OLYMPUS HDL CHOLESTEROL REAGENT, OSR6195/OSR6295 AND OLYMPUS HDL CHOLESTEROL CALIBRATOR ODC0023

K040692 · Olympus America, Inc. · JIT · Apr 8, 2004 · Clinical Chemistry

Device Facts

Record IDK040692
Device NameOLYMPUS HDL CHOLESTEROL REAGENT, OSR6195/OSR6295 AND OLYMPUS HDL CHOLESTEROL CALIBRATOR ODC0023
ApplicantOlympus America, Inc.
Product CodeJIT · Clinical Chemistry
Decision DateApr 8, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1150
Device ClassClass 2

Intended Use

A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens.

Device Story

Olympus HDL Cholesterol Reagent & Calibrator is an in vitro diagnostic system for quantitative determination of HDL cholesterol in human serum and plasma. Used in clinical laboratory settings by trained personnel. Reagents react with patient samples to produce measurable signals proportional to HDL concentration; calibrators establish reference points for value determination. Results assist clinicians in diagnosing and managing lipid-related metabolic conditions, atherosclerosis, and organ-specific diseases. System provides objective data for clinical decision-making regarding patient lipid management.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

In vitro diagnostic reagent and calibrator system. Quantitative measurement of HDL cholesterol in serum/plasma. Standard clinical chemistry assay format.

Indications for Use

Indicated for measurement of lipoproteins in serum and plasma to aid in diagnosis and treatment of lipid disorders (e.g., diabetes mellitus), atherosclerosis, and liver/renal diseases.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/2 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three stripes extending from its head, representing the department's commitment to health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. APR = 8 2004 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mr. Michael Campbell Senior Manager, Regulatory Affairs/ Quality Assurance Olympus America Inc. 3131 West Royal Lane Irving TX, 75063-3104 k040692 Re: Trade/Device Name: Olympus HDL Cholesterol Reagent & Calibrator Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIT, LBS Dated: March 8, 2004 Received: March 17, 2004 Dear Mr. Campbell We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices markcted in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencics. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {1}------------------------------------------------ ## Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Jean M. Cooper, MS, DVM. Yean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: Olympus HDL Cholesterol Reagent & Calibrator Indications For Use: A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) 0101 Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Carol C. Binson Division Sign-Off **Division Sign-Off** Office of In Vitro Diagnostic Device Evaluation and Safety 5100K0 K040692 Page 1 of *_*_
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%