ACON BAR II ONE STEP BARBITURATES TEST STRIP & DEVICE

K050593 · ACON Laboratories, Inc. · DIS · May 6, 2005 · Clinical Toxicology

Device Facts

Record IDK050593
Device NameACON BAR II ONE STEP BARBITURATES TEST STRIP & DEVICE
ApplicantACON Laboratories, Inc.
Product CodeDIS · Clinical Toxicology
Decision DateMay 6, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3150
Device ClassClass 2

Intended Use

The ACON BAR II One Step Barbiturates Test Strip and ACON BAR II One Step Barbiturates Test Device are rapid chromatographic immunoassays for the qualitative detection of Barbiturates in urine at a cutoff concentration of 300 ng/mL. These tests are used to provide only a preliminary analytical result. All positive test results obtained with these devices must be confirmed by another test method, preferably GC/MS analysis. They are intended for healthcare professionals including professionals at point-of-care sites.

Device Story

The ACON BAR II One Step Barbiturates Test Strip and Test Device are lateral flow, competitive binding immunochromatographic assays. The device uses mouse monoclonal antibodies to detect Barbiturates in urine samples at a 300 ng/mL cutoff. The test is performed manually without instrumentation. A positive result is indicated by the absence of a colored line in the test region, while a negative result or a sample below the cutoff generates a colored line. A control line confirms proper sample volume and membrane wicking. Used in point-of-care settings by healthcare professionals, the device provides a preliminary screening result. Positive results require confirmation via GC/MS. The device aids in clinical decision-making by providing rapid, qualitative drug screening, allowing for immediate preliminary assessment of potential barbiturate presence.

Clinical Evidence

Clinical evaluation of 300 urine specimens compared the device against a predicate test and GC/MS. Results showed 99% positive agreement and 98% negative agreement with GC/MS. Overall agreement with GC/MS was 98.67% (95% CI: 97%-99%). A point-of-care (POL) study across three sites demonstrated 97% (262/270) correct interpretation by non-laboratory personnel, comparable to trained technicians.

Technological Characteristics

Lateral flow immunochromatographic assay; competitive binding principle; mouse monoclonal antibody-based detection; visual qualitative readout; no instrumentation required; manual operation; urine specimen matrix; 300 ng/mL cutoff concentration.

Indications for Use

Indicated for the qualitative detection of Barbiturates in human urine at a 300 ng/mL cutoff concentration. Intended for use by healthcare professionals, including those at point-of-care sites, to provide preliminary analytical results. Requires confirmation by a more specific chemical method, preferably GC/MS.

