HEMOGLOBIN ALC CONTROL
K060570 · Compass Bioscience · JJX · Apr 13, 2006 · Clinical Chemistry
Device Facts
| Record ID | K060570 |
| Device Name | HEMOGLOBIN ALC CONTROL |
| Applicant | Compass Bioscience |
| Product Code | JJX · Clinical Chemistry |
| Decision Date | Apr 13, 2006 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1660 |
| Device Class | Class 1 |
Intended Use
The Compass Bioscience Hemoglobin A1c Assayed Controls are intended for use as quality control materials to assess the accuracy and precision of assay procedures for HbA1c used in individual laboratories.
Device Story
Hemoglobin A1c Assayed Control consists of human whole blood matrix adjusted to target HbA1c concentrations; includes stabilizers and preservatives. Supplied as frozen liquid in two levels; ready-to-use; no reconstitution required. Used in clinical laboratories to monitor performance of HbA1c assay procedures; validates measurement accuracy and precision within clinically significant ranges. Healthcare providers use control results to verify instrument/reagent performance before reporting patient HbA1c values; ensures reliability of diagnostic testing for diabetes management.
Clinical Evidence
Bench testing only. Stability studies performed to validate shelf life and open-vial stability. Closed vial stability: 2 years at -15°C to -25°C (accelerated) and 12 months at 2-8°C. Open vial stability: 90 days at 2-8°C and 7 days at 20-25°C. Failure criterion: >10% deviation from initial analyte recovery.
Technological Characteristics
Human whole blood matrix; contains stabilizers and preservatives. Supplied as frozen liquid in two levels. Storage requirements: -15°C to -25°C (long term) or 2-8°C (refrigerated). No software or electronic components.
Indications for Use
Indicated for use as a quality control material to assess accuracy and precision of laboratory test methods for measuring Hemoglobin A1c levels in patient samples. Intended for clinical laboratory use.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
- MAS Diabetes Control (Medical Analysis Systems)
Related Devices
- K023032 — LIQUITROL, LIQUID HUMAN GLYCOSYLA,TED HEMOGLOBIN (A1C) CONTROL - NORMAL/ABNORMAL · Bioresource Technology, Inc. · Oct 8, 2002
- K103744 — DROPPER A1C DIABETES CONTROL · Quantimetrix Corp. · May 12, 2011
- K131999 — HBA1C CONTROL, LEVEL 1 AND LEVEL 2 · Randox Laboratories, Ltd. · Jul 26, 2013
- K032791 — GLYCOHEMOSURE HBA1C CONTROL · Quantimetrix Corp. · Oct 30, 2003
- K103099 — PRECICONTROL HBA1C NORM; PRECICONTROL HBA1C PATH · Roche Diagnostics · Dec 6, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
K06057\$
Prepared: 2/22/2006 Submitted by: Evy K. Johnson Establishment Address: Compass Bioscience 1850 Evergreen Street Duarte CA 91010 phone: 626.359.9645 626.359.7697 fax:
APR 1 3 2006
Establishment Registration Number: 2032652
Contact Person: Dr. Carter Grandjean
Proprietary (Trade) Name:
Common Name: Hemoglobin A1c Assayed Control
Classification Name: Single Analyte Controls, All Kinds (assayed and unassayed) 75 JJX
Classification: Class I
### Substantial Equivalence:
The Compass Bioscience Hemoglobin A1c Control is supplied as frozen liquid in two levels and consists of a human whole blood matrix adjusted to target concentrations of glycated hemoglobin (HbA1c) and containing preservatives and stabilizers. Assayed values are determined from interlaboratory data.
The Compass Bioscience controls are substantially equivalent to other such controls in general use, such as the MAS Diabetes Control, sold by Medical Analysis Systems, which is supplied liquid in two levels as a human whole blood matrix adjusted to specific concentrations of glycated hemoglobin.
#### Description:
Hemoglobin A1c Assayed Controls are supplied in two levels, 3 x 1 mL each level per box . The controls are supplied as a ready-to-use frozen liguid, requiring no reconstitution or dilution. They are prepared in a human whole blood matrix adjusted to target concentrations of HbA1c. Stabilizers and preservatives have been added to inhibit microbial growth.
