FREELITE HUMAN KAPPA AND LAMBDA FREE KITS FOR USE ON THE SPAPLUS ANALYZER
K062372 · The Binding Site, Ltd. · DFH · Jun 21, 2007 · Immunology
Device Facts
Record ID
K062372
Device Name
FREELITE HUMAN KAPPA AND LAMBDA FREE KITS FOR USE ON THE SPAPLUS ANALYZER
Applicant
The Binding Site, Ltd.
Product Code
DFH · Immunology
Decision Date
Jun 21, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5550
Device Class
Class 2
Indications for Use
This kit is intended for the quantitation of kappa free light chains in serum on The Binding Site SpaPlus. Measurement of the various amounts of the different types of light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom's macroglobulinemia, amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus in conjunction with other laboratory and clinical findings.
Device Story
In vitro diagnostic assay for quantitative measurement of kappa and lambda free light chains in human serum; utilizes nephelometry on The Binding Site SPAplus™ analyzer. Input: serum sample; reagent containing polyclonal monospecific sheep antibody coated onto polystyrene latex. Principle: antibody-antigen reaction forms insoluble immune complexes; light scattering measured as signal proportional to antigen concentration. Output: quantitative concentration of free light chains. Used in clinical laboratory settings by trained personnel. Results aid clinicians in diagnosing and monitoring monoclonal gammopathies and related systemic diseases. Prozone detection flags potential antigen excess to prevent falsely low values.
Clinical Evidence
No clinical data provided in the document; the submission relies on bench testing and performance validation for the intended use.
Technological Characteristics
Quantitative nephelometric assay. Reagents: polyclonal monospecific sheep antibody coated on polystyrene latex. Detection: light scattering by immune complexes. Instrument: SPAplus™ analyzer. Matrix: serum. Calibration: multi-point curve using provided calibrators. Software: automated kinetic analysis for prozone detection.
Indications for Use
Indicated for the quantitation of kappa and lambda free light chains in serum to aid in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom's macroglobulinemia, amyloidosis, light chain deposition disease, and systemic lupus erythematosus in patients requiring clinical laboratory assessment.
Regulatory Classification
Identification
An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.
Related Devices
K150658 — Optilite FreeLite Kappa Free Kit, Optilite Freelite Lambda Free Kit · The Binding Site Group , Ltd. · Aug 26, 2015
K140396 — FREELITE HUMAN LAMBDA FREE KIT · The Binding Site Group , Ltd. · Apr 16, 2014
K070900 — FREELITE HUMAN KAPPA AND LAMBDA FREE DIAGNOSTIC TEST KITS FOR USE ON THE ROCHE COBAS INTEGRA 400/400 PLUS ANALYZER KIT · The Binding Site, Ltd. · Nov 30, 2007
K120056 — HUMAN KAPPA AND LAMBDA FREE KITFOR USE ON THE SPA PLUS ANALYSER · The Binding Site Group , Ltd. · Apr 1, 2013
K173732 — Optilite Freelite Mx Kappa Free Kit, Optilite Freelite Mx Lambda Free Kit · The Binding Site Group , Ltd. · Aug 23, 2018
Submission Summary (Full Text)
{0}
Page 1 of 6
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k062372
B. Purpose for Submission:
New analyzer
C. Analyte:
Kappa (κ) free light chains and Lambda (λ) free light chains
Type of Test:
Quantitative nephelometry
Applicant:
The Binding Site, Ltd
Proprietary and Established Names:
FREELITE™ Human Kappa Free Kit for use on the SPAplus™
FREELITE™ Human Lambda Free Kit for use on the SPAplus™
D. Regulatory Information:
1. Regulation section:
21CFR §866.5550 Immunoglobulin (light chain specific) immunological test system
2. Classification:
Class II
3. Product Code:
DFH - Kappa antigen, antiserum, control
DEH - Lambda antigen, antiserum, control
4. Panel:
Immunology (82)
Intended Use:
5. Intended use(s):
The intended use for each kit is the same as the predicate with the exception of the matrices and analyzer.
