VACU LAB PLAIN TUBE, VACU LAB GEL & CLOT ACTIVATOR TUBE, VACU LAB COAGULATION TUBE

K060364 · Teco Diagnostics · JKA · Jun 15, 2006 · Clinical Chemistry

Device Facts

Record IDK060364
Device NameVACU LAB PLAIN TUBE, VACU LAB GEL & CLOT ACTIVATOR TUBE, VACU LAB COAGULATION TUBE
ApplicantTeco Diagnostics
Product CodeJKA · Clinical Chemistry
Decision DateJun 15, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Intended Use

The Vacu Lab Plain Tube is a sterile, plastic, evacuated blood collection tube with a silica clot activator that provides a means of collecting, transporting, separating, and processing blood in a closed tube. The specimens are used for clinical laboratory assays involving the use of patient serum. The Vacu Lab Gel & Clot Activator Tube is a sterile, plastic, evacuated blood collection tube with a silica clot activator and a barrier gel that provides a means of collecting, transporting, separating, and processing blood in a closed tube. The specimens are used for clinical laboratory assays involving the use of patient serum.

Device Story

Teco Blood Specimen Collection Devices consist of sterile, evacuated plastic tubes used for venous blood collection. The Vacu Lab Plain Tube contains a silica clot activator and silicone surfactant coating; the Vacu Lab Gel & Clot Activator Tube adds a barrier gel. Used in clinical settings by healthcare personnel to collect and transport blood samples. Following collection, tubes are centrifuged to separate serum from cellular components. The resulting serum is used for various clinical chemistry assays. The closed system design facilitates safe handling and processing. The device benefits patients by providing a standardized, stable medium for serum-based diagnostic testing, ensuring sample integrity for laboratory analysis.

Clinical Evidence

Bench testing and clinical comparison studies performed. Four studies compared Teco tubes to BD Vacutainer Plus SST tubes using 182 total paired samples (66 in Study I, 25 in Study II, 70 in Study III, 25 in Study IV). Assays included total protein, albumin, bilirubin, AST, ALT, ALP, GGT, cholesterol, electrolytes, glucose, creatinine, CK, HCG, magnesium, phosphorus, triglycerides, uric acid, BUN, TSH, Free T4, ferritin, and prolactin. All studies demonstrated correlation coefficients ≥ 0.95. Stability studies confirmed no significant differences between fresh samples and those stored for 72 hours at 2-8°C. Shelf-life validated via accelerated aging (50 days at 60°C).

Technological Characteristics

Sterile, evacuated plastic tubes. Components: closure assembly, silica clot activator, silicone surfactant coating, and barrier gel (in Gel & Clot Activator version). Dimensions/form factor: standard evacuated tube. Connectivity: N/A. Sterilization: Not specified. Materials: Plastic tube, silica, silicone, barrier gel. Standards: NCCLS EP9-A2 (Method Comparison), NCCLS H1-A5 (Tubes and Additives).

