K042152 · The Binding Site · CFN · Nov 22, 2004 · Immunology
Device Facts
Record ID
K042152
Device Name
BINDING SITE IGG SUBLCASS OLYMPUS KITS
Applicant
The Binding Site
Product Code
CFN · Immunology
Decision Date
Nov 22, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5510
Device Class
Class 2
Intended Use
This kit is intended for quantifying human IgG subclasses 1, 2, 3, and 4 immunoglobulins in serum on the Olympus™ AU series analyzers. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body’s lack of ability to resist infectious agents.
Device Story
The Binding Site Human IgG Subclass Kit is an in vitro diagnostic reagent system designed for use on Olympus AU series clinical chemistry analyzers. The device utilizes immunoturbidimetric technology to quantify IgG subclasses (1, 2, 3, and 4) in human serum samples. The assay involves the reaction of specific antibodies with the target IgG subclasses, resulting in turbidity proportional to the concentration of the analyte, which is measured photometrically by the analyzer. The system is intended for use by laboratory professionals in clinical settings to assist in diagnosing protein metabolism disorders and immune deficiencies. The output provides quantitative concentration values, which clinicians use to assess patient immune status and guide therapeutic decisions.
Clinical Evidence
Bench testing only. Precision/reproducibility demonstrated with within-run CVs (1.2-5.5%) and between-run CVs (1.2-5.1%). Linearity confirmed via serial dilutions (r² > 0.99). Method comparison with predicate (n=100) showed strong correlation (r² 0.98-0.9889). Interference testing showed no significant effect from bilirubin, hemoglobin, or chyle.
Technological Characteristics
Turbidimetric assay. Reagents: monospecific sheep IgG antisera (liquid); latex-coated antisera for IgG3/4. Reaction accelerators included. Calibration: 6-point curve. Instrument: Olympus AU series analyzers. Software: standard instrument parameters for turbidimetric analysis.
Indications for Use
Indicated for the quantification of human IgG subclasses 1, 2, 3, and 4 in serum to aid in the diagnosis of abnormal protein metabolism and immune deficiency states in patients.
Regulatory Classification
Identification
An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
Predicate Devices
Binding Site IgG subclass kit for use on the Behring BNII analyzer (K012291)
Related Devices
K021081 — HUMAN IGG SUBCLASS LIQUID REAGENT KITS FOR USE ON THE HITACHI 911/7070 TURBIDIMETRIC ANALYSER · The Binding Site, Ltd. · Jun 17, 2002
K022083 — HUMAN IGG SUBCLASS KIT FOR USE ON THE BECKMAN IMMAGE ANALYSER · The Binding Site, Ltd. · Jul 16, 2002
K012291 — HUMAN IGG SUBCLASS LIQUID REAGENT KITS, PRODUCT CODE LK001.TA · The Binding Site, Ltd. · Aug 20, 2001
K982023 — PELICALSS HUMAN IGG SUBCLASS NEPHLOMETRIC ARRAY KIT · Inova Diagnostics, Inc. · Sep 4, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the symbol. The caduceus is a common symbol for healthcare and medicine, and the logo is used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
NOV 2 2 2004
The Binding Site, Ltd. c/o Mr. Jay H. Geller West Tower, Suite 4000 2425 West Olympic Boulevard Santa Monica, CA 90404
Re: k042152
Trade/Device Name: Binding Site Human IgG Subclass Kit for use on the Olympus AU Series of Instruments Regulation Number: 21 CFR 866.5510 Regulation Name: Immunoglobulins A, G, M, D, E Immunological Test Systems Regulatory Class: Class II Product Code: CFN Dated: August 4, 2004 Received: August 10, 2004
Dear Mr Geller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Page 2 – Mr. Jay H. Geller
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours,
Robert Becker/
Robert L. Becker, Jr., M.D., PhD Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known): K042152
Device Name: Human IgG Subclass Kit for Use on the Olympus AU Series
Indications for Use: This kit is intended for quantifying human IgG subclasses 1, 2, 3 and 4 immunoglobulins in serum on the Olympus AU series analysers. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mare In Yar
---
Division Sign-Off
Page 1 of 1
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K042152
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.