#### Intended Use:
The Compass Bioscience Hemoglobin A1c Assayed Controls are intended for use as quality control materials to assess the accuracy and precision of assay procedures for HbA1c used in individual laboratories.
## Technological Characteristics Compared to Predicate Devices:
The Compass Bioscience control product employs a similar matrix and constituent formulation to the equivalent predicate device listed above: human whole blood matrix with adjusted (HBA1c) levels containing stabilizers and preservatives. The Compass Bioscience Control also has similar storage and stability requirements as the equivalent devices.
{1}------------------------------------------------
## Performance Characteristics:
The closed vial stability claim made for this product is 2 years when stored at -15° C to -25° C, based on accelerated stability studies.2 The Hemoglobin A1c control was stored at 2-8°C to simulate 2 years storage at -15°C to -25° C. An increase or decrease of >10% of analyte recovery compared to the initial test value ± the highest allowable instrument/reagent imprecision was used as the analyte failure criterion for determining shelf life. Real time stability testing is ongoing on multiple lots of product.
The closed vial stability claim for this product when stored at 2-8° C is 12 months. Accelerated stability testing at 20 - 25° C was used to determine the closed vial refrigerated shelf life. The Hemoglobin A1c Control was stored at 20 - 25° C and the recovery (vs. day 0) was measured in intervals of 7 days. The product passed stability if its recovery was within ±10% of the day 0 value after 35 days of storage at 20 - 25° C.
The open vial stability claim for this product is 90 days when stored at 2 - 8° C. Real time testing of vials opened, warmed to room temperature, sampled, and returned to the refrigerator, and re-used at various intervals in the same manner was performed and measured against freshly opened vials. The product passed stability if its recovery was within ±10% of the Day 0 value after 88 days of storage at 2 - 8° C. Multiple lots of product were tested with no significant difference in performance.
The open vial stability claim for this product is 7 days when stored at 20 - 25° C (Room Temperature). Real time testing of vials opened, sampled, re-capped, and re-used at various intervals in the same manner was performed and measured against freshly opened vials. The product passed stability if its recovery was within ±10% of the day 0 value after 10 days of storage at 20 - 25° C. Multiple lots of product were tested with no significant difference in performance.
The equivalent predicate device, MAS Diabetes Control, claims a 2 year shelf life for storage at -20° C and a 14 day opened/closed vial stability at refrigerated (2 - 8° C) storage.
# Assayed Values
Assay values were established from interlaboratory data using instrument manufacturers' reagents. Mean values and expected ranges for the listed lots of controls were calculated from reagent lots available at the time of assay.
<sup>2</sup> L. Kennon, Stability Prediction Model, Journal of Pharmaceutical Sciences 53:7, 815-818, 1964.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers. The eagle is positioned to the right of a circular border containing the text "U.S. Department of Health and Human Services". The text is arranged around the circumference of the circle.
Public Health Service
APR 1 3 2006
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Evy Johnson Chief Operating Officer Compass Bioscience 1850 Evergreen Street Duarte CA, 91010
k060570 Re:
> Trade/Device Name: Hemoglobin Alc Assayed Control Regulation Number: 21 CFR§ 862.1660 Regulation Name: Quality control material Regulatory Class: Class I Product Code: JJX Dated: February 28, 2006 Received: March 7, 2006
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Alberto Gutierrez
Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K060570
Device Name: Hemoglobin A1c
Indications For Use:
The Compass Bioscience Hemoglobin A1c Control is to be used as a quality control material to assess the accuracy and precision of laboratory test methods used to measure Hemoglobin A1c levels. It is intended to validate the measurement of Hemoglobin A1c in patient samples.
Two levels of control are provided to allow the performance of the analyte test methods to be monitored within the clinically significant range.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Carol Benson
Division Sign-Off
Office of in Vitro Diagnostic Device Evaluation and Safety
510(k) K060570
Page 1 of 1