HOMECHECK MULTIPLE DRUG CUP TEST
K071489 · Tianjin New Bay Bioresearch Co., Ltd. · DKZ · Mar 28, 2008 · Clinical Toxicology
Device Facts
| Record ID | K071489 |
| Device Name | HOMECHECK MULTIPLE DRUG CUP TEST |
| Applicant | Tianjin New Bay Bioresearch Co., Ltd. |
| Product Code | DKZ · Clinical Toxicology |
| Decision Date | Mar 28, 2008 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.3100 |
| Device Class | Class 2 |
Intended Use
The assay provides a simple and rapid analytical screening procedure to detect single or multiple different abused drugs (Amphetamine, Methamphetamine, Benzoylecgonine, Benzodiazepine, Marijuana, Morphine, Phencyclidine, Methadone, Oxycodone, Tricyclic Antidepressant, and Barbiturates in human urine. The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography /mass spectrometry (GC/MS) is the preferred confirmatory method. HPLC is preferred confirmatory method for Tricyclic Antidepressant. This test is intended for over the counter (OTC) consumer use as the first step in a 2-step process to provide consumers with information concerning the presence or absence of the above stated drugs in a urine sample. Information regarding the confirmatory testing - the second step in the process-, along with materials for shipping the urine specimen to the laboratory is included with the test. There are no uniformly recognized levels for Benzodiazepine, Oxycodone, Tricyclic Antidepressant, and Barbiturates. The test is not intended to screen individuals who are prescribed these drugs by a physician; the test my yield positive results for individuals taking such drugs, as prescribed.
Device Story
HomeCheck Rapid Multiple(X) Abuse Drug Screen Test Cup is a lateral flow chromatographic immunoassay for qualitative detection of drugs of abuse in human urine. Device consists of absorbent strips containing immobilized drug conjugates and colloidal gold-labeled antibodies. Sample flows through strip; drug in specimen competes with immobilized conjugate for antibody sites. Absence of magenta band in test zone indicates positive result (drug concentration above cutoff); presence of two bands indicates negative result. Control zone confirms proper wicking. Intended for OTC home use as first step in 2-step process; positive results require confirmatory testing via GC/MS or HPLC. Provides rapid preliminary screening; helps consumers identify potential drug presence. No instrumentation required; visual interpretation only.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing comparing the subject device to the predicate device and analytical performance verification.
Technological Characteristics
Lateral flow immunochromatographic assay; colloidal gold-labeled antibody conjugate; porous membrane support. Visual read, single-use test cup. No instrumentation, electronics, or software. Storage below 28°C.
Indications for Use
Indicated for OTC qualitative screening of Amphetamine, Methamphetamine, Benzoylecgonine, Benzodiazepine, Marijuana, Morphine, Phencyclidine, Methadone, Oxycodone, Tricyclic Antidepressant, and Barbiturates in human urine. Not intended for screening individuals prescribed these medications by a physician.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
- New Bay Forsure Multiple Drug Screen Test Device (K052882)
Related Devices
- K052882 — FORSURE RAPID ONE STEP MULTIPLE(X) ABUSE DRUG SCREEN TEST CUP DEVICE · Tianjin New Bay Bioresearch Co., Ltd. · Feb 15, 2006
- K131645 — INSTANT-VIEW MULTI-DRUG OF ABUSE URINE TEST (CUP, PANEL CASSETTE) · Alfa Scientific Designs, Inc. · Oct 7, 2013
- K250727 — AllTest Multi-Drug Rapid Urine Test Cup; AllTest Multi-Drug Urine Test Cup · Hangzhou AllTest Biotech Co., Ltd. · Apr 4, 2025
- K193480 — BIOEASY Multi-Drug Test Cup · Shenzhen Bioeasy Biotechnology Co., Ltd. · Jan 31, 2020
- K232659 — Dochek® Multi-Drug Urine Test Cup Rx, Dochek® Multi-Drug Urine Test Cup · Guangzhou Decheng Biotechnology Co., Ltd. · Dec 13, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
510(k) Number: Designed K #: k071489
#### Submitter:
Tianjin New Bay Bioresearch Co., Ltd. #3 Jian She Rd, Ba Li Tai Industry Area Jin Nan District, Tianjin, China Telephone: 86-22-28751515 Facsimile: 86-222-875-1516
#### Contact Person:
Armando Torrescano New Product Development Director Telephone: (760) 822-6517 Facsimile: (760) 602-2999
### Preparation Date:
Jan 23, 2008
#### Device Information:
Device Classification Name: Immunoassay of Amphetamine, Methamphetamine, Benzoylecgonine, Morphine, Phencyclidine. Benzodiazepine, Barbiturates, Marijuana. Methadone , Oxycodone, and Tricyclic Antidepressant.
Common/Usual Name: Immunoassay Test System for Detection of Single and Multiple(X) Abuse Drug Screen Test Cup Device in Human Urine.
