K Number
K100144

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-02-11

(388 days)

Product Code
Regulation Number
862.3840
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QMS® Everolimus assay is intended for the quantitative determination of everolimus, the active ingredient of Zortress®, in human whole blood on automated clinical chemistry analyzers. The results obtained are used as an aid in the management of kidney transplant patients receiving Everolimus therapy. This in vitro diagnostic device is intended for clinical laboratory use only.

The QMS Everolimus Calibrators set is intended for use in calibration of the QMS Everolimus Assay.

The QMS® Everolimus Controls set is intended for use in quality control of the QMS® Everolimus Assay.

Device Description

The QMS® Everolimus Assay system is a homogeneous assay utilizing particle agglutination technology and competitive binding principles.

AI/ML Overview

The acceptance criteria and study proving device performance are detailed below:

1. Table of Acceptance Criteria and Reported Device Performance

Test TypeAcceptance CriteriaReported Device Performance
Functional Sensitivity (LOQ)Not explicitly stated as a numerical threshold in the provided text.The QMS Everolimus Assay Functional Sentivity results in a C.V. of 20%. The claimed LOQ is 2.0 ng/mL.
PrecisionTotal run %CV less than or equal to 15.0%.In the study, the total run %CV was less than or equal to 15.0%.
Method ComparisonNot explicitly stated as a numerical threshold (e.g., specific correlation coefficient).Samples were tested in the Everolimus Assay and compared to LC/MS. The method comparison exhibited correlated well with LC/MS as follows: System 1 = 0.93x + 0.03 R2 = 0.94; System 2 = 1.00x - 0.08 R2 = 0.95.
Dilution Recovery (Linearity)Not explicitly stated as a numerical threshold for recovery or linearity.Results showed that the assay performs in a linear fashion.
On-Board Open Vial Reagent StabilityReagents are stable for up to 30 days when stored on the analyzer.Reagents stored on the analyzer are stable for up to 30 days.

2. Sample Size and Data Provenance

  • Test Set Sample Size: The document does not explicitly state the specific number of samples used for each performance test (e.g., Functional Sensitivity, Precision, Method Comparison, Dilution Recovery, Stability). For Method Comparison, it mentions "Samples were tested in the Everolimus Assay and compared to LC/MS."
  • Data Provenance: The document does not specify the country of origin of the data or whether the data was retrospective or prospective.

3. Number and Qualifications of Experts for Ground Truth

  • This information is not provided. The ground truth for comparative studies (e.g., Method Comparison) was established using LC/MS, which is an analytical method rather than relying on human expert consensus.

4. Adjudication Method

  • Not applicable as the ground truth for the Method Comparison was LC/MS, an analytical technique, not expert consensus requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This device is an in vitro diagnostic assay, and its performance is evaluated against reference analytical methods (like LC/MS) or established precision protocols, not against human reader performance.

6. Standalone Performance Study

  • Yes, a standalone performance study was done for the algorithm (assay). The performance metrics like Functional Sensitivity, Precision, Method Comparison, Dilution Recovery, and Stability are all assessments of the device's standalone analytical performance.

7. Type of Ground Truth Used

  • For the Method Comparison study, the ground truth was LC/MS (Liquid Chromatography/Mass Spectrometry), which is a gold standard analytical method for quantifying Everolimus in biological samples.

8. Sample Size for the Training Set

  • This information is not provided. The document describes an assay system, not a machine learning algorithm that requires a separate "training set" in the conventional sense. The "calibration" of the assay refers to establishing a standard curve with known concentrations, but the sample size for developing this calibration is not specified.

9. How the Ground Truth for the Training Set Was Established

  • For the assay's calibration (which can be considered analogous to training in a general sense for an analytical device), the QMS® Everolimus Assay's calibrators are "Traceable to calibration based on minimization of bias between the QMS assay and an LCMSMS assay for a specified adult renal transplant sample set." This means the ground truth for these calibrators was established by comparing them to results from an LC/MS/MS assay using a specific set of clinical samples.

{0}------------------------------------------------

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Devices Act of 1990 and 21 CFR 807,92

The assigned 510(k) number is: K100144

COMPANY/CONTACT PERSON

Lisa Charter Manager, Requlatory Affairs Microgenics Corporation Thermo Fisher Scientific, Clinical Diagnostics Division 46360 Fremont Blvd. Fremont, CA 94538-6406 (510) 979-5142 Office (510) 979-5422 Fax Lisa. Charter@ThermoFisher.com

DATE PREPARED

January 12, 2011

DEVICE NAME

Trade Name:Thermo Scientific QMS® Everolimus Assay
Thermo Scientific QMS® Everolimus Calibrators
Thermo Scientific QMS® Everolimus Controls
Common Name:QMS® Everolimus Assay System
Device Classification:Sirolimus test system
Regulation number:21 CFR 862.3840
Product Code:NRP

INTENDED USE

The QMS® Everolimus assay is intended for the quantitative determination of everolimus, the active ingredient of Zortress®, in human whole blood on automated clinical chemistry analyzers. The results obtained are used as an aid in the management of kidney transplant patients receiving Everolimus therapy. This in vitro diagnostic device is intended for clinical laboratory use only.

