DIMERTEST
K080069 · American Diagnostica, Inc. · DAP · Feb 13, 2008 · Hematology
Device Facts
| Record ID | K080069 |
| Device Name | DIMERTEST |
| Applicant | American Diagnostica, Inc. |
| Product Code | DAP · Hematology |
| Decision Date | Feb 13, 2008 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 864.7320 |
| Device Class | Class 2 |
Intended Use
The DIMERTEST® Latex Assay is intended for the rapid qualitative or semi-quantitative evaluation of circulating derivatives of cross-linked fibrin degradation products (D-dimer) in human plasma.
Device Story
DIMERTEST® is a latex agglutination slide test for detecting cross-linked fibrin degradation products (D-dimer) in human plasma. The device consists of latex particles coated with D-dimer-specific monoclonal antibodies. When mixed with a plasma sample on a reaction card, the presence of D-dimer antigen causes visible agglutination of the latex particles, observable by the unaided eye. Semi-quantitative estimation is achieved by preparing serial dilutions of the plasma sample and comparing results against a titration matrix. The test is performed in a clinical laboratory setting by trained personnel. The output provides a qualitative or semi-quantitative concentration of D-dimer, which assists clinicians in evaluating fibrinolysis and thrombotic conditions. The modified version features a lower diagnostic cut-off of 200 ng/mL compared to the predicate's 250 ng/mL.
Clinical Evidence
Bench testing only. Analytical performance included precision/reproducibility studies (intra-assay and inter-assay over 28 days) and analytical specificity (no interference from bilirubin, lipids, hemoglobin, or rheumatoid factor). Method comparison studies included a specificity study (N=170 healthy donors) and a sensitivity study (N=145 patients with high probability of thrombotic conditions). The new 200 ng/mL cut-off was validated against 373 patient samples using the FDA-cleared Dimertest Gold EIA method (K945642).
Technological Characteristics
Latex immuno-assay using monoclonal antibodies attached to polystyrene latex particles. Qualitative/semi-quantitative visual agglutination on disposable black background reaction cards. Reportable range: 200–3,200 ng/mL. Storage: 2-8°C. No electronic or software components.
Indications for Use
Indicated for the rapid qualitative or semi-quantitative evaluation of circulating D-dimer in human plasma to assess fibrin degradation products.
Predicate Devices
Reference Devices
- Dimertest Gold EIA (K945642)
Related Devices
- K974596 — DIMERTEST LATEX ASSAY · Agen Biomedical , Ltd. · May 26, 1998
- K032419 — QUANTEX D-DIMER · Instrumentation Laboratory CO · Oct 24, 2003
- K974776 — MDA D-DIMER · Organon Teknika Corp. · Apr 16, 1998
- K992957 — ADVANCED D-DIMER · Dade Behring, Inc. · Jan 28, 2000
- K030740 — MODIFICATION TO TINA-QUANT D-DIMER TEST SYSTEM · Roche Diagnostics Corp. · Apr 1, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
K080069
Image /page/0/Picture/1 description: The image shows the logo for American Diagnostica Inc. along with their slogan "creating products for life". The logo is a stylized "ai" with horizontal lines resembling an American flag. The date "FEB 13 2008" is printed at the top of the image.
#### 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990 and 21 CFR part 807.92.
A. 510(k) Number: K974596
#### B. Purpose for the Submission:
To seek clearance for the existing DIMERTEST® product that was manufactured by AGEN Biomedical Ltd., but is now manufactured by American Diagnostica Inc. (ADI) due to the transfer of ownership of the DIMERTEST® product-line to ADI. The performance characteristics of the modified DIMERTEST® product and a summary of the modifications made to the DIMERTEST® product that had been cleared in 1988 (K882944) are described in the DIMERTEST® 510(k) submission K974596.
