HEMOGLOBIN CAPILLARY COLLECTION SYSTEM (HCCS), 1 TEST, HEMOGLOBIN CAPILLARY COLLECTION SYSTEM (HCCS), 5 TESTS, HEMOGLOBI

K090737 · Bio-Rad Laboratories, Inc. · JKA · Jun 3, 2009 · Clinical Chemistry

Device Facts

Record IDK090737
Device NameHEMOGLOBIN CAPILLARY COLLECTION SYSTEM (HCCS), 1 TEST, HEMOGLOBIN CAPILLARY COLLECTION SYSTEM (HCCS), 5 TESTS, HEMOGLOBI
ApplicantBio-Rad Laboratories, Inc.
Product CodeJKA · Clinical Chemistry
Decision DateJun 3, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Intended Use

The Hemoglobin Capillary Collection System (HCCS) is intended for the collection of human whole blood for the percent determination of hemoglobin A1c using Bio-Rad HPLC methods. For in vitro diagnostic use.

Device Story

HCCS facilitates collection/preparation of capillary or venous whole blood for HbA1c analysis; system includes microvials with reagent (EDTA/potassium cyanide), plastic capillaries, capillary holder, and labels. Professional users collect blood via capillary; sample is diluted in reagent vial at time of collection. Prepared samples are stable for transport/storage (4 days at 42°C, 2 weeks at 15-30°C, or 4 weeks at 2-8°C). Samples are analyzed using Bio-Rad VARIANT II or D-10 Hemoglobin A1c Programs (HPLC). Output is percent HbA1c; used by clinicians to monitor long-term glucose control in diabetic patients.

Clinical Evidence

Method comparison study conducted at an external site. Compared HbA1c values from capillary samples (HCCS) vs. venous whole blood samples (predicates). Results showed no significant difference in HbA1c values between HCCS-prepared samples and predicate-diluted samples.

Technological Characteristics

Aqueous reagent containing EDTA and potassium cyanide. Components: 1.5 mL microvials with pierceable caps, 5 uL plastic capillaries, capillary holder. Manual sample collection and dilution. Compatible with Bio-Rad HPLC systems.