The QMS Everolimus Calibrators set is intended for use in calibration of the QMS Everolimus Assay.

The QMS® Everolimus Controls set is intended for use in quality control of the QMS® Everolimus Assay.

SUBSTANTIALLY EQUIVILANT PREDICATE DEVICE

Thermo Scientific QMS® Everolimus Assay is substantially equivalent to the previously cleared CEDIA Sirolimus Assay (K034069).

PRINCIPLE OF THE ASSAY

The QMS® Everolimus Assay system is a homogeneous assay utilizing particle agglutination technology and competitive binding principles.

{1}------------------------------------------------

In particle agglutination assays, the degree of agglutination (detected by optical method at 700m) is inversely proportional to the quantity of free drug in the reaction well. Henrio music pressming he on the particle which will cause it to agglutinate and will result in higher absorbance. If increased amount of competing drug is present in the sample, this will result in decreased binding of bound drug by the antibody, resulting in a relative decrease in particle agglutination. This in turn results in lowler absorbance.

The precise relationship between particle agglutination of the unlabeled drug in the sample is established by measuring the absorbance values of calibrators with known concentration of the The absorbance of unknown samples can be interpolated from the absorbance values of the drug. calibration curve and the concentration of the drug present in the sample can be calculated.

ComparisonPredicate DeviceThermo ScientificQMS® Everolimus Assay
CEDIA® Sirolimus Assay (K034069)
Intended UseThe CEDIA® Sirolimus Assay is an in vitrodiagnostic medical device intended for thequantitative determination of sirolimus in humanwhole blood using automated clinical chemistryanalyzers as an aid in the management ofsirolimus therapy in renal transplant patientstaking sirolimus.The QMS® Everolimus Assay is intended forthe quantitative determination of everolimus,the active ingredient of Certican®, in humanwhole blood on automated clinical chemistryanalyzers. The results obtained are used asan aid in the management of kidney transplantpatients receiving Everolimus therapy.
Test PrincipleThe CEDIA® Sirolimus Assay uses recombinantDNA technology (US Patent No. 4708929) toproduce a unique homogenous enzymeimmunoassay system.16 The assay is basedon the enzyme β-galactosidase, which hasbeen genetically engineered into two inactivefragments. These fragments spontaneouslyreassociate to form fully active enzymes that, inassay format, cleave a substrate, generating acolor change that can be measuredspectrophotometrically.In the assay, analyte in the specimen competeswith analyte conjugated to one inactivefragment of β-galactosidase for antibodybinding site. If analyte is present in the sample,it binds to antibody, leaving the inactive enzymefragments free to form active enzymes.If analyte is not present in the sample, antibodybinds to analyte conjugated to the inactivefragment, inhibiting the reassociation of inactiveβ-galactosidase fragments, and no activeenzyme is formed. The amount of activeenzyme formed and resultant absorbancechange are directly proportional to the amountof analyte present in the sample.The QMS® Everolimus Assay system is ahomogeneous assay utilizing particleagglutination technology and competitivebinding principles.In particle agglutination assays, the degree ofagglutination (detected by optical method at700nm) is inversely proportional to thequantity of free drug in the reaction well.Hence, if no drug is present in the sample, theantibodies in the QMS® Everolimus AntibodyReagent (R1) will bind only to the bound drugon the particle which will cause it toagglutinate and will result in higherabsorbance. If increased amount ofcompeting drug is present in the sample, thiswill result in decreased binding of bound drugby the antibody, resulting in a relativedecrease in particle agglutination. This in turnresults in lower absorbance. The preciserelationship between particle agglutination andconcentration of the unlabeled drug in thesample is established by measuring theabsorbance values of calibrators with knownconcentration of the drug. The absorbance ofunknown samples can be interpolated fromthe absorbance values of the calibration curveand the concentration of the drug present inthe sample can be calculated.

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

{2}------------------------------------------------

Sample MatrixHuman Whole BloodHuman Whole Blood
gentsTwo Lyophilized reagents (R1 and R2)One Liquid Ready-to-Use Reagent(Precipitation reagent)Three Liquid Ready-to-Use reagents(R1,R2 and Precipitation reagent)
Reagent storageCondition2-8°C2-8°C
CalibratorsLyophilized(0 and 30 ng/mL)Liquid ready-to-use(0, 1.5, 3.0, 6.0,12.0 and 20.0 ng/mL)
ControlsLyophilized(Low, Medium, High)Liquid ready-to-use(Low, Medium, High)
Calibrator andControl StorageCondition5°C ±3°C (2-8°C)-20°C ± 5°C (15-25°C)
CalibratorTraceabilityTraceable to purified sirolimusTraceable to calibration based on minimizationof bias between the QMS assay and anLCMSMS assay for a specified adult renaltransplant sample set.