C. Measurand: D-Dimer
D. Type of Test: Latex Immuno Assay
E. Applicant:
Submitted by: American Diagnostica Inc. 500 West Avenue Stamford, CT 06902 Tel. 203 602-7777 Ext. 14 Fax 203 602-5553
Contact: Leigh Avres Director of Regulatory Affairs and Quality Assurance 203-602-7777 x 14
Summary Prepared: December 17, 2007
F. Proprietary and Established Name: DIMERTEST®
> 500 West Avenue, P.O. Box 110215, Stamford, CT 06911-0215 USA · Tel. +1.203.602.7777 · Fax +1.203.602.5553 2007DEC17 DIMERTEST® 510(K) Summary.doc Page 1 of 5
{1}------------------------------------------------
# G. Regulatory Information:
- 1. Regulation section: 864.7320
- 2. Classification: Class II
- 3. Product code: DAP
- 4. Panel: Hematology
# H. Intended Use:
The DIMERTEST® Latex Assay is intended for the rapid qualitative or semi-quantitative evaluation of circulating derivatives of cross-linked fibrin degradation products (D-dimer) in human plasma.
# I. Device Description:
The DIMERTEST® Latex Kit is a qualitative and semi-quantitative latex agglutination slide test for cross-linked fibrin degradation products in human plasma. The active ingredient in the DIMERTEST® product is the latex reagent. This reagent consists of highly specific D-dimer monoclonal antibodies attached to polystyrene latex particles. When mixed with the latex reagent, the presence of antigen (Ddimer) in a plasma sample results in agglutination of the latex particles which can be seen with the unaided eye. A semi-quantitative estimate of the Ddimer concentration can be made by preparing dilutions of a plasma sample. The concentration is determined according to a titration matrix.
Plasma from normal individuals is not expected to agglutinate DIMERTEST® Latex. Reactive fibrinolysis will be demonstrated by latex agglutination at a plasma concentration of approximately 200 ng/mL D-dimer.
# J. Substantial Equivalence Information:
- 1. Predicate device name(s): DIMERTEST®
- 2. Predicate 510(k) number: K882944
# 3. Comparison of the Modified DIMERTEST® Kit to the Predicate DIMERTEST® Kit:
| SEQ | Kit Component | Is there a<br>difference from<br>the Predicate<br>DIMERTEST® kit? | Description of the Change that Was Made<br>to the Modified DIMERTEST® Kit |
|-----|---------------|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Latex Reagent | Yes | The supplier for the latex starting material<br>was changed, the latex reagent in the kit<br>is now supplied with a dropper bottle, and<br>the cut-off value was changed from 250<br>ng/mL to 200 ng/mL. |
500 West Avenue, P.O. Box 110215, Stamford, CT 06911-0215 USA · Tel. +1.203.602.7777 · Fax +1.203.602.5553 2007DEC17 DIMERTEST® 510(K) Summary.doc Page 2 of 5
{2}------------------------------------------------
| SEQ | Kit Component | Is there a<br>difference from<br>the Predicate<br>DIMERTEST® kit? | Description of the Change that Was Made<br>to the Modified DIMERTEST® Kit |
|-----|------------------------|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| 2 | Positive Control | Yes | The formulation was not changed, but the<br>bottle in the kit is now provided with a<br>dropper tip. |
| 3 | Negative<br>Control | Yes | This was not provided in the predicate<br>DIMERTEST® kit. It is a liquid formulation<br>ready to use and it has a dropper tip. |
| 4 | Buffer Solution | No | No changes |
| 5 | Dropper Pipette | Yes | Deleted from the modified<br>DIMERTEST® kit. No longer necessary<br>because the latex reagent is now supplied<br>with a dropper tip. |
| 6 | Reading Cards | Yes | Changed to a pack of 10 disposable black<br>background cards with 8 reaction wells<br>marked on each card. |
| 7 | Plastic Stirrers | Yes | There was a minor change to the shape of<br>the stirrers. |
| 8 | Intended Use | No | No changes |
| 9 | Operating<br>Principle | No | No changes |
| 10 | Capture<br>Antibody | No | No changes |
| 11 | Assay Sample | No | No changes |
# K. Standard/Guidance Document Referenced (if applicable): N/A
# L. Test Principle:
The assay is based on a visible agglutination that occurs when a patient plasma containing D-Dimer is mixed with latex particles coated with monoclonal antibodies.
# M. Performance Characteristics (if/when applicable):
- 1. Analytical performance:
# a. Precision/Reproducibility:
Intra-assay (within run) reproducibility was determined by testing 10 replicates from three plasma pools that contained D-dimer titers ranging from 0 to 4. The test results for all ten replicates from each pool had the same values.
Inter-assay (run-to-run) reproducibility was determined by testing 10 plasma pools that had Ddimer titers ranging from 1 to 16. The assay values of the replicates from 10 runs of each plasma pool tested over a 28-day period did not vary by more than 1 titer.
