K Number
K100211

Validate with FDA (Live)

Date Cleared
2010-05-11

(106 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. The device is single use only.

Device Description

The Pearlescent Black Textured Powder- Free Latex Examination Gloves, Coated with Aloe Vera at pH 5.0-5.5, with (Protein Content Labeling Claim) Contains 50 micrograms or less of Total Water Extractable Protein per gram which had been submitted and cleared under 510(k) number K013793. The difference in this submission is: a) Change of color to be Pearlescent Black.

AI/ML Overview

The provided document describes the acceptance criteria and performance of "Pearlescent Black Textured Powder-Free Latex Examination Gloves, Coated with Aloe Vera at pH 5.0-5.5, with Protein Content Labeling Claim". This is a medical device (examination glove) and the testing performed is for product specifications and biocompatibility, not for a diagnostic AI algorithm. Therefore, many of the requested categories (sample size for test/training sets, experts for ground truth, adjudication, MRMC studies, standalone performance, data provenance) are not applicable in this context.

Here's the relevant information that can be extracted:

1. Table of Acceptance Criteria and Reported Device Performance:

CharacteristicsAcceptance Criteria (Standards)Device Performance
DimensionsASTM D 3578-05e1Meets
Physical PropertiesASTM D 3578-05e1Meets
Freedom from pinholesASTM D 5151-99 (AQL)Meets (AQL 2.5)
Powder ResidualASTM D 6124-06Meets (<2mg/glove)
Protein LevelASTM D 5712-05Meets (<50 ug/g)
BiocompatibilityPrimary Skin Irritation in RabbitsPasses (Not a primary skin irritant)
BiocompatibilityDermal SensitizationPasses (Not a primary skin irritant)
pH of Aloe Vera CoatpH 5.0-5.5Confirmed pH 5.0-5.5

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not explicitly stated for each test, but the methods are based on ASTM standards (e.g., AQL for pinholes), which involve specific sampling plans. For biocompatibility, "Rabbits" are mentioned for Primary Skin Irritation, implying a biological study with an animal test set.
  • Data Provenance: Not explicitly stated, however, the testing is performed to meet US standards (ASTM) for FDA clearance. The document is a 510(k) summary submitted to the FDA in the USA.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable as this is a device performance test based on laboratory measurements and biological reactions, not expert interpretation of data.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable. The tests involve quantifiable measurements against established standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI algorithm. The device itself (the glove) is the "standalone" component being tested for its physical, chemical, and biological properties.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • The "ground truth" here is defined by objective, measurable criteria established by recognized national and international standards (ASTM). For example, a glove either meets the specified AQL for pinholes or it does not, based on the statistical sampling plan and water leak test. For protein level, it's a quantitative measurement of protein content verified against a threshold. For biocompatibility, it's the observed biological response (or lack thereof) in test subjects.

8. The sample size for the training set:

  • Not applicable. This is not an AI device that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable.

{0}------------------------------------------------

MAY 1 1 2010

K10021/5002

510(k) SUMMARY

1.0 Submitter

ﺎ ﻓ

NameShen Wei (USA) Inc.
Street Address33278 Central Ave., Suite 102
Union City, CA. 94587
Phone No.(510)429-8692
Fax No.(510)487-5347

Date of Summary Prepared: 1/20/2010

2.0 Contact Person:

Name:Mr. Albert T Li
Phone No.(510)429-8692
Fax No.(510)487-5347

3.0 Device Identification:

Glove Proprietary or Trade Name: (Multiple Names) Pearlescent Black Textured Powder- Free Latex Examination Gloves, Coated with Aloe Vera at pH 5.0-5.5, with Protein Content Labeling Claim (50 micrograms or less) Common Name: Exam gloves

Classification Name: Patient examination glove (per 21 CFR 880.6250) Classification Information: Class I Latex patient examination glove 80LYY, powder-free and meeting all the requirements of ASTM D 3578-05e1 Standard Specification for Latex Examination Gloves for Medical Application.

