K Number
K081269

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-03-20

(319 days)

Product Code
Regulation Number
864.7470
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Avie™ Alc test is a point of care system that quantitatively measures %A1C (glycated hemoglobin) in capillary or venous whole blood samples. The test is for prescription use and physician-directed home use to monitor glycemic control in people with diabetes mellitus. The device cannot be used in patients with hemoglobinopathies of HbF, HbC, and HbD.

Device Description

The Avie A1C test system is for the quantitative measurement of %A1C in fingerstick whole blood at the point-of-care (Professional or Physician Directed Home Use). The system consists of a small instrument ("Reader), a single-use diluent solution vial, single-use reagent test cartridges, and control levels I and II solutions. To perform one test, the Reader is turned on by the push of a button on its top surface, and a self-check (optics and software) is performed automatically. If a malfunction is detected, the Reader displays an error message on the LCD display. When the LCD display denotes "Ready" a test cartridge is inserted into an unambiguous slot in the reader. About four uL of whole blood is delivered into a vial containing wash solution or diluent, via a transfer pipette, and the vial is inverted five times for mixing. At the instruction of the reader (LCD read-out), three drops of blood mixture are applied to a well in the cartridge: the reactions then proceed automatically. There are no further procedural steps, and results in %A1C are displayed on the reader's LCD screen in approximately three minutes. The concentration of A1C is expressed as a percentage of the total hemoglobin as follows: %HBA1C = (HBA1C ÷ TOTAL HB) ×100

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the Avie™ A1C Test System, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
PrecisionPercent coefficient of variation within 5%
Accuracy (estimated bias)Versus NGSP: Y=0.992 ± 0.001, R=0.94 N=155 (This is from the comparison table against the predicate, likely representing the target for the new device. The clinical study results provide the actual performance.) Professional: Avie = 0.95 (NGSP) + 0.004 r = 0.94 Self: Avie = 0.99 (NGSP) + 0.001 r = 0.94 Combined professional vs. self-testing (109 subjects): Y=0.98S+0.003 r=0.94 Trained vs. Untrained operators (42 subjects): Y=1.05x -0.1899 r²=0.95
LinearityLinear between 5 and 14 %A1C
InterferenceInsensitive to high physiological levels of triglyceride and bilirubin, high therapeutic levels of various over-the-counter pharmaceuticals, and hemoglobin levels ranging from approximately 9 g/dL to 20 g/dL. (However, sensitive to high levels of Metformin and glybenclamide, and high levels of hemoglobinopathies).
Operational Condition LimitsOperable between temperatures of 18°C to 28°C
NGSP CertificationCertified

2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Study Test Set:
    • Approximately 150 subjects for the main clinical study (professional vs. self-testing analysis).
    • 109 subjects for the combined professional vs. self-testing sensitivity study.
    • 42 subjects for the untrained operator study.
    • 60 subjects for the comparison against the A1C Now+ predicate.
    • 60 subjects for the comparison of venous to capillary samples.
  • Data Provenance: The document does not explicitly state the country of origin for the clinical data. It mentions "four sites" and "three sites" for the clinical studies and "NGSP laboratory in Columbia. MO" for ground truth testing. This suggests the studies were likely conducted within the United States. The studies are described as clinical studies, implying they are prospective as participants actively performed tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The ground truth for the test set was established by sending collected venous blood samples to an NGSP laboratory in Columbia, MO.
  • The document does not specify the number of experts (e.g., lab technicians, certified professionals) within the NGSP laboratory who processed and analyzed these samples.
  • The qualifications of these experts are implied by the NGSP (National Glycohemoglobin Standardization Program) certification, which signifies adherence to standardization and quality control for HbA1c testing. Specific individual qualifications (e.g., years of experience as a clinical chemist) are not provided.

4. Adjudication Method for the Test Set

  • The document does not describe an explicit "adjudication method" in the context of expert consensus for image or diagnostic interpretation typically found in AI/ML device studies.
  • Instead, the ground truth was derived from a single, standardized laboratory method performed by an NGSP-certified laboratory. The device's performance was then compared to this reference standard. Therefore, there was no multi-reader adjudication process as such.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

  • No, an MRMC comparative effectiveness study was not done in the context of human readers improving with or without AI assistance.
  • This device (Avie™ A1C Test System) is a diagnostic instrument for quantitative measurement of A1C, not an AI or imaging diagnostic aid that assists human readers.
  • The studies involved comparing the device's readings to laboratory reference methods and evaluating performance by professional and self-testing users, and trained vs. untrained operators, but not human interpretation aided by AI.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

  • Yes, the primary evaluation of the Avie™ A1C Test System's performance (its "algorithm only") was done in a standalone manner against the NGSP reference method. The "Professional" and "Self" testing studies reflect the device's performance when used by different types of operators (humans-in-the-loop), but still measure the device's output against the ground truth.
  • The regression statistics (e.g., "Avie = 0.95 (NGSP) + 0.004 r = 0.94") directly reflect the standalone performance of the device's measurement compared to the reference.

