VITEK 2 GRAM POSITIVE MOXIFLOXACIN

K032399 · bioMerieux, Inc. · LON · Sep 10, 2003 · Microbiology

Device Facts

Record IDK032399
Device NameVITEK 2 GRAM POSITIVE MOXIFLOXACIN
ApplicantbioMerieux, Inc.
Product CodeLON · Microbiology
Decision DateSep 10, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Intended Use

VITEK 2® Gram Positive Moxifloxacin is designed for antimicrobial susceptibility testing of Staphylococcus aureus (methicillin-susceptible strains only). It is intended for use with the VITEK 2® System as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.

Device Story

VITEK 2 Gram Positive Moxifloxacin is an in vitro diagnostic test card used with the VITEK 2 System. The device accepts bacterial isolates diluted in 0.45% saline to rehydrate antimicrobial media within the card. The VITEK 2 system automatically fills, seals, and incubates the card, monitoring bacterial growth in each well for up to 18 hours. The system utilizes a miniaturized doubling dilution technique to determine the minimum inhibitory concentration (MIC). The output is a report providing the MIC value and interpretive category for the antibiotic. Used in clinical laboratories by technicians to assist healthcare providers in determining antimicrobial susceptibility, thereby informing clinical decision-making for patient treatment.

Clinical Evidence

Bench testing only. External evaluation compared VITEK 2 Gram Positive Moxifloxacin performance against the NCCLS agar dilution reference method using fresh and stock clinical isolates and stock challenge strains. Results demonstrated 99.2% overall Essential Agreement. Reproducibility and Quality Control testing yielded acceptable results.

Technological Characteristics

Automated antimicrobial susceptibility testing system. Uses miniaturized doubling dilution methodology in disposable cards. System performs automated filling, sealing, incubation, and optical monitoring of bacterial growth. Connectivity via VITEK 2 System. Software-based analysis of growth kinetics to determine MIC values.

Indications for Use

Indicated for antimicrobial susceptibility testing of Staphylococcus aureus (methicillin-susceptible strains only) in clinical laboratory settings.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K032399 SEP 1 0 2003 Image /page/0/Picture/2 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a stylized font, with a vertical line running through the center of the word. Above the word is a black circle with a series of lines on the left side, and the vertical line extends from the top of the circle. # 510(k) SUMMARY ### VITEK® 2 Gram Positive Moxifloxacin #### 510(k) Submission Information: | Submitter's Name: | bioMérieux, Inc. | |----------------------|--------------------------------------------------------------------------------------------------------| | Address: | 595 Anglum Road<br>Hazelwood, MO 63042 | | Contact Person: | Nancy Weaver<br>Staff Regulatory Affairs Specialist | | Phone Number: | 314-731-8695 | | Fax Number: | 314-731-8689 | | Date of Preparation: | July 18, 2003 | | B. Device Name: | | | Formal/Trade Name: | VITEK® 2 Gram Positive Moxifloxacin | | Classification Name: | Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility Device,<br>21 CFR 866.1645 | | Common Name: | VITEK 2 AST-GP Moxifloxacin | | C. Predicate Device: | VITEK 2 Gram Positive AST Panel for<br>Sparfloxacin (N50510/S141) | #### D. 510(k) Summary: VITEK® 2 Gram Positive Moxifloxacin is designed for antimicrobial susceptibility testing of Staphylococcus aureus (methicillin-susceptible strains only). It is intended for use with the VITEK® 2 System as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology. The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 automatically fills, seals and places the card into the incubator/reader. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up {1}------------------------------------------------ to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card. VITEK 2 Gram Positive Moxifloxacin demonstrated substantially equivalent performance when compared with the NCCLS reference agar dilution method, as defined in the FDA DRAFT document "Guidance on Review Criteria for Assessment of Antimicrobial susceptibility Devices", dated March 8, 2000 and Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA. Issued Feb. 5, 2003. The Premarket Notification (510[k]) presents data in support of VITEK 2 Gram Positive Moxifloxacin. An external evaluation was conducted with fresh and stock clinical isolates and stock challenge strains. The external evaluations were designed to confirm the acceptability of VITEK 2 Gram Positive Moxifloxacin by comparing its performance with the NCCLS agar dilution reference method. VITEK 2 Gram Positive Moxifloxacin demonstrated acceptable performance of 99.2% overall Essential Agreement when compared to the agar dilution reference method. Reproducibility and Quality Control demonstrated acceptable results. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is written around the perimeter of the circle. SEP 1 0 2003 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Nancy Weaver Staff Regulatory Affairs Specialist BioMérieux, Inc. 595 Anglum Road Hazelwood, MO 63042-2320 Re: k032399 Trade/Device Name: VITEK 2% Gram Positive Moxifloxacin (0.25-8 µg/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: July 29, 2003 Received: August 4, 2003 Dear Ms. Weaver: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ #### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement 510(k) Number (if known): K032399 Device Name: VITEK 2® Gram Positive Moxifloxacin (0.25 - 8 µg/ml) Indications for Use: VITEK 2® Gram Positive Moxifloxacin is designed for antimicrobial susceptibility testing of Staphylococcus aureus (methicillin-susceptible strains only). It is intended for use with the VITEK 2® System as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division Sign-Off the text in the image is Office of In Vitro Diagnostic Device Evaluation and Safety 510(k)_ 奖心32399 \$\checkmark\$ Perscription Use Over the country USA
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%