AWBAT-S PLUS, AWBAT-D PLUS, AWBAT-M PLUS
K091863 · Aubrey, Inc. · FRO · Jan 21, 2010 · SU
Device Facts
| Record ID | K091863 |
| Device Name | AWBAT-S PLUS, AWBAT-D PLUS, AWBAT-M PLUS |
| Applicant | Aubrey, Inc. |
| Product Code | FRO · SU |
| Decision Date | Jan 21, 2010 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Intended Use
AWBAT-S Plus: is intended for clean Superficial burn wounds. AWBAT-D Plus: are intended for Donor sites after hemostasis has been established. AWBAT-M Plus: are intended to be used as a protective covering for Meshed autografts.
Device Story
Temporary wound dressing for coverage of superficial burns, donor sites, and meshed autografts. Device consists of knitted nylon fabric bonded to a thin porous silicone membrane. Nylon side coated with non-toxic mixture of porcine collagen peptides, chondroitin-4-sulfate, chondroitin-6-sulfate, Aloe polysaccharides (Immuno-10), vitamin C, vitamin E, and polysorbate 20. Applied by physicians in hospital settings. Provides moist wound healing environment. Available in various sizes and configurations (S, D, M) with varying porosity levels to manage wound exudate. Sterile, single-use device.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility testing per ISO 10993-1:2003.
Technological Characteristics
Knitted nylon fabric (15/2 or 12/1 denier) bonded to porous silicone membrane (~0.001 inch). Coated with porcine collagen, chondroitin sulfates, Aloe polysaccharides, vitamins C/E, and polysorbate 20. Sterilized via electron beam radiation. Shelf life: 3 years. Porosity ranges from 6.0% to 12.0%.
Indications for Use
Indicated for patients with superficial burn wounds (AWBAT-S Plus), donor sites after hemostasis (AWBAT-D Plus), or as a protective covering for meshed autografts (AWBAT-M Plus). For prescription use only.
Predicate Devices
Reference Devices
Related Devices
- K082869 — AWBAT-S, AWBAT-D, AWBAT-M · Aubrey, Inc. · Feb 6, 2009
- K242146 — BIOBRANE Temporary Wound Dressing; BIOBRANE Glove · Smith & Nephew Medical Limited · Dec 17, 2024
- K973170 — SIS WOUND DRESSING · Cook Biotech, Inc. · Apr 30, 1998
- K070269 — MODIFICATION TO COLLAWOUND DRESSING · Collamatrix Co., Inc. · Mar 2, 2007
- K090894 — COLLAWOUND WOUND DRESSING · Collamatrix Co., Inc. · Jun 25, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K091863
page 1 of 6
AWBAT Plus Wound Dressings Traditional 510(k)
## SECTION 5: 510(k) Summary
## JAN 9 1 2010
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 5807.92.
### 5.1 Name, Address, Phone and Fax Number of the Applicant
Aubrey Inc.
5930 Sea Lion Place, Suite 100
Carlsbad, CA 92010
Phone: (760) 602-8303
Fax: (760) 602-8304
#### Contact Person 5.2
Aubrey Woodroof, PhD, MBA Chairman and CEO Aubrey Inc. 5930 Sea Lion Place, Suite 100 Carlsbad, CA 92010 Phone: (760) 602-8303 Fax: (760) 602-8304
5.3 Date Prepared
22 June 2009
5.4 Device
Name: AWBAT Plus
Trade Names: AWBAT-S Plus, AWBAT-D Plus, AWBAT-M Plus Common Name: Wound Dressing Collagen Classification Name: Unclassified Product Code: FRO
CONFIDENTIAL
{1}------------------------------------------------
K091863
page 2 of 6
A WBAT Plus Wound Dressing Traditional 510(k)
### 5.5 Predicate Device
Name: AWBAT
Trade Names: AWBAT-S, AWBAT-D, AWBAT-M
Common Name: Wound Dressing, Collagen
Classification Name: Unclassified
Product Code: KGN
#### 5.6 Device Description
A Temporary Wound Dressing for coverage of Superficial burns, Donor sites and Meshed autografts until healing occurs. AWBAT Plus is composed of a single or multiple filesned knitted nylon fabric, bonded to a thin porous silicone membrane (enproximately 0.01 inch). The nylon side of these components is coated with a non-toxic mixture of porcine collagen. peptides, chondroitin-4-sulfate, chondroitin-6-sulfate, Immuno-10 (Aloe polysiaconarides), vitamin C, vitamin E, and polysorbate 20.