Regulatory Classification

Identification

A barbiturate test system is a device intended to measure barbiturates, a class of hypnotic and sedative drugs, in serum, urine, and gastric contents. Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). A barbiturate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) SUMMARY 8. This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The Assigned 510(k) number is KO50593 ## Submitter: ACON Laboratories, Inc. 4108 Sorrento Valley Boulevard San Diego, California 92121 Tel.: 858-535-2030 Fax: 858-535-2038 ## Date: March 7, 2005 ## Contact Person: Edward Tung, Ph.D. ## Product Names: ACON® BAR II One Step Barbiturates Test Strip ACON® BAR II One Step Barbiturates Test Device ## Common Name: Immunochromatographic test for the qualitative detection of Barbiturates in urine. ## Regulation Name: Barbiturates test system. ## Product Code: DIS {1}------------------------------------------------ ## Classification Number: ## 21.CFR. 862.3150 ## Device Classification: The Barbiturates test systems have been classified as Class II devices with moderate complexity. The ACON BAR II One Step Barbiturates Test Strip and the ACON BAR II One Step Barbiturates Test Device are similar to another FDA-cleared device for the qualitative detection of Barbiturates in urine specimens. These tests are used to provide only a preliminary analytical result. All positive test results obtained with these devices must be confirmed by another test method, preferably GC/MS analysis. ## Intended Use: The ACON BAR II One Step Barbiturates Test Strip and ACON BAR II One Step Barbiturates Test Device are rapid chromatographic immunoassays for the qualitative detection of Barbiturates in urine at a cutoff concentration of 300 ng/mL. These tests are used to provide only a preliminary analytical result. All positive test results obtained with these devices must be confirmed by another test method, preferably GC/MS analysis. They are intended for healthcare professionals including professionals at point-of-care sites. ### Description: The ACON BAR II One Step Barbiturates Test Strip and the ACON BAR II One Step Barbiturates Test Device are competitive binding, lateral flow immunochromatographic assays for the qualitative screening of Barbiturates in a urine sample. The test is based on the principle of antibody immunochemistry. It utilizes the mouse monoclonal antibody to selectively detect elevated levels of Barbiturates and its metabolite in urine at a cutoff concentration of 300 ng/mL. These tests can be performed without the use of an instrument. A drug-positive urine specimen will not generate a colored-line in the designated test region, while a negative urine specimen or a urine specimen containing Barbiturates at the concentration below the cutoff level will generate a colored-line in the test region. To serve as a procedural control, a coloredline should always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred. {2}------------------------------------------------ #### Comparison to a Predicate Device: A comparison of the features of the ACON BAR II One Step Barbiturates Test Strip and the ACON BAR II One Step Barbiturates Test Device versus a FDA-cleared Barbiturates test with 300 ng/mL Barbiturates cutoff is shown below: - Both tests are assays intended for the qualitative detection of Barbiturates in urine samples. . - Both tests are intended as a screening method that provides a preliminary analytical test . result. - Both tests are immunochromatographic, lateral flow assays for the rapid detection of . Barbiturates with a visual, qualitative end result. - Both tests utilize the same basic immunoassay principles that rely on antigen/ antibody . interactions to indicate a positive or negative result. - Both tests have a cutoff Barbiturates concentration of 300 ng/mL. . #### Safety and Effectiveness Data: #### Accuracy A clinical evaluation was conducted using 300 clinical urine specimens including approximately 10% of the specimens containing Barbiturates concentration fell between -25% cutoff to +25% cutoff range. This evaluation compared the test results between the ACON BAR II One Step Barbiturates Test Strip and the ACON BAR II One Step Barbiturates Test Device with a FDA-cleared Barbiturates test; as well as compared against data obtained from the customary Gas Chromatography/Mass Spectrometry analysis. These comparisons of data yielded the following results: ### ACON BAR II One Step Barbiturates Test Strip versus a FDA-cleared BAR Test: Positive Agreement: 120 / 120 > 99% (97% - 100%) ** Negative Agreement: 156 / 180 = 87% (81% - 91%)* Overall Agreement: 276 / 300 = 92% (88% - 95%)* ** Since the proportion cannot go above 100%, this is really a 97.5% confidence interval. - * 95% confidence intervals ### ACON BAR II One Step Barbiturates Test Device versus a FDA-cleared BAR Test: Positive Agreement: 120 / 120 > 99% (97% - 100%) ** Negative Agreement: 156 / 180 = 87% (81% - 91%)* Overall Agreement: 276 / 300 = 92% (88% - 95%)* ** Since the proportion cannot go above 100%, this is really a 97.5% confidence interval. * 95% confidence intervals {3}------------------------------------------------ ACON BAR II One Step Barbiturates Test Strip versus data obtained with GC/MS at the cutoff concentration of 300 ng/mL: | | Test<br>Result | Specimen Cutoff Range by GC/MS Data | | | | | | |---------------------------------|----------------|-------------------------------------|------------------|-------------------|-------------------|------------------|-------------------------------| | | | Negative+ | < -25%<br>Cutoff | -25% to<br>Cutoff | Cutoff to<br>+25% | > +25%<br>Cutoff | % Agreement | | ACON<br>BAR II<br>Test<br>Strip | Positive | 0 | 1 | 2 | 9 | 132 | 99% (141/142)<br>(96% - 99%)* | | | Negative | 149 | 0 | 6 | 0 | 1 | 98% (155/158)<br>(95% - 99%)* | ## ACON BAR II One Step Barbiturates Test Strip versus GC/MS. Total agreement with GC/MS: 296/300 = 98.67% (97%- 99%)* * Denotes 95% confidence interval. + Negative specimens were confirmed using GC/MS analysis by pooling these samples in groups of న్ | | | ACON BAR II One Step Barbiturates Test Device versus GC/MS. | |--|--|-------------------------------------------------------------| |--|--|-------------------------------------------------------------| | | Test<br>Result | Specimen Cutoff Range by GC/MS Data | | | | | | |----------------------------------|----------------|-------------------------------------|------------------|-------------------|-------------------|------------------|-------------------------------| | | | Negative+ | < -25%<br>Cutoff | -25% to<br>Cutoff | Cutoff to<br>+25% | > +25%<br>Cutoff | % Agreement | | ACON<br>BAR II<br>Test<br>Device | Positive | 0 | 1 | 2 | 9 | 132 | 99% (141/142)<br>(96% - 99%)* | | ACON<br>BAR II<br>Test<br>Device | Negative | 149 | 0 | 6 | 0 | 1 | 98% (155/158)<br>(95% - 99%)* | Total agreement with GC/MS: 296/300 = 98.67% (97%- 99%)* * Denotes 95% confidence interval. † Negative specimens were confirmed using GC/MS analysis by pooling these samples in groups of న్. ## Performance Characteristics and Other information: The performance characteristics of the ACON BAR II One Step Barbiturates Test Strip and the ACON BAR II One Step Barbiturates Test Device were verified by analytical sensitivity study, specificity and cross reactivity study, interference studies, precision study, read time flex study, temperature flex study, specimen storage and stability study. Study results indicate that these test devices are robust and can perform satisfactorily when used according to the "Indication for Use" statement specified in their package inserts. ## POL Study Summary: Test results obtained from three POL study sites indicated that personnel at different doctor's offices with various educational background and working experience could perform the ACON® BAR II One Step Barbiturates tests properly and interpret test results correctly in most cases (97%, 262/270). The POL study results are also comparable to those obtained from a trained lab technician (97%, 87/90). {4}------------------------------------------------ ## Conclusion: These clinical studies demonstrated substantial equivalency on performance among the ACON BAR II One Step Barbiturates Test Strip, the ACON BAR II One Step Barbiturates Test Device and a FDA-cleared Barbiturates test with the same Barbiturates cutoff concentration. It is also demonstrated that these tests are safe and effective in qualitatively detecting Barbiturates at a concentration of 300 ng/mL. The POL study demonstrated that these tests are suitable for healthcare professionals including professionals at point-of-care sites. {5}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircles the eagle. The text is arranged in a circular fashion around the eagle, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" at the top and "USA" at the bottom. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 MAY - 6 2005 Edward Tung, Ph. D. Regulatory Affairs ACON Laboratories 4108 Sorrento Valley Blvd San Diego, CA 92121 Re: k050593 Trade/Device Name: ACON BAR II One Step Barbiturates Test Strip ACON BAR II One Step Barbiturates Test Device Regulation Number: 21 CFR 862.3150 Regulation Name: Barbiturate test system Regulatory Class: Class II Product Code: DIS Dated: March 7, 2005 Received: March 8, 2005 Dear Dr. Tung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for accession to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The 1 ou may controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act that I DA has intact and regulations administered by other Federal agencies. You must or any I vatures all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {6}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, Jean M. Cooper MS, DUM Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ #### INDICATIONS FOR USE 11. 510(k) Number (if known): K0505959 ACON BAR II One Step Barbiturates Test Strip Device Name: ACON BAR II One Step Barbiturates Test Device Indications for Use: The ACON BAR II One Step Barbiturates Test Strip and the ACON BAR II One Step The ACON DAR II One Bep Bail chromatographic immunoassays for the qualitative Barbiturates Test Device are rapia virus at a designated cutoff concentration of 300 ng/mL delection of Darontales for healthcare professionals including professionals at point-of-care sites. This assay provides only a preliminary analytical test result. A more specific alternate Themical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used. × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Albert Satz Division Sign-Off Office of In Vitro Dlagnostic Device Evaluntion and Safety K050593 Page 1 of 1 __
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%