TELITHROMYCIN 15UG BBL SENSI-DISC

K052876 · Becton, Dickinson & CO · JTN · Nov 14, 2005 · Microbiology

Device Facts

Record IDK052876
Device NameTELITHROMYCIN 15UG BBL SENSI-DISC
ApplicantBecton, Dickinson & CO
Product CodeJTN · Microbiology
Decision DateNov 14, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1620
Device ClassClass 2

Intended Use

Antimicrobial Susceptibility Test Disks are used for semi-quantitative in vitro susceptibility testing by standardized agar diffusion test procedures. Telithromycin 15ug BBL™ is intended for use in determining the susceptibility to telithromycin of a wide range of bacteria, as described in the "Indications for Use" section. Zone sizes used for interpretation of tests, including control organism limits, were determined by the antimicrobic manufacturer and received FDA approval under NDA Number 21-144.

Device Story

Device consists of paper disks impregnated with 15µg telithromycin; used in clinical microbiology laboratories for semi-quantitative in vitro susceptibility testing via agar diffusion (Bauer-Kirby method). Disks applied to inoculated Mueller Hinton Agar (or specialized media for specific organisms); plates incubated; zones of inhibition measured. Healthcare providers compare zone diameters against established CLSI/NCCLS standards to categorize isolates as susceptible, intermediate, or resistant. Results guide antimicrobial therapy selection for patients with bacterial infections. Benefits include standardized, reliable determination of antibiotic susceptibility to inform clinical decision-making.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on established standardized agar diffusion methodology and performance standards defined by CLSI/NCCLS and FDA-approved drug inserts.

Technological Characteristics

Impregnated paper disks; 15µg telithromycin concentration. Principle: Agar diffusion (Bauer-Kirby). Media: Mueller Hinton Agar, Haemophilus Test Medium Agar, or Mueller Hinton Agar with 5% Sheep Blood. Manual measurement of inhibition zones. No software, electronics, or energy source.

Indications for Use

Indicated for in vitro agar diffusion susceptibility testing to determine bacterial susceptibility to telithromycin. Active against: S. aureus (methicillin/erythromycin susceptible), S. pneumoniae (including multi-drug resistant), H. influenzae, S. pyogenes (erythromycin susceptible), Streptococci (groups C and G), and Viridans group streptococci.

