AP-1000 BLOOD GLUCOSE MONITORING SYSTEM, MODEL AP-1000, MAJOR GLUCOSE CONTROL SOLUTION
K100437 · Bestgen Biotech Corp. · NBW · Jan 12, 2012 · Clinical Chemistry
Device Facts
| Record ID | K100437 |
| Device Name | AP-1000 BLOOD GLUCOSE MONITORING SYSTEM, MODEL AP-1000, MAJOR GLUCOSE CONTROL SOLUTION |
| Applicant | Bestgen Biotech Corp. |
| Product Code | NBW · Clinical Chemistry |
| Decision Date | Jan 12, 2012 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The AP-1000 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, calf, and thigh) for self testing by persons with diabetes. The AP-1000 Blood Glucose Monitoring Systems is intended to be used by a single person and should not be shared. The AP-1000 Blood Glucose Monitoring Systems is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The AP-1000 Meter is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, upper arm, calf, and thigh) for self testing by persons with diabetes. AP-1000 Blood Glucose Test Strips must be used with the AP-1000 Meter is intended to be used by a single person and should not be shared. AP-1000 Meter is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The AP-1000 Blood Glucose Test Strips are intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, upper arm, calf, and thigh) for self testing by persons with diabetes. AP-1000 Blood Glucose Test Strips must be used the AP-1000 Blood Glucose Meter. AP-1000 Blood Glucose Test Strips are intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The AP-1000multi Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, calf, and thigh). The AP-1000multi Blood Glucose Monitoring Systems is intended for testing outside the body (in vitro diagnostic use) and is intended for multi-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system is only used with single-use, auto-disabling lancing device. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The AP-1000multi Meter is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, upper arm, calf, and thigh). AP-1000multi Blood Glucose Test Strips must be used with the AP-1000multi Meter is intended for testing outside the body (in vitro diagnostic use) and is intended for multi-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The AP-1000multi Blood Glucose Test Strips are intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, upper arm, calf, and thigh). AP-1000multi Blood Glucose Test Strips must be used with the AP-1000multi Blood Glucose Test Strips are intended for testing outside the body (in vitro diagnostic use) and are intended for multi-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
Device Story
System comprises glucose meter, test strips, and control solutions. Input: fresh capillary whole blood sample applied to test strip. Principle: electrochemical measurement of glucose concentration. Output: digital glucose reading on meter display. AP-1000 intended for single-patient home self-testing; AP-1000multi intended for multi-patient professional use with auto-disabling lancing devices. Healthcare providers/patients use output to monitor diabetes control effectiveness. Benefits: provides rapid, quantitative glucose data to assist in diabetes management.
Clinical Evidence
Performance evaluated in laboratory and clinical settings by healthcare professionals and lay users. Results demonstrated suitability for intended use.
Technological Characteristics
Electrochemical glucose monitoring system. Includes meter, test strips, and control solutions. AP-1000multi requires single-use, auto-disabling lancing devices for professional multi-patient use.
Indications for Use
Indicated for quantitative blood glucose measurement in capillary whole blood (finger, palm, forearm, upper arm, calf, thigh). AP-1000 for single-patient home use; AP-1000multi for multi-patient professional healthcare use. Contraindicated for neonates, diabetes diagnosis, or screening.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- AP-1000 Blood Glucose Monitoring System (K090389)
Related Devices
- K103230 — AP-3000 BLOOD GLUCOSE MONITORING SYSTEM, MAJOR GLUCOSE CONTROL SOLUTION · Bestgen Biotech Corp. · Apr 24, 2012
- K123007 — VERI-Q MGD-1001 BLOOD GLUCOSE MONITORING SYSTEM, VERI-Q PLUS MGD-1001 BLOOD GLUCOSE MONITORING SYSTEM · Mico Biomed Co., Ltd. · Jun 19, 2013
- K132072 — TRUE METRIX SELF-MONITORING BLOOD GLUCOSE SYSTEM, TRUE METRI PRO PROFESSIONAL MONITORING BLOOD GLUCOSE SYSTEM · Nipro Diagnostics, Inc. · Aug 15, 2013
- K203562 — Finetest Lite Blood Glucose Monitoring System · Osang Healthcare Co. , Ltd. · Jul 13, 2021
- K133509 — EASY PLUS II BLOOD GLUCOSE MONITORING SYSTEM,EASY PLUS II MULTI BLOOD GLUCOSE MONITORING SYSTEM,SMARTEST PERSONA II BLOO · Biotest Medical Corp. · Feb 21, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
K100437
# 510(k) Summary
JAN 1 2 2012