> 500 West Avenue, P.O. Box 110215, Stamford, CT 06911-0215 USA · Tel. +1.203.602.7777 · Fax +1.203.602.5553 2007DEC17 DIMERTEST® 510(K) Summary.doc Page 3 of 5
{3}------------------------------------------------
#### b. Linearity/assay reportable range:
The reportable range for the semi-quantitative assay is 200 to 3,200 ng/mL Ddimer.
#### c. Traceability, Stability, and Expected Values:
The DIMERTEST® kit contains a positive control and a negative control.
The vial labels contain the expiration dates. Reagents are stored between 20 C and 8 €.
Normal plasma samples give negative results.
#### d. Analytical specificity:
The DIMERTEST® reagent showed no assay interference when plasma samples were spiked with interferents at the following concentrations:
Bilirubin 0.2 mg/dL Lipids (triglycerides) 30 mg/dL Hemoglobin 5.0 mg/dL Protein (gamma globulin) 0.06 g/mL
The DIMERTEST® reagents are insensitive to rheumatoid factor.
#### e. Assay cut-off:
200 ng/mL D-dimer
2. Comparison studies:
# a. Method Comparison of the Modified DIMERTEST® Reagent versus the Predicate DIMERTEST® Reagent:
#### Specificity Comparison
An in-house comparative study of 170 Blood Bank donor plasma samples from ostensibly healthy volunteers was performed. 95.3% of the normal results were negative by the modified DIMERTEST® reagent compared to 97.6% negative by the predicate DIMERTEST® reagent.
#### Sensitivity Comparison
145 plasmas from hospital patients who had a high probability of thrombotic conditions were analyzed in an in-house comparative study.
> 500 West Avenue, P.O. Box 110215. Stamford, CT 06911-0215 USA · Tel. +1.203.602.7777 · Fax +1.203.602.5553 2007DEC17 DIMERTEST® 510(K) Summary.doc Page 4 of 5
{4}------------------------------------------------
#### Clinical Cut-off Change
The new value of the diagnostic cut-off for the modified DIMERTEST® reagent was calculated to be 200 ng/ml. D-dimer with the linear curve generated from the N=145 sensitivity study plasma samples. The D-dimer values that were determined with the predicate DIMERTEST® reagent were plotted as the x-axis and the D-dimer values that were determined with the modified DIMERTEST® reagent were plotted as the y-axis. With a slope of 1.19, the new cutoff was calculated to be approximately 20% less than the predicate DIMERTEST® reagent cut-off of 250 ng/mL D-dimer.
The new cut-off value was validated with an additional 373 patient plasmas that were tested by the modified DIMERTEST® reagent and also tested concurrently by the FDA cleared Dimertest Gold ElA method (K945642).
#### N. Conclusion
The modified DIMERTEST® product is substantially equivalent to the predicate DIMERTEST® product based on the comparison summary and the performance characteristics.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Public Health Service
FEB 1 8 2008
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
American Diagnostica, Inc. C/O Leigh Ayres 500 West Avenue Stamford, Connecticut 06902
Re: k080069 Trade/Device Name: Dimertest Regulation Number: 21 CFR 864.7320 Regulation Name: Fibrinogen/Fibrin Degradation Products Assay Regulatory Class: Class II Product Code: DAP Dated: December 19, 2007 Received: January 11, 2008
Dear Ms. Ayres:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket
{6}------------------------------------------------
Page 2 -- American Diagnostica
notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), plcase contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Katt Beckerl
Robert L. Becker, Jr., MA., Ph.D. Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number:
K080069
K974596
Device name:
DIMERTEST®
Indications for Use:
The DIMERTEST® Latex Assay is intended for the rapid qualitative or semi-quantitative evaluation of circulating derivatives of cross-linked fibrin degradation products (D-dimer) in human plasma.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/7/Picture/9 description: The image shows a square with a check mark inside. The square is outlined in black, and the check mark is also black. The check mark starts at the bottom left corner of the square, goes up to the top right corner, and then goes down to the bottom right corner.
.
Prescription Use (Per Title 21 CFR part 801.109)
OR
Over-The-Counter-Use
Josephine Bautista
Office of In Vitro Diagnostic Device Evaluation and Sai
510(k) KD80069