AMEDICHECK DRUG SCREEN THC, COC, OPI, AMP, MET, PCP, MDMA,BAR, BZ
K080872 · Amedica Biotech, Inc. · LDJ · Aug 19, 2008 · Clinical Toxicology
Device Facts
| Record ID | K080872 |
| Device Name | AMEDICHECK DRUG SCREEN THC, COC, OPI, AMP, MET, PCP, MDMA,BAR, BZ |
| Applicant | Amedica Biotech, Inc. |
| Product Code | LDJ · Clinical Toxicology |
| Decision Date | Aug 19, 2008 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.3870 |
| Device Class | Class 2 |
Intended Use
The AmediCheck Drug Screen Test THC/COC/OPVAMP/MET/PCP/BAR/BZQ/MDMA/OXY/MTD/ PPX/TCA is an in vitro diagnostic test for the rapid detection of THC, benzoylecgonine, morphine, amphetamine, methamphetamine, phencyclidine, barbiturates, benzodiazepines, MDMA, oxycodone, methadone, propoxyphene and tricyclic antidepressants in human urine at the following cutoff concentration: [Table of cutoffs provided]. This test kit is used to obtain a visual, qualitative result and is intended for professional use. It is not intended for over the counter sale. This assay provides only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometers (GCMS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
Device Story
AmediCheck Drug Screen Test is an in vitro diagnostic device for rapid, qualitative detection of multiple drugs of abuse in human urine. Device utilizes lateral flow immunoassay technology; urine sample applied to test device; presence of specific drugs indicated by visual color change on test strips. Intended for professional use in clinical settings; provides preliminary results only. Positive results require confirmation via alternative chemical methods, preferably GC/MS. Healthcare providers use visual output to inform clinical decision-making; results must be interpreted alongside clinical judgment.
Clinical Evidence
No clinical data provided. Bench testing only.
Technological Characteristics
Lateral flow immunoassay; qualitative visual readout; urine specimen; multi-panel test format for simultaneous detection of THC, COC, OPI, AMP, MET, PCP, BAR, BZO, MDMA, OXY, MTD, PPX, and TCA at specified ng/mL cutoff concentrations.
Indications for Use
Indicated for professional use for the rapid, qualitative detection of drugs of abuse (THC, benzoylecgonine, morphine, amphetamine, methamphetamine, phencyclidine, barbiturates, benzodiazepines, MDMA, oxycodone, methadone, propoxyphene, and tricyclic antidepressants) in human urine samples.
Regulatory Classification
Identification
A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.
Special Controls
*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Related Devices
- K042975 — AMEDITECH IMMUTEST MULTI-DRUG SCREEN PANEL II · Ameditech, Inc. · Dec 29, 2004
- K063015 — AMEDITECH IMMUTEST MULTI-DRUG SCREEN BUP/PPX/COC150 · Ameditech, Inc. · Dec 20, 2006
- K063379 — AMEDICA DRUG SCREEN THC, COC, OPI, AMP, MET, PCP, MDMA, BAR, BZO, M, MODELS C1210, C1220 · Amedica Biotech, Inc. · Dec 11, 2006
- K061556 — AMEDICA DRUG SCREEN THC, COC,OPI,AMP,MET,PCP,MDMA,BAR,BZO,; MODEL C1010,C1015,C1020 · Amedica Biotech, Inc. · Jul 21, 2006
- K012159 — RAPID DRUG SCREEN · American Bio Medica Corp. · Jul 25, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a circular seal with text around the perimeter and a symbol in the center. The symbol in the center appears to be a stylized depiction of an eagle or other bird in flight. The text around the perimeter is difficult to read due to the image quality, but it seems to be in a circular arrangement.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Amedica Biotech, Inc. c/o Mr. Jeff Chen President 28301 Industrial Blvd. Suite K Hayward, CA 94545
AUG 1 9 2008
Re: K080872
> Trade/Device Name: Amedicheck Drug Screen Test. THC/COC/OPI/AMP/MET/PCP/ BAR/BZO/MDMA/OXY/MTD/PPX/TCA Regulation Number: 21 CFR 862.3870 Regulation Name: Cannabinoid Test System. Regulatory Class: Class II Product Code: LDJ, DIG, DKZ, LAF, LCM, DJC, DIS, JXM, DJR, MLR, JXN Dated: July 18, 2008 Received: July 21, 2008
Dear Mr. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{1}------------------------------------------------
## Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.V.M.
Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number: K080872
## Device Name: AmediCheck Drug Screen Test THC/COC/OPVAMP/MET/PCP/BAR/BZO/MDMA/OXY/MTD/PPX/TCA
Indications For Use:
The AmediCheck Drug Screen Test THC/COC/OPVAMP/MET/PCP/BAR/BZQ/MDMA/OXY/MTD/ PPX/TCA is an in vitro diagnostic test for the rapid detection of THC, benzoylecgonine, morphine, amphetamine, methamphetamine, phencyclidine, barbiturates, benzodiazepines, MDMA, oxycodone, methadone, propoxyphene and tricyclic antidepressants in human urine at the following cutoff concentration:
| THC | 11-nor-Δ9-Tetrahydrocannabinol-9-carboxylic | 50 ng,ml |
|------|---------------------------------------------|------------|
| COC | Benzoylecgonine | 300 ng,ml |
| OPI | Morphine | 2000 ng,ml |
| OPI | Morphine | 300 ng,ml |
| AMP | Amphetamine | 1000 ng,ml |
| MET | Methamphetamine | 1000 ng,ml |
| PCP | Phencyclidine | 25 ng,ml |
| MDMA | 3,4 methylenedioxymethamphetamine | 500 ng/ml |
| BAR | Secobarbital | 300 ng/ml |
| BZO | Oxazepam | 300 ng/ml |
| MTD | Methadone | 300 ng/ml |
| TCA | Nortriptyline | 1000 ng/ml |
| PPX | Propoxyphene | 300 ng/ml |
| OXY | Oxycodone | 300 ng/ml |
This test kit is used to obtain a visual, qualitative result and is intended for professional use. It is not intended for over the counter sale.
This assay provides only a preliminary result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometers (GCMS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
| Prescription Use | X | AND/OR | Over-The-Counter Use |
|-----------------------------|---|--------|------------------------|
| (Part 21 CFR 801 Subpart D) | | | (21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign Off
Office of In Vitro Diagnostic Device
Evaluation and Safety
| 6 | 510(k) | K080872 |
|---|--------|---------|
|---|--------|---------|