、
:
## 1. Submitter Information
| Company name | Bestgen Biotech Corpration |
|----------------|--------------------------------|
| Contact person | Steven Shen |
| Address | 7F., No.186, Jian-Yi Rd., 235 |
| | Jung-He City, Taipei, Taiwan |
| Phone | +886-2-8227-2188 |
| FAX | +886-2-8227-2178 |
| E-mail | stevenshen@mail.bestgen.com.tw |
| Date prepared | Feb. 1st, 2010 |
## 2. Name of Device
:
| Trade name | AP-1000 and AP-1000multi Blood Glucose Monitoring System |
|---------------------|--------------------------------------------------------------------|
| Common name | Blood Glucose Monitoring System |
| | Blood Glucose Test Strips |
| Classification name | Class II devices, 21 CFR Section 862.1345, Glucose Test System |
| | Class I devices, 21 CFR Section 862.1660, Quality Control Material |
| Product Code: | CGA, NBW, JJX |
## 3. Predicate Device
| Trade name | AP-1000 Blood Glucose Monitoring System<br>Blood Glucose Monitoring System |
|---------------------|----------------------------------------------------------------------------|
| Common name | Blood Glucose Test Strip |
| Classification name | Class II devices<br>21 CFR Section 862.1345, Glucose Test System |
| 510(k) Number | K090389 |
{1}------------------------------------------------
#### 4. Device Description
The AP-1000 Blood Glucose Monitoring System consists of three main products: the meter, test strip, and control solutions. These products have been designed, tested, and proven to work together as a system accurate blood glucose test results. Use only AP-1000 test strips and MAJOR control solution with the AP-1000 Blood Glucose Monitoring System.
The AP-1000multi Blood Glucose Monitoring System consists of three main products: the meter, test strip, and control solutions. These products have been designed, tested, and proven to work together as a system accurate blood glucose test results. Use only AP-1000multi test strips and MAJOR control solution with the AP-1000multi Blood Glucose Monitoring System.
#### 5. Intended Use
#### 5.1 AP-1000 Blood Glucose Monitoring System
The AP-1000 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, calf, and thigh) for self testing by persons with diabetes. The AP-1000 Blood Glucose Monitoring Systems is intended to be used by a single person and should not be shared. The AP-1000 Blood Glucose Monitoring Systems is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
The AP-1000 Meter is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, upper arm, calf, and thigh) for self testing by persons with diabetes. AP-1000 Blood Glucose Test Strips must be used with the AP-1000 Meter is intended to be used by a single person and should not be shared. AP-1000 Meter is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
The AP-1000 Blood Glucose Test Strips are intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, upper arm, calf, and thigh) for self testing by persons with diabetes. AP-1000 Blood Glucose Test Strips must be used the AP-1000 Blood Glucose Meter. AP-1000 Blood Glucose Test Strips are intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
#### 5.2 AP-1000multi Blood Glucose Monitoring System
The AP-1000multi Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, calf, and thigh). The AP-1000multi Blood Glucose Monitoring Systems is intended for testing outside the body (in vitro diagnostic use) and is intended for multi-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system is only used with single-use, auto-disabling lancing device. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
The AP-1000multi Meter is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, upper arm, calf, and thigh). AP-1000multi Blood Glucose Test Strips must be used with the AP-1000multi Meter is intended for testing outside the body (in vitro diagnostic use) and is intended for multi-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
The AP-1000multi Blood Glucose Test Strips are intended for use in the quantitative measurement of glucose in
{2}------------------------------------------------
fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, upper arm, calf, and thigh). AP-1000multi Blood Glucose Test Strips must be used with the AP-1000multi Blood Glucose Test Strips are intended for testing outside the body (in vitro diagnostic use) and are intended for multi-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
#### 6. Comparison to Predicate Device
The AP-1000/AP-1000multi Blood Glucose Monitoring System (k100437) has equivalent technological characteristics as the predicate AP-1000 Blood Glucose Monitoring System (K090389). The two devices have same trade name, because of the new added AST function of k100437 device that is based k090389 device to re-test. There are no any changes between the two devices.