4.0 Identification of the Legally Marketed Device:

Class I latex patient examination gloves 80LYY, powder free and meeting all the requirements of ASTM D 3578-05e1 Standard Specifications for Latex Examination Gloves for Medical Application.

5.0 Description of the Device:

The Pearlescent Black Textured Powder- Free Latex Examination Gloves, Coated with Aloe Vera at pH 5.0-5.5, with (Protein Content Labeling Claim) Contains 50 micrograms or less of Total Water Extractable Protein per gram which had been submitted and cleared under 510(k) number K013793

1

{1}------------------------------------------------

The difference in this submission is:

  • a) Change of color to be Pearlescent Black. (MSDS Attachment C)
    Reference Documents:

  • a) SOP for Random Sampling of pH 5.0-5.5 Gloves to Ensure pH 5.0-5.5 Coating of Gloves (Attachment A)
  • b) Testing to Confirm pH 5.0-5.5 of gloves (Attachment B)
  • c) Characterization and Application of Aloe Vera at pH 5.0-5.5 (Attachment Four)

The modification of the device does not affect the intended use of the device as well it does not affect its safety and effectiveness. The indication for use and proposed labeling for the device are illustrated in subsequent sections

The Pearlescent Black Textured Powder- Free Latex Examination Gloves, Coated with Aloe Vera at pH 5.0-5.5, with Protein Content Labeling Claim meets all the requirements of ASTM D 3578-05e1 Standard Specification for Latex Examination Gloves for Medical Application.

6.0 Intended Use of Device:

A disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. The device is single use only.

7.0 Summary of Technological Characteristics for Modified Device:

The Pearlescent Black Textured Powder-Free Latex Examination Gloves, Coated with Aloe Vera at pH 5.0-5.5, with Protein Content Labeling Claim are summarized with the following technological characteristics compared to ASTM or equivalent standards.

CharacteristicsStandardsDevice Performance
DimensionsASTM D 3578-05e1Meets
Physical PropertiesASTM D 3578-05e1Meets
Freedom from pinholesASTM D 5151-99Meets (AQL 2.5)
Powder ResidualASTM D 6124-06Meets (<2mg/glove)
Protein LevelASTM D 5712-05Meets (<50 ug/g)
BiocompatibilityPrimary Skin Irritation inRabbitsPasses(Not a primary skinirritant)
Dermal SensitizationPasses(Not a primary skinirritant)

{2}------------------------------------------------

8.0 Conclusion:

The Pearlescent Black Textured Powder- Free Latex Examination Gloves, Coated with Aloe Vera at pH 5.0-5.5, with Protein Content Labeling Claim will perform according to the glove performance standards reference in section 7 above and meet ASTM standards, and FDA requirements for water leak test on pinhole AQL. Consequently, this device is substantially equivalent to currently marketed devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the circle is a stylized symbol consisting of three curved lines that resemble a person with outstretched arms.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Albert Li Project Manager Shen Wei (USA), Incorporated 33278 Central Avenue, Suite 102 Union City, California 94587

MAY 1 1 2010

Re: K100211

Trade/Device Name: Pearlescent Black Textured Powder-Free Latex Examination Gloves, Coated with Aloe Vera at pH 5.0-5.5, with Protein Content Labeling Claim (50 micrograms or less) Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: 1 Product Code: LYY Dated: April 26, 2010 Received: April 28, 2010

Dear Mr. Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Mr. Li

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Susan Runoi

$\xi$

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital,

Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K100211/5002

Attachment Two

INDICATION FOR USE

Shen Wei (USA) Inc. Applicant:

Device Name: Pearlescent Black Textured Powder- Free Latex Examination Gloves, Coated with Aloe Vera at pH 5.0-5.5, with Protein Content Labeling Claim (50 micrograms or less)

Indication For Use:

A disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

Prescription Use

Per 21CFR 801.109

Or Over-The Counter X (Optional Format 1-2-96)

signature

(Division Sign-Off) (Division Sign-on)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices

KI00211 510(k) Number: ر

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.