FREELITE™ Human Kappa Free Kit for use on the SPAplus™: The kit is intended for quantitation of kappa free light chains in serum on The Binding Site SpaPlus. Measurement of the different types of light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom's macroglobulinemia, amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus in conjunction with other laboratory and clinical findings.
FREELITE™ Human Lambda Free Kit for use on the SPAplus™: The kit is intended for quantitation of lambda free light chains in serum on The Binding Site SpaPlus. Measurement of the different types of light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom's macroglobulinemia, amyloidosis, light chain deposition disease and connective
{1}
Page 2 of 6
tissue diseases such as systemic lupus erythematosus in conjunction with other laboratory and clinical findings.
6. Indication(s) for use:
Same as above
7. Special condition for use statement(s):
Prescription use only.
8. Special instrument Requirements:
The Binding Site $\mathrm{SPA}_{\mathrm{plus}}^{\mathrm{TM}}$
## Device Description:
Latex reagent consists of polyclonal monospecific sheep antibody coated onto polystyrene latex in the presence of preservatives. The calibrator and controls consists of human sera that contain free polyclonal kappa or lambda free light chains, stabilized in a liquid form with preservatives.
## E. Substantial Equivalence Information:
1. Predicate device name(s):
FREELITE™ Human Kappa Free Kit for use on the Dade Behring Nephelometer™ II
FREELITE™ Human Lambda Free Kit for use on the Dade Behring Nephelometer™ II
2. Predicate K number(s):
k040009
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indications for Use | Measurement of the different types of light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom’s macroglobulinemia, amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus in conjunction with other laboratory and clinical findings. | Same |
| Detection Method | Nephelometry | Same |
| Differences | | |
| Stability | Device | Predicate |
| Instrument | The Binding Site $\mathrm{SPA}_{\mathrm{plus}}^{\mathrm{TM}}$ | Dade Behring Nephelometer™ BN II |
| Matrix | Serum only | Serum or urine |
{2}
Page 3 of 6
# Standard/Guidance Document Referenced (if applicable):
CLSI document EP5-A: Evaluation of Precision Performance of Clinical Chemistry; CLSI document EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures - A Statistical Approach.
# Test Principle:
A test sample is added to a solution containing the appropriate antibody in a reaction cuvette. A beam of light is passed through the cuvette and is increasingly scattered by the formation of insoluble immune complexes as the antibody-antigen reaction proceeds. The antibody in the cuvette is in excess so that the amount of immune complex formed is proportional to the antigen concentration. A series of calibrators of known antigen concentration are assayed to generate a calibration curve which is then used to determine the unknown concentration of the sample. The polyclonal, sheep antibody that is specific for the free light chain of interest, is coated onto polystyrene latex to increase the relative light-scattering signal of the immune complex formation, and thereby increase the sensitivity of the assay.
# F. Performance Characteristics (if/when applicable):
## 1. Analytical performance:
### a. Precision/Reproducibility:
A precision study was performed according to CLSI document EP5-A: Evaluation of Precision Performance of Clinical Chemistry Approved Guideline. Three different serum samples representing the low, medium and high levels of the measuring range were tested over 21 days, 2 runs per day, in duplicate. Acceptance criteria ±20% were met.
| Kappa Precision | Within run | | Between-run | | Between-day | | Total Precision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Low (mean 7.21 mg/L) | 0.24 | 3.3 | 0.30 | 4.2 | 0.81 | 11.2 | 0.90 | 12.5 |
| Medium (mean 35.72 mg/L) | 0.56 | 1.6 | 0.69 | 1.9 | 3.17 | 9.0 | 3.30 | 9.3 |
| High (mean 123.77 mg/L) | 2.22 | 1.8 | 2.90 | 2.3 | 8.22 | 6.6 | 8.99 | 7.3 |
| | | | | | | | | |
| Lambda Precision | Within run | | Between-run | | Between-day | | Total Precision | |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Low (mean 10.35 mg/L) | 0.35 | 3.4 | 0.23 | 2.2 | 0.74 | 7.2 | 0.85 | 8.2 |
| Medium (mean 35.09 mg/L) | 0.85 | 2.4 | 0.0 | 0.0 | 1.55 | 4.4 | 1.76 | 5.0 |
| High (mean 142.10 mg/L) | 2.81 | 2.0 | 3.40 | 2.4 | 9.62 | 6.8 | 10.58 | 7.4 |
κ/λ ratio Precision: A precision study was also performed testing two levels of a control serum sample for both kappa and lambda for 10 separate runs in one day, in replicates of three, on 3 different SPAplus™ analyzers to determine the precision of the contrived κ/λ ratios: SD< 0.08, %CV <6.65%.