#### 7. Performance Studies
The performance of the AP-1000/AP-1000multi Blood Glucose Monitoring System was studied in the laboratory and in clinical settings by healthcare professionals and lay users. The studies demonstrated that the above blood glucose monitoring system is suitable for its intended use.
### 8. Conclusion
The AP-1000/AP-1000multi Blood Glucose Monitoring System demonstrates satisfactory performance and is suitable for its intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol of an eagle or bird with three wing-like shapes.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Bestgen Biotech Corporation c/o Steven Shen Quality Assurance Manager 7F., No. 186, Jian 1st Road, Jhonghe City Taipei County, Taiwan, 23511
JAN 1 2 2012
k100437 Re: AP-1000 Blood Glucose Monitoring System Trade/Device Name: AP-1000multi Blood Glucose Monitoring System Regulation Number: 21 CFR §862.1345 Regulation Name: Blood Glucose Test System Regulatory Class: Class H Product Code: NBW, CGA Dated: December 23, 2011 Received: January 6, 2012
Dear Mr. Shen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{4}------------------------------------------------
### Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and
Biometrics/Division of Postmarket Surveillance...
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely yours,
N
Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
#### 510(k) Number: k100437
### Device Name: AP-1000 Blood Glucose Monitoring System
#### Indications for Use:
The AP-1000 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in The All 2000 Blood Clubod from the alternative sites (palm, forearm, upper arm, calf, and Tresh Lapilaly while blood Tron the Tinger The AP-1000 Blood Glucose Monitoring Systems is intended to be thigh for sell testing by persons with albected. The AP-1000 Blood Glucose Monitoring Systems is intended for used by a single person and should not be shares. He by people with diabetes at home as aid to monitor the sell cosing outlike the body (in this ot intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
The AP-1000 Meter is intended for use in the quantitative measurement of glucose in fresh capillary whole blood The Ar-1000 Meter is Intended for ass in the quilm, forearm, upper arm, calf, and thigh for self testing by mom the finger and Trom the Shornearle Starts must be used with the AP-1000 Meter. AP-1000 Meter is persons with diabetes. Al 1000 blood Should not be shared. AP-1000 Meter is intended for self testing intended to be used by a single person and enome as aid to monitor the effectiveness of butside the body (in vitro dignosis of pospil if in a creening for diabetes mellitus, and is not intended for use on neonates.
The AP-1000 Blood Glucose Test Strips are intended for use in the quantitative measurement of glucose in fresh The Ar-1000 Blood from the finger and from the alternative sites (palm, forearm, upper arm, calf, and thigh for capillary whole blood wom the high the now tom the below the used the AP-1000 Blood Glucose Sell testing of persons with diaces intended for self testing outside the body (in vitro diagnostic use) by Metel. Ar-1000 blood Glood Glacose rest strips are man as a months and on the not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
| Prescription Use | |
|-----------------------------|---|
| (21 CFR Part 801 Subpart D) | |
| And/Or | |
| Over the Counter Use | V |
| (21 CFR Part 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
k100437 510(k)
Page 1 of 2-
{6}------------------------------------------------
## Indications for Use
#### 510(k) Number: _k100437
#### Device Name: AP-1000mutli Blood Glucose Monitoring System
#### Indications for Use:
The AP-1000multi Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, upper arm, calf, and thigh). The AP-1000muti Blood Glucose Monitoring Systems is intended for testing outside the body (in vitro diagnostic use) and is intended for multi-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system is only used with single-use, auto-disabling lancing device. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
The AP-1000multi Meter is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, upper arm, caff, and thigh). AP-1000multi Blood Glucose Test Strips must be used with the AP-1000multi Meter is intended for testing outside the body (in vitro diagnostic use) and is intended for multi-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
The AP-1000multi Blood Glucose Test Strips are intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and from the alternative sites (palm, forearm, upper arm, calf, and thigh). AP-1000multi Blood Glucose Test Strips must be used with the AP-1000multi Blood Glucose Test Strips are intended for testing outside the body (in vitro diagnostic use) and are intended for multi-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.
Prescription Use_ V (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use__________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Chief Officer
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
k100437 510(k)