BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-NITROFURANTOIN-8-512 UG/ML
K031530 · Becton, Dickinson & CO · LON · Jul 10, 2003 · Microbiology
Device Facts
| Record ID | K031530 |
| Device Name | BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-NITROFURANTOIN-8-512 UG/ML |
| Applicant | Becton, Dickinson & CO |
| Product Code | LON · Microbiology |
| Decision Date | Jul 10, 2003 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.1645 |
| Device Class | Class 2 |
Intended Use
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus. This premarket notification is for the antimicrobial agent nitrofurantoin at concentrations of 8-512 ug/mL to Gram-negative ID/AST or AST only Phoenix panels. Nitrofurantoin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Device Story
BD Phoenix Automated Microbiology System performs rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. System components include instrument, software, molded polystyrene panels with 136 micro-wells, ID broth, AST broth, and redox indicator. Clinicians prepare pure culture isolates in broth (0.5 McFarland standard), inoculate panels, and load into instrument. Instrument incubates panels at 35°C, taking readings every 20 minutes. System uses redox indicator and turbidity measurements to detect bacterial growth in presence of antimicrobial agents. Output includes organism identification, MIC values, and category interpretations (S, I, R). Results assist healthcare providers in selecting appropriate antibiotic therapy for patients with bacterial infections.
Clinical Evidence
Performance evaluated using clinical, stock, and challenge isolates across multiple U.S. sites. Comparison made against NCCLS reference broth microdilution method. For nitrofurantoin (8-512 ug/mL) with Gram-negative panels, study included 1646 isolates. Results showed 96.1% Essential Agreement (EA) and 82.9% Category Agreement (CA). Reproducibility testing at three sites demonstrated intra-site reproducibility >90% and inter-site reproducibility >95%.
Technological Characteristics
System uses broth microdilution in a 136-well molded polystyrene panel. Sensing principle: redox indicator and turbidity measurement. Energy source: electrical (automated instrument). Connectivity: integrated system software. Incubation: 35°C continuous. Software: automated interpretation of growth curves to determine MIC and S/I/R categories.
Indications for Use
Indicated for in vitro antimicrobial susceptibility testing (MIC) of Gram-negative aerobic/facultative anaerobic bacteria (Enterobacteriaceae/Non-Enterobacteriaceae) and Gram-positive bacteria (Staphylococcus/Enterococcus) from pure culture. Specifically indicated for testing nitrofurantoin (8-512 ug/mL) against E. coli, Citrobacter spp., and Klebsiella spp.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
- VITEK® System (N50510)
- BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321)
- BD Phoenix™ Automated Microbiology System with Ofloxacin (K020323)
- BD Phoenix™ Automated Microbiology System with Levofloxacin (K020322)
Related Devices
- K123404 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM- IMIPENEM (0.0625-32 UG/ML) · Becton, Dickinson & CO · Jun 27, 2013
- K020322 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM · Becton, Dickinson & CO · Mar 27, 2002
- K052269 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM -CIPROFLOXACIN GN · Becton, Dickinson & CO · Oct 3, 2005
- K060444 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CEFEPIME (GN) , CEFTRIAXONE (GN), CEFOTAXIME (GN), 0.5-64 UG/ML · Becton, Dickinson & CO · Apr 12, 2006
- K082852 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-NITROFURANTOIN (GP) 4-128 UG/ML · Becton, Dickinson & CO · Jan 29, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 1 0 2003
## K031530 510(K) SUMMARY Becton, Dickinson and Company SUBMITTED BY: 7 Loveton Circle Sparks, MD 21152 Phone: (410) 316-4287 410-316-4499 Fax: Monica Evelyn Giguere, RAC CONTACT NAME: Regulatory Affairs Specialist May 14, 2003 DATE PREPARED: BD Phoenix™ Automated Microbiology System -DEVICE TRADE NAME: Nitrofurantoin 8-512 ug/mL Antimicrobial susceptibility test system-short incubation DEVICE COMMON NAME: Fully Automated Short-Term Incubation Cycle Antimicrobial DEVICE CLASSIFICATION: Susceptibility Device, 21 CFR 866.1645 VITEK® System (PMA No. N50510) and BD Phoenix™ PREDICATE DEVICES: Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002). The BD Phoenix™ Automated Microbiology System is INTENDED USE: intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Grampositive bacteria of human origin.
### DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- . BD Phoenix instrument and software.
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
- . BD Phoenix AST Broth used for performing AST tests only.
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth ● determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
# DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the NCCLS reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram-negative ID/AST or AST only Phoenix panels with this antimicrobial agent.
## SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the NCCLS reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000.
### Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of Gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram-negative Phoenix panels containing this antimicrobial agent and associated reagents.
The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the Gram-negative isolates tested.
{2}------------------------------------------------
## Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Gram-negative Phoenix panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the NCCLS reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within + one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).
Table 1 summarizes the performance for the isolates tested in this study.
#### Table 1: Performance of BD Phoenix System for Gram-negative Organisms by Drug
| Antimicrobial | Concentration | ---------------------------------------------------------------------------------------------------------------------------<br>The first of the first of the country of the status and the states of the status of the status of the status of the status of the status of the status of the status of the st<br>EA (n)<br>Acres of Children, cases career of the program and the comments of<br>Company of Children | EA (%)<br>The Research of the Research Acres of the Research and Children | I CA (n) | I CA (%)<br>Company of the country of the county of the county of the county of the county of |
|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|----------|-----------------------------------------------------------------------------------------------|
| and the control of the country of the province of the comments of the comments of the comments of<br> Nitrofurantoin | program and the control of the control control control control control control control controller of the consistence of the consisted to the consistence of the consistences o<br>8-512 ug/mL | 1646 | 96.1 | 1646 | 82.9 |
### Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized abstract image of a human figure with three flowing lines representing the body and head. The figure is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
# JUL 1 0 2003
Ms. Monica Evelyn Giguere Regulatory Affairs Specialist Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
Re: k031530
Trade/Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent nitrofurantion (8-512 ug/mL) - Gram-negative ID/AST or AST only Phoenix panels Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle AST Device Regulatory Class: Class II
Product Code: LON Dated: May 14, 2003
Received: May 15, 2003
Dear Ms. Giguere:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number: K03 15 36
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent nitrofurantoin (8-512 ug/mL) - Gram-negative ID/AST or AST only Phoenix panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.
This premarket notification is for the antimicrobial agent nitrofurantoin at concentrations of 8-512 ug/mL to Gram-negative ID/AST or AST only Phoenix panels. Nitrofurantoin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
### Active In Vitro and in Clinical Infections Against:
Escherichia coli
### Active In Vitro Against:
Citrobacter amalonaticus Citrobacter koseri (diversus) Citrobacter freundii Klebsiella oxytoca Klebsiella pneumoniae (ozaenae)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Freddie H. Pook
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
| Prescription Use | <span style="text-decoration:underline;">✓</span> |
|----------------------|---------------------------------------------------|
| (Per 21 CFR 801.109) | |
| 510(k) | K031530 |
|--------|---------|
|--------|---------|
| | Over-the-Counter Use |
|--|----------------------|
|--|----------------------|
BD Diagnostic Systems