PERFORMANCE TESTING SUMMARY

Functional Sensitivity (LOQ)

Functional Sensitivity determines the lowest concentration which results at a C.V. of 20% that has been measured over an extended period. The LOQ claim will be 2.0 ng/mL.

Precision and Accuracy

Everolimus samples tested for precision following a CLSI protocol. In the study, the total run %CV was less than or equal to 15.0%.

Method Comparison

Samples were tested in the Everolimus Assay and compared to LC/MS. The method comparison exhibited correlated well with LC/MS as follows: System 1 = 0.93x + 0.03 R2 = 0.94 System 2 = 1.00x - 0.08 R2 = 0.95

Dilution Recovery

Samples were tested to demonstrate linearity throughout the assay range. Results showed that the assay performs in a linear fashion.

On-Board Open Vial reagent Stability

Uncapped reagents were stored in an analyzer and all calibrators and controls were tested in the assay. Reagents stored on the analyzer are stable for up to 30 days.

CONCLUSION

As summarized, the Thermo Scientific QMS® Everolimus Assay is substantially equivalent to the CEDIA® Sirolimus Assay. Substantial equivalence has been demonstrated through performance testing to verify that the device functions as intended and that design specifications have been satisfied.

{3}------------------------------------------------

  • Brignol N, McMahon LM, Luo S, et al. High-throughput semi-automated 96-well liguid/liquid 7. extraction and liquid chromatography/mass spectrometric analysis of everolimus (RAD001) and cyclosporine A (CsA) in whole blood. Rapid Commun Mass Spectrom 2002; 15: 1-10.
  • Streit F, Armstrong VW, Oellerich M, et al. Rapid liquid chromatography-tandem mass spectrometry 8. routine method for simultaneous determination of sirolimus, tacrolimus, and cyclosporine A in whole blood. Clin Chem 2002; 48 (6): 955-958.
    1. Bablok W, Passing H, Bender R, Schneider B. A general regression procedure for method transformation. Application of linear regression procedures for method comparison studies in clinical chemistry. Part III J Clin Chem Clin Biochem 1988; 26 (11): 783-790.
    1. Tholen DW, Kallner A, Kennedy JW, et al. Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition (EP5-A2). Clinical and Laboratory Standards Institute (CLSI), Wayne, PA, 2004.
    1. McEnroe RJ, Burritt MF, Powers DM, et al. Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition (EP7-A2). Clinical and Laboratory Standards Institute (CLSI), Wayne, PA, 2005.

TRADEMARKS

Zortress® is a registered trademark of Novartis®. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.

Microgenics Corporation 46360 Fremont Blvd. Fremont, CA 94538-6406 USA

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol on the right. To the left of the symbol, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Image /page/4/Picture/4 description: The image shows the date February 1, 2011. The month is abbreviated as FEB. The day is the first day of the month. The year is 2011.

Microgenics Corporation c/o Ms. Lisa Charter Manager, Regulatory Affairs 46360 Fremont Blvd. Fremont, CA 94538

Re: K100144

Trade Name: Thermo Scientific QMS® Everolimus Reagents, Thermo Scientific QMS® Calibrators and Thermo Scientific QMS® Controls Regulation Number: 21 CFR 862.3840 Regulation Name: Sirolimus test system. Regulatory Class: Class II Product Codes: OUF, DLJ, LAS Dated: February 3, 2011 Received: February 4, 2011

Dear Ms. Charter:

.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device · reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2 –

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolli the (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

C.C.

Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indication for Use

510(k) Number (if known): K100144

Device Name:

Thermo Scientific QMS® Everolimus Reagents Thermo Scientific QMS® Everolimus Calibrators Thermo Scientific QMS® Everolimus Controls

Indication For Use:

The QMS® Everolimus assay is intended for the quantitative determination of everolimus, the active ingredient of Zortress® in human whole blood on automated clinical chemistry analyzers. The results obtained are used as an aid in the management of kidney transplant patients receiving Everolimus therapy. This in vitro diagnostic device is intended for clinical laboratory use only.

The QMS® Everolimus Calibrator set is intended for use in calibration of the QMS® Everolimus Assay.

The QMS® Everolimus Control set is intended for use in quality control of the QMS® Everolimus Assay.

And/Or

Prescription Use × (21 CFR Part 801 Subpart D) Subpart C)

Over the Counter Use (21 CFR Part 801

Page 1 of 1 .

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Carol C. Benson

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K100144

§ 862.3840 Sirolimus test system.

(a)
Identification. A sirolimus test system is a device intended to quantitatively determine sirolimus concentrations in whole blood. Measurements are used as an aid in management of transplant patients receiving therapy with sirolimus.(b)
Classification. Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Sirolimus Test Systems.” See § 862.1(d) for the availability of this guidance document.