AWBAT Plus is manufactured in 12"x12" sheets that are cut to sizes of 6"x6", 6"x12" and 12"x12" and ink stamped AWBAT-S Plus, AWBAT-D Plus, AWBAT-M Plus accordingly.
AWBAT-S Plus has pores in the silicone membrane at 1/4" centers and the approximate area of an AWBAT-S Plus pore is 0.0036 square inch each.
AWBAT-D Plus has pores in the silicone membrane at 1/4" centers and the approximate area of an AWBAT-D Plus pore is 0.0073 square inch each.
AWBAT-M Plus has pores in the silicone membrane at 1/4" centers and the approximate area of an AWBAT-M Plus pore is 0.0070 square inch each.
The following table addresses the Design and Use of the Device:
| Question | Yes | No |
|-----------------------------------------------------------------------------------------|-------|-----|
| Is the intended device for prescription use? | Yes* | |
| Is the intended device for over-the-counter use? | | No |
| Does the device contain components derived from a tissue or other<br>biological source? | Yes** | |
| Is the device provided sterile? | Yes | |
| Is the device intended for single use? | Yes | |
| Is the device a reprocessed single use device? | | No |
| If yes, does this device type require reprocessed validation data? | | N/A |
| Does the device contain a drug? | | No |
| Does the device contain a biologic? | Yes** | |
| Does the device use software? | | No |
### Table 1: Design and Use of the Device
{2}------------------------------------------------
K091863
page 3 of 6
AWBAT Plus Wound Dressing Traditional 510(k)
| Question | Yes | No |
|---------------------------------------------------|-----|----|
| Does the submission include clinical information? | | No |
| Is the device implanted? | | No |
*AWBAT is intended for application by a physician.
** AWBAT contains biological components: porcine type 1 collagen (gelatin), chondroitin sulfate, Immuno-10 (Aloe polysaccharides), vitamin C and vitamin E.
AWBAT S Plus is composed of a 15/2 denier multiple filament nylon knitted fabric, bonded to a thin porous sillcone membrane. The nylon side of these components is coated with a non-toxic mixture of porcine collagen peptides, chondroitin-4-sulfate, chondroitin-sulfate, lmmuno-10 (Aloe polysaccharides), vitamin C, vitamin E, and polysorbate 20.
AWBAT-D Plus has the same composition as AWBAT-S Pius, except it is more porous to enable the clinician to evacuate a clot through the membrane into an outler sterior wasn
for is is m . . Si fabric is monofilament (thinner & less adherent).
AWBAT Plus, like AWBAT, can be manufactured in flat sheets or shaped into gloves, socks or masks.
#### 5.7 Device Intended Use
Temporary wound dressing for coverage of Superficial burns, Donor sites and Meshed autografts.
AWBAT-S Plus: is intended for clean Superficial burn wounds.
AWBAT-D Plus: is intended for Donor sites after hemostasis has been established.
AWBAT-M Plus: is intended to be used as a protective covering for Meshed autografts.