Regulatory Classification

Identification

An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Submission K052876 Telithromycin 15pg BBLTM Sensi-Disc™ # NOV 1 4 2005 ### 510(k) SUMMARY | SUBMITTED BY: | Becton Dickinson and Company 7 Loveton Circle Sparks, MD 21152 Phone 410-316-4905 Fax: 410-316-4499 | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------| | CONTACT NAME: | Dainelle N. Clark, Regulatory Affairs Specialist | | DATE PREPARED: | November 08, 2005 | | DEVICE TRADE NAME: | Telithromycin 15µg, BBL <sup>TM</sup> Sensi-Disc <sup>TM</sup> Antimicrobial Susceptibility Test Disks | | DEVICE COMMON NAME: | Antimicrobial Susceptibility Test Disks | | DEVICE CLASSIFICATION: | 21 CFR§866.1620, Class II (Product Code JTN), Susceptibility Test Disks, Antimicrobial | | PREDICATE DEVICE: | Other BBL <sup>TM</sup> Sensi-Disc <sup>TM</sup> (eg, Ciprofloxacin 5 µg, BBL <sup>TM</sup> Sensi-Disc <sup>TM</sup> ) | ### INTENDED USE: Antimicrobial Susceptibility Test Disks are used for semi-quantitative in vitro susceptibility testing by standardized agar diffusion test procedures. Telithromycin 15ug BBL™ is intended for use in determining the susceptibility to telithromycin of a wide range of bacteria, as described in the "Indications for Use" section. Zone sizes used for interpretation of tests, including control organism limits, were determined by the antimicrobic manufacturer and received FDA approval under NDA Number 21-144. {1}------------------------------------------------ ## 510(k) SUMMARY ### Indications for Use: Use of Telithromycin 15ug, BBL™ Sensi-Disc™ for in vitro agar diffusion susceptibility testing is indicated when there is a need to determine the susceptibility of bacteria to telithromycin. The concentration of 15ug has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA approved drug insert for this antimicrobic. ### Active In Vitro and in clinical infections against: #### Aerobic gram-positive microorganisms Staphylococcus aureus (methicillin and erythromycin susceptible isolates only) Streptococcus pneumoniae (including multi-drug resistant isolates) # Aerobic gram-negative microorganisms Haemophilus influenzae ### Active In Vitro Against: #### Aerobic gram-positive microorganisms Streptococcus pyogenes (erythromycin susceptible isolates only Streptococci (Lancefield groups C and G) Viridans group streptococci {2}------------------------------------------------ # DEVICE DESCRIPTION: Telithromycin 15ug BBL™ Sensi-Disc™ is prepared by impregnating high quality paper with accurately determined amounts of Telithromycin supplied by the drug manufacturer. Each Telithromycin disk is clearly marked on both sides with the agent and drug content. Telithromycin cartridges each contain 50 impregnated disks that are packed as either a single cartridge in a single box, or in a package containing ten cartridges. Telithromycin disks are used for semi-quantitative in vitro susceptibility evaluations by the agar diffusion test method. Agar diffusion susceptibility methods employing dried filter paper disks impregnated with specific concentrations of antimicrobial agents were developed in the 1940s. In order to eliminate or minimize variability in the testing, Bauer et al. developed a standardized procedure in which Mueller Hinton Agar was selected as the test medium. Various regulatory agencies and standards-writing organizations subsequently published standardized reference procedures based on the Bauer-Kirby method. Among the earliest and most widely accepted of these standardized procedures were those published by the U.S. Food and Drug Administration (FDA) and the World Health Organization (WHO). The procedure was adopted as a consensus standard by the Clinical and Laboratory Standards Institute (CLSI) [Formerly National Committee for Clinical Laboratory Standards (NCCLS)] and is periodically updated. ## DEVICE PRINCIPLE: Disks containing a wide variety of antimicrobial agents are applied to the surface of Mueller Hinton Agar plates [or Haemophilus Test Medium Agar for Haemophilus influenzae or Mueller Hinton Agar with 5% Sheep Blood for Streptococcus species] inoculated with pure cultures of clinical isolates. Following incubation, the plates are examined and the zones of inhibition surrounding the disks are measured and compared with established zone size ranges for individual antimicrobial agents in order to determine the agent(s) most suitable for use in antimicrobial therapy. The categorical interpretation [susceptible (S), intermediate (I), or resistant (R)] for the organism being tested with the antimicrobial agent is made by comparing zone diameters to those found in the respective organism tables of CLSI/NCCLS Document M2 ("Performance Standards for Antimicrobial Disk Susceptibility Tests) and of CLSI/NCCLS Document M100 ("Performance Standards for Antimicrobial Susceptibility Testing"). {3}------------------------------------------------ # DEVICE COMPARISON: The BBLTM Sensi-Disc™ Antimicrobial Susceptibility Test Disks - Telithromycin 15ug is similar to the BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks - Ciprofloxacin 5 ug in that: - Both methods are for antimicrobial susceptibility testing using paper disks impregnated with . an antimicrobial agent. - . Both methods have the same intended use. - . Both methods provide the user with antimicrobic minimum inhibitory concentration (MIC) results based on measurements of zone diameters. - . Both methods require the user to determine categorical interpretations (S/I/R) using the measured zone diameters against CLSI/NCCLS Approved Standards M2 and M100. - . Both methods use pure cultures of bacterial isolates. The BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks - Telithromycin 15µg differs from the BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks - Ciprofloxacin 5 µg in that: - BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks Telithromycin 15ug is a . susceptibility test that uses disks impregnated with the antimicrobic Telithromycin at a concentration of 15ug while the BBLTM Sensi-Disc™ Antimicrobial Susceptibility Test Disks - Ciprofloxacin 5 µg is a susceptibility test that uses disks impregnated with the antimicrobic Ciprofloxacin at a concentration of 5 ug. - . BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disk - Telithromycin 15 ug is a susceptibility test used to test a different battery of microorganisms than the BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disk - Ciprofloxacin 5 ug. # SUBSTANTIAL EQUIVALENCE TESTING DATA: See the Telithromycin drug package insert, "Susceptibility Tests: Diffusion Techniques" (Appendix 1). {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure with three horizontal lines above it, possibly representing the concept of health and well-being. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 NOV 1 4 2005 Ms. Dainelle N. Clark Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 k052876 Re: Trade/Device Name: BBL™ Sensi-Disc ™ Antimicrobial Susceptibility Test Disks, Telithromycin 15 ug Regulation Number: 21 CFR 866.1620 Regulation Name: Antimicrobial Susceptibility Test Disc Regulatory Class: Class II Product Code: JTN Dated: October 11, 2005 Received: October 12, 2005 Dear Ms. Clark: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {5}------------------------------------------------ #### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Sale, a Hopp Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # INDICATIONS FOR USE K052876 510(k) Number (if known): Device Name: BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks, Telithromycin 15ug ### Indications for Use: Use of Telithromycin 15μg, BBL™ Sensi-Disc™ for in vitro agar diffusion susceptibility testing is indicated when there is a need to determine the susceptibility of bacteria to telithromycin. The concentration of 15μg has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA approved drug insert for this antimicrobic. Active In Vitro and in clinical infections against: Aerobic gram-positive microorganisms Staphylococcus aureus (methicillin and erythromycin susceptible isolates only) Streptococcus pneumoniae (including multi-drug resistant isolates) Aerobic gram-negative microorganisms Haemophilus influenzae Active In Vitro Against: Aerobic gram-positive microorganisms Streptococcus pyogenes (erythromycin susceptible isolates only Streptococci (Lancefield groups C and G) Viridians group streptococci Prescription Use V (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (Per 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Freddie th. Poole Division Sign Off Division Sign-Off )ffice of In Vitro Diagnostic Device Evaluation and Safety Becton, Dickinson and Company 510(k)
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%