ACCESS ESTRADIOL CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS, MODEL 33545
K080166 · Beckman Coulter, Inc. · JIT · Mar 19, 2008 · Clinical Chemistry
Device Facts
Record ID
K080166
Device Name
ACCESS ESTRADIOL CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS, MODEL 33545
Applicant
Beckman Coulter, Inc.
Product Code
JIT · Clinical Chemistry
Decision Date
Mar 19, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
The Access Estradiol Calibrators are intended to calibrate the Access Estradiol assay for the quantitative determination of estradiol levels in human serum and plasma using the Access Immunoassay Systems.
Device Story
Device is a re-standardized estradiol calibrator for Beckman Coulter Access immunoassay systems. Modification involves shifting calibration traceability from USP reference material to isotope dilution gas chromatography/mass spectroscopy (ID-GC/MS) method. Used in clinical laboratory settings by laboratory technicians to calibrate estradiol assays. Output provides standardized reference values for patient sample quantification, ensuring consistency in estradiol measurement. Benefits include improved accuracy and traceability of clinical test results.
Clinical Evidence
Bench testing only. Verification and validation activities performed per design control procedures (21 CFR 820.30) to confirm performance following re-standardization of the calibrator.
Technological Characteristics
Liquid calibrators containing estradiol at levels of approximately 106, 570, 1800, 3100, and 4800 pg/mL. Traceability is established via ID-GC/MS (isotope dilution-gas chromatography/mass spectrometry) reference method. The device is used with the Access Immunoassay Systems, which utilize a competitive binding immunoenzymatic assay.
Indications for Use
Indicated for the calibration of the Access Estradiol assay for quantitative determination of estradiol levels in human serum and plasma using Access Immunoassay Systems.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K012900 — ACCESS UNCONJUGATED ESTRIOL CALIBRATORS · Beckman Coulter, Inc. · Oct 16, 2001
K142021 — Elecsys Estradiol III CalSet · Roche Diagnostics · Aug 12, 2014
K163053 — Lumipulse G E2-III Calibrators · Fujirebio Diagnostics,Inc. · Dec 1, 2016
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification
ODE Review Memorandum (Decision Making Document is Attached)
To: THE FILE
RE: DOCUMENT NUMBER K080166
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) k097343
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for re-standardization of the estradiol calibrator for the Beckman Coulter Access from a USP traceable estradiol reference material to an isotope dilution gas chromatography/mass spectroscopy (ID-GC/MS) method.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, linearity, and reference range.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis FMEA.
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.