Indications for Use

Indicated for monitoring long-term glucose control in individuals with diabetes mellitus via measurement of percent hemoglobin A1c.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### Summary of Safety and Effectiveness ... ... This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: Submitter: Bio-Rad Laboratories, Inc., Clinical Systems Division, 4000 Alfred Nobel Drive, Hercules. California 94547 Phone: (510) 741-6114; Fax: (510) 741-3954 Contact Person: Jolene Bartilson, Regulatory Affairs Representative Device Name: Hemoglobin Capillary Collection System (HCCS) Classification Name: Blood specimen collection device, JKA Assay, Glycosylated hemoglobin, LCP Predicate Devices: VARIANT™ II Hemoglobin A1c Program, Bio-Rad Laboratories, Inc., K070452 -Wash/Diluent Solution D-10TM Hemoglobin A1c Program, Bio-Rad Laboratories, Inc., K031043 - Wash/Diluent Solution Intended Use: The Hemoglobin Capillary Collection System (HCCS) is intended for the collection of human whole blood for the percent determination of hemoglobin A1c using Bio-Rad HPLC methods. For in vitro diagnostic use. Indications For Use: The measurement of percent hemoglobin A1c is effective in monitoring long-term glucose control in individuals with diabetes mellitus. {1}------------------------------------------------ ## Description of the Device: Each Hemoglobin Capillary Collection System (HCCS) contains a combination of the following components (quantity is dependent upon kit size): 1. Sample Preparation Vials: clear microvials with blue pierceable caps, each contains 1.5 mL of the HCCS reagent. 2. Capillaries: plastic capillaries (5 uL) in a dispenser. 3. Capillary Holder: holder for manipulating the capillaries. 4. Labels: to label prepared samples. The HCCS provides the necessary materials for a professional user to collect and prepare whole blood samples (capillary or venous blood) for analysis with the VARIANT™ II Hemoglobin A1c Program run on the VARIANT™ II Hemoglobin Testing System or the D-10™ Hemoglobin A1c Program run on the D-10™ Hemoglobin Testing System. Technical Characteristics Compared to the Predicates: The Hemoglobin Capillary Collection System (HCCS) reagent is intended to dilute whole blood samples in the same way that the predicate Wash/Diluent Solutions of the VARIANT™ II Hemoglobin A1c Program and D-10™ Hemoglobin A1c Program are used to dilute whole blood samples; in preparation for the percent determination of hemoglobin A I c in human whole blood using ion-exchange high-performance liquid chromatography (HPLC). The similarities and differences between the HCCS and the predicates are summarized in the following table: | Characteristic | New Device:<br>Hemoglobin<br>Capillary<br>Collection System<br>(HCCS) | Predicate Device:<br>VARIANT™ II<br>HbA1c Program,<br>Wash/Diluent Solution<br>(K070452) | Predicate Device: D-<br>10™ HbA1c<br>Program,<br>Wash/Diluent<br>Solution (K031043) | |----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The Hemoglobin<br>Capillary<br>Collection System<br>(HCCS) is intended<br>for the collection of<br>human whole blood<br>for the percent<br>determination of<br>hemoglobin A1c<br>using Bio-Rad<br>HPLC methods. For | The Bio-Rad<br>VARIANT II<br>Hemoglobin A1c<br>Program is intended for<br>the percent<br>determination of<br>hemoglobin A1c in<br>human whole blood<br>using ion-exchange<br>high performance liquid<br>chromatography | The Bio-Rad D-10<br>Hemoglobin A1c<br>Program is intended for<br>the percent<br>determination of<br>hemoglobin A1c in<br>human whole blood<br>using ion-exchange<br>high-performance<br>liquid chromatography<br>(HPLC). The Bio-Rad | | | in vitro diagnostic<br>use. | (HPLC). The Bio-Rad<br>VARIANT II<br>Hemoglobin A1c<br>Program is intended for<br>Professional Use Only. | D-10 Hemoglobin A1c<br>Program is intended for<br>Professional Use Only. | | Indications<br>For Use | Measurement of<br>percent hemoglobin<br>A1c is effective in<br>monitoring long-<br>term glucose<br>control in<br>individuals with<br>diabetes mellitus. | Measurement of<br>percent hemoglobin<br>A1c is effective in<br>monitoring long-term<br>glucose control in<br>individuals with<br>diabetes mellitus. | Measurement of<br>percent hemoglobin<br>A1c is effective in<br>monitoring long-term<br>glucose control in<br>individuals with<br>diabetes mellitus. | | Formulation | Aqueous solution<br>of EDTA and<br>potassium cyanide. | Deionized water with<br><0.05% sodium azide<br>as a preservative. | Deionized water with<br><0.05% sodium azide<br>as a preservative. | | Sample Type | Capillary or<br>venous blood from<br>plastic capillary. | Venous whole blood<br>sample collected in an<br>EDTA vacuum<br>collection tube. | Venous whole blood<br>sample collected in an<br>EDTA vacuum<br>collection tube. | | Purpose of<br>Reagent | Sample dilution for<br>hemoglobin A1c<br>analysis using Bio-<br>Rad HPLC<br>methods. | Sample dilution for<br>hemoglobin A1c<br>analysis using Bio-Rad<br>HPLC method. | Sample dilution for<br>hemoglobin A1c<br>analysis using Bio-Rad<br>HPLC method. | | Timing of<br>Dilution Step | Sample