GEM HEART ONE/ONE+

K092576 · Gem-Med S.L. · DPS · Sep 4, 2009 · Cardiovascular

Device Facts

Record IDK092576
Device NameGEM HEART ONE/ONE+
ApplicantGem-Med S.L.
Product CodeDPS · Cardiovascular
Decision DateSep 4, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2340
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

Gem Heart One/One+ is a standard 12 leads electrocardiograph designed to register, analyze and display electrocardiograms at rest. It is capable of displaying up to 12 leads at a time on the screen and print 3/6 channels by means of its thermal printer of 112mm width. It incorporates the Hannover Electrocardiogram System (HES) automatic interpretation program for cardiac signal analysis and has the possibility of introducing the patient's data, store the records and send them to a PC by serial communication (RS-232, USB and LAN). The interpretation by the Gem Heart One/One+ is only significant if it is considered and analysed by a qualified physician and used for patients over 16 years old. No therapy or drugs can be administered based solely on the interpretations statements obtained from the HES interpretation program. For non-interpretative applications, the Gem Heart One/One+ can be used for patients with no limitations with respect to age, sex and race of the patient. It is intended for use in routine ECG recording by trained physicians in hospital environment. - The Gem Heart One/One+ is a standard portable electrocardiograph that measures and records 12 channel electrocardiographic waveforms, measures ST segment level changes and gives real time heart rate measurements. - Intended use is equivalent to the intended use of the predicate Cardioline AR 1200 and CorScreen electrocardiograph.

Device Story

Portable 12-channel electrocardiograph; records, analyzes, and displays resting ECG waveforms. Inputs: 12-lead ECG signals via patient electrodes. Processing: Hannover ECG System (HES) software performs automatic cardiac signal analysis; measures ST segment changes and real-time heart rate. Outputs: ECG waveforms on high-resolution screen; 3/6 channel thermal printouts. Used in hospital environments by trained physicians. Connectivity: RS-232, LAN, USB 2.0 for data storage and PC transmission. Clinical utility: aids physicians in cardiac assessment; interpretation requires physician review; not for sole diagnostic or therapeutic decision-making.

Clinical Evidence

Bench testing only. Compliance with EN 60601-1, IEC 60601-2-25, and EN 60601-1-2 EMC standards demonstrated.

Technological Characteristics

Portable 12-lead ECG; 16-bit A/D conversion; 112mm thermal printer (16 dots/mm); lithium-ion battery; 240x320 pixel display. Connectivity: RS-232, LAN (10/100 Base T), USB 2.0. Integrated defibrillation protection. Software: Hannover ECG System (HES) for interpretation. Standards: IEC 60601-2-51:2003, EN 60601-1, IEC 60601-2-25, EN 60601-1-2.

Indications for Use

Indicated for routine resting ECG recording in hospital environments by trained physicians. Automatic interpretation program indicated for patients >16 years old. Non-interpretative ECG recording indicated for all patients regardless of age, gender, or race. Contraindicated for use as the sole basis for administering therapy or drugs.