# 5.8 Technological Characteristics of Device Compared to Predicate Device
The following Table describes the comparative characteristics of AVBAT Plus and AVBAT.
| Characteristic | AWBAT Plus | Predicate(s) |
|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use/Indications for<br>Use | AWBAT-S Plus is intended for<br>clean, Superficial burn wounds. | Same intended use as AWBAT-<br>S standard (K082869). |
| | AWBAT-D Plus is intended for<br>Donor sites after hemostasis<br>has been established. | Same intended use as AWBAT-<br>D standard (K082869). |
| Characteristic | AWBAT Plus | Predicate(s) |
| | AWBAT-M Plus is intended for<br>use as a protective covering for | Same intended use as AWBAT-<br>M standard (K082869). |
| | Meshed autografts. | |
| Target Population | Burn Victims (AWBAT-S Plus).<br>Patients who have had skin<br>removed (donor site) to apply to<br>another body site (AWBAT-D<br>Plus). | Same as AWBAT-S standard<br>(K082869).<br>Same as AWBAT-D standard<br>(K082869). |
| | Patients who have had skin<br>grafts (AWBAT-M Plus) | Same as AWBAT-M standard<br>(K082869). |
| Where Used | Hospital | Same as AWBAT standard<br>(K082869). |
| Design | Thin porous silicone membrane,<br>mechanically bonded to a<br>knitted bifilament nylon fabric,<br>coated with a non-toxic mixture<br>of porcine collagen peptide, and<br>hydrophilic components (see<br>materials, below) (AWBAT-S<br>Plus and AWBAT-D Plus) | Similar to AWBAT-S and<br>AWBAT-D standard (K082869):<br>Thin porous silicone membrane<br>mechanically bonded to a<br>knitted bifilament nylon fabric,<br>coated with a non-toxic mixture<br>of porcine collagen peptide. |
| | Thin porous silicone membrane,<br>mechanically bonded to a<br>knitted monofilament nylon<br>fabric, coated with a non-toxic<br>mixture of porcine collagen<br>peptide, and hydrophilic<br>components (see materials,<br>below) (AWBAT-M Plus) | Similar to AWBAT-M standard<br>(K082869): Thin porous silicone<br>membrane, mechanically<br>bonded to a knitted<br>monofilament nylon fabric,<br>coated with a non-toxic mixture<br>of porcine collagen peptide. |
| Materials: | | |
| Nylon | Nylon fabric made from 15/2<br>denier yarn (AWBAT-S Plus<br>and AWBAT-D Plus) | Same as AWBAT-S and<br>AWBAT-D standard (K082869) |
| | Nylon fabric made from 12/1<br>denier yarn (AWBAT-M Plus) | Same as AWBAT-M standard<br>(K082869) |
| Silicone | Silicone (approx 0.001 inch) | Same as AWBAT standard<br>(K082869) |
| Collagen | Porcine gelatin | Same as AWBAT standard<br>(K082869) |
| Chondroitin sulfate | Chondroiting-4-sulfate and<br>chondoitin-6-sulfate | Chondroitin-6-sulfate same as<br>Integra (P900033) |
| Immuno-10, Vitamin C,<br>vitamin E and Polysorbate 20 | Immuno-10, Vitamin C,<br>vitamin E and Polysorbate 20 | None |
| Porosity | 6.0% (AWBAT-S Plus)<br>12.0% (AWBAT-D Plus) | Same as AWBAT-S standard<br>(K082869)<br>Same as AWBAT-D standard<br>(K082869) |
| | 11.0% (AWBAT-M Plus) | Same as AWBAT-M standard<br>(K082869) |
| Occlusiveness | 94.0% (AWBAT-S Plus) | Same as AWBAT-S standard<br>(K082869) |
| Characteristic | AWBAT Plus | Predicate(s) |
| | 88.0% (AWBAT-D Plus) | Same as AWBAT-S standard<br>(K082869) |
| | 89.0% (AWBAT-M Plus) | Same as AWBAT-S standard<br>(K082869) |
| Sterilization Method | Electron Beam Radiation | Same as AWBAT standard<br>(K082869). |
| Shelf Life. | 3 years | Same as AWBAT standard<br>(K082869). |
| Sizes | 6"x6", 6"x12", 12"x12" | Same as AWBAT standard<br>(K082869). |
### Table 2: Comparative Characteristics
{3}------------------------------------------------
K091863
pose 496
A WBAT Plus Wound Dressing Traditional 510(k)
{4}------------------------------------------------
K091863
page 506
AWBAT Plus Wound Dressing Traditional 510(k)
#### Device Testing 5.9
Biocompatibility testing is being conducted per FDA recognized consensus standard ise 10993-1:2003 for the materials used to manufacture the AVBAT product family to assure they are safe for its intended use. Aubrey Inc. certifies that it will not distribute AWBAT Plus until this testing is completed and demonstrates compliance with the requirements of FDA recognized consensus standard ISO 10993-1:2003.
