S40 CREATINE KINASE (CK)
K082226 · Alfa Wassermann Diagnostic Technologies, Inc. · JHS · Sep 18, 2008 · Clinical Chemistry
Device Facts
| Record ID | K082226 |
| Device Name | S40 CREATINE KINASE (CK) |
| Applicant | Alfa Wassermann Diagnostic Technologies, Inc. |
| Product Code | JHS · Clinical Chemistry |
| Decision Date | Sep 18, 2008 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1215 |
| Device Class | Class 2 |
Intended Use
The S-Test Creatine Kinase Reagent is intended for the quantitative determination of Creatine Kinase activity in serum or heparin plasma using the S40 Clinical Analyzer. Measurements of creatine phosphokinase activity are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchennetype muscular dystrophy. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Device Story
S-Test CK reagent cartridge used with S40 Clinical Analyzer for quantitative in vitro diagnostic determination of Creatine Kinase (CK) activity in serum or heparin plasma. Principle: photometric test measuring formation of NADPH (absorbs at 340 nm). Reagents include hexokinase, glucose-6-phosphate dehydrogenase, adenosine-5'-diphosphoric acid, D-glucose, NADP, and creatine phosphate. Used in clinical laboratories or physician office laboratories by trained personnel. Output: quantitative CK activity levels. Healthcare providers use results to diagnose and monitor myocardial infarction and muscle diseases. Benefits: provides rapid, accurate diagnostic data for clinical decision-making.
Clinical Evidence
Bench testing only. Precision studies (n=3 levels, 22 days) showed within-run CV 1.4-2.0%, total CV 5.1-6.1%. POL site precision studies showed within-run CV 0.5-3.7%, total CV 0.8-3.7%. Accuracy correlation study (n=95 samples, 32-1181 U/L) yielded correlation coefficient 0.998. Additional POL site correlation studies showed correlation coefficients of 1.00. Detection limit: 24 U/L.
Technological Characteristics
Bi-reagent cartridge system. Reagent 1: Hexokinase, glucose-6-phosphate dehydrogenase, adenosine-5'-diphosphoric acid, D-glucose, NADP. Reagent 2: Creatine phosphate. Sensing principle: Photometric measurement of NADPH formation at 340 nm. System: S40 Clinical Analyzer.
Indications for Use
Indicated for quantitative determination of Creatine Kinase activity in serum or heparin plasma for diagnosis and treatment of myocardial infarction and muscle diseases (e.g., Duchenne-type muscular dystrophy). Intended for use in clinical or physician office laboratories.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
Predicate Devices
- Alfa Wassermann ACE plus ISE/Clinical Chemistry System ACE Creatine Kinase Reagent (k931786)
Related Devices
- K962247 — HICHEM CK/NAC REAGENT KIT · Elan Pharma, Inc. · Aug 15, 1996
- K191296 — Pointe Scientific Creatinine Kinase (CK) Reagent Set · Medtest DX · Aug 11, 2020
- K182702 — SEKURE Creatine Kinase Assay · Sekisui Diagnostics P.E.I., Inc. · Apr 17, 2019
- K983070 — CK · Abbott Laboratories · Feb 16, 1999
- K122083 — ELITECH CLINICAL SYSTEMS CK NAC SL · Elitechgroup · Aug 22, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Submission S40 Clinical Analyzer
S-Test CK
## 510(k) SUMMARY
| 510(k) Owner: | Alfa Wassermann Diagnostic Technology, LLC<br>4 Henderson Drive<br>West Caldwell, NJ 07006 | SEP 1 8 2008 |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Contact: | Dennis Taschek<br>Phone: 973-852-0177<br>Fax: 973-852-0237 | K082226 |
| Date Summary Prepared: | July 14, 2008 | |
| Device: | | |
| | Trade Names: S-Test CK Reagent cartridge | |
| | Classification: Class II | |
| | Common/Classification Name: CK: Creatine phosphokinase/creatine<br>kinase or isoenzymes test system<br>(21 C.F.R. § 862.1215)<br>Product Code JHS | |
| Predicate Device: | Manufacturer for analyzer/reagent system predicate is: | |
| | Alfa Wassermann ACE plus ISE/Clinical Chemistry System<br>ACE Creatine Kinase Reagent (k931786) | |
| Device Description: | The S-Test Creatine Kinase (CK) reagent cartridge, used with the S40 Clinical<br>Analyzer, is intended for quantitative <i>in vitro</i> diagnostic determination of CK<br>activity in serum or heparin plasma based on a photometric test measuring the<br>formation of NADPH, which absorbs strongly at 340 nm. | |
