REAADS VON WILLEBRAND FACTOR ACTIVITY TEST KIT, MODEL 10826
K032139 · Corgenix, Inc. · GGP · Apr 5, 2004 · Hematology
Device Facts
| Record ID | K032139 |
| Device Name | REAADS VON WILLEBRAND FACTOR ACTIVITY TEST KIT, MODEL 10826 |
| Applicant | Corgenix, Inc. |
| Product Code | GGP · Hematology |
| Decision Date | Apr 5, 2004 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 864.7290 |
| Device Class | Class 2 |
Intended Use
The REAADS von Willebrand Factor Activity Test Kit is an enzyme-linked immunosorbent assay (ELSA) for the quantitative determination of von Willebrand Factor Activity (vWF:Act) in citrated human plasma. The REAADS von Willebrand Factor Activity Test Kit is intended to be used by clinical (hospital and reference) laboratories.
Device Story
REAADS von Willebrand Factor Activity Test Kit is an ELISA-based diagnostic assay for citrated human plasma. Process: patient plasma incubated in 96-microwell plates coated with monoclonal antibody; unbound proteins washed; bound antigen quantitated via HRP-conjugated anti-human vWF antibody; chromogenic reaction developed using TMB and H2O2. Output: optical density (O.D.) measured at 450nm via spectrophotometer; relative percent concentration calculated against reference plasma curve. Used in clinical/reference laboratories by trained personnel. Results assist clinicians in diagnosing and differentiating von Willebrand disease subtypes (Type I vs Type II) to guide treatment decisions.
Clinical Evidence
Clinical study compared REAADS kit to predicate device using 262 patient plasma samples. Statistical analysis via Pearson Product Moment Correlation demonstrated a positive correlation (r=0.943) with significant correlation (P<0.001) between the two test populations.
Technological Characteristics
Enzyme-linked immunosorbent assay (ELISA) in 96-microwell plate format. Utilizes monoclonal antibody for antigen capture and HRP-conjugated anti-human vWF antibody for detection. Chromogenic substrate: TMB and H2O2. Measurement: spectrophotometric optical density at 450nm. Total incubation time: 40 minutes at room temperature.
Indications for Use
Indicated for quantitative determination of von Willebrand Factor Activity (vWF:Act) in citrated human plasma for clinical laboratory use in the diagnosis of von Willebrand disease subtypes (Type I and Type II).
Predicate Devices
- Shield von Willebrand Factor Activity Kit (K000398)
Related Devices
- K972005 — REAADS VON WILLEBRAND FACTOR ANTIGEN TEST KIT · Reaads Medical Products, Inc. · Aug 13, 1997
- K983107 — VFE · Ramco Laboratories, Inc. · Oct 22, 1998
- K962675 — STA - LIATEST VWF TEST KIT · American Bioproducts Co. · Oct 25, 1996
- K040843 — HEMOSIL VON WILLEBRAND ACTIVITY · Instrumentation Laboratory CO · Apr 30, 2004
- K133005 — HEMOSIL VON WILLEBRAND ACTIVITY ASSAY · Instrumentation Laboratory CO · Mar 13, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
KO32139
APR = 5 2004
Image /page/0/Picture/2 description: The image shows the word "corgenix" in a simple, sans-serif font. A dark, circular shape obscures part of the "c" and extends upwards and to the left with a thin, curved line. The letters are evenly spaced and the word is presented in a straightforward manner.
## SUMMARY OF 510(K) SAPETY AND EFFECTIVENESS REAADS von Willebrand Factor Activity Test Kit March 31, 2004
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92,
The REAADS von Willebrand Factor Activity Test Kit is compured to a legally warketed prodicate device and a substantial equivalence claim made. The predicate device is Shield von Willebrand Factor Activity Kit (K000398) currently manufactured and marketed by Shield Diagnostics., Dundee, Scotland.
The REAADS von Willebrand Factor Activity Test Kit is an enzyme-linked immunosorbent assay (ELISA), utilizing the 96-microwell plate formal, similar to the predicate device. Diluted patient plasma is incubated in the wells, allowing any available antigen to bind to the monoclonal antibody on the microwell surface. The plates are washed to remove unbound proteins or other plasma molecules. Bound antigen is quantitated using horseradish peroxidase (HRP) conjugated anti-human von Willebrand Factor antibody. Following incubation, unbound conjugate is removed by washing. A chromogenic substrate of tetramethylbenzidine (TMB) and hydrogen peroxide (H2O2) is added to develop a colored reaction. The intensity of the color is measured in optical density (O.D.) units with a spectrophotometer at 450mm. Patient von Willebrand Factor Activity in relative percent concentration is determined against a curve made from the reference plasma provided with the kit. The total incubation time (at room temperature) of the assay is 40 minutes.
The intended use of the device is for the quantitative determination of von Willebrand Factor Activity (vWF: Act) in citrated human plasma. The laboratory diagnosis of von Willebrand disease may require both ( 11 the ) and qualitative (functional) determinations to differentiate the two predominant subtypes of the disease, type I and type II. The classification of von Willebrand disease into subtypes is important in determining the course of clinical treatment.
Clinical studies demonstrate that REAADS von Willebrand Factor Activity Test Kit is safe and effective. These clinical studies also indicate that REAADS von Willebrand Factor Activity Test Kit and the Shield I von Willebrand Factor Activity Kit are equivalent. A statistical evaluation of the entire group of clinical samples (n=262) was done to compare the two tests. Using the Pearson Product Moment Correlation statistical analysis, the correlation coefficient (r) demonstrates a positive correlation (r=0.943) and the P value (P<0.001) demonstrates a significant correlation of individual values between both populations.
Name Dept
2004-03-31
Nanci Dexter Director, Quality Assurance and Regulatory Affairs
Date
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The seal is black and white.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
APR - 5 2004
Ms. Nanci Dexter Director, Quality and Regulatory Affairs Corgenix, Inc. 12061 Tejon Street Westminster, CO 80234
Re: . k032139
> Trade/Device Name: REAADS von Willebrand Factor Activity Test Kit Regulation Number: 21 CFR 864.7290 Regulation Name: Factor deficiency Test Regulatory Class: Class II Product Code: GGP Dated: March 4, 2004 Received: March 5, 2004
Dear Ms. Dexter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR). Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{2}------------------------------------------------
## Page 2
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Joseph L. Archellett
Joseph L. Hackett, Ph.D. Acting Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use Statement
**510(k) Number:** K032139
Device Name: REAADS von Willebrand Factor Activity Test Kit
Indications for Use:
The REAADS von Willebrand Factor Activity Test Kit is an enzyme-linked immunosorbent assay (ELSA) for the quantitative determination of von Willebrand Factor Activity (vWF:Act) in citrated human plasma.
The REAADS von Willebrand Factor Activity Test Kit is intended to be used by clinical (hospital and reference) laboratories.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription use ✓
Josephine Buttedi
Division Sign-Off
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K032139