dilution at<br>time of sample<br>collection. No<br>further preparation<br>is required; the<br>VARIANT II<br>Hemoglobin<br>Testing System and<br>D-10 Hemoglobin<br>Testing System do<br>not perform<br>dilution on samples<br>withdrawn from<br>sample vials. | Dilution with the<br>Wash/Diluent Solution<br>is performed<br>automatically by the<br>VARIANT II<br>Hemoglobin Testing<br>System at time of<br>analysis; when vacuum<br>collection tubes are<br>used. | Dilution with the<br>Wash/Diluent Solution<br>is performed<br>automatically by the D-<br>10 Hemoglobin Testing<br>System at time of<br>analysis; when vacuum<br>collection tubes are<br>used. | | Stability of<br>Sample | Prepared samples<br>can be passed on<br>for analysis or<br>shipped to another | Whole blood<br>specimens may be<br>stored up to 7 days at 2<br>to 8 °C. | Whole blood<br>specimens may be<br>stored up to 7 days at 2<br>to 8 °C or 3 days at | {2}------------------------------------------------ : . : {3}------------------------------------------------ | location. Samples<br>are stable for: 4<br>days at 42 °C, 2<br>weeks at 15 to 30<br>°C, or 4 weeks at 2<br>to 8 °C. | room temperature (15<br>to 30 °C). | |--------------------------------------------------------------------------------------------------------------------|------------------------------------| |--------------------------------------------------------------------------------------------------------------------|------------------------------------| Testing to Establish Substantial Equivalence: # Method Comparison: A method comparison study was conducted at an external site to determine if there is a significant difference in the hemoglobin A1c values of capillary samples collected and prepared with the Hemoglobin Capillary Collection System (HCCS) versus the values obtained from the corresponding venous whole blood samples when analyzed with the VARIANT™ II Hemoglobin A1c Program on the VARIANT™ II Hemoglobin Testing System and with the D-10™ Hemoglobin A1c Program on the D-10™ Hemoglobin Testing System. One HCCS capillary sample and one venous EDTA primary tube sample were collected from each patient; the samples were then analyzed with the VARIANT™ II Hemoglobin A1c Program or the D-10TM Hemoglobin A1c Program. The results of the study are summarized in the tables below: # HCCS Comparison Study on the VARIANT™ II Hemoglobin A1c Program | · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·<br>M. A . A | | | | | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|--|--|-------|-----------| | Method | nle Value Kange | | | SIDDE | Intercent | | : VII HhAlc Program | 126%HhAlc<br>y to | | | | | # HCCS Comparison Study on the D-10™ Hemoglobin A1c Program | ということです。 そして、 そして、 その他の人気をしている。 この時には、 2005年には、 2005年には、 2005年には、 2005年に、 2005年に、 2005年に、 2005年に、 2005年に<br>Method | ample Value Kange | | ntercen | |--------------------------------------------------------------------------------------------------------------------|-------------------|--|---------| | : D-10 HbAlc Program<br>and an any and sells and address and | x to 14.3 %HbAIc | | -0.06X | ## Conclusion: The results of the method comparison study and performance evaluation studies of the Hemoglobin Capillary Collection System (HCCS) show that there is no significant difference in the hemoglobin Alc values for whole blood samples prepared with the HCCS versus whole blood samples diluted with the predicate Wash/Diluent Solutions of the VARIANT II Hemoglobin A1c Program and D-10 Hemoglobin A1c Program. Therefore, we trust that the information provided in this 510(k) Submission will support a decision of substantial equivalence of the Hemoglobin Capillary Collection System (HCCS). {4}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure. #### Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Bio-Rad Laboratories, Inc c/o Ms. Jolene Bartilson 4000 Alfred Nobel Dr Hercules, CA 94547 JUN - 8 2009 k090737 Re: Trade/Device Name: Hemoglobin Capillary Collection System Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylated Hemoglobin Assay Regulatory Class: Class II Product Code: LCP and JKA Dated: March 19, 2009 Received: March 20, 2009 Dear Ms. Bartilson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally {5}------------------------------------------------ Page - 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely vours. Corg C. He Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health {6}------------------------------------------------ # Indication for Use ### KO9073 510(k) Number (if known): Device Name: Hemoglobin Capillary Collection System (HCCS) Indication For Use: The Hemoglobin Capillary Collection System (HCCS) is intended for the collection of human whole blood for the percent determination of hemoglobin A1c using Bio-Rad HPLC methods. For in vitro diagnostic use. The measurement of percent hemoglobin A1c is effective in monitoring long-term glucose control in individuals with diabetes mellitus. Prescription Use (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K090737 Page 1 of 2
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%