S-TEST CREATININE (CRE)

K080073 · Alfa Wassermann, Inc. · JFY · Jun 30, 2008 · Clinical Chemistry

Device Facts

Record IDK080073
Device NameS-TEST CREATININE (CRE)
ApplicantAlfa Wassermann, Inc.
Product CodeJFY · Clinical Chemistry
Decision DateJun 30, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1225
Device ClassClass 2

Intended Use

The S-Test Creatinine Reagent is intended for the quantitative determination of Creatinine concentration in serum or heparin plasma using the S40 Clinical Analyzer. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

Device Story

S-Test Creatinine (CRE) reagent cartridge used with S40 Clinical Analyzer; performs quantitative in vitro diagnostic determination of creatinine in serum or heparin plasma. Principle of operation: photometric test measuring reddish-purple pigment formation via coupled enzymatic reaction. Used in clinical laboratories or physician office laboratories by laboratory personnel. Output: creatinine concentration values; used by clinicians to diagnose/treat renal disease and monitor dialysis. Benefits: provides diagnostic data for renal function assessment.

Clinical Evidence

Bench testing only. Precision studies (22 days, 3 levels) showed within-run CV 1.3-9.6%, total CV 4.3-17.2%. POL site precision (5 days) showed within-run CV 1.0-10.1%, total CV 1.3-10.8%. Accuracy correlation study (n=65, range 0.6-14.6 mg/dL) yielded r=0.998, slope 1.002-1.031. POL site correlation studies yielded r=0.997-0.999. Detection limit 0.3 mg/dL.

Technological Characteristics

Bi-reagent cartridge. Reagent 1: Creatinine amidinohydrolase, sarcosine oxidase, N-ethyl-N-sulfobutyl-m-toluidine, catalase, TES buffer. Reagent 2: Creatinine amidinohydrolase, 4-aminoantipyrine, peroxidase, TES buffer. Photometric sensing principle. Standalone reagent system for S40 Clinical Analyzer.

Indications for Use

Indicated for quantitative determination of creatinine in serum or heparin plasma for diagnosis/treatment of renal disease and monitoring renal dialysis. Intended for use in clinical or physician office laboratories.

Regulatory Classification

Identification

A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Kon0073 ## JUN 3 0 2008 ## 510(k) SUMMARY | 510(k) Owner: | Alfa Wassermann Diagnostic Technology, LLC<br>4 Henderson Drive<br>West Caldwell, NJ 07006 | | | |---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|--| | | Contact: | Dennis Taschek<br>Phone: 973-852-0177<br>Fax: 973-852-0237 | | | Date Summary<br>Prepared: | June 19, 2008 | | | | Device: | Trade Name: | S-Test CRE Reagent cartridge | | | | Common/Classification Names: | Creatinine test system<br>(21 C.F.R. § 862.1225)<br>Product Code JFY | | | | Classification: | Class II | | | Predicate<br>Device: | Manufacturer for analyzer/reagent system predicate is: | | | | | Alfa Wassermann ACE plus ISE/Clinical Chemistry System<br>ACE Creatinine Reagent (K931786) | | | | Device<br>Description: | The S-Test Creatinine (CRE) reagent cartridge, used with the S40<br>Clinical Analyzer, is intended for quantitative <i>in vitro</i> diagnostic<br>determination of CRE in serum or heparin plasma based on a<br>photometric test measuring the formation of a reddish-purple pigment<br>in a coupled enzymatic reaction. | | | | Intended Use: | The S-Test Creatinine Reagent is intended for the quantitative<br>determination of Creatinine concentration in serum or heparin plasma<br>using the S40 Clinical Analyzer. Creatinine measurements are used in<br>the diagnosis and treatment of renal diseases, in monitoring renal<br>dialysis, and as a calculation basis for measuring other urine analytes.<br>This test is intended for use in clinical laboratories or physician office<br>laboratories. For <i>in vitro</i> diagnostic use only. | | | {1}------------------------------------------------ | Technological<br>Characteristics: | The S-Test CRE Reagent is contained in a bi-reagent cartridge.<br>Reagent 1 contains: Creatinine amidinohydrolase, sarcosine oxidase, N-<br>ethyl-N-sulfobutyl-m-toluidine, catalase, and N-tris (hydroxymethyl)<br>methyl-2-aminoethane sulfonic acid (TES) buffer. Reagent 2 contains:<br>Creatinine amidohydrolase, 4-aminoantipyrine, peroxidase, and N-tris<br>(hydroxymethyl) methyl-2-aminoethane sulfonic acid (TES) buffer. | |-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Performance<br>Data: | Performance data on the S-Test CRE Reagent included precision,<br>accuracy, and sensitivity data.<br><br>Precision: In testing at three CRE levels for 22 days, the within-run<br>CV ranged from 1.3 to 9.6%, and total CV ranged from 4.3 to 17.2%.<br>In precision studies at three separate Physician Office Laboratory<br>(POL) sites and in-house over five days, the within-run CV ranged<br>from 1.0 to 10.1% and total CV ranged from 1.3 to 10.8%.<br><br>Accuracy: In the correlation study, 65 samples with CRE values<br>ranging from 0.6 to 14.6 mg/dL were assayed on the S40 Clinical<br>Analyzer using S-Test CRE (y) and a comparative method (x). Least-<br>squares regression analysis yielded a correlation coefficient of 0.998, a<br>standard error estimate of 0.16, a confidence interval slope of 1.002 to<br>1.031, and a confidence interval intercept of -0.20 to -0.10. In patient<br>correlation studies at four separate POL sites using the S40 Clinical<br>Analyzer and a comparative method, least-squares regression analysis<br>yielded correlation coefficients ranged from 0.997 to 0.999, standard<br>error estimates of 0.25 to 0.34, confidence interval slopes of 1.003 to<br>1.060 and a confidence interval intercepts of -0.38 to 0.10.<br><br>Sensitivity: The detection limit was 0.3 mg/dL. | | Conclusions: | Based on the foregoing data, the device is safe and effective. These<br>data also indicate substantial equivalence to the predicate device. | · {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The emblem is positioned in the center of a circular border, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircling it. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Alfa Wassermann Diagnostic Technology, Inc. c/o Mr. Dennis Tascheck Vice President, Reagent & Instrument Technologies 4 Henderson Drive West Caldwell, NJ 07006 JUN 3 0 2008 Re: k080073 Trade Name: S-Test Creatinine (CRE) Regulation Number: 21 CFR 862.1225 Regulation Name: Creatinine Regulatory Class: Class II Product Codes: JFY Dated: June 16, 2008 Received: June 17, 2008 Dear Mr. Tascheck: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jean M. Cooper, M.S., D.V.M. Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indication for Use 510(k) Number (if known): k080073 Device Name: S-Test Creatininc (CRE) Indication For Use: The S-Test Creatinine Reagent is intended for the quantitative determination of Creatinine concentration in serum or heparin plasma using the S40 Clinical Analyzer. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Prescription Use _ X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) spetins Diyision Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K080073
Innolitics
510(k) Summary
Decision Summary
Classification Order