Per ISO 10993-1:2003, AWBAT Plus is categorized as follows (from Table 1 of the standard):
- . Contact Duration: (B), Prolonged
- . Body Contact Category: Surface Device
- . Body Contact Type: Breached or compromised surface
Per Table 1 of ISO 10993-1:2003, the following testing is indicated:
- . Cytotoxicity
- Sensitization .
- Irritation or intracutaneous reactivity. .
The manufacturing process of the AWBAT product family complies with the United States Food and Drug Administration and European Standards for the manufacturing of medical devices.
## 5.10 Substantial Equivalence Summary
The submitted material in this Premarket Notification demonstrates that AWBAT Plus wound dressing (AWBAT-S Plus, AWBAT-D Plus and AWBAT-M Plus) is substantially equivalent to the AWBAT wound dressing (K082869) in terms of intended use, materials, design, function
Confidential
{5}------------------------------------------------
K091863
page 6 of 6
AWBAT Plus Wound Dressing Traditional 510(k)
Both AWBAT and AWBAT Plus are Temporary Wound and operating characteristics. Dressings for coverage of Superficial burns, Donor sites and Meshed autografts until healing occurs. Both AWBAT and AWBAT Plus are composed of the same three base components in the same configurations: nylon, silicone, and collagen peptides. Additionally, AVØRT Plus contains hydrophilic components Immuno-10 (Aloe polysaccharides) and chondroitin-4sulfate/chondroitin-6-sulfate for enhancing the creation of a noist wound healing environment. In conjunction, small amounts of anti-oxidant components, vitamin E oil and vitamin C crystals, are added to protect the hydrophilic components during the manufacturing process, with a small amount polysorbate 20 included to emulsify the vitamin E oil.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling an emblem. The emblem is a stylized representation of an eagle with its wings spread, symbolizing protection and care.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
## JAN 2 1 2010
Aubrey, Inc. % Shepard G. Bentley 5930 Sea Lion Place, Suite 100 Carlsbad, California 92010
Re: K091863 Trade/Device Name: AWBAT Plus (AWBAT-S Plus, AWBAT-D Plus, AWBAT-M Plus) Regulation Number: Unclassified Product Code: FRO Dated: December 17, 2009 Received: December 28, 2009
Dear Shepard G. Bentley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act, The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
## Page 2 - Shepard G. Bentley
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm.for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807:97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
**Sincerely yours,**
Sincerely yours,
Fid
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
AWBAT Plus Wound Dressing Traditional 510(k)
SECTION 4: Indications for Use Statement INDICATIONS FOR USE STATEMENT
Kog1863 510(k) Number (if known): __ Device Name: AWBAT Plus (AWBAT-S Plus, AWBAT-D Plus, AWBAT-M Plus)
Indications for Use:
AWBAT-S Plus: is intended for clean Superficial burn wounds.
AWBAT-D Plus: are intended for Donor sites after hemostasis has been established. AWBAT-M Plus: are intended to be used as a protective covering for Meshed autografts.
Prescription Use ____ X Over-The-Counter Use (Part 21 CFR 801 Subpart D) AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page _1_ of _1_
(Posted November 13, 2003)
Daniel Krause for MXM
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K091863
CONFIDENTIAL