Regulatory Classification

Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} SEP 04 2009 K092576 +1/4 # 6. 510(k) Summary Date of application: Applicant's name and address: Gem-Med S.L. Calle Cartagena 245, 5é, of. 5-6 08025, Barcelona (Spain) Contact Person: Mr. Wilson Tseng Tel. (+34) 935 309 650 Fax. (+34) 934 366 144 E-mail: wtseng@gem-med.com Date of submission: December, 15th 2008 Device Trade Name: Gem Heart One/One+ Device Common Name: Electrocardiograph Device Classification Name: CFR 870.2340 Electrocardiograph Product Code: DPS Class II | Predicate Device | Primary Notification | | --- | --- | | CARDIOLINE AR1200 | K051534 | | CORSCREEN | K070614 | ## Summary of Safety and Effectiveness | Parameter | Cardioline AR1200 | Gem Heart One | CorScreen | | --- | --- | --- | --- | | Input dynamic range | +/-300mV @ DC +/-10.0mV within the bandpass | +/-300mV @ DC +/-18.0mV within the band-pass | +/-300mV @ DC | | Frequency response | 0.05-150Hz (-3dB) | 0.05-150Hz | 0.05-150Hz | | A/D conversion | 12 bits | 16 bits | 24 bits | | Leads | 12 Standard / 12 Cabrera | 12 Standard / 12 Cabrera Orthogonal | 12 Standard | - 32 - {1} K092576 p2/4 SEP 04 2009 | Sensitivity | 2.5 5 10 20 mm/mV +/-5% | 2.5 5 10 20 mm/mV | 5 10 20 mm/mV | | --- | --- | --- | --- | | Writing system | Thermal head 108mm 8 dots/mm | Thermal head 110mm 16dots/mm - 8dots/mm | Ink Printer US letter and DIN-A4 size | | Printed channels | 3/4/6 | 1/3/6 | 1/2/6/12 | | Paper speed | 5mm/s +/-10% 50mm/s +/-5% | 10 25 50 mm/s | 25 50 mm/s | | Thermal paper | DOTCARD 120mm | DOTCARD 120mm | US letter and DIN-A4 | | Mode of operation | Manual and Automatic recording | Manual and Automatic recording | Manual | | Input/output | Infrared digital interface | RS 232 Ethernet USB | SD Memory card | | DISPLAY | AR 1200 | Gem Heart One | CG Screens | | Size | 120x32 pixels / 240x320 pixels | 240X320 pixels | 240X320 pixels | | N° of displayed channels | 3/6 | 1/3/6/12 | 1/3/6/12 | | Traces speed | 12.5 25 50 mm/s | 10 25 50 mm/s | 5 10 25 50 mm/s | | Sensitivity | 5 10 20 mm/mV | 2.5 5 10 20 mm/mV | 5 10 20 40 mm/mV | ## Device description The Gem Heart One/One+ is a portable electrocardiograph of 12 channels with simultaneous leads recording, with a high quality thermal printing of 112 mm, a lithium ion battery, high resolution screen, and connectivity through serial port (RS-232), local network LAN (10/100 Base T) and USB 2.0. The device includes protection against defibrillation integrated. Provides the automatic interpretation based on the program Hannover Ecg System HES® developed by Biosigna GmbH under the standards IEC 60601-2-51:2003. Uses the format SCP (Standard Communications Protocol) to import and export the ECG records in the framework of the CSE (Common Standards for Quantitative Electrocardiography) of the European Union. - 33 - {2} K092576 P 3/4 SEP 04 2009 The Gem Heart One/One+ can operate in Spanish, Catalan and English and other languages can be implemented. It includes downloading and ECG viewing PC software gem heart viewer. ## Intended use Gem Heart One/One+ is a standard 12 leads electrocardiograph designed to register, analyze and display electrocardiograms at rest. It is capable of displaying up to 12 leads at a time on the screen and print 3/6 channels by means of its thermal printer of 112mm width. It incorporates the Hannover Electrocardiogram System (HES) automatic interpretation program for cardiac signal analysis and has the possibility of introducing the patient's data, store the records and send them to a PC by serial communication (RS-232, USB and LAN). The interpretation by the Gem Heart One/One+ is only significant if it is considered and analysed by a qualified physician and used for patients over 16 years old. No therapy or drugs can be administered based solely on the interpretations statements obtained from the HES interpretation program. For non-interpretative applications, the Gem Heart One/One+ can be used for patients with no limitations with respect to age, sex and race of the patient. It is intended for use in routine ECG recording by trained physicians in hospital environment. - The Gem Heart One/One+ is a standard portable electrocardiograph that measures and records 12 channel electrocardiographic waveforms, measures ST segment level changes and gives real time heart rate measurements. - Intended use is equivalent to the intended use of the predicate Cardioline AR 1200 and CorScreen electrocardiograph. ## Summary of Device Testing The Gem Heart One/One+ has been tested under the same standards to its predicate device the Cardioline AR1200 and CorScreen, in accordance with the EN 60601-1 and IEC 60601-2-25, as well as for EMC tests EN60601-1-2 standards. The device has completely passed the tests and has shown full compliance to the standards. - 34 - {3} K092576 +4/4 SEP 04 2009 # Conclusions The intended use, indication for use and principle of operation are similar to the predicate devices with some distinct features that have been previously cleared which do not raise any new types of safety and effectiveness questions. Furthermore it has been demonstrated that the Gem Heart One/One+ has shown to have passed the same safety and effectiveness standards to its predicate device, Cardioline AR1200 and the CorScreen. Based on the performance and characteristics of the Gem Heart One/One+ described above, it is demonstrated that it is substantially equivalent to the predicate device Cardioline AR 1200 and CorScreen electrocardiographs. - 35 - {4} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SEP - 4 2009 Gem-Med S. L. c/o Mr. Casey Conry Underwriters Laboratories, Inc. 1285 Walt Whitman Road Melville, NY 11747 Re: K092576 Trade/Device Name: Gem Heart One/One+ Regulatory Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (Two) Product Code: DPS Dated: August 19, 2009 Received: August 21, 2009 Dear Mr. Conry: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5} Page 2 - Mr. Casey Conry Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health’s (CDRH’s) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ![img-0.jpeg](img-0.jpeg) Erichsure {6} K092576 # Indication for Use 510(k) Number (if known): Device Name: Gem Heart One/One+ ## Indications for Use: Gem Heart One/One+ is a standard 12 leads electrocardiograph designed to register, analyze and display electrocardiograms at rest. It is capable of displaying up to 12 leads at a time on the screen and print 3/6 channels by means of its thermal printer of 112mm width. It incorporates the Hannover Electrocardiogram System (HES) automatic interpretation program for cardiac signal analysis and has the possibility of introducing the patient's data, store the records and send them to a PC by serial communication (RS-232, USB and LAN). The interpretation by the Gem Heart One/One+ is only significant if it is considered and analysed by a qualified physician and used for patients over 16 years old. No therapy or drugs can be administered based solely on the interpretations statements obtained from the HES interpretation program. For non-interpretative applications, the Gem Heart One/One+ can be used for patients with no limitations with respect to age, sex and race of the patient. It is intended for use in routine ECG recording by trained physicians in hospital environment. - The Gem Heart One/One+ is a standard portable electrocardiograph that measures and records 12 channel electrocardiographic waveforms, measures ST segment level changes and gives real time heart rate measurements. - Intended use is equivalent to the intended use of the predicate Cardioline AR 1200 and CorScreen electrocardiograph. Prescription Use ☑ (21 CFR Part 801 Subpart D) And/Or Over the Counter Use ☐ (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) ![img-1.jpeg](img-1.jpeg) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K092576 Page 1 of 1
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%