Indications for Use

Indicated for the collection, transport, separation, and processing of venous blood in a closed system for clinical laboratory assays requiring serum specimens. Intended for use by healthcare professionals in clinical settings.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K060364 # TECO DIAGNOSTICS Image /page/0/Picture/3 description: The image shows a black hexagon with the letters "TC" in white inside. The letters are in a simple, sans-serif font. The hexagon is solid black, providing a strong contrast to the white letters. The overall design is simple and modern. 1268 N. Lakeview Ave., Anaheim, CA 92807, U.S.A. Phone: (714) 693-7788 Toll Free: 1-800-222-9880 Fax: (714) 693-3838 www.tecodiag.com Email:tecodiag@tecodiag.com ## 510 K Summary Teco Vacu Lab. Plain Tube & Vacu Lab Gel & Clot Activator Tube ## Device Name The device trade names and common/classifications name are: | Device Trade Name | Common/Classification Name | |------------------------------------------------------------------------------------------------------|-----------------------------------------------------------| | Teco Blood Specimen Collection Devices:<br>Vacu lab Plain Tube<br>Vacu Lab Gel & Clot Activator Tube | Tubes, Vials, Systems, Serum Separators, Blood Collection | ## Address and Registration The address and registration number of the manufacturer site for Teco Blood Specimen Collection Devices. TECO Diagnostics 1268 N. Lakeview Ave. Anaheim, CA 92807, U.S.A. FDA Registration # 1832216 Contact Person: Jian Vaeches Prepared Date: 03/12/2006 #### Device Class Teco Blood Specimen Collection Devices have been classified as Class II with Product Code JKA"in vitro" diagnostics devices having the classification number: 21 CFR. 862.1675. This is the description available from the classification names listed in the " CDRH Home Page- Listing Database." ### Labeling and Intended Use Draft labels and Instructions for use can be found in appendix A. #### Intended Use Teco Blood Specimen Collection Devices: Vacu lab Plain Tube and Vacu Lab Gel & Clot Activator Tube provide a means for collection, processing and transportation of blood in a closed system. Blood collected in these {1}------------------------------------------------ tubes is primarily used for clinical laboratory chemistry assays, but may be used for other assays requiring serum specimens as determined by the laboratory. #### Device Description The Teco Blood Collection Devices include: Vacu Lab Plain Tube; Vacu Lab Gel & Clot Activator Tube; - 1. The Vacu Lab Plain Tube is sterile, plastic, evacuated blood collection tube. The tube consists of (1) a closure assembly, (2) a silica clot activator, and (3) a silicone surfactant coated plastic tube. The specimens are used for clinical laboratory assays involving the use of patient serum. - 2. The Vacu Lab Gel & Clot Activator Tube is sterile, plastic, evacuated blood collection tube. The tube consists of (1) a closure assembly, (2) a silica clot activator, (3) a Barrier Gel and (4) a silicone surfactant coated plastic tube. The specimens are used for clinical laboratory assays involving the use of patient serum. Method Comparison of Vacu Lab Plain Tube; Vacu Lab Gel & Clot Activator Tube to BD Vacutainer ™ Plus SST ™ Tube following the guidelines of NCCLS Guideline EP9-A2 was conducted. #### Substantial Equivalence Base on a comparison of the device features, materials and intended use, the Teco Blood Specimen Collection Device: Vacu lab Plain Tube and Vacu Lab Gel & Clot Activator Tube is substantially equivalent to the predicate devices. The predicate device is BD Vacutainer TM Plus SST TM Tube The 510 (K) approval letter is provided in Appendix I. 510(K) #: K023075 Approval Date: 11-25-2002 #### Synopsis of Test Methods and Results #### Comparison Studies: Clinical evaluations were performed to determine the safety and efficiency of Teco Blood Specimen Collection Devices: Vacu lab Plain Tube and Vacu Lab Gel & Clot Activator Tube. The devices were compared to the predicate devices BD Vacutainer ™ Plus SST ™ Tube. The results of the clinical evaluation demonstrated that the Teco Blood Specimen Collection Devices: Vacu lab Plain Tube and Vacu Lab Clot Activator Tube provide clinically equivalent chemistry analyte results when compared to the BD Vacruainer TM Plus SST TM Tube. Four studies were conducted to evaluate the use of Teco Blood Specimen Collection Devices: Vacu lab Plain Tube and Vacu Lab Gel & Clot Activator Tube in chemistry assays. Study I was conducted to evaluate the use of the Vacu Lab Plain Tube. The objective of the studies was to demonstrate substantial equivalence to the Becton Dickinson (BD) Vacutainer TM Plus SSS ™ Tube (K023075) when samples from these tubes are used in chemistry assays included total protein, Albumin, Total Bilirubin, Direct Bilirubin, AST, ALT, ALP, r-GT, Cholesterol. A total 66 paired samples were collected from outpatients and samples were tested on Hitachi 717. The Vacu Lab Plain Tube demons "rated equivalent results to the predicate device. Study II was continuous conducted to evaluate the use of the Vacu Lab Plain Tube. Randomly collect outpatient blood samples 25. Draw the same amount blood from each patient into the Vacu Lab Plain Tube and BD Vacutainer TM Plus SST ™ Tube and Centrifye 10 minutes. Test the supermatant. Assays performed on Beckman CX7, Hitachi 912 and Elisa Reader TC98 included potassium, glucose, creatinine, calcium, chloride, creatinine kinase (CK), human chorionic gonadotropin (HCG), magnesium, phosphorous, triglyceride, uric acid, blood urea nitrogen (BUN), sodium, CK-MB, thyroid stimulating homone (TSS), Free T4, ferritin, prolactin, total protein, albumin, total bilirubin, direct bilirubin, AST, ALT, J.J. J.