BIOGEL SENSOR SURGEON'S GLOVE WITH NON-PYROGENIC STATEMENT

K071700 · Molnlycke Health Care Us, LLC · KGO · Sep 11, 2007 · General, Plastic Surgery

Device Facts

Record IDK071700
Device NameBIOGEL SENSOR SURGEON'S GLOVE WITH NON-PYROGENIC STATEMENT
ApplicantMolnlycke Health Care Us, LLC
Product CodeKGO · General, Plastic Surgery
Decision DateSep 11, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4460
Device ClassClass 1

Intended Use

The Biogel® Sensor™ Surgeon's Glove is a sterile disposable device made of natural rubber latex that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.

Device Story

Biogel® Sensor™ Surgeon's Glove is a sterile, powder-free, polymer-coated natural rubber latex surgical glove. Designed for use in surgical settings by clinicians to provide a protective barrier against infectious materials and contaminants. Device features low protein content (≤ 50 micrograms/gram) and non-pyrogenic properties. It is a disposable, single-use device.

Clinical Evidence

No clinical data was required. Substantial equivalence is supported by bench testing, including compliance with ASTM D3577 for dimensions, physical properties, and freedom from holes; ISO 10993-1 for biocompatibility; and LAL testing for endotoxin concentration (≤ 0.50 EU/ml).

Technological Characteristics

Natural rubber latex surgical glove; polymer coated; powder-free. Meets ASTM D3577 (dimensions, physical properties, freedom from holes). Biocompatibility per ISO 10993-1. Non-pyrogenic with endotoxin concentration ≤ 0.50 EU/ml. Protein content ≤ 50 micrograms/gram.

Indications for Use

Indicated for use as a protective barrier on the hands of clinicians in surgical settings to prevent exposure to infectious materials and contaminants.

Regulatory Classification

Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## K071700 | 510(k) SUMMARY | | | | | | | | | | | | | | |-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|----------|------------|------------------|---------------------|------------------|--------------------|------------------|------------------|-------------------|-----------------------------------------|-------------| | 1. Applicant: | Mölnlycke Health Care US, LLC<br>5550 Peachtree Parkway Suite 500<br>Norcross, GA 30092 | | | | | | | | | | | | | | 2. Contact Person: | Steven Dowdley, RAC<br>Director of Regulatory Affairs<br>Tel.: 678-250-7930<br>Fax: 678-250-7979 | | | | | | | | | | | | | | | or | | | | | | | | | | | | | | | MKA Consulting<br>Marcia K. Ash, RAC<br>Tel.: 404-906-0492 | | | | | | | | | | | | | | 3. Device Name: | Biogel® Sensor™ Surgeon's Glove with non-pyrogenic statement | | | | | | | | | | | | | | Common Name: | Surgical Glove (CFR 878.4460) | | | | | | | | | | | | | | Classification: | Class I | | | | | | | | | | | | | | 4. Predicate Device: | K053443 - Eclipse, Non-pyrogenic, Latex, Powder-free, Glove | | | | | | | | | | | | | | 5. Device Description: | The Biogel® Sensor™ Surgeon's Glove is a sterile powder free, polymer coated latex surgical glove. The glove contains 50 micrograms or less of total water extractable protein per gram. | | | | | | | | | | | | | | 6. Intended Use of the Device: | The Biogel® Sensor™ Surgeon's Glove is a disposable device made of natural rubber latex that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants. | | | | | | | | | | | | | | 7. Technological Characteristics of the Device: | The Biogel® Sensor™ Surgeon's Glove characteristics are summarized below as compared to ASTM requirements and to the predicate devices. | | | | | | | | | | | | | | | <table><tr><td>Characteristic</td><td>Standard</td></tr><tr><td>Dimensions</td><td>Meets ASTM D3577</td></tr><tr><td>Physical Properties</td><td>Meets ASTM D3577</td></tr><tr><td>Freedom from Holes</td><td>Meets ASTM D3577</td></tr><tr><td>Biocompatibility</td><td>Meets ISO 10993-1</td></tr><tr><td>LALTest - final endotoxin concentration</td><td>≤ .50 EU/ml</td></tr></table> | Characteristic | Standard | Dimensions | Meets ASTM D3577 | Physical Properties | Meets ASTM D3577 | Freedom from Holes | Meets ASTM D3577 | Biocompatibility | Meets ISO 10993-1 | LALTest - final endotoxin concentration | ≤ .50 EU/ml | | Characteristic | Standard | | | | | | | | | | | | | | Dimensions | Meets ASTM D3577 | | | | | | | | | | | | | | Physical Properties | Meets ASTM D3577 | | | | | | | | | | | | | | Freedom from Holes | Meets ASTM D3577 | | | | | | | | | | | | | | Biocompatibility | Meets ISO 10993-1 | | | | | | | | | | | | | | LALTest - final endotoxin concentration | ≤ .50 EU/ml | | | | | | | | | | | | | | 8. Performance Data | The performance data are summarized above. | | | | | | | | | | | | | | 9. Clinical Data | No clinical data was required. | | | | | | | | | | | | | | 10. Conclusion: | The Biogel® Sensor™ Surgeon's Glove meets the technological characteristics of ASTM D3577 and is substantially equivalent to the predicate devices cited in this 510(k) summary. | | | | | | | | | | | | | SEP 11 2007: {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a person with outstretched arms, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Steven Dowdley Director of Regulatory Affairs Molnlkcke Healthcare, US LLC 5550 Peachtree Parkway, Suite 500 Norcross, Georgia 30092 SEP 1 1 2007 Re: K071700 Trade/Device Name: Biogel Sensor, Latex Surgeon's Glove with Non-pyrogenic Claim and Protein Content Labeling Claim (50 micrograms or less) Regulation Number: 878,4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: August 22, 2007 Received: August 27, 2007 Dear Mr. Dowdley: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indication for Use 510(k) Number: K071700 Device Name: Biogel Sensor, Latex Surgeon's Glove with Non-pyrogenic Claim and Protein Content Labeling Claim (50 micrograms or less) Indication For Use: The Biogel® Sensor™ Surgeon's Glove is a sterile disposable device made of natural rubber latex that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants. Prescription Use (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Shuse R. Murphy (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Device 510(k) Number: Concurrence of CDRH, Office of Device Evaluation
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%