### b. Linearity/assay reportable range:
Linearity studies were performed according to CLSI document EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures - A Statistical Approach. Both monoclonal (myeloma) and polyclonal (SLE) samples previously identified as containing high levels of free kappa or
{3}
Page 4 of 6
lambda light chains were diluted in a normal serum sample. Polyclonal samples were serially diluted 1:10 and linearity measured across the measuring range (kappa 4.0-180 mg/L; lambda 6.0-220mg/L).
| Light Chain | Type | Slope (95% CI) | Y-intercept (95% CI) | R² | % recovery range |
| --- | --- | --- | --- | --- | --- |
| Kappa (224161 kit lot 7699) | Polyclonal | 1.0283 (0.9814 to 1.075) | -0.8611 (-4.513 to 2.7907) | 0.9969 | 83.19 to 109.20 |
| Lambda (224161 kit lot 7695) | Polyclonal | 1.0035 (0.9925 to 1.015) | 0.1024 (-0.731 to 0.936) | 0.9998 | 95.56 to 104.98 |
Myeloma (monoclonal antibody) samples were diluted to give final percentages of 100%, 80%, 60%, 40%, 20%, 10% and 0% which covers the measuring ranges. Three samples for each kappa and lambda were evaluated and % recovery ranged 85.68 to 102.92% for kappa and 84.65 to 115.39% for lambda. The linear regression data below represents data from one myeloma sample. Because linearity for monoclonal samples varies, a limitation was added to the package insert with regard to linearity of free light chains. A limitation was added to the package insert with regard to linearity of free light chains.
| Sample | Slope (95% CI) | Y-intercept (95% CI) | R² |
| --- | --- | --- | --- |
| Kappa (monoclonal sample) | 1.0036 (0.963 to 1.044) | 0.543 (-4.1595 to 5.2455) | 0.9975 |
| Lambda (monoclonal sample) | 0.9942 (0.9766 to 1.0118) | -2.4962 (-8.6025 to 3.61) | 0.9994 |
c. Traceability (controls, calibrators, or method):
There are no reference standards for this method. A calibrator and two controls are provided with each kit. No changes were made to the calibrator and controls from the previous cleared submission.
d. Detection limit:
The limit of detection was evaluated by assaying a blank 20 times for both kappa and lambda free light chain assays. The mean plus 3SD was calculated. The lower limit of detection equates to -0018 delta absorbance units for kappa free and 0.0006 delta absorbance units for lambda free. The limit of detection was demonstrated to be below the measuring range of the assay.
e. Analytical specificity:
i. Interference- Susceptibility to interference was assessed by adding high concentrations of triglycerides (Intralipid), hemoglobin and bilirubin to a test serum sample containing known concentrations of polyclonal free kappa and free lambda light chains and compared to matched samples without interferent. Samples tested in triplicate.
{4}
Page 5 of 6
| Analyte (mg/L.) | % Interference | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | κ (~1.4 mg/L) | κ (~10.0 mg/L) | λ (~2.5 mg/L) | λ (~6.8 mg/L) | κ high control (~31.2 mg/L) | κ low control (~16.5 mg/L) | λ high control (~59.4 mg/L) | λ low control (~28.5 mg/L) |
| Bilirubin 200 mg/L | | -5.0% | | 2.7% | | | | |
| Hemoglobin 3 g/L | 2.10% | | -1.55% | | -2.68% | -2.5% | -9.01% | -5.28% |
| Triglycerides (Intralipid) | -9.1% (0.1% Intralipid) | | | -3.0% (0.3% Intralipid) | | | | |
Interference by whole immunoglobulins IgG, IgA and IgM serum samples and rheumatoid arthritis was evaluated by spiking into human sera containing low and high levels of kappa or lambda, and compared to saline spiked matched samples.