7. The Type of Ground Truth Used

  • The ground truth used was laboratory reference measurements of %A1C from venous blood samples, performed by an NGSP-certified laboratory. This is a highly standardized and accepted method for determining HbA1c levels.

8. The Sample Size for the Training Set

  • The document does not provide information regarding a separate training set. This is typical for a diagnostic device like the Avie™ A1C Test System, which is based on a fixed immunoassay methodology rather than a machine learning or AI model that requires explicit training data. The device is likely calibrated at the factory.

9. How the Ground Truth for the Training Set Was Established

  • As there's no mention of a traditional "training set" for an AI/ML model, the concept of establishing ground truth for it is not applicable here. The device's calibration would be established by the manufacturer using a separate, internal process traceable to NGSP standards, but this is distinct from an AI model's training data.

{0}------------------------------------------------

KOD269

MAR 2 0 2009

MEC 2 oynamics. MEC Dynamics, Corp. 2225 Martin Ave., Suite I Santa Clara, Ca 95050

SECTION 9 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is K081269.

MEC Dynamics Corporation 807.92 (a)(1): Name:

Address:2225 Martin Avenue, Suite ISanta Clara, CA 95050
Phone:408-844-9279
FAX:408 844-9285
Contact:Mr. Emmanuel Mpock

807.92 (a)(2): Device name- trade name and common name, and classification

Trade name: Avie™ A1C Test System

Common Name: Glycosylated hemoglobin assay

Classification: CFR §21.864.7470

807.92 (a)(3): Identification of the legally marketed predicate device

The Avie™ A1C Test System is substantially equivalent to other A1C test systems, and specifically, the G5 I/II HbA1C Test System (Provalis Diagnostics, Ltd.), most recently cleared under premarket notification K041635 on August 16, 2004.

807.92 (a)(4): Device Description

The Avie A1C test system is for the quantitative measurement of %A1C in fingerstick whole blood at the point-of-care (Professional or Physician Directed Home Use). The system consists of a small instrument ("Reader), a single-use diluent solution vial, single-use reagent test cartridges, and control levels I and II solutions.

To perform one test, the Reader is turned on by the push of a button on its top surface, and a self-check (optics and software) is performed automatically. If a

{1}------------------------------------------------

S MEC DYNAMICS MEC Dynamics, Corp. 2225 Martin Ave., Suite I Santa Clara, Ca 95050

malfunction is detected, the Reader displays an error message on the LCD display. When the LCD display denotes "Ready" a test cartridge is inserted into an unambiguous slot in the reader. About four uL of whole blood is delivered into a vial containing wash solution or diluent, via a transfer pipette, and the vial is inverted five times for mixing. At the instruction of the reader (LCD read-out), three drops of blood mixture are applied to a well in the cartridge: the reactions then proceed automatically. There are no further procedural steps, and results in %A1C are displayed on the reader's LCD screen in approximately three minutes. The concentration of A1C is expressed as a percentage of the total hemoglobin as follows: %HBA1C = (HBA1C ÷ TOTAL HB) ×100

807.92 (a)(5): Intended Use

The Avie™ Alc test is a point of care system that quantitatively measures %A1C (glycated hemoglobin) in capillary whole blood samples. The test is for Professional and Physician Directed home use to monitor glycemic control in people with diabetes.

{2}------------------------------------------------

807.92 (a)(6): Technological Similarities and Differences to the Predicate

Comparison between the Bayer A1C Now+™ the MEC Dynamics Avie™ A1C Systems

CharacteristicBayer A1CNow+ K051321MEC Dynamics Avie™ AIC K081269
Intended UseQuantitative measurement of the percent ofglycated hemoglobinQuantitative measurement of the percent ofglycated hemoglobin
Indications for UseUsed in the management and treatment ofdiabetes, for monitoring long-term glycemiccontrolUsed in the management and treatment ofdiabetes, for monitoring long-term glycemiccontrol
Risk to patientNot a critical analyte - reflects glucose levelover timeNot a critical analyte - reflects glucose levelover time
SampleWhole bloodWhole blood
Visual displayLCD readoutLCD readout
Hemolysate preparationManual via sample dilution kitManual, pipette and diluent supplied
MethodologyImmunoassayImmunoassay
Detection methodFour-channel reflectance photometerTransmission
CalibrationFactory calibratedFactory calibrated
Recommended testingenvironmentProfessional usePrescription and Physician directed home use
Throughput5 minutes per sample3 minutes per sample
Reagent storageRoom temperature up to 120 days thenrefrigerateRoom temperature
Quality control requirementsExternal controls testing recommended:- With each new shipment- With each new lot- With each new operator- Whenever problems (storage, operator,instrument, or other) are identified- To ensure the storage conditions have notaffected the product, run a control samplebefore running a patient sample if the testkit has been stored for more than a monthand it has been a least a month since thelast control testing.External controls testing recommended:-Prior to home testing or at the start of eachtesting day.-Upon receipt of each new shipment or use ofa new lot of cartridges-Whenever storage room conditions have beenabove 28°C (82°F).-To become familiar with the process or toperform training or retraining of testingpersonnel.-Whenever Avie™ Alc results do not matchother clinical findings or symptoms.
PrecisionPercent coefficient of variation within 5%Percent coefficient of variation within 5%
Accuracy (estimated bias)Versus NGSPY= 1.02x- 0.23R=0.95 N=189Versus NGSPY=0.992±0.001R=0.94 N=155
NGSP certification statusCertifiedCertified