Proprietary Name: HomeCheck Rapid Multiple(X) Abuse Drug Screen Test Cup Device for Amphetamine, Methamphetamine, Benzoylecgonine, Benzodiazepine. Marijuana. Morphine, Phencyclidine, Methadone, Oxycodone, Tricyclic Antidepressant, and Barbiturates.
Regulation Number: 21 CFR§862.3650
Regulatory Name:
{1}------------------------------------------------
Amphetamine, Methamphetamine, Benzoylecgonine, Benzodiazepine, Marijuana, Morphine, Phencyclidine, Methadone, Oxycodone , Tricyclic Antidepressant, and Barbiturates test system.
Product Code: DJG Regulatory Class: Class II
#### Predicate Devices:
HomeCheck Rapid Single and Multiple (X) Abuse Drug Screen Test Cup Device are substantially equivalent to New Bay Forsure Multiple Drug Screen Test Device ( Professional Use ), cleared by FDA( K052882) and GC/MS for its stated intended use.
# Device Description:
New Bay HomeCheck Rapid Single and Multiple(X) Abuse Drug Screen Test Cup Device consists of a or multiple chromatographic absorbent strip in which the drug or drug metabolites in the sample compete with a drug conjugate immobilized on a porous membrane support for the limited antibody sites. As the test sample flows through the absorbent strip, the Colloidal Gold labeled antibody- conjugate binds to the free drug in the specimen forming an antibody-antigen complex. This complex competes with immobilized antigen conjugate in the Test reaction zone and will not produce a magenta color band when the drug is above the detection level of 1000ng/ml of Amphetamine, 1000ng/ml of Methamphetamine, 50ng of THC, 2000ng/ml of Morphine, 300ng of Benzoylecognine, 25ng/ml of Phencyclidine, 300ng/ml of Benzodiazepine, 300ng/ml of Methadone, 100ng/ml of Oxycodone, 1000ng/ml of Tricyclic Antidepressant, and 300ng/ml of Barbiturates. Unbound colloidal gold-labeled antibody conjugate binds to the reagent in the negative control zone, producing a magenta color band, demonstrating that the reagents and device are functioning correctly. A NEGATIVE specimen produces two distinct color bands in both the specific drug test region and control area. A POSITIVE specimen produces only one color band in the control area and no color band on the specific drug test region. There is no meaning attributed to color or its intensity for either line. To serve as a procedural control, a colored line will always appear at the control line region, indicating that proper volume of specimen has been added and membrane wicking has occurred.
# Intended Use:
The HomeCheck Single or Multiple (X) Drug Screen Strip of Amphetamine, Methamphetamine, Benzoylecgonine, Benzodiazepine, Marijuana, Morphine, Phencyclidine, Methadone , Oxycodone , Tricyclic Antidepressant, and Barbiturates Test device are a Chromatographic immunoassay for qualitative determination of the presence of Amphetamine at a cutoff concentration of 1000ng/ml, Methamphetamine at a cutoff concentration of 1000ng/ml, THC at a cutoff concentration of 50ng/ml, Morphine at a cutoff concentration of 2000ng/ml, Benzovlecgonine at a cutoff concentration of 300ng/ml. Phencyclidine at a cutoff concentration of 25ng/ml, Benzodiazepine at cutoff concentration of 300ng/ml for Oxazepam, Methadone at cutoff concentration of 300ng/ml. Oxycodone at cutoff concentration of 100ng/ml. Tricvelic Antidepressant at cutoff concentration of 1000ng/ml for Nortriptyline HCl. and Barbiturates at cutoff concentration of 300ng/ml for Secobarbital. The assay provides a simple and rapid analytical screening procedure to detect Amphetamine, THC, Morphine, Methamphetamine.
{2}------------------------------------------------
Benzoylecognine, Phencyclidine, Benzodiazepine, Methadone, Oxycodone, Tricyclic Antidepressant, and Barbiturates in human urine. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method.
# Comparison to Predicate Device(s):
Tianjin New Bay HomeCheck Multiple Rapid Drug Screen Test ( K 071489 ) is substantially equivalent to Tianjin New Bay ForSure Multiple Drug Screen Test Device cleared by FDA(510K 052882 ).
