distribution.

This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.

Traditional 510(k) Submission Lin-X Linearity Control Beckman Coulter, Inc.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of a human figure embracing a bird.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Beckman Coulter, Inc. c/o Ms. Lourdes Coba Staff Regulatory Affairs Specialist 11800 SW 147" Avenue Miami, FL 33196

DEC 2 1 2010

Re: K100790

Trade/Device Name: COULTER® LIN-X Linearity Control Regulation Number: 21CFR§864.8625 Regulation Name: Hematology quality control mixture Regulatory Class: Class II Product Code: JPK Dated: December 10, 2010 Received: December 13, 2010

Dear Ms. Coba:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice

{3}------------------------------------------------

Page 2 – Ms. Lourdes Coba

requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

enei m. chan

Maria M. Chan, Ph.D Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indication for Use

510(k) Number (if known):

DEC 2 1 2010

Device Name: COULTER® LIN-X Linearity Control

Indication For Use:

COULTER® LIN-X Linearity Controls are intended to assess calibration and verify the reportable range of Coulter® Cellular Analysis Systems listed in the TABLE OF EXPECTED RESULTS in conjunction with specific COULTER reagents. Refer to Instructions for Use.

Prescription Use × (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K100790

§ 864.8625 Hematology quality control mixture.

(a)
Identification. A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).(b)
Classification. Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.