-GT {2}------------------------------------------------ and Cholesterol. Test conducted within 8 hours after collecting samples. The Vacu Lab Plain Tube demonstrated equivalent results to the predicate device. Study III was conducted to evaluate the use of Vacu Lab Gel & Clot Activator Tube. The objective of the studies was to demonstrate substantial equivalence to the Becton Dickinson (BD) Vacutainer ™ Plus SST TM Tube (K023075) when samples from these tubes are used in chemistry assays included total protein, Albumin, Total Bilirubin, Direct Bilirubin, AST, ALT, ALP, r-GT, Cholesterol. A total 70 paired samples were collected from outpatients and samples were tested on Hitachi 717. The Vacu Lab Gel & Clot Activator Tube demonstrated equivalent results to the predicate device. Study IV was continuous to evaluate the use of Vacu Lab Gel & Clot Activator Tube. Randomly collect outpatient blood samples 25. Draw the same amount blood from each patient into Vacu Lab Gel & Clot Activator Tube and BD Vacutainer "" Plus SST TM Tube and Centrifuge 10 minutes. Test the supernatiant. Assays performed on Beckman CX7, Hitachi 912 and Elisa Reader TC98 included potassium, glucose, creatinine, calcium, chloride, creatinine kinase (CK), human chorionic gonadotropin (HCG), magnesium, phosphorous, triglyceride, uric acid, blood urea nitrogen (BUN), sodium, CK-MB, thyroid stimulating hormone (TSH), Free T4, ferritin, prolactin, albumin, total bilirubin, direct bilirubin, AST, ALT, ALP,r-GT and Cholesterol. Test conducted within 8 hours after collecting samples. The Vacu Lab Gel & Clot Activator Tube demonstrated equivalent results to the predicate device. All comparison studies yield correlation coefficient ≥ 0.95. An acceptable correlation coefficient for all comparison studies is ≥ 0.95 #### Stability Studies of Specimens Studies were conducted to compare fresh specimens (8 hours after collecting samples) and 72 hours stored at 2-8 ℃. All results showed no significant different from fresh specimens and 72 hours stored at 2-8 ℃. All comparison studies yield correlation coefficient > 0.95. An acceptable correlation coefficient for all comparison studies is ≥ 0.95 #### Shelf- life Studies The shelf studies are based on accelerated aging theory, Q 10 method. Put tubes in 60 centigrade oven for 50 days, to achieve equivalency at least 24 months of real life aging of Vacu Lab Plain Tube and Vacu Lab Gel & Clot Activator Tube at 4-30 centigrade. The calculation as below: The real time shelf life = Oven aging times (days) X Q (0) (0)en aging temperature-storage temperature) 10-50 x 2 (60-20) 10 =50 x 2 4 =800 days. So we claimed that the shelf life is 2 years. Note: Q 10: reaction-rate coeffiecient 2. Average Storage temperature: 20 centigrade. Procedure: Put Vacu Lab Plain Tube and Vacu Lab Gel & Clot Activator Tube into 60 centigrade oven for 50 days, After 50 days, we collected 25 outpatient blood specimens. We drew each patient blood into viou Lab Plain Tubes, Vacu Lab Gel & Clot activator Tubes and BD vacutainer TM Plus SST Tm Tubes to compare studies. All tests were undertaken on Hitach912 and Elisa Reader TC 98. The results of comparison studies did not show any significant difference from the results of BD vacutainer TM Plus SUT Tm Tubes. All comparison studies yield correlation coefficient ≥ 0.95. An acceptable correlation coefficient for all comparison studies is ≥ 0.95 ## Standard/ Guidance Document Referenced (if applicable) NCCLS Guideline EP9-A2 -- Method Comparison and Bias Estimation Using Patient Samples. NCCLS HI-A5-Tubes and Additives for Venous Blood Specimen Collection: Anproved Standard-5 th, ed. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. The text is in all capital letters and is evenly spaced around the circle. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mr. Jian Vaeches Official FDA Correspondent TECO Diagnostics 1268 N. Lakeview Anaheim, CA 92807 JUN 15 2006 Re: k060364 > Trade/Device Name: Teco Diagnostics Vacu Lab Plain Tube and Vecu Lab Gel and Clot Activator Tube Regulation Number: 21 CFR§862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: JKA Dated: May 30, 2006 Received: May 30, 2006 Dear Mr. Vaeches: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CliR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ ## Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Alberto Gutierrez, Ph.D. Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K 060 364 Device Name: Teco Diagnostics Vacu Lab Plain Tube and Vecu Lab Gel and Clot Activator Tube Indications For Use: - 1. The Vacu Lab Plain Tube is a sterile, plastic, evacuated blood collection tube with a silica clot activator that provides a means of collecting, transporting, separating, and processing blood in a closed tutter The specimens are used for clinical laboratory assays involving the use of patient serum. - 2. The Vacu Lab Gel & Clot Activator Tube is a sterile, plastic, evacuated blood collection tube with a silica clot activator and a barrier gel that provides a means of collecting, transporting, separating, and processing blood in a closed tube. The specimens are used for clinical laboratory assays involving the use of patient serum. Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) G.C.H. Division Sign-Off Office of In Vitro De one alle Device Evaluation and Safety KIDNEY Page 1 of I
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%