| | % Difference | | | |
| --- | --- | --- | --- | --- |
| Sample (conc. in mg/L) | IgG to 10 g/L | IgA to 2 g/L | IgM to 1 g/L | RF to 320 IU/mL |
| Kappa high (~148.43) | 2.77% | 2.21% | 0.96% | 15.52% |
| Kappa low (~21.31) | 14.9% | 19.8% | -1.85% | -2.68% |
| Lambda high (~148.29) | -0.80% | -3.76% | 0.00% | -3.63% |
| Lambda low (~19.35) | 4.67% | 8.07% | -0.75% | 1.33% |
ii. Cross-Reactivity-cross-reactivity to other disease conditions was not evaluated.
iii. Antigen excess- The $\mathrm{SPA}_{\mathrm{plus}}^{\mathrm{TM}}$ analyzer is set up to flag the user to prozone effects based on the kinetics of the reaction. 147 myelomas (monoclonal) were used to establish prozone parameters. A limitation warns that due to the unpredictable nature of monoclonals, falsely low values may be reported. Antigen excess for polyclonal samples does not occur up to $330~\mathrm{mg / L}$ for both free kappa and lambda light chains.
f. Assay cut-off:
See Expected values
2. Comparison studies:
a. Method comparison with predicate device:
The Binding Site FREELITE™ assays for kappa and lambda were evaluated on the new analyzer $\mathrm{SPA}_{\mathrm{plus}}^{\mathrm{TM}}$ and compared to the results obtained with the kits on the predicate analyzer Dade Behring BNII. A total of 176 samples were assayed spanning the measuring ranges (kappa sample range 5.38 to $1978~\mathrm{mg / L}$ ; lambda sample range 2.1 to $6620~\mathrm{mg / L}$ ) of which 119 were normal serum samples and 56 were known/suspected monoclonal gammopathy patients (including multiple myeloma) or systemic lupus erythematosus (SLE) patients. Samples were obtained from a university clinical lab and a blood transfusion service located in the UK (patient age 18-65). All samples were stored at $-20^{\circ}\mathrm{C}$ prior to use. Passing-Bablock linear regression analysis was performed and yielded the following data:
{5}
| | Sample | n | Slope (95% CI) | Y-intercept (95% CI) | r |
| --- | --- | --- | --- | --- | --- |
| Kappa | Normal | 119 | 1.000(0.945 to 1.063) | -1.700(-2.525 to -1.088) | 0.9564 |
| | Clinical | 56 | 1.162(1.092 to 1.212) | 1.800(0.278 to 2.911) | 0.9912 |
| | Overall | 175 | 1.091(1.043 to 1.162) | -2.426(-3.281 to -1.902) | 0.9564 |
| Lambda | Normal | 120 | 0.912(0.842 to 0.984) | 1.192(0.157 to 2.222) | 0.9612 |
| | Clinical | 56 | 1.026(0.983 to 1.067) | 0.483(-0.042 to 1.833) | 0.9975 |
| | Overall | 176 | 1.003(0.982 to 1.032) | 0.095(0.982 to 1.032) | 0.9977 |
| κ/λ Ratio | All | 176 | 0.892(0.841 to 0.934) | 0.003(-00.23 to 0.049) | 0.9954 |
b. Matrix comparison:
Serum is the only sample specimen.
3. Clinical studies:
a. Clinical sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when $a$ and $b$ are not applicable):
Not applicable
4. Clinical cut-off:
Refer to expected values
5. Expected values/Reference range:
Adult serum ranges were obtained by measuring the light chain concentrations of 282 normal adult sera (age 20-90 years old) using the assays on the Dade Behring BNII.
| Normal Adult serum | Mean conc. | Median conc. | 95th percentile range |
| --- | --- | --- | --- |
| Free Kappa | 8.36 mg/L | 7.30 mg/L | 3.30 to 19.40 mg/L |
| Free Lambda | 13.43 mg/L | 12.40 mg/L | 5.71 to 26.30 |
| | Mean | Median | Total Range |
| κ/λ ratio | 0.63 | 0.60 | 0.26 to 1.65 |
# N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
# O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.