| Intended Use: | The S-Test Creatine Kinase Reagent is intended for the quantitative<br>determination of Creatine Kinase activity in serum or heparin plasma using the<br>S40 Clinical Analyzer. Measurements of creatine phosphokinase activity are<br>used in the diagnosis and treatment of myocardial infarction and muscle<br>diseases such as progressive, Duchenne-type muscular dystrophy. This test is<br>intended for use in clinical laboratories or physician office laboratories. For <i>in<br/>vitro</i> diagnostic use only. | |
| Technological Characteristics: | The S-Test CK Reagent is contained in a bi-reagent cartridge. Reagent 1<br>contains: Hexokinase, glucose-6-phosphate dehydrogenase, adenosine-5'-<br>diphosphoric acid, D-glucose, and nicotinamide adenine dinucleotide phosphate<br>(oxidized form). Reagent 2 contains: Creatine phosphate. | |
.
.
{1}------------------------------------------------
Performance data on the S-Test CK reagent included precision, accuracy, and Performance sensitivity data. Data:
> Precision: In testing at three CK levels for 22 days, the within-run CV ranged from 1.4 to 2.0%, and total CV ranged from 5.1 to 6.1%. In precision studies at three separate Physician Office Laboratory (POL) sites and in-house over five days, the within-run CVs ranged from 0.5 to 3.7% and total CVs ranged from 0.8 to 3.7%.
> Accuracy: In a correlation study, 95 samples with CK values ranging from 32 to 1181 U/L were assayed on the S40 Clinical Analyzer using S-Test CK (y) and a comparative method (x). Least-squares regression analysis yielded a correlation coefficient of 0.998, a standard error estimate of 14.5, a confidence interval slope of 1.027 to 1.053, and a confidence interval intercept of -12.5 to -5.2. In patient correlation studies at three separate POL sites using the S40 Clinical Analyzer and a comparative method, least-squares regression analysis yielded correlation coefficients of 1.00, standard error estimates of 5.0 to 8.5, confidence interval slopes of 1.006 to 1.059, and a confidence interval intercepts of -10.8 to -3.5.
Sensitivity: The detection limit was 24 U/L.
Based on the foregoing data, the device is safe and effective. These data also Conclusions: indicate substantial equivalence to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a serif font. The text is black and the background is white. The text is centered in the image.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Alfa Wassermann Diagnostic Technologies, Inc. c/o Mr. Dennis Taschek Vice President, Reagent and Instrument Technology 4 Henderson Drive West Caldwell, NJ 07006
SEP 1 8 2008
Re: K082226
> Trade/Device Name: S-Test CK Reagent Cartridge Regulation Number: 21 CFR 862.1215 Regulation Namc: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: Class II Product Code: JHS Dated: August 14, 2008 Received: August 18, 2008
Dear Mr. Taschek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
Image /page/2/Picture/11 description: The image shows the seal of the Department of Health & Human Services, USA. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is simple and monochromatic.
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.V.M.
Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Submission S40 Clinical Analyzer S-Test CK
## Indications for Use
K082226
510(k) Number (if known):
Device Name: . S-Test Creatine Kinase (CK)
Indications for Use: The S-Test Creatine Kinase Reagent is intended for the quantitative determination of Creatine Kinase activity in serum or heparin plasma using the S40 Clinical Analyzer. Measurements of creatine phosphokinase activity are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchennetype muscular dystrophy. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
V Prescription Use _ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED
| Concurrence of CDRH, Office of Device Evaluation (ODE) | <div> Division Sign-Off </div> |
|--------------------------------------------------------|-------------------------------------------------------------------------------|
| | Acting |
| August 6, 2008 | Office of In Vitro Diagnostic Device<br>Evaluation and Safety<br>CONFIDENTIAL |
Page 66