{3}------------------------------------------------

MEC / DYNAMICS . MEC Dynamics, Corp. 2225 Martin Ave., Suite I Santa Clara, Ca 95050

807.92 (b)(1): Brief Description of Nonclinical Data

Studies were done that evaluated precision, linearity, interference, operational condition limits, and NGSP traceability for the Avie A1C Test System. The Avie A1C Test System demonstrates precision within 5 %CV at each of two levelsnormal and abnormal %A1C. The Avie A1C Test System is linear between 5 and 14 %A1C. The Avie A1C Test System is insensitive to high physiological levels of triglyceride and bilirubin. high therapeutic levels of various over-the-counter pharmaceuticals, and hemoglobin levels ranging from approximately 9 g/dL to 20 g/dL. The system is sensitive to high levels of Metformin and glybenclamide (greater than 0.39 mg/dL and 0.02 mg/dL, respectively), as well as high levels of hemoglobinopathies. The Avie A1C Test System may be operated between temperatures of 18°C to 28°C. NGSP issued a certificate of traceability granting manufacturer certification to MEC Dynamics, Corporation.

807.92 (b)(2): Brief Description of Clinical Data

Clinical studies were conducted across four sites with approximately 150 subjects. Subjects first performed one Avie A1C test on themselves, and this was followed by a second Avie AlC test performed by a health care professional at the site. Venous blood was collected and sent to the NGSP laboratory in Columbia. MO. The regression statistics from this testing are described below.

Professional: Avie = 0.95 (NGSP) + 0.004 r = 0.94 Self: Avie = 0.99 (NGSP) + 0.001 r = 0.94

Clinical sensitivity testing of professional vs. self testing conducted at four sites with 109 subjects within the linearity range combined results were:

$$\text{Y=0.98S+0.003 r=0.94}$$

The ability of untrained non-laboratory physician office staff was evaluated in an additional clinical sensitivity study conducted at three sites with 42 subjects. Testing comparing at least 2 untrained operators given only documented instructions and one trained operator per site resulted in the following combined values for trained vs. untrained operators:

$$\text{Y=1.05x -0.1899 } \text{r}^{2\text{\textquotedblleft 0.95}}$$

807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing

Nonclinical and clinical testing was performed for the Avie™ A1C Test System. The test system was shown to be safe and effective for its intended use.

{4}------------------------------------------------

MEC Dynamics, Corp. 2225 Martin Ave., Suite I Santa Clara, Ca 95050

A comparison test of the Avie™ A1C Test System to the A1C Now+ was run at one site with 60 subjects with the following results:

Y=1.000x +0.004 $r^{2}$=0.981 venous Y=0.993x +0.020 $r^{2}$=0.946 capillary

A comparison of venous to capillary samples using the Avie™ A1C Test System on 60 subjects results were:

$$\text{Y=0.972x +0.191 r}^{2=} \text{0.947}$$

807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing

Nonclinical and clinical testing was performed for the AvieTM A1C Test System. The test system was shown to be safe and effective for its intended use.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and features an eagle with its wings spread. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the top of the circle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAR 2 0 2009

MEC Dynamics Corporation c/o Mr. Emmanuel Mpock 2225 Martin Ave, Suite 1 Santa Clara, CA 95050

Re: K081269

Trade/Device Name: Avie A1C Test System Regulation Number: 21 CFR 864.7470 Regulation Name: Glycosylated hemoglobin assay Regulatory Class: Class II Product Code: LCP Dated: February 24, 2009 Received: February 25, 2009

Dear Mr. Mpock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{6}------------------------------------------------

Page - 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Corg C. He

Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

{7}------------------------------------------------

MEC Dynamics Corporation Avie™ AIC Test System Premarket Notification

Indications for Use

510(K) Number (if known): K081269

Device Name: Avie™ A1C Test System

Indications for Use:

The AvieTM A1c test is a point of care system that quantitatively measures %A1C (glycated hemoglobin) in capillary or venous whole blood samples. The test is for prescription use and physician-directed home use to monitor glycemic control in people with diabetes mellitus. The device cannot be used in patients with hemoglobinopathies of HbF, HbC, and HbD.

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Ruta club

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

§ 864.7470 Glycosylated hemoglobin assay.

(a)
Identification. A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.(b)
Classification. Class II (performance standards).