| Device<br>Characteristics | Subject Device (s)<br>Tianjin New Bay HomeCheck<br>Multiple (X) Drug Screen Test device<br>.(K 071489) | Predicate Device(s)<br>Tianjin New Bay ForSure Multiple<br>(X)Drug Screen Test Device<br>(K 052882) |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | HomeCheck Multiple Drug Screen<br>Immunochromatographic<br>Qualitative test.<br>The assay provides a simple and rapid<br>analytical screening procedure to detect<br>different abuse drugs in human urine for<br>the home use. | ForSure Multiple Drug Screen<br>Immunochromatographic<br>Qualitative test.<br>The assay provides a simple and rapid<br>analytical screening procedure to detect<br>different abuse drugs in human urine<br>for the professional use |
| Analytes | Amphetamine, Methamphetamine,<br>Benzoylecgonine, Benzodiazepine,<br>Marijuana, Morphine, Phencyclidine,<br>Methadone, Oxycodone, Tricyclic<br>Antidepressant, and Barbiturates | Amphetamine, Methamphetamine,<br>Benzoylecgonine, Benzodiazepine,<br>Marijuana, Morphine, Phencyclidine,<br>Methadone, Oxycodone, Tricyclic<br>Antidepressant, and Barbiturates<br>Propoxphene. |
| Cutoff | AMP: 1000 ng/ml, MET:1000 ng/ml,<br>BEG: 300ng/ml, THC: 50ng/ml,<br>MOR: 2000ng/ml, PCP: 25ng/ml, BZD:<br>300ng/ml, MAD: 300ng/ml,<br>OXY: 100ng/ml, TCA: 1000ng/ml | AMP: 1000 ng/ml, MET:1000 ng/ml,<br>BEG: 300ng/ml, THC: 50ng/ml,<br>MOR: 2000ng/ml, PCP: 25ng/ml, BZD:<br>300ng/ml, MAD: 300ng/ml,<br>OXY: 100ng/ml, TCA: 1000ng/ml<br>PPX: 300 ng/ml. |
| Assay time | 5 minutes | 5 minutes |
| Preliminary Positive<br>Reconfirm by GC/Mass | Yes | Yes |
| Matrix | Urine | Urine |
| Calibrator | None | None |
| Instrument | None, Visual read single use | None, Visual Read single use |
| Calibration of Reagent | None | None |
| Storage | Below 28 °C until expiration | Below 28 °C until expiration |
{3}------------------------------------------------
### Summary:
The information provided in this pre-market notification demonstrates that HomeCheck Rapid single or Multiple (X) Abuse Drug Screen Test Cup Device (Home Use) is substantially equivalent to Tianjin New Bay Forsure Multiple (X) Drug Screen Cassette Test system (Professional Use) and GC/MS or HPLC.
Substantial equivalence was demonstrated through comparison of intended use and physical properties to the commercially available and analytical predicate devices. The information supplied in this pre-market notification provides reasonable assurance that the HomeCheck Rapid Single or Multiple(X) Abuse Drug Screen Test Cup Device is safe and effective for its stated intended use.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" around the perimeter. Inside the circle is an emblem featuring a stylized eagle with its wings spread.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR 2 8 2008
Tianjin New Bay Bioresearch Co., Ltd. c/o Mr. Armando Torrescano Aventir Biotech, New Product Development 3108 Avenida Olmeda Carlsbad, CA 92009
Re: k071489
Trade/Device Name: HomeCheck™ Multiple Drug Cup Test Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine Test System. Regulatory Class: Class II Product Code: DKZ, DIS, JXM, LDJ, DIO, DJC, DJR, DJG, LCM, LFH Dated: March 14. 2008 Received: March 17, 2008
Dear Mr. Torrescano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Jean M. Cooper, M.S., D.v.M.
Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (k071489):
Device name: HomeCheck™ Multiple Drug cup test device for Amphetamine (cutoff at 1000ng/ml), Methamphetamine (cutoff at 1000ng/ml), Benzoylecgonine (cutoff at 300ng/ml), Benzodiazepine (cutoff at 300ng/ml for Oxazepam), Marijuana (cutoff at 50ng/ml), Morohine (cutoff at 2000ng/ml), Phencyclidine (cutoff at 25ng/ml ), Methadone (Cutoff at 300ng/ml), Oxycodone (Cutoff at 100ng/ml), Tricyclic Antidepressant (Cutoff at 1000ng/ml for Nortriptyline HCl), and Barbiturates (Cutoff at 300ng/ml for Secobarbital).
#### Indications for Use:
The assay provides a simple and rapid analytical screening procedure to detect single or multiple different abused drugs (Amphetamine, Methamphetamine, Benzoylecgonine, Benzodiazepine, Marijuana, Morphine, Phencyclidine, Methadone, Oxycodone, Tricyclic Antidepressant, and Barbiturates in human urine.
The assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography /mass spectrometry (GC/MS) is the preferred confirmatory method. HPLC is preferred confirmatory method for Tricyclic Antidepressant.
# Special Conditions for Use
This test is intended for over the counter (OTC) consumer use as the first step in a 2-step process to provide consumers with information concerning the presence or absence of the above stated drugs in a urine sample. Information regarding the confirmatory testing - the second step in the process-, along with materials for shipping the urine specimen to the laboratory is included with the test. There are no uniformly recognized levels for Benzodiazepine, Oxycodone, Tricyclic Antidepressant, and Barbiturates. The test is not intended to screen individuals who are prescribed these drugs by a physician; the test my yield positive results for individuals taking such drugs, as prescribed.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR AND/OR
Over-The-Counter Use X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Carol G